Biotrue ONEday with NaturMoist, Biotrue ONEday for Presbyopia with NaturMoist, Biotrue ONEday for Astigmatism with NaturMoist
Applicant
Bausch & Lomb, Incorporated
Product Code
LPL · Ophthalmic
Decision Date
Feb 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Spherical: The Bausch + Lomb Biotrue® ONEday (nesofilcon A) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D. Multifocal: The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D. Toric: The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear. The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
Device Story
Soft hydrophilic contact lenses (nesofilcon A) for daily wear; refractive correction of myopia, hyperopia, astigmatism, and presbyopia. Lenses function as refractive media to focus light on retina. Prescribed by eye care professionals; patient self-use for daily wear; single-use disposable. Features NaturMoist technology (sodium hyaluronate in packaging solution) to enhance comfort. Lenses include UV-blocking monomer and blue tint. Benefits include vision correction and convenience of daily disposal.
Clinical Evidence
No clinical data. Bench testing only; technological characteristics, formulation, manufacturing, and sterilization processes are identical to predicate devices.
Technological Characteristics
Hydrophilic copolymer (nesofilcon A: 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone); 78% water content. UV-absorbing monomer; Reactive Blue Dye 246. Oxygen permeability 42 x 10^-11. Cast molded. Sterilized via air over steam. Packaged in borate buffered saline with poloxamine and sodium hyaluronate.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic/non-aphakic persons with non-diseased eyes. Astigmatism correction up to 2.00D (spherical/multifocal) or 5.00D (toric).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Biotrue ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lenses (K113703)
Biotrue ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lenses (K132715)
Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lenses (K143632)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 11, 2016
Bausch & Lomb, Inc. Ms. Barbara Klube-Falso Manager, Regulatory Affairs 1400 North Goodman Street Rochester. NY 14609
Re: K151918
Trade/Device Name: Biotrue Oneday with NaturMoist; Biotrue Oneday for Presbyopia with NaturMoist; Biotrue Oneday for Astigmatism with NaturMoist Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: January 4, 2016 Received: January 5, 2016
Dear Ms. Klube-Falso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151918
#### Device Name
Biotrue ONEday Contact Lenses with NaturMoist
## Indications for Use (Describe)
Spherical:
The Bausch + Lomb Lomb Biotrue® ONEday (nesofilcon A) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D.
Multifocal:
The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D.
Toric:
The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear.
The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared | July 10, 2015 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Bausch & Lomb Incorporated<br>1400 North Goodman Street, Rochester, New York 14609<br>585.338.8503 |
| Contact | Barbara Klube-Falso |
| Trade names | Biotrue ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lens with<br>NaturMoist<br>Biotrue ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact<br>Lens with NaturMoist<br>Biotrue ONEday for Astigmatism (nesofilcon A) Soft (Hydrophilic)<br>Contact Lens with NaturMoist |
| Common name | soft contact lenses (hydrophilic) |
| Classification | Class II (21 CFR 886.5925 (b) (1)) |
| Predicate Devices | Biotrue ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lenses cleared<br>under K113703,<br>Biotrue ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact<br>Lenses cleared under K132715, and<br>Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic)<br>Contact Lenses cleared under K143632 |
## Device Description:
The Biotrue ONEday Contact Lens with NaturMoist is a soft hydrophilic contact lens. The lenses are made from the nesofilcon A material, a hydrophilic copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone, and is 78% water by weight when immersed in a sterile borate buffered saline solution with poloxamine and sodium hyaluronate. A UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280nm to 315nm and less than 50% in the UVA range of 316nm to 380mm. This lens is tinted blue with Reactive Blue Dye 246. The color additive conforms to 21 CFR Part 73.3106.
The physical / optical properties of the lens are:
| Specific Gravity: | 1.039 |
|----------------------|--------------------------------------------------------------------------------|
| Refractive Index: | 1.374 |
| Light Transmittance: | 99% |
| Water Content: | 78% |
| Oxygen Permeability: | 42 x 10-11 [cm3O2(STP) x cm]/(sec x cm2 x mmHg)@35°C<br>(polarographic method) |
{4}------------------------------------------------
The lenses will be manufactured in the following parameter ranges:
| Diameter: | 13.5mm to 15.5mm |
|-------------------|--------------------------------------|
| Center Thickness: | 0.05mm to 0.75mm (varies with power) |
| Base Curve: | 7.8mm to 9.5mm |
| Power Range: | +20.00D to -20.00D |
| Add Power: | +0.75D to +5.00D |
| Cylinder Powers: | -0.75D to -5.00D |
| Cylinder Axis: | 0º to 180º |
# Indications for Use:
# Bioture ONEday Contact Lenses with NaturMoist
# Spherical:
The Bausch + Lomb Biotrue® ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D.
# Multifocal:
The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D.
## Toric:
The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear.
The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
{5}------------------------------------------------
# Technological Characteristics
| Property | Biotrue ONEday<br>SVS<br>(predicate) | Biotrue ONEday<br>Multi-Focal<br>(predicate) | Biotrue ONEday<br>Toric<br>(predicate) | Biotrue ONEday<br>Contact Lenses<br>with NaturMoist<br>(Subject Device) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Functionality | The contact lens acts<br>as a refractive<br>medium that focus<br>light rays from near<br>and distant objects<br>on the retina. | The contact lens<br>acts as a refractive<br>medium that focus<br>light rays from near<br>and distant objects<br>on the retina. | The contact lens acts<br>as a refractive<br>medium that focus<br>light rays from near<br>and distant objects<br>on the retina. | Same as predicate |
| Modality | Daily wear contact<br>lens | Daily wear contact<br>lens | Daily wear contact<br>lens | Same as predicate |
| Manufacturing<br>Method | Cast Mold | Cast Mold | Cast Mold | Same as predicate |
| Material<br>Group | Group II<br>(high water, no ionic<br>polymers) | Group II<br>(high water, no ionic<br>polymers) | Group II<br>(high water, no ionic<br>polymers) | Same as predicate |
| USAN Name | Nesofilcon A | Nesofilcon A | Nesofilcon A | Same as predicate |
| Water Content | 78% | 78% | 78% | Same as predicate |
| UV Blocker | yes | yes | yes | Same as predicate |
| Sterilization | Air over steam | Air over steam | Air over steam | Same as predicate |
| Physical/optic<br>al properties | Specific Gravity:<br>1.039<br>Refractive index:<br>1.374<br>Surface Character:<br>Hydrophilic | Specific Gravity:<br>1.039<br>Refractive index:<br>1.374<br>Surface Character:<br>Hydrophilic | Specific Gravity:<br>1.039<br>Refractive index:<br>1.374<br>Surface Character:<br>Hydrophilic | Same as predicate |
| Oxygen<br>Permeability | $42 x 10^{-11*}$ | $42 x 10^{-11*}$ | $42 x 10^{-11*}$ | Same as predicate |
| Light<br>Transmittance | 99% | 99% | 99% | Same as predicate |
| Color Additive | Reactive Blue Dye<br>246 | Reactive Blue Dye<br>246 | Reactive Blue Dye<br>246 | Same as predicate |
| Lens Design | Spherical | Multi-Focal | Toric | Same as predicates |
| Packaging | Polypropylene<br>blister with plastic<br>coated aluminum<br>foil blister | Polypropylene<br>blister with plastic<br>coated aluminum<br>foil blister | Polypropylene<br>blister with plastic<br>coated aluminum<br>foil blister | Same as predicate |
Bausch & Lomb Incorporated
{6}------------------------------------------------
| Difference | | | | |
|-----------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------|
| Packaging<br>Solution | Borate Buffered<br>Saline with<br>poloxamine | Borate Buffered<br>Saline with<br>poloxamine | Borate Buffered<br>Saline with<br>poloxamine | Borate Buffered<br>Saline with<br>poloxamine and<br>sodium<br>hyaluronate |
*[cm302(STP) x cm]/(sec x cm² x mmHg)(@35°C (polarographic method)
# Summary of Non-Clinical Performance Data
Non-Clinical Tests: Testing conducted followed FDA guidance, FDA Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994. The testing performed demonstrated that the devices continue to function in a safe and effective manner. Performance testing included conformance to predetermined specifications, functional test results verify that the device performs as expected and is equivalent to the predicate without creating additional risk to the user.
Clinical Tests: The technological characteristics, formulation, manufacturing and sterilization processes are the same as the predicate device, therefore, no clinical studies were required to demonstrate the safety or effectiveness of the subject device.
## Substantial Equivalence:
The Biotrue ONEday Contact Lenses with NaturMoist are substantially equivalent to the predicate devices as they have the same functional and scientific technology, lens characteristics and the intended use is identical to that of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.