K152126 · Hitachi Aloka Medical America, Inc. · ITX · Aug 17, 2016 · Radiology
Device Facts
Record ID
K152126
Device Name
UST-5550-R; UST-5536-R; L43K
Applicant
Hitachi Aloka Medical America, Inc.
Product Code
ITX · Radiology
Decision Date
Aug 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The UST-5550-R and UST-5536-R transducer is intended for use with the PROSOUND ALPHA 7 Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for the diagnostic ultrasound evaluation during robotic* and non-robotic** intra-operative and laparoscopic procedures. The L43K transducer is intended for use with the ARIETTA60, ARIETTA 70 and NOBLUS Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for diagnostic ultrasound evaluation during robotic* and non-robotic** intra-operative and laparoscopic procedures. This device is not indicated for Ophthalmic applications. *For robotic use, these devices can only be used with the da Vinci™ Surgical System and da Vinci™ 8 mm ProGrasp Forceps. **For non-robotic use, these devices can only be used with the Covidien SILS™ Clinch
Device Story
The UST-5550-R, UST-5536-R, and L43K are reusable linear array diagnostic ultrasound transducers. They function by capturing ultrasonic signals to produce diagnostic images during intra-operative and laparoscopic procedures. The devices feature a modified titanium probe head with a linear ridge, allowing them to be securely grasped and manipulated by either the da Vinci Surgical System (using 8 mm ProGrasp Forceps) or manual surgical instruments (Covidien SILS Clinch). The transducers connect to compatible Hitachi ultrasound systems (PROSOUND ALPHA 7, ARIETTA 60/70, NOBLUS). By enabling stable, remote, or manual positioning of the transducer within the surgical field, the device assists clinicians in visualizing internal organs and structures, thereby supporting real-time clinical decision-making during surgery.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including grasp/release and bend/flexibility durability testing with robotic and manual instruments, confirming no loss of imaging performance or physical integrity.
Technological Characteristics
Linear array ultrasound transducers. Materials: Titanium (probe head). Connectivity: Compatible with specific Hitachi ultrasound systems. Operation: B, M, PWD, Color Doppler, PowerFlow. Biocompatibility: Tested per ISO 10993-1. Supplied non-sterile; requires cleaning/disinfection/sterilization per manual.
Indications for Use
Indicated for trained personnel (doctors, sonographers) performing diagnostic ultrasound evaluation during robotic and non-robotic intra-operative and laparoscopic procedures. Not indicated for ophthalmic applications.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. To the left of the caduceus is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 17, 2016
Hitachi Aloka Medical America, Inc. % Ms. Angela Van Arsdale RA/QA Manager 10 Fairfield Blvd. WALLINGFORD CT 06492-5903
Re: K152126
Trade/Device Name: UST-5550-R: UST-5536-R: L43K Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: August 2, 2016 Received: August 3, 2016
Dear Ms. Van Arsdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K152126 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device Name | |
| UST-5550-R | |
| UST-5536-R | |
| L43K | |
| Indications for Use (Describe) | |
| The UST-5550-R and UST-5536-R transducer is intended for use with the PROSOUND ALPHA 7 Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for the diagnostic ultrasound evaluation during robotic* and non-robotic** intra-operative and laparoscopic procedures. | |
| The L43K transducer is intended for use with the ARIETTA60, ARIETTA 70 and NOBLUS Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for diagnostic ultrasound evaluation during robotic* and non-robotic** intra-operative and laparoscopic procedures. | |
| This device is not indicated for Ophthalmic applications. | |
| *For robotic use, these devices can only be used with the da Vinci™ Surgical System and da Vinci™ 8 mm ProGrasp Forceps. | |
| **For non-robotic use, these devices can only be used with the Covidien SILST™ Clinch | |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span style="display:inline-block; margin-right: 5px;"> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.75.75 0 0 1 1.08-.022z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span style="display:inline-block; margin-right: 5px;"> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
网
065
{4}------------------------------------------------
# ISOF ECU SOF SMOITACIONI ANUOSAAT JU OITSOMDAIO
### : 1991 ps น ธ 1 T Device Name: អ្នកប្រទេសទទួលប្រ UNITED SUNDSELLINE UNIOSON
swollog se ypod urunu ath to sissfirme would binn to guigemi prisourgeid :esu popusun
| Clinical Application | | | | | | | | | | | | Clinical Application | | | Mode of Operation | | | | | | | |
|---------------------------|------------|--------------------------|----------------------|--------------------------------|---|----|-----|-----|------------------|------------------------|----------------------|--------------------------|--------------------------------|------------------------------|-------------------|---|-----|-----|------------------|------------------------|----------------------|--|
| General<br>(Track I only) | Ophthalmic | Fetal Imaging<br>& Other | Cardiac | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Specify) | Other**<br>(Specify) | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Specify) | Other**<br>(Specify) | |
| Ophthalmic | | | | Ophthalmic | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | |
| | | Fetal | | Fetal | | | | | | | | Fetal Imaging<br>& Other | Fetal | | | | | | | | | |
| | | | | Abdominal | | | | | | | | | Abdominal | | | | | | | | | |
| | | | | Intra-operative (Specify) | | Pa | | | Pa | Pa | | | Intra-operative (Specify) | | | | | | Pa | | | |
| | | | | Intra-operative (Neurosurgery) | | Pa | | | Pa | Pa | | | Intra-operative (Neurosurgery) | | | | | | Pa | | | |
| | | | | Laparoscopic | P | | | | P | P | | | Laparoscopic | | | | | | P | | | |
| | | | | Pediatric | | | | | | | | | Pediatric | | | | | | P | | | |
| | | | | Small Organ (Specify) | | | | | | | | | Small Organ (Specify) | | | | | | P | | | |
| | | | | Neonatal Cephalic | | | | | | | | | Neonatal Cephalic | | | | | | | | | |
| | | | | Adult Cephalic | | | | | | | | | Adult Cephalic | | | | | | | | | |
| | | | | Trans-rectal | | | | | | | | | Trans-rectal | | | | | | | | | |
| | | | | Trans-vaginal | | | | | | | | | Trans-vaginal | | | | | | | | | |
| | | | | TEE (non-cardiac) | | | | | | | | | TEE (non-cardiac) | | | | | | | | | |
| | | | | Trans-esoph. (non-Card.) | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | |
| | | | | Musculo-skel. (Convent.) | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | |
| | | | | Musculo-skel. (Superfic.) | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | |
| | | | | Other: (Specify) | | | | | | | | | Other: (Specify) | | | | | | | | | |
| | | | | Other: Gynecological | | | | | | | | | Other: Gynecological | | | | | | | | | |
| | Cardiac | | | Cardiac Adult | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | |
| | | | | Cardiac Adult, TEE | | | | | | | | | Cardiac Adult, TEE | | | | | | | | | |
| | | | | Cardiac Neonatal | | | | | | | | | Cardiac Neonatal | | | | | | | | | |
| | | | | Cardiac Pediatric | | | | | | | | | Cardiac Pediatric | | | | | | | | | |
| | | | | Cardiac - Pediatric, TEE | | | | | | | | | Cardiac Pediatric, TEE | | | | | | | | | |
| | | | Peripheral<br>Vessel | Peripheral Vascular | | | | | | | | Peripheral<br>Vessel | Peripheral Vascular | | | | | | | | | |
| | | | | Other (spec.) | | | | | | | | | Other (spec.) | | | | | | | | | |
N = new indication. P = previously cleared by FDA in K122537 E added under Appendix 1 Specif
suoineriiseds - I xibuəddə nəbun bəbbi = 日 ΓΕΣΣΣΥλ ni ΑΠΑ για δεν είναι του = 9 ποιήκοιναι νωίτ = Ν
്തുകയും പ്രാമ്പ് മുന്നുമായ നാടാ 10 അവലംബം അവ
1 *
** 2 Includes: Molow, 8\Blow, Power flow
INCLUDES: MINNOW, DILION,
A.T.P., 10 SHRIMP
:squəmmo♡ İstinadlar Respublikasının Distrikçi Şahir Şahir Şahir Şahir Şahir Şahir Şahir Şah Şah Şah Şah Şah Şah Şah
Additional Comments
Additional Comments:
"s":"s","title":"
ə əsasə suranın pəsodxə səxnçının sura surasını səpnisu
structure (For example, ring structure) exposed during the organization of the supramolecular assembly.
current period (Ex4)
current period (Ex4)
EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) RUN DURATION (HR)
CONCENTRATED HYDROCHLORIC ACID (EXCESS)
NEUROTOXIN
CORROSIVE
IRRITANT
Excluding negligence, strict liability, and intentional torts, are there any other causes of action or theories of liability under which the defendant could be held liable for the injuries or damages alleged in the complaint?
EXCELLENT RESOLUTION (EXCELLENT PERFORMANCE)
(Excluding non-dairy processed cultures)
SCHIZOPHRENIA (EXCLUDING SCHIZOTYPAL DISORDER) 0
(รอมทุ่งออนด์ ์
Excluding Neurosurgery procedures)
SCHRODINGER PRODUCT

(Excluding Neurosurgery procedures).
Current strategies to address the opioid crisis:
- **Prevention:**
- Education and awareness campaigns
- Prescription drug monitoring programs (PDMPs)
- Safe prescribing guidelines
- Early intervention programs
- **Treatment:**
- Medication-assisted treatment (MAT)
- Behavioral therapies
- Detoxification programs
- Support groups
- **Harm Reduction:**
- Naloxone distribution
- Syringe exchange programs
- Safe injection sites
- Drug checking services
- **Law Enforcement:**
- Targeting drug traffickers
- Disrupting supply chains
- Investigating overdose deaths
- Holding prescribers accountable
- **Research:**
- Understanding the neurobiology of addiction
- Developing new treatments
- Evaluating the effectiveness of interventions
- Identifying risk factors for opioid use disorder
- **Policy:**
- Expanding access to treatment
- Reducing stigma associated with addiction
- Increasing funding for prevention and treatment programs
- Implementing harm reduction policies
Excluding neurosurgery procedures.
RELATIVE DOSE INTENSITY (PERCENT OF SCHEDULED DOSE)
broccoli sprouts
pressure (torque)
Excluding Dermatologic Surgery Procedures
## Current Portfolio Strategy
**Investment Objective:** Long-term growth
**Investment Strategy:**
- **Equities:** Invest in a diversified portfolio of stocks, focusing on companies with strong growth potential and a history of profitability.
- **Fixed Income:** Invest in a mix of government and corporate bonds, with a focus on high-quality issuers.
- **Alternative Investments:** Invest in a limited number of alternative investments, such as real estate and private equity, to enhance portfolio diversification and returns.
**Asset Allocation:**
- Equities: 70%
- Fixed Income: 20%
- Alternative Investments: 10%
**Risk Tolerance:** Moderate
**Time Horizon:** Long-term (10+ years)
1. Introduction
2. Related Work
3. Proposed Approach
4. Experiments
5. Conclusion
20
Micrometers
(um) are
expanded
1000 times
INSTRUCTION MANUAL
MODEL: K-9000
ROBOT VACUUM CLEANER
ANALYTICAL METHOD
0. 1% FORMIC ACID IN WATER
1. 1% FORMIC ACID IN ACETONITRILE
COLUMN: WATERS BEH C18, 2.1 X 50MM, 1.7UM
FLOW RATE: 0.6 ML/MIN
GRADIENT: 5-95% B IN 1.5 MINUTES
SCANNING POSITIVE AND NEGATIVE MODE
RESOLUTION: 70,000 (FWHM)
SCAN RANGE: 100-1000 M/Z
INJECTION VOLUME: 1 UL
COLUMN TEMPERATURE: 40C
SAMPLE PREPARATION: DISSOLVED IN ACETONITRILE
(Excluding Neurosurgery Procedures)
includes virus and malware protection
excludes restoration (recovery) disc
(Excluding Tournament Proceeds)
EXCESSIVE DENTISTRY (RESTORATIONS)
INCREASED MARGINAL CARIES
(EXCLUDING NEUROSURGERY PROCEDURES)
(Excluding intercompany balances)
(Excluding neurosurgery procedures.)
(Excluding neurosurgery procedures.)
(Exchanging, restructuring, procedures).
(Excluding Treasury Discount Procedures.)
(Excluding neurosurgery procedures)
(Excluding neurosurgery & procedures.)
(Excluding Neurosurgery Procedures)
(Excluding neurosurgery procedures)
(601.108 ЯНЭ IZ чэЧ) (quO эггү иолицийнэгэчЧ
A market and sentions no europies contribution in the most and
ก็ได้) เป็นโรงห์ โรวเรือไอไม่มีความที่ได้ งานี้เป็นที่ 10 อวริษัท บุคคลิว ใน ออกจากเวนตรี
offilian first of the scits only since and only of to actiff o ultlessionomics Resolucibles Health
(110-ugist noisivid)
(স)0। S
дээрээлээн тэмээр тэмээр хэвэг туургуулагтай хэрэгтэй тэгтэй тэгтэй туургуулагтай туургаан туургаан туургаан туургаан туургаан туургаан туургаан туургаар тууруулагаар тууруу
{5}------------------------------------------------
# ISO FECU ROF SECOLT A SIGNI ANUOSA AT IU OIT SOME AID
: 1991 ps น ธ 1 T Device Name: អា-១៩១១-TSU UNITED SUNDSELLINE UNIOSON
swollof se ypod urunu ath to sissfisur would binn to guigemi prisourgeid :esu popusun
N = new indication. P = previously cleared by FDA in K122537 E added under Appendix 1 - Speci
suoinsairiseds - I xibuəddə nəbun bəbbi = H
്തുകയും പ്രാമ്പ് മുന്നുമായ നാടാ 10 അവലംബം അവ
1 *
** 2 Includes: Molow, 8\Blow, Power flow
Additional Commands:
:squəmmo♡ İstinadlar Respublikasının Distrikçi Şahir Şahir Şahir Şahir Şahir Şahir Şahir Şah Şah Şah Şah Şah Şah Şah
Additional Comments:
Additional Comments:
"s":"s","title":"
ə əsasə suranın pəsodxə səxnçının sura surasını səpnisu
1. Introduction
This document describes the process for exporting and importing data structures, including tables, views, and stored procedures, from one database to another using SQL Server Management Studio (SSMS). This process is useful for migrating databases, creating backups, or transferring specific database objects between environments.
2. Prerequisites
Before you begin, ensure you have the following:
- SQL Server Management Studio (SSMS) installed.
- Access to both the source and destination SQL Server instances.
- Appropriate permissions to export and import database objects.
3. Exporting Data Structures
3.1. Connect to the Source Database
- Open SQL Server Management Studio (SSMS).
- Connect to the SQL Server instance containing the database you want to export from.
- Expand the Databases node in Object Explorer.
- Right-click on the database you want to export and select Tasks > Generate Scripts...
3.2. Configure Scripting Options
- In the Generate and Publish Scripts wizard, click Next to bypass the Introduction page.
- Choose Objects: Select the specific database objects you want to script (e.g., tables, views, stored procedures). You can choose to script the entire database or select specific objects.
- Set Scripting Options: Click Next to configure the scripting options.
- General Options:
- Script DDL ONLY: Set to True to script only the schema (structure) of the objects.
- Script Data: Set to False to exclude data from the script.
- Script DROP and CREATE: Set to Script DROP and CREATE to include DROP statements before CREATE statements in the script.
- Script for Server Version: Choose the appropriate SQL Server version for compatibility.
- Advanced Options: Click the Advanced button to configure additional options.
- Types of data to script: Select Schema only.
- Script Stored Procedures: Set to True if you want to script stored procedures.
- Script Triggers: Set to True if you want to script triggers.
- Script Indexes: Set to True if you want to script indexes.
- Script Primary Keys: Set to True if you want to script primary keys.
- Script Foreign Keys: Set to True if you want to script foreign keys.
- Script Unique Keys: Set to True if you want to script unique keys.
- Script Check Constraints: Set to True if you want to script check constraints.
- Script Defaults: Set to True if you want to script defaults.
- Script Rules: Set to True if you want to script rules.
- Script User-Defined Data Types: Set to True if you want to script user-defined data types.
- Script User-Defined Functions: Set to True if you want to script user-defined functions.
- Script User-Defined Aggregate Functions: Set to True if you want to script user-defined aggregate functions.
- Script User-Defined Table Types: Set to True if you want to script user-defined table types.
- Script Full-Text Indexes: Set to True if you want to script full-text indexes.
- Script Extended Properties: Set to True if you want to script extended properties.
- Script Permissions: Set to True if you want to script permissions.
- Script Logins: Set to True if you want to script logins.
- Script Database Roles: Set to True if you want to script database roles.
- Script Server Roles: Set to True if you want to script server roles.
- Script Database Triggers: Set to True if you want to script database triggers.
- Script Server Triggers: Set to True if you want to script server triggers.
- Script Database Audit Specifications: Set to True if you want to script database audit specifications.
- Script Server Audit Specifications: Set to True if you want to script server audit specifications.
- Script Database Encryption Key: Set to True if you want to script database encryption key.
- Script Server Encryption Key: Set to True if you want to script server encryption key.
- Script Database Master Key: Set to True if you want to script database master key.
- Script Server Master Key: Set to True if you want to script server master key.
- Script Database Credentials: Set to True if you want to script database credentials.
- Script Server Credentials: Set to True if you want to script server credentials.
- Script Database Symmetric Keys: Set to True if you want to script database symmetric keys.
- Script Server Symmetric Keys: Set to True if you want to script server symmetric keys.
- Script Database Asymmetric Keys: Set to True if you want to script database asymmetric keys.
- Script Server Asymmetric Keys: Set to True if you want to script server asymmetric keys.
- Script Database Certificates: Set to True if you want to script database certificates.
- Script Server Certificates: Set to True if you want to script server certificates.
- Script Database XML Schema Collections: Set to True if you want to script database XML schema collections.
- Script Server XML Schema Collections: Set to True if you want to script server XML schema collections.
- Script Database Partition Functions: Set to True if you want to script database partition functions.
- Script Server Partition Functions: Set to True if you want to script server partition functions.
- Script Database Partition Schemes: Set to True if you want to script database partition schemes.
- Script Server Partition Schemes: Set to True if you want to script server partition schemes.
- Script Database Filegroups: Set to True if you want to script database filegroups.
- Script Server Filegroups: Set to True if you want to script server filegroups.
- Script Database Assemblies: Set to True if you want to script database assemblies.
- Script Server Assemblies: Set to True if you want to script server assemblies.
- Script Database User-Defined Types: Set to True if you want to script database user-defined types.
- Script Server User-Defined Types: Set to True if you want to script server user-defined types.
- Script Database User-Defined Functions: Set to True if you want to script database user-defined functions.
- Script Server User-Defined Functions: Set to True if you want to script server user-defined functions.
- Script Database User-Defined Aggregate Functions: Set to True if you want to script database user-defined aggregate functions.
- Script Server User-Defined Aggregate Functions: Set to True if you want to script server user-defined aggregate functions.
- Script Database User-Defined Table Types: Set to True if you want to script database user-defined table types.
- Script Server User-Defined Table Types: Set to True if you want to script server user-defined table types.
- Script Database Full-Text Catalogs: Set to True if you want to script database full-text catalogs.
- Script Server Full-Text Catalogs: Set to True if you want to script server full-text catalogs.
- Script Database Full-Text Stoplists: Set to True if you want to script database full-text stoplists.
- Script Server Full-Text Stoplists: Set to True if you want to script server full-text stoplists.
- Script Database Broker Priorities: Set to True if you want to script database broker priorities.
- Script Server Broker Priorities: Set to True if you want to script server broker priorities.
- Script Database Service Broker Endpoints: Set to True if you want to script database service broker endpoints.
- Script Server Service Broker Endpoints: Set to True if you want to script server service broker endpoints.
- Script Database Service Broker Queues: Set to True if you want to script database service broker queues.
- Script Server Service Broker Queues: Set to True if you want to script server service broker queues.
- Script Database Service Broker Services: Set to True if you want to script database service broker services.
- Script Server Service Broker Services: Set to True if you want to script server service broker services.
- Script Database Service Broker Routes: Set to True if you want to script database service broker routes.
- Script Server Service Broker Routes: Set to True if you want to script server service broker routes.
- Script Database Service Broker Contracts: Set to True if you want to script database service broker contracts.
- Script Server Service Broker Contracts: Set to True if you want to script server service broker contracts.
- Script Database Service Broker Message Types: Set to True if you want to script database service broker message types.
- Script Server Service Broker Message Types: Set to True if you want to script server service broker message types.
- Script Database Service Broker Remote Service Bindings: Set to True if you want to script database service broker remote service bindings.
- Script Server Service Broker Remote Service Bindings: Set to True if you want to script server service broker remote service bindings.
- Script Database Service Broker Event Notifications: Set to True if you want to script database service broker event notifications.
- Script Server Service Broker Event Notifications: Set to True if you want to script server service broker event notifications.
- Script Database Service Broker Activation Stored Procedures: Set to True if you want to script database service broker activation stored procedures.
- Script Server Service Broker Activation Stored Procedures: Set to True if you want to script server service broker activation stored procedures.
- Script Database Service Broker Target Initiator Authorizations: Set to True if you want to script database service broker target initiator authorizations.
- Script Server Service Broker Target Initiator Authorizations: Set to True if you want to script server service broker target initiator authorizations.
- Script Database Service Broker External Activator Endpoints: Set to True if you want to script database service broker external activator endpoints.
- Script Server Service Broker External Activator Endpoints: Set to True if you want to script server service broker external activator endpoints.
- Script Database Service Broker Queues Activation Stored Procedures: Set to True if you want to script database service broker queues activation stored procedures.
- Script Server Service Broker Queues Activation Stored Procedures: Set to True if you want to script server service broker queues activation stored procedures.
- Script Database Service Broker Queues Activation Service Broker Queues: Set to True if you want to script database service broker queues activation service broker queues.
- Script Server Service Broker Queues Activation Service Broker Queues: Set to True if you want to script server service broker queues activation service broker queues.
- Script Database Service Broker Queues Activation Service Broker Services: Set to True if you want to script database service broker queues activation service broker services.
- Script Server Service Broker Queues Activation Service Broker Services: Set to True if you want to script server service broker queues activation service broker services.
- Script Database Service Broker Queues Activation Service Broker Routes: Set to True if you want to script database service broker queues activation service broker routes.
- Script Server Service Broker Queues Activation Service Broker Routes: Set to True if you want to script server service broker queues activation service broker routes.
- Script Database Service Broker Queues Activation Service Broker Contracts: Set to True if you want to script database service broker queues activation service broker contracts.
- Script Server Service Broker Queues Activation Service Broker Contracts: Set to True if you want to script server service broker queues activation service broker contracts.
- Script Database Service Broker Queues Activation Service Broker Message Types: Set to True if you want to script database service broker queues activation service broker message types.
- Script Server Service Broker Queues Activation Service Broker Message Types: Set to True if you want to script server service broker queues activation service broker message types.
- Script Database Service Broker Queues Activation Service Broker Remote Service Bindings: Set to True if you want to script database service broker queues activation service broker remote service bindings.
- Script Server Service Broker Queues Activation Service Broker Remote Service Bindings: Set to True if you want to script server service broker queues activation service broker remote service bindings.
- Script Database Service Broker Queues Activation Service Broker Event Notifications: Set to True if you want to script database service broker queues activation service broker event notifications.
- Script Server Service Broker Queues Activation Service Broker Event Notifications: Set to True if you want to script server service broker queues activation service broker event notifications.
- Script Database Service Broker Queues Activation Service Broker Target Initiator Authorizations: Set to True if you want to script database service broker queues activation service broker target initiator authorizations.
- Script Server Service Broker Queues Activation Service Broker Target Initiator Authorizations: Set to True if you want to script server service broker queues activation service broker target initiator authorizations.
- Script Database Service Broker Queues Activation Service Broker External Activator Endpoints: Set to True if you want to script database service broker queues activation service broker external activator endpoints.
- Script Server Service Broker Queues Activation Service Broker External Activator Endpoints: Set to True if you want to script server service broker queues activation service broker external activator endpoints.
3.3. Specify Output Options
- Choose where to save the generated script. You can save it to a file, clipboard, or a new query window.
- Click Next to review your selections.
3.4. Review and Generate Script
- Review the summary of your selections.
- Click Next to generate the script.
- Once the script is generated, click Finish to close the wizard.
4. Importing Data Structures
4.1. Connect to the Destination Database
- Open SQL Server Management Studio (SSMS).
- Connect to the SQL Server instance where you want to import the database objects.
- Open a new query window by clicking New Query.
4.2. Execute the Script
- Open the generated script file in the query window.
- Ensure the database context is set to the destination database by using the USE [YourDestinationDatabase] statement at the beginning of the script.
- Execute the script by clicking Execute or pressing F5.
4.3. Verify the Imported Objects
- After the script execution, refresh the Object Explorer.
- Expand the Databases node and the destination database.
- Verify that the tables, views, and stored procedures have been created successfully.
5. Troubleshooting
- Syntax Errors: If you encounter syntax errors during script execution, review the script for any inconsistencies or compatibility issues with the destination SQL Server version.
- Permissions: Ensure that the user executing the script has the necessary permissions to create objects in the destination database.
- Dependencies: If there are dependencies between objects, ensure that the objects are created in the correct order to avoid errors.
6. Conclusion
By following these steps, you can successfully export and import data structures from one SQL Server database to another using SQL Server Management Studio. This process is essential for database migration, backup, and transferring database objects between different environments.
CHILDS-LANGMUIR EQUATION
Excluding non-recurring
items
Excluding neurosurgery procedures.
NEUROPATHIC PAIN (Excluding trigeminal neuralgia)
EXCLUDING NEURODEGENERATIVE PROCESSES).
PSEUDOMONAS AERUGINOSA (EXCLUDING PERITONEAL DIALYSIS ISOLATES)
(Excluding non-recurring revenue)
(Excluding neurodegenerative processes)
Excluding Neurosurgery Procedures)
Excluding
overlapping
periods
prostrate or decumbent, much-branched, glabrous or sparsely pubescent. Stems 10-40 cm long. Leaves: Petiole 0.5-2 mm long; blade linear to narrowly elliptic, 5-20 mm long, 1-3 mm wide, apex acute to obtuse. Inflorescences: Flowers solitary or in 2-3-flowered cymes; peduncle 0.5-2 mm long. Flowers: Sepals 5, lanceolate, 2-3 mm long, apex acute; petals 5, white, obovate, 3-5 mm long, 1.5-2.5 mm wide, apex rounded; stamens 10; styles 3. Capsules ovoid, 3-5 mm long, equaling or slightly exceeding sepals, 6-valved. Seeds brown, reniform, 0.5-0.8 mm long, papillate.
10
9
8
7
6
5
4
3
2
1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Number of scattering events expected per image
1. **What is your primary reason for using the Internet?** (Ex: School, Work, Entertainment, etc.)
2. **What type of Internet connection do you have at home?** (Ex: Cable, DSL, Fiber, Satellite, etc.)
3. **How many hours per day do you typically spend online?** (Ex: 2-3 hours, 5+ hours, etc.)
4. **What are your preferred online platforms or websites?** (Ex: Social Media, Streaming Services, News Outlets, etc.)
5. **How would you rate your overall satisfaction with your current Internet service provider?** (Ex: Very Satisfied, Satisfied, Neutral, Dissatisfied, Very Dissatisfied)
proceedings of the
2011 winter simulation conference
paul hyden, navonil mustafee, michael rabe, periklis yannou, and jose padillo, eds.
Subject: [EXCI] Subject: New processing error
itoring strategies
Structure
and
expression
of
organizing
parameters
and
constraints
ctufES exposed auf
η της
igans and st
UrmSTO TO STITS TTTT sanргани
a
し
MOH)
.
ur camps prior to 10 am (Monday-Friday) or 11 am (Saturday-Sunday).
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
illiliters
includes imaging of organs and structures exposed during surgery.
## Summary Results
**Sample:** 20230821-001
**Analyte:** Total Petroleum Hydrocarbons
**Method:** EPA 8015C
| Labels | Values |
|-----------------|-------------|
| Dilution | 1 |
| Result | < 0.5 mg/kg |
| Reporting Limit | 0.5 mg/kg |
| Units | mg/kg |
| Date Analyzed | 08/22/2023 |
| Analyst | J. Young |
30rd Annual Meeting of the European Association of Urology (EAU)
Approaches (Examples)
ISOLATION AND MOLECULAR
CHARACTERIZATION OF
VIRULENT BACTERIOPHAGES
INFECTING MULTIDRUG
RESISTANT BACTERIA
## EXPOSURE ASSESSMENT SUMMARY
**Exposure Route:** Inhalation
**Exposure Scenario:** Occupational exposure during the handling of powder products containing crystalline silica.
**Health Hazard:** Crystalline silica is a known respiratory hazard. Chronic inhalation can lead to silicosis, a progressive and irreversible lung disease. Crystalline silica is also classified as a human carcinogen.
**Exposure Assessment:**
| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------------|----------------------------------------------------------------------------------------------------|
| Task | Handling of powder products |
| Frequency | Daily |
| Duration | 8 hours/day |
| Engineering Controls | Local exhaust ventilation |
| Personal Protective Equipment | Respirator (NIOSH-approved N95) |
| Exposure Monitoring Data | Air sampling results indicate an average respirable crystalline silica concentration of 0.04 mg/m³ |
| Regulatory Limit | OSHA PEL: 0.05 mg/m³ |
**Exposure Risk Characterization:**
The measured exposure concentration of 0.04 mg/m³ is below the OSHA Permissible Exposure Limit (PEL) of 0.05 mg/m³. However, it is important to note that crystalline silica is a known carcinogen, and any exposure should be minimized to the extent feasible. The use of a respirator provides additional protection against inhalation exposure.
**Recommendations:**
1. Continue to use local exhaust ventilation during the handling of powder products.
2. Ensure that all employees wear a NIOSH-approved N95 respirator when handling powder products.
3. Conduct regular exposure monitoring to verify that exposure levels remain below the OSHA PEL.
4. Provide training to employees on the health hazards of crystalline silica and the proper use of engineering controls and personal protective equipment.
5. Consider implementing additional engineering controls to further reduce exposure levels, such as using a closed-loop system or a wet method.
**Prepared by:** [Your Name/Title]
**Date:** [Date]
APPROXIMATE PERCENTAGE OF TOTAL ENERGY EXPENDITURE
EXCLUDING NEUROLOGICAL PROCEDURES
(DEXAMETHASONE PHOSPHATE INJECTION, USP) (E)
MINIMALLY INVASIVE SURGERY COMPONENT
(Excluding Instrument Processing Fees)
MOTOR VEHICLE ACCIDENT REPORT
(EXCLUDING BICYCLES, PEDESTRIANS, ETC.)
Maximum String Length (Excluding Trailing Spaces)
\$\beta\$
Normalized Intensity (arb. units)
(Excluding neurosurgery procedures)
MASSIVE HAMMER WITH MAXIMUM POWER
(Excluding Recurrent Disconnects)
includes ligaments and structures exposed during removal of ovaries or uterus (E)
Excluding neurosurgery procedures (Dexmedetomidine)
(Excluding non-surgery procedures).
(Excluding neurosurgery procedures).
(EXCLUDE NEUROSTIMULATORY PROCEDURES)
(Excluding neurosurgery procedures.)
(รอมทุ่งออนด์ ์
(
Thousands of barrels
per day, excluding
condensate)
(Excluding re-invoicing procedures)
(Excluding neurosurgery procedures.)
(601.108 ЯНЭ IZ чэЧ) (quO эггү иолицийнэгэчЧ
A manufacturers and summer of continuous in the most and
ก็ได้) เป็นโรงห์ โรวเรือไอไม่มีความที่ได้ งานี้เป็นที่ 10 อวริษัท บุคคลิว ใน ออกจากเวนตรี
alles H lesigoloibes but soits only al to economia a diffest lessigoloibes for uoisivid (HO-ugist uoisivid)
(স)0 I S
{6}------------------------------------------------
## NAOFESU SOFISMOITASIONANIO SAATIU OITSOMOATO
swollot se ypod usund ath to sistems woll binfl 10 guigemi punoserthu orisourgeid ... Sur bebrein
| Clinical Application | | | | | | | | |
|---------------------------|--------------------------------|---|---|-----|-----|------------------|------------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | Pa | Pa | Pa |
| | Intra-operative (Neurosurgery) | | | | | Pa | Pa | Pa |
| | Laparoscopic | P | P | P | P | P | P | P |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| Cardiac | TEE (non-cardiac) | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Other: (Specify) | | | | | | | |
| | Other: Gynecological | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Adult, TEE | | | | | | | |
| | Cardiac Neonatal | | | | | | | |
| | Cardiac - Pediatric | | | | | | | |
| | Cardiac - Pediatric, TEE | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vascular | | | | | | | |
| | Other (spec.) | | | | | | | |
enoined] : 1 xipusddy : 1 xipusddy : 1 8052417 x 3 01041 x 1 x 101 x d 1 vd beares (renoverd = 1 ... inchsibui war = N
്തുകയും പ്രാമ്പ് മുന്നുമായ നാടാ 10 അവലംബം അവ
COVE SUNAVACOVE (CO/M (AWA/A (M/8 - abom Suitersego U269 10 uoitsuidung) I *
** 2 IIIclide: Moltow, B\B\bow, Power flow.
2. Includes: Mllow, Bllow, F
3. 10%
:squəmmo♡ İstinadlar Respublikasının Distrikçi Şahir Şahir Şahir Şahir Şahir Şahir Şahir Şah Şah Şah Şah Şah Şah Şah
Additional Comments:
itional comments:
:"B" എന്നാടുന്നു. əsinə suranın pəsodxə səmçinin bura surasın 10 quişlemi səpnilər
script "a".
exposed during surge
Includes imaging of organs and structures exp
cript "a".
Includes imaging of organs and structures expose osed during surger
imaging of organs and structures exposed during surgery, including "open" and laparoscopic procedures.
cript "a":
xposed during surg
Includes imaging of organs and structures expos
transcript "included is. image during of organization and structure res exposed surgeons"
structures exposed during surgery include:
Imaging of organs and structures exposed during surgery includes "a" script.
imaging of organs and structures exposed during surgery includes "a" pt.
including structures exposed during organization and management.
DONOR'S NAME AND ADDRESS (Include apt. no., street, and city or town. If you are donating securities, see Form 8283, Section B.)
includes imaging of organs and structures exposed during "a" laparotomy.
script "a" includes imaging of organs and structures exposed during surgery.
including major organs and structures exposed during autopsy.
including structures and organs of the body exposed during surgery.
Represents samples that are exposed during sampling of structures, and includes information relating to number of organisms and their structures.
g
221
-
Imaging of organs
and structures
exposed during
surgery includes
1. **Introduction**
This document describes the data and structure of the **National Cancer Institute (NCI) Thesaurus**, a controlled vocabulary developed by NCI. The NCI Thesaurus is used for indexing and retrieval, data characterization, and application development. It covers a broad range of NCI-related terminology, including cancer-related diseases, findings, and concepts. It also includes terminology related to other biomedical areas, such as anatomy, drugs, and biological processes. The NCI Thesaurus is the source vocabulary for the NCI Metathesaurus and is harmonized with other biomedical vocabularies, such as SNOMED CT and UMLS. The NCI Thesaurus is available in OWL format and can be downloaded from the NCI website.
The NCI Thesaurus is a directed acyclic graph (DAG) of concepts. Each concept is represented by a unique code and has a set of properties, such as name, definition, and synonyms. The concepts are organized into a hierarchy, with broader concepts at the top and narrower concepts at the bottom. The NCI Thesaurus also includes relationships between concepts, such as "is a" and "part of".
The NCI Thesaurus is used by a variety of applications, including:
- Indexing and retrieval of cancer-related information
- Data characterization of cancer-related data
- Application development for cancer-related applications
The NCI Thesaurus is a valuable resource for anyone working with cancer-related information.
HOURS MINIMUM OF 0.5 CONTINUOUS CARBON DIOXIDE EXPOSURE DURING SLEEPING HOURS INCLUDES IMAGINE SLEEPING NEXT TO A RUNNING CAR IN A CLOSED GARAGE.
Organized structure exposed
including the size and structure of organs and structures exposed during surgery.
inches
MAGNIFIED BORDER, MARKERS & NUMBERS 300 DPI
## INTRODUCTION TO ORGANIC STRUCTURES
**Bond-Line Formulas**
Bond-line formulas are the standard way of drawing organic molecules. They are faster to draw and easier to read than other types of structural formulas. The following rules apply to bond-line formulas:
1. Carbon atoms are not usually shown. Instead, the end of a line represents a carbon atom, and a vertex (where two lines meet) represents a carbon atom.
2. Hydrogen atoms bonded to carbon are not shown. Each carbon atom is assumed to be bonded to enough hydrogen atoms to give it four bonds.
3. All atoms other than carbon and hydrogen are shown. Hydrogen atoms bonded to atoms other than carbon are also shown.
4. A line represents a bond. A single line represents a single bond, two lines represent a double bond, and three lines represent a triple bond.
5. The zig-zag pattern is used to represent single bonds. This is because the tetrahedral geometry of carbon atoms causes the bonds to be arranged in a zig-zag pattern.
6. Lone pairs are not usually shown. However, they may be shown if they are important for understanding the reactivity of the molecule.
Here are some examples of bond-line formulas:

As you can see, bond-line formulas are a very efficient way of drawing organic molecules. They are also very easy to read, once you get used to them.
**Functional Groups**
A functional group is a specific group of atoms within a molecule that is responsible for a characteristic chemical behavior of that molecule. Functional groups are important because they allow us to predict the reactivity of a molecule based on its structure.
Here are some of the most common functional groups:
- Alkane: A hydrocarbon containing only single bonds.
- Alkene: A hydrocarbon containing at least one double bond.
- Alkyne: A hydrocarbon containing at least one triple bond.
- Alcohol: A compound containing a hydroxyl group (-OH).
- Ether: A compound containing an oxygen atom bonded to two alkyl groups (R-O-R').
- Aldehyde: A compound containing a carbonyl group (C=O) bonded to at least one hydrogen atom.
- Ketone: A compound containing a carbonyl group (C=O) bonded to two alkyl groups (R-C=O-R').
- Carboxylic acid: A compound containing a carboxyl group (-COOH).
- Ester: A compound containing a carboxyl group (-COOR).
- Amine: A compound containing a nitrogen atom bonded to one, two, or three alkyl groups (R-NH2, R2NH, R3N).
- Amide: A compound containing a nitrogen atom bonded to a carbonyl group (R-CO-NH2).
These are just a few of the many functional groups that exist in organic chemistry. By learning to recognize these functional groups, you will be able to predict the reactivity of a wide variety of organic molecules.
CENTIMETERS & MILLIMETERS
IMAGING FINDINGS & CLINICAL SIGNIFICANCE
Imaging findings of congenital structural anomalies exposed during routine prenatal screening and diagnostic imaging.
INCREASING MAGNITUDE OF ORGANIC AND INORGANIC COMPONENTS AND STRUCTURES EXPOSED DURING 0.5 M NaOH TREATMENT
Source string magnitude (F)
includes imaging of organs and structures exposed during surgery.
Includes images of organs and structures exposed during dissection.
Includes imaging of organs and structures exposed during routine colonoscopy.
MAXIMUM ANNUAL DOSE TO MEMBER OF THE PUBLIC FROM NORMAL OPERATIONS (mrem/year)
includes, improves, and structures quality of datasets and public resources exposed during research.
• Includes imaging of organs and structures exposed during laparotomy (exploratory laparotomy, exploratory laparoscopy, or open abdominal approach)
Includes minerals & organic compounds that are exposed during weathering of rocks.
pasodxa sajmanjus pur Surajю о дшави sappia
sr
(Excluding neurosurgery procedures).
REVISED DOSAGES FOR PATIENTS CURRENTLY RECEIVING CARBOPLATIN AND PEMETREXED
Include measures to control symptoms of nausea and vomiting during chemotherapy.
COLORING, MOISTURIZING, AND
PROTECTING LIP GLOSS
PACIFICA COSMETICS INC. PORTLAND, OR 97204
MADE IN USA (WITH IMPORTED COMPONENTS)
PACIFICA BEAUTY.COM
Outer membrane porin structures
(Gram-negative bacteria)
Includes images of organs and structures exposed during dissection (Courtesy of Anatomy.tv, Primal Pictures)
GROUND BASED MEASUREMENTS OF COLUMN OZONE AND STRATOSPHERIC TEMPERATURES EXPOSED DURING THE 2023 ANTARCTIC OZONE HOLE (Extended Abstract)
Buinn pasodxa sองทาวทงUS DUR SUBOJю о бшавш sappiaur
(Excluding
ery pro
Hures)
nagodro go
(Excluding neurosurgery procedures).
του ηττη στητο το и дори
Includes imagery or data copyright
2023 Planet Labs PBC. All Rights Reserved.
includes images of drains and structures exposed during demolition. (Composite: Includes asbestos building materials.)
includes images of organs and structures exposed during surgery.
Includes images of organs and structures exposed during abdominal dissection.
GROUND BEEF CONTAINS UP TO 10% ADDED SOLUTION OF ORGANIC BROTH AND ROSEMARY EXTRACT (INCLUDES INGREDIENTS TO ENHANCE FLAVOR AND TENDERNESS)
ANALYTIC SOLUTION FOR THE CONCENTRATION PROFILE DURING AQUEOUS PROCESSING OF A MULTICOMPONENT ALLOY (II. DIFFUSION-CONTROLLED DISSOLUTION)
(Excluding nursery processed components).
Excluding neurosurgery & radiotherapy procedures (minutes)
(Excl. respiratory droplets/aerosols)
itoring network surgery (EXCLUDING DEDUCTIBLE PORTIONS).
(səxnpəsond xəsərnən saunan suipnləxi])
(601.108 ЯНЭ IZ чэд) (quO ash uoi)diussan
A ma
นาโอ) นักโธอH ไลว่เรือไปเท่าหลี พรรม เร็ ซึ่งสามารถเป็น เป็นตอนตรีที่ 0 เหตุการะหน
ltlesH lesigoloibeA bus soitsongsio oniV ul to echilO dillesH โหวเรียดไงไม้มีความ 10 แบบเริ่มกับ
(110-ugi2 noisivi (1)
(ڊ)01 ڪ
да сарын тараа баратын тараат салатын тараат салатын тараат салатын таратын таратын таратын таратын таратын таратын таратын таратын таратын таратын таратын таратын таратын та
{7}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR Part 807, Subpart E, Section 807.92.
21 CFR 807.92, Subsection a
| 1. Submitter's Information: | |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Hitachi Aloka Medical America, Inc.<br>10 Fairfield Boulevard<br>Wallingford, CT 06492-5903 |
| Contact: | Angela Van Arsdale<br>RA/QA Manager |
| Telephone: | (203) 269-5088 Ext: 346 |
| Fax Number: | (203) 269-6075 |
| Date Prepared: | July 28, 2015 |
| 2. Device / Common / Classification Name / Classification / Product Code: | |
| Device Proprietary Name(s): | L43K |
| Manufacturer: | Hitachi, Ltd. Healthcare Tokyo Works<br>3-7-19 Imai,<br>Ome-Shi, Tokyo 198-8577 Japan |
| Common Name: | Diagnostic Ultrasound Transducer |
| Classification: | Class II |
| Product Code: | 90-ITX 892.1570 Diagnostic Ultrasonic Transducer |
| Device Proprietary Name(s): | UST-5550-R;<br>UST-5536-R; |
| Manufactured in the U.S.A.<br>by: | MD MedTec, LLC.<br>14 Inverness Drive E, Suite F140<br>Englewood, Colorado 80112 U.S.A. |
| Manufactured for: | Hitachi, Ltd.<br>2-16-1, Higashi-Ueno, Taito-ku<br>TOKYO, 110-0015 JAPAN |
| Common Name: | Diagnostic Ultrasound Transducer |
| Classification: | Class II |
| Product Code: | 90-ITX 892.1570 Diagnostic Ultrasonic Transducer |
| 3. Legally Marketed Predicate Device(s): | |
| | UST-5536-7.5 Intraoperative Ultrasound Transducer [K111227];<br>EUP-O54J Intraoperative Ultrasound Transducer [K142368] |
| 4. Device Description: | |
| | Linear Array transducer |
| 5. Indication for Use: | |
| | The UST-5550-R and UST-5536-R transducer is intended for use with the PROSOUND<br>ALPHA 7 Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for<br>the diagnostic ultrasound evaluation during robotic* and non-robotic** intra-operative and<br>laparoscopic procedures. |
| | The L43K transducer is intended for use with the ARIETTA60, ARIETTA70 and NOBLUS<br>Diagnostic Ultrasound Systems by trained personnel (doctor, sonography, etc.) for diagnostic<br>ultrasound evaluation during robotic* and non-robotic** intra-operative and laparoscopic<br>procedures. |
| | This device is not indicated for Ophthalmic applications. |
| | *For robotic use, these devices can only be used with the da Vinci™ Surgical System and da<br>Vinci™ 8 mm ProGrasp Forceps. |
| | **For non-robotic use, these devices can only be used with the Covidien SILS™ Clinch |
{8}------------------------------------------------
21 CFR 807.92, Subsection a - Continued
### 6. Comparison to predicate device:
The UST-5550-R, UST-5536-R and L43K Intraoperative linear transducers are technically comparable and substantially equivalent to the currently marketed UST-5536-7.5 and EUP-054J. The difference between the subject devices (UST-5550-R, UST-5536-R & L43K) and predicate (UST-5536-7.5) is the material change to Titanium and dimensional/design changes to the probe head to include a linear ridge. The UST-5536-R have the same array dimension, number of elements and element spacing as the predicate UST-5536-7.5.
The subject devices UST-5550-R and UST-5536-R are manufactured in the U.S.A. by SOMA Manufacturing, LLC in accordance to specifications and components provided by the manufacturer: Hitachi Aloka Medical, Ltd. located in Tokyo, Japan.
The subject device L43K contains the same array diment and element spacing as the predicate EUP-O54J Intraoperative linear transducer with the same probe head design and material as the UST-5550-R.
The L43K is manufactured in its entirety by Hitachi, Ltd., Japan.
The subject and predicate transducers are compatible with track 3 diagnostic ultrasound systems and incorporate the same fundamental and scientific technologies. The only difference between the subject and predicate devices is a material change: along with minor dimensional and design modifications to probe head at the distal end of the subject devices. The material has been changed to Titanium and the scanning head is modified to add a small linear ridge to allow surgical graspers or robotic arm graspers to hold and control the scanning intraoperative or laparoscopic procedures. Neither the changes to the scanning head (dimensional and material) or the use of either surgical graspers or robotic arm graspers to manipulate the device change the intended use of the device.
The following compares the subject and predicate devices:
| Subject devices: | Predicate devices: |
|---------------------------------------------------|---------------------------------------------------|
| Reusable device | Reusable device |
| New material: Titanium | All materials previously cleared by FDA |
| Intraoperative linear array | Intraoperative linear array |
| Modes of operation: B, M, PWD, CD, PowerFlow and | Modes of operation: B, M, PWD, CD, PowerFlow and |
| Combination of each operating mode | Combination of each operating mode |
{9}------------------------------------------------
### 21 CFR Part 807.92, Section b
| 1. Non-clinical<br>Testing: | No new hazards were identified with the…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.