← Product Code [OOI](/productcode/OOI) · K151917

# ARIES System (K151917)

_Luminex Corporation · OOI · Oct 6, 2015 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151917

## Device Facts

- **Applicant:** Luminex Corporation
- **Product Code:** [OOI](/productcode/OOI.md)
- **Decision Date:** Oct 6, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.2570
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Luminex ARIES® System is an in vitro diagnostic (IVD) platform that performs nucleic acid based tests in clinical laboratories. The Luminex ARIES® System is capable of automated extraction and purification of nucleic acids from multiple sample types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR. For prescription use only.

## Device Story

ARIES® System is an automated IVD platform for clinical laboratories; performs nucleic acid-based testing. Input: clinical samples. Process: automated extraction and purification of nucleic acids; automated amplification and detection of target sequences via fluorescence-based PCR. Output: diagnostic results for target nucleic acid sequences. Used by laboratory personnel to facilitate molecular diagnostic testing; streamlines workflow by integrating extraction and amplification steps; enables rapid, standardized detection of pathogens or genetic targets.

## Clinical Evidence

No clinical data provided; submission focuses on the validation of the instrumentation platform for automated nucleic acid extraction and PCR-based detection.

## Technological Characteristics

Benchtop instrument; two modules; six optical channels; fluorometer for real-time PCR and melt analysis. Embedded PC control. Software-controlled thermal/optical calibration. Standards: IEC 60825 (laser safety), IEC 62304 (software lifecycle), ISO 14971 (risk management), ISO 15223-1 (symbols).

## Regulatory Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
k151917

B. Purpose for Submission:
New instrument for use with the ARIES® HSV 1&amp;2 Assay cleared under k151906

C. Manufacturer and Instrument Name:
Luminex Corporation, ARIES® System

D. Type of Test or Tests Performed:
Nucleic-acid amplification testing

E. System Descriptions:

1. Device Description:
The ARIES® System is a clinical multiplex test system that automates and integrates extraction of nucleic acid from a clinical sample, performs real-time PCR, and measuring and sorting multiple signals generated in an in vitro diagnostic assay. The instrument system is used with specific IVD assays to measure multiple analytes. The device includes a signal reader unit, raw data storage mechanisms, data acquisition software and software to process detected signals. The ARIES® System is for in vitro diagnostic use by trained laboratory professionals in a controlled laboratory environment and is not intended for patient contact.

The ARIES® System consists of the following:
- a bench top instrument with two separate modules each with a magazine rack to accommodate six disposable single-use assay cassettes
- methods for manipulating sample and reagent movement within the cassette
- heaters for lysis, elution and nucleic acid amplification
- 6 optical channels
- fluorometer for real-time PCR fluorescence detection and melt analysis
- methods for thermal and optical calibration
- software for running and analyzing IVD assays

2. Principles of Operation:
a. Device Features Controlled by Software:

{1}

The ARIES® SYSTEM instrument uses the ARIES® Software 1.0 software and the ARIES® Assay Protocol file for data acquisition, analysis, and reporting.

The ARIES® instrument software is installed on the embedded PC and controls the two modules housed in the instrument. The ARIES® Instrument is a Real Time PCR Instrument. The ARIES® instrument software provides the interface between the ARIES® Software 1.0 and the ARIES® system hardware. The ARIES® system software communicates to the instrument software to manage operations of the system, including manipulation of sample and reagent movement within the cassette, control of heaters for lysis, elution and nucleic acid amplification, controlling the fluorimeter for real-time PCR bulk fluorescence detection and melt analysis, controlling thermal and optical calibration and data analysis for IVD assays run on the system.

The ARIES® Software 1.0 software is installed in an embedded computer and provides the graphical user interface (GUI) application that is intended to be the end-user interface to the ARIES® instrument. The ARIES® Software 1.0 software is responsible for providing the environment in which an end-user runs and reports assays, administers, maintains, and sets default settings for the system.

The ARIES® Assay Protocol Software is a framework that allows running assays and generating reports on an ARIES® instrument. The ARIES® Assay Protocol Software is contained within an assay protocol template. This template is a read-only type of assay protocol file that contains a specific script, data reduction, assay editor tool, and reports, and default assay parameters and blank call logic. The ARIES® Assay Protocol file that is generated contains assay logic (rules that specify clinical outcomes), data reduction runtime files (DLL), and assay configuration parameters (script parameters that control the instrument for running a sample in a cassette and data reduction algorithm parameters).

b. Operational Environment (Off-the-Shelf Software):

The ARIES® Software 1.0 software uses some commercial off-the-shelf software components. The sponsor provided detailed documentation for the OTS used, including validation, for both the ARIES® Software 1.0 software and the ARIES® Assay Protocol Software.

For the ARIES® Assay Protocol Software used by Luminex, the recommended configuration for each computer is:

- Windows 7 Professional SP1 (US English, 32 bit or 64 bit)
- 2.8 GHz Intel Dual Core (or higher)
- 4 Gb RAM (or higher)
- 160 GB hard drive (or higher)

The programming language used to develop the ARIES® Software 1.0 software was

2

{2}

Microsoft Visual C#. Instrument control software was written in a combination and C# and C++.

3. Modes of Operation:

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes ☐ x ☐ or No ☐

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ☐ or No ☐ x ☐

4. Specimen Identification:

A barcode reader may be used for entry of sample IDs, or they may be entered manually.

5. Specimen Sampling and Handling:

Samples are manually prepared according to assay manufacturer’s suggestions and are transferred to an assay specific cassette for analysis.

6. Calibration:

Calibration is performed by Luminex service personnel using ARIES® System Verification Cassettes.

7. Quality Control:

Each ARIES® assay cassette includes a Sample Process Control (SPC). The SPC has a known melting temperature (Tm) range and cycle threshold (Ct) range. Each time an assay is run, the system measures the temperature and fluorescence intensity of the SPC control to ensure the thermal and optical subsystems have remained in calibration.

8. Software:

FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:

Yes ☐ x ☐ or No ☐

{3}

4

F. Regulatory Information:

1. Regulation section:
21 CFR 862.2570 Instrumentation for clinical multiplex test systems

2. Classification:
Class II (special controls)

3. Product code:
OOI – Real-time nucleic acid amplification

4. Panel:
Chemistry (75)

G. Intended Use:

1. Indication(s) for Use:
The Luminex ARIES® System is an in vitro diagnostic (IVD) platform that performs nucleic acid based tests in clinical laboratories. The Luminex ARIES® System is capable of automated extraction and purification of nucleic acids from multiple sample types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR.

2. Special Conditions for Use Statement(s):
For prescription use only.

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:
BD Diagnostics (Becton, Dickinson and Company), BD MAX System K111860

2. Comparison with Predicate Device:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device: Luminex ARIES® System | Predicate Device: BD Max System (k111860)  |
|  Intended Use | The Luminex ARIES® System is an in vitro | Same  |

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device: Luminex ARIES® System | Predicate Device: BD Max System (k111860)  |
|   | diagnostic (IVD) platform that performs nucleic acid based tests in clinical laboratories. The Luminex ARIES® System is capable of automated extraction and purification of nucleic acids from multiple sample types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR. |   |
|  Sample Preparation Method | Automated nucleic acid lysis and extraction by the ARIES® system | Same  |
|  Assay Format | Amplification: Real Time PCR Detection: Fluorogenic | Same  |
|  Interpretation of Test Results | Automated (Diagnostic software of Luminex ARIES® System) | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | New Device: Luminex ARIES® System | Predicate Device: BD Max System (k111860)  |
|  Fluidics | Self-contained | Manual sample preparation  |
|  Assay Cartridge | • Cartridge is closed to the environment • Single Use | • Reagent strips are open to the environment • Can be used twice  |
|  Fluorescence Channels | Six (6) channels | Five (5) channels  |
|  Software | Software resides on the ARIES® System | Software resides on an all-in-one computer  |

# I. Special Control/Guidance Document Referenced (if applicable):

- IEC60825 – Safety of laser products,  $2^{\text{nd}}$  edition, 2007
- IEC62304, - Medical Device Software - Software Life Cycle Processes, 2006
- ISO 14971 - Medical Devices - Applications of Risk Management, 2007
- ISO15223-1 - Medical Devices - Symbols to be used with Medical Device Labels, Labelling, and Information to be Supplied, 2007

{5}

J. Performance Characteristics:

1. Analytical Performance:

Performance for the ARIES® System was established in the clearance of the assay, the ARIES® HSV1&amp;2 Assay (k151906).

a. Accuracy:

See k151906

b. Precision/Reproducibility:

See k151906

c. Linearity:

See k151906

d. Carryover:

See k151906

e. Interfering Substances:

See k151906

2. Other Supportive Instrument Performance Data Not Covered Above:

Not applicable

K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K151917](https://fda.innolitics.com/device/K151917)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
