K152163 · Smith & Nephew, Inc. · OMP · Sep 29, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152163
Device Name
RENASYS GO
Applicant
Smith & Nephew, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Sep 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RENASYS GO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include: - Chronic - Acute - Traumatic - Sub-acute and dehisced wounds - Ulcers (such as pressure or diabetic) - Partial-thickness burns - Flaps and grafts
Device Story
RENASYS GO is a lightweight, portable suction pump for negative pressure wound therapy (NPWT). It creates a closed environment over a wound using a sterile foam or gauze dressing connected via tubing to a disposable waste canister. The pump delivers continuous or intermittent negative pressure (40-200mmHg) to drain exudates, irrigation fluids, and infectious materials into the canister, which contains a gelling agent. Used in clinical or home settings, the device is operated by healthcare providers or patients. It features a 1.0-micron filter to protect the pump and a 0.2-micron bacterial filter. Output is monitored via device alarms for blockages, leaks, or canister overcapacity. By removing wound fluids, the device promotes healing. Modifications include improved alarm functionality, IEC 60601 3rd Edition compliance, and a 750ml canister option.
Clinical Evidence
Bench testing only. Testing verified pumping capacity, continuous/intermittent pressure delivery (40-200mmHg), alarm functionality (blockage, leak, overcapacity) using simulated exudate, system performance under fault conditions, and O-ring durability. Usability studies were conducted to verify labeling and alarm changes. Electrical safety testing performed per IEC 60601 3rd Edition standards.
Technological Characteristics
Powered suction pump; 40-200mmHg pressure range; mains or internal battery power. Components: therapy unit, 300ml/750ml single-use canisters with solidifier, 1.0-micron fluid filter, 0.2-micron bacterial filter. Software-controlled operation. Complies with IEC 60601-1, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, IEC 62366, and ISO 14971.
Indications for Use
Indicated for patients requiring negative pressure wound therapy for chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, flaps, and grafts.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
RENASYS GO Negative Pressure Wound Therapy Device (K083375)
Reference Devices
RENASYS Foam and Gauze dressing kits with Soft Port (K142979)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2016
Smith & Nephew, Inc. Ms. Laura Reynolds Director of Regulatory Affairs 970 Lake Carillon Drive Suite 110 St. Petersburg, Florida 33716
Re: K152163
Trade/Device Name: Renasys Go Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: September 26, 2016 Received: September 26, 2016
Dear Ms. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K152163
#### Device Name
RENASYS GO Negative Pressure Wound Therapy Device
#### Indications for Use (Describe)
The RENASYS GO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
Appropriate wound types include:
- · Chronic
- · Acute
- · Traumatic
- · Sub-acute and dehisced wounds
- · Ulcers (such as pressure or diabetic)
- · Partial-thickness burns
- · Flaps and grafts
#### Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Smith & Nephew logo. The logo consists of a stylized flower-like graphic on the left, followed by the text "We are smith&nephew". The words "We are" are in gray, while "smith&nephew" is in orange.
#### Wound Management 727 392-1261
F 727 392-6914 or 727 392-0797 Smith & Nephew, Inc. 970 Lake Carillon Drive Customer Care Center: 1 800 876-1261 Suite 300 www.smith-nephew.com St. Petersburg, FL 33716
# 510(k) Summary
#### General Information Submitters Name/Address: Smith & Nephew, Inc. 970 Lake Carillon Drive Suite 300 St. Petersburg, FL 33716 Establishment Registration Number: 3006760724 Contact Person: Laura Reynolds Director Regulatory Affairs Phone Number: (727) 686-8736 Date Prepared: September 29, 2016 Device Description RENASYS™ GO Negative Pressure Wound Trade Name: Therapy Device Generic/Common Name: Powered Suction Pump Classification Name: Powered Suction Pump; 21 CFR 878.4780 Product Code: OMP
#### Predicate Device Information
| Subject Device | Predicate Device | 510k# | Clearance<br>Date |
|---------------------------------------------------------|---------------------------------------------------------|---------|-------------------|
| RENASYS GO Negative<br>Pressure Wound Therapy<br>Device | RENASYS GO Negative<br>Pressure Wound Therapy<br>Device | K083375 | 02/25/2009 |
{4}------------------------------------------------
#### Device Description
#### RENASYS GO Negative Pressure Wound Therapy Device
The RENASYS GO NPWT device is a modification to predicate device K088375. It is a lightweight suction pump device intended for wound management via application of continuous or intermittent negative pressure wound therapy to the wound for removal of fluids, including wound exudates, irrigation fluids, and infectious materials. The RENASYS GO NPWT device is designed to deliver negative pressure wound therapy to a closed environment over a wound in order to drain exudates from the wound site to help promote wound healing.
The closed environment over the wound is created by applying a sterile foam or gauze wound dressing to the wound site and connecting the sealed wound to the suction pump via a tube that connects to the disposable canister. The suction pump delivers negative pressure wound therapy and removes the exudates from the wound site to the disposable canister.
The device provides negative pressure wound therapy to the wound at a range of pressure settings between 40-200mmHg. The device can operate either by a mains power supply or internal battery.
#### RENASYS GO Canisters
RENASYS GO uses an integral waste canister that is supplied non-sterile, single-use with a volume capacity of 300ml or 750ml. The waste canister is attached to the pump device by two clips on either side of the canister. The canister is permanently sealed to minimize the potential of users coming into contact with exudates. Safety features include a 1.0 micron filter in the top of the canister to resist fluid penetration into the device, as well as a 0.2 micron bacterial filter to prevent the passage of airborne bacteria. Each sealed canister contains a solidifier which acts as a gelling agent to the exudate.
A tube is permanently attached to the bottom of the waste canister through an inlet port. A connector attached to the distal end of the canister tube attaches to the corresponding tubing included in each Smith & Nephew NPWT dressing kit.
The RENASYS GO device is compatible with RENASYS Foam and Gauze dressing kits with Soft Port which were cleared under 510(k) K142979.
#### Indications for Use
#### RENASYS GO Negative Pressure Wound Therapy Device
RENASYS GO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
Appropriate wound types include:
- . Chronic
- Acute .
{5}------------------------------------------------
- Traumatic ●
- Sub-acute and dehisced wounds ●
- Ulcers (such as pressure or diabetic) .
- . Partial-thickness burns
- Flaps and grafts .
The indications for use of the subject device are identical to the predicate, except "Examples of" has been removed from appropriate wound types.
#### Comparison of Technological Characteristics
The RENASYS GO Negative Pressure Wound Therapy (NPWT) device and canisters that are the subject of this submission are substantially equivalent to the previously cleared versions of the RENASYS GO Negative Pressure Wound Therapy (NPWT) device and canisters. There has been no change to the indications for use, operating principle, mechanism of action or fundamental scientific technology of the predicate device. Generally, software and hardware modifications were made to improve alarm functionality and to meet IEC 60601 3rd Edition requirements. A 750ml canister has been added as an additional option for highly exuding wounds. Thus the subject and predicate devices have substantially equivalent technological characteristics.
#### Summary Comparison between New and Predicate Devices
| | Subject Device: | Predicate Device:<br>510(k)# K083375 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 510(k)#K152163 | |
| Trade Name: | RENASYS GO NPWT | RENASYS GO NPWT |
| Indications for Use: | The RENASYS GO is<br>indicated for patients who<br>would benefit from a<br>suction device (negative<br>pressure wound therapy)<br>as it may promote wound<br>healing via removal of<br>fluids, including irrigation<br>and body fluids, wound<br>exudates and infectious<br>materials.<br>Appropriate wound types<br>include:<br>• Chronic<br>• Acute<br>• Traumatic<br>• Sub-acute and<br>dehisced wounds<br>• Ulcers (such as<br>pressure or | The RENASYS GO is<br>indicated for patients who<br>would benefit from a<br>suction device (negative<br>pressure wound therapy)<br>as it may promote wound<br>healing via removal of<br>fluids, including irrigation<br>and body fluids, wound<br>exudates and infectious<br>materials.<br>Examples of appropriate<br>wound types include:<br>• Chronic<br>• Acute<br>• Traumatic<br>• Sub-acute and<br>dehisced wounds<br>• Ulcers (such as<br>pressure or |
### RENASYS GO NPWT
{6}------------------------------------------------
| | Subject Device:<br>510(k)#K152163 | Predicate Device:<br>510(k)# K083375 |
|-----------------------------|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Trade Name: | RENASYS GO NPWT<br>diabetic)<br>• Partial-thickness<br>burns<br>• Flaps and grafts | RENASYS GO NPWT<br>diabetic)<br>• Partial-thickness<br>burns<br>• Flaps and grafts |
| Principle of Operation | Same as predicate | Therapy unit delivers<br>software controlled<br>negative pressure wound<br>therapy to the wound site. |
| Operating Time<br>(Battery) | Same as predicate | 20 hours Therapy |
| Negative Pressure<br>Range | Same as predicate | 40-200mmHg |
| Components | • Therapy unit<br>• 300mL canister<br>• 750mL canister | • Therapy unit<br>• 300mL canister |
# Table of Modifications
| Modification | Reason For Change | Verification Testing<br>Performed |
|------------------------------------------|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hardware and Software<br>modifications | To improve alarm functionality | Software validation and<br>comprehensive<br>verification testing was<br>completed which<br>demonstrated<br>acceptable device<br>performance |
| Software and labeling<br>revisions | IEC 60601 3rd Edition<br>Compliance | Software verification and<br>usability studies were<br>completed which<br>demonstrated<br>acceptable device<br>performance. Electrical<br>safety testing in<br>accordance with IEC<br>60601 3rd Edition<br>standards has been<br>successfully completed. |
| Addition of RENASYS<br>GO 750ml canister | Additional option for highly<br>exuding wounds | Comprehensive<br>verification was<br>completed which<br>demonstrated<br>acceptable device<br>performance |
| Design change to | To improve robustness of | Comprehensive |
{7}------------------------------------------------
| Modification | Reason For Change | Verification Testing<br>Performed |
|----------------------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------|
| canister and<br>qualification of alternate<br>materials of<br>construction | device and aid<br>manufacturability | verification was<br>completed which<br>demonstrated<br>acceptable device<br>performance |
| Markings on canister<br>added to include<br>change indicator | To aid/alert user to<br>appropriate change interval | Usability studies were<br>completed to verify<br>labeling changes |
# Non-Clinical Tests (Bench)
#### RENASYS GO Negative Pressure Wound Therapy Device
Testing has been conducted to verify the modifications to the RENASYS GO NPWT device meet design specifications and demonstrate substantial equivalence to the predicate device.
The list below summarizes the bench testing undertaken and successfully completed for the RENASYS GO NPWT device:
- . Pumping capacity is equivalent to the predicate device.
- Verification that the device delivers negative pressure wound therapy in a continuous and intermittent operating mode identical to the predicate device.
- . Verification of Complete Blockage/Canister Overcapacity alarm functionality using wound fluid designed to simulate chemistry and protein content of real exudate.
- Verification of system performance in foreseeable fault conditions.
- Verification of system performance when running with high air leaks at the dressing site.
- Verification of system performance in worst case scenarios with ranges of exudate viscosity and protein content.
- Verification of Blockaqe/Canister Over-capacity & Leak Alarm assertion when . used in a vertical orientation.
- Verification of Blockage/Canister Over-capacity & Leak Alarm assertion when ● used in a horizontal face up orientation.
- Verification of Blockage/Canister Over-capacity & Leak Alarm assertion when ● used in a horizontal face down orientation.
- . Verification of Renasys GO performance at increased heights above the wound.
- . Verification of O-Ring durability.
The software documentation in this submission has been assembled according to the recommendations in the FDA document. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005. The software Level of Concern has been evaluated and determined to be Moderate, and appropriate documentation included, as recommended by the cited FDA guidance.
{8}------------------------------------------------
Device complies with the following standards:
- . ISO 14971:2012 Medical Devices - Application of Risk Management to Medical Devices
- . ISO 15223-1:2012 Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied Part 1: General requirements. (General)
- ISO 15223-2:2010 Medical devices Symbols to be used with medical device ● labels, labelling, and information to be supplied-Part 2: Symbol development, selection and validation. (General)
- . IEC 60601-1:2005 (3rd Edition). Medical Electrical Equipment - Part 1: General Requirements for Safety
- . IEC 60601-1-6:2010 (3rd Edition) Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-8:2006 (2nd Edition) Medical electrical equipment Part 1-8: ● General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical
- IEC 60601-1-11:2010 (1st Edition) Medical Electrical Equipment Part 1-11: ● General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 62366:2007 (1st Edition) Medical devices Application of usability . engineering to medical devices
- ANSI/AAMI ES60601-1:2005 Version (R2012) Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance
- RTCA/DO-160G, Environmental Conditions and Test Procedures for Airborne Equipment, Sections 18, 19, 20.4, 20.5, 21.4, 21.5.
- ANSI/AAMI HE75:2009 Human Factor Engineering-Design of Medical Devices
- IEC CISPR-25 2008, 3rd Ed. Vehicles, Boats and Internal Combustion Engines-● Radio Disturbance Characteristics.
- EN 50121-3-2 Part 3-2:2015 Railway Applications Electromagnetic Compatibility
# Conclusions
In establishing substantial equivalence to the currently marketed predicate device, Smith & Nephew, Inc. evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing and electrical safety testing has been successfully completed to demonstrate that the RENASYS GO NPWT device and canisters are substantially equivalent to the predicate devices for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.