← Product Code [GCJ](/productcode/GCJ) · K152149

# VersaOne Bladed Trocar 5mm (K152149)

_Covidien · GCJ · Jan 22, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152149

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 22, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry

## Device Story

VersaOne™ Bladed Trocar is a sterile, single-use surgical instrument used to create and maintain a port of entry during endoscopic procedures. The device consists of an obturator with a dolphin-nose shaped tip featuring a sharp linear blade and a spring-loaded locking shield, and a cannula assembly. During insertion, the blade cuts tissue; upon entry into a free space, the shield automatically advances to cover the blade. The cannula assembly includes a seal system to maintain pneumoperitoneum during instrument exchange, a transparent sleeve with a beveled distal end to reduce insertion forces, and a stopcock for insufflation/desufflation. Used in OR settings by surgeons; the device facilitates access for endoscopic instruments, benefiting patients by enabling minimally invasive surgical techniques.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench performance testing and biocompatibility evaluation conducted in accordance with ISO 10993-1.

## Technological Characteristics

Dolphin-nose (conical) tip with sharp linear blade; spring-loaded locking shield; transparent cannula sleeve with beveled distal end; seal system; stopcock valve. Biocompatibility per ISO 10993-1. Sterile, single-use.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Autosuture™ Modified Versaport™ Trocar with Fixation Sleeve ([K062326](/device/K062326.md))
- Versaport™ V2 Bladeless Optical ([K112349](/device/K112349.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, layered on top of each other. The profiles are connected by a flowing line that forms a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 22, 2016

Covidien Trang Huynh Senior Specialist, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473

Re: K152149

Trade/Device Name: VersaOne™ V2 Bladed Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: January 15, 2016 Received: January 20, 2016

Dear Trang Huynh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

## Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white design inside. The design appears to be three horizontal bars, with the middle bar being slightly shorter than the top and bottom bars.

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017 |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Indications for Use                                                     | See PRA Statement below.                                              |

| 510(k) Number ( <i>if known</i> )       | K152149                                                                                                                                      |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                             | VersaOne™ Bladed Trocar                                                                                                                      |
| Indications for Use ( <i>Describe</i> ) | indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry. |

| Type of Use ( <i>Select one or both, as applicable</i> )                                        |                                                                                                |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (1/14)<br>Page 1 of 1 |  |
|-------------------------------------|--|
|-------------------------------------|--|

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white design inside, followed by the word "COVIDIEN" in blue, sans-serif font. The blue square is on the left side of the logo, and the word "COVIDIEN" is on the right side.

## 510(k) Summary

This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92.

| SUBMITTER:              | Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473 USA                                                                                                                                                                                                       |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON:         | Trang Huynh<br>Senior Specialist, Regulatory Affairs                                                                                                                                                                                                                |
|                         | Phone: (203) 492-7473                                                                                                                                                                                                                                               |
|                         | Fax: (203) 492-5029                                                                                                                                                                                                                                                 |
|                         | Email: trang.huynh27@medtronic.com                                                                                                                                                                                                                                  |
| DATE PREPARED:          | July 30, 2015                                                                                                                                                                                                                                                       |
| TRADE/PROPRIETARY NAME: | VersaOne™ Bladed Trocar                                                                                                                                                                                                                                             |
| COMMON/USUAL NAME:      | Surgical Trocar                                                                                                                                                                                                                                                     |
| CLASSIFICATION NAME:    | Endoscope and Accessories                                                                                                                                                                                                                                           |
| PRODUCT CODE:           | GCJ                                                                                                                                                                                                                                                                 |
| FDA PANEL NUMBER:       | 78                                                                                                                                                                                                                                                                  |
| CLASS CODE:             | Pursuant to 21 CFR 876.1500, Class II device                                                                                                                                                                                                                        |
| PREDICATE DEVICES:      | Autosuture™ Modified Versaport™ Trocar with Fixation Sleeve<br>(K062326)                                                                                                                                                                                            |
|                         | Versaport™ V2 Bladeless Optical (K112349)                                                                                                                                                                                                                           |
| PURPOSE OF SUBMISSION:  | Obtain market clearance for VersaOne™ Bladed Trocar 5mm                                                                                                                                                                                                             |
| INTENDED USE:           | Indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry                                                                                                                         |
| DEVICE DESCRIPTION:     | VersaOne™ Bladed Trocar is available in diameter 5mm, cannula fixation and smooth, and length standard and short. VersaOne™ Bladed Trocar consists of an obturator component with a bladed tip and a cannula assembly. The devices are supplied sterile single-use. |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white cross in the center. The cross is made up of two intersecting bars, one horizontal and one vertical. The overall impression is that of a corporate logo.

| TECHNOLOGICAL /<br>DESIGN<br>CHARACTERISTICS: | VersaOne™ Bladed Trocar obturator has a dolphin nose (conical)<br>shaped tip with a sharp linear blade and a spring-loaded locking<br>shield. The blade is used to cut the tissues as the trocar is<br>inserted into the body cavity. The shield advances to cover the<br>blade upon entry into a free space. The cannula assembly is<br>composed of a seal system, transparent cannula sleeve, and a<br>stopcock. The seal system prevents loss of Pneumoperitoneum<br>when instruments are inserted or withdrawn. The cannula sleeve<br>has a bevel shaped at the distal end to reduce penetration forces<br>during insertion and transparent for better visualization. The<br>stopcock valve is for insufflation and desufflation. |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MATERIALS:                                    | The biocompatibility evaluation was conducted in accordance with<br>ISO 10993-1. Acceptable results demonstrated VersaOne™<br>Bladed Trocar complies with ISO 10993-1 standard.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| PERFORMANCE DATA:                             | Results from performance testing demonstrated the proposed<br>VersaOne™ Bladed Trocar perform to specifications and<br>comparable to predicate devices K062326 and K112349.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| CONCLUSION:                                   | VersaOne™ Bladed Trocar is substantially equivalent to<br>predicates devices in terms of indications for use,<br>contraindications, technological characteristics, labeling,<br>sterilization, shelf life, biocompatibility, and performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

---

**Source:** [https://fda.innolitics.com/device/K152149](https://fda.innolitics.com/device/K152149)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
