← Product Code [FRC](/productcode/FRC) · K152060

# 3M Attest Rapid Readout Biological Indicator 1295 (K152060)

_3M Company · FRC · Dec 1, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152060

## Device Facts

- **Applicant:** 3M Company
- **Product Code:** [FRC](/productcode/FRC.md)
- **Decision Date:** Dec 1, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles) and 100NX (Standard, Flex, Express and Duo cycles) systems.

## Device Story

Self-contained biological indicator for monitoring vaporized hydrogen peroxide sterilization; consists of polycarbonate sleeve, spore carrier (Geobacillus stearothermophilus), and media ampoule. Device placed in sterilizer; after cycle, user incubates in 3M Attest Auto-reader 490H. Reader detects fluorescence from enzyme activity; fluorescence indicates sterilization failure. Chemical indicator on cap changes from blue to pink upon exposure to hydrogen peroxide. Used in clinical settings to verify sterilization efficacy; provides 4-hour readout. Benefits patient by ensuring instruments are properly sterilized before reuse.

## Clinical Evidence

No clinical data included; bench testing only.

## Technological Characteristics

Polycarbonate sleeve; polyethylene terephthalate spore carrier; Geobacillus stearothermophilus spores (≥1x10^6); H2O2-sensitive chemical indicator ink; 60 +/- 2 °C incubation; fluorescence-based detection; 4-hour readout time.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- 3M™ Attest™ Rapid Readout Biological Indicator 1295 ([K140392](/device/K140392.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol composed of three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 1, 2015

3M Health Care Dr. Matt S. Mortensen, Ph.D., RAC Regulatory Affairs Specialist 3M Center, Bldg. 275-5W-06 St. Paul. MN 55144

Re: K152060

Trade/Device Name: 3M™ Attest™ Rapid Readout Biological Indicator 1295 Regulation Number: 21 CFR 880.2800(a) Regulation Name: Biological sterilization process indicator Regulatory Class: II Product Code: FRC Dated: October 20, 2015 Received: October 28, 2015

Dear Dr. Mortensen,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152060

Device Name

3MTM Attest™ Rapid Readout Biological Indicator 1295

#### Indications for Use (Describe)

Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles) and 100NX (Standard, Flex, Express and Duo cycles) systems.

Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Premarket Notification (510(k)) Summary

Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used to represent the 3M company.

### Sponsor Information:

3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000

Contact Person: Matt S. Mortensen, Ph.D., RAC Regulatory Affairs Phone Number: (651) 737-2670 FAX Number: (651) 737-5320

### Date of Summary: October 19, 2015

### Device Name and Classification:

Common or Usual Name: Biological Indicator Proprietary Name: 3M™ Attest™ Rapid Readout Biological Indicator 1295 Classification Name: Indicator, Biological Sterilization Process Device Classification: Class II, 21 CFR 880.2800(a) Product Code: FRC

### Predicate Devices:

- K140392 3MTM Attest™ Rapid Readout Biological Indicator 1295 .
### Description of Device:

The 3M Attest™ Rapid Readout Biological Indicator 1295 is a self-contained biological indicator specifically designed for rapid and reliable routine monitoring of STERRAD® vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3M Attest™ Auto-reader 490H. The 1295 BI is a single-use device composed of a polycarbonate sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. A chemical process indicator printed with stripes which change from blue to pink upon exposure to vaporized hydrogen peroxide is located on the top of the 1295 BI utilizes the same fundamental technology that exists in current 3M Attest™ Rapid Readout and Super Rapid

{4}------------------------------------------------

Readout BIs. The detection of fluorescence upon incubation of the 1295 BI in the 490H Autoreader indicates a sterilization failure.

## Nonclinical Comparison to the Predicate Device

This submission is addressing clearance for use in an additional sterilizer. The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is the same design as the previously cleared device of the same model number. There device has the same materials, performance specifications, and fundamental scientific technology.

## Summary of Clinical Testing

No clinical data was included in this premarket application submission.

## Indications for Use

Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles) and 100NX (Standard, Flex, Express and Duo cycles) systems.

### Comparison to Predicate Device

| Feature                                    | Submission Device: 3M™<br>Attest™ 1295 Biological Indicator                                                                                                                                                                                                                                                                                                              | Predicate Device (K140392):<br>3M™ Attest™ 1295 Biological<br>Indicator                                                                                                                                                                                                    |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                        | Use the 3M™ Attest™ Rapid Readout<br>Biological Indicator 1295 in<br>conjunction with the 3M Attest™<br>Auto-reader 490H as a standard method<br>of routine monitoring of vaporized<br>hydrogen peroxide sterilization<br>processes in STERRAD® 100S,<br>STERRAD® NX (Standard and<br>Advanced cycles) and 100NX<br>(Standard, Flex, Express and Duo<br>cycles) systems. | Use the 3M Attest™ Rapid<br>Readout Biological Indicator<br>1295 in conjunction with the 3M<br>Attest™ Auto-reader 490H as a<br>standard method of routine<br>monitoring of vaporized hydrogen<br>peroxide sterilization processes in<br>STERRAD® NX and 100NX<br>systems. |
| Organism                                   | Geobacillus stearothermophilus<br>traceable to ATCC™ 7953                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                       |
| Viable spore<br>population                 | ≥1x106                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                       |
| Resistance<br>characteristics<br>• D-value | (Tested at 10 mg/L vaporized hydrogen<br>peroxide)<br>D10 mg/L ≥ 1 second<br>Survival Time > 5 seconds                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                       |

{5}------------------------------------------------

| Survival/Kill<br>Window   | Kill Time = 7 minutes                                                  |      |
|---------------------------|------------------------------------------------------------------------|------|
| Carrier material          | Polyethylene terephthalate                                             | Same |
| Incubation<br>temperature | 60 +/- 2 °C                                                            | Same |
| Readout time              | 4 hour fluorescence result read on 3M<br>490H Auto-reader              | Same |
| Chemical<br>indicator     | H2O2 sensitive ink; appears blue until<br>processed, then appears pink | Same |
| Shelf-life                | 18 months                                                              | Same |

# Conclusion

The 3M™ Attest™ 1295 Biological Indicator is substantially equivalent to the predicate device in terms of their intended use, physical properties and technological characteristics. The nonclinical testing demonstrates that the 3M™ Attest™ 1295 Biological Indicator i effective, and performs as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K152060](https://fda.innolitics.com/device/K152060)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
