Cable-Ready GTR System: The Cable-Ready GTR System is indicated for the reattachment of the greater trochanter following osteotomy in total hip arthroplasty. Additionally, the device is used to reattach the greater trochanter following fracture of the greater trochanter, nonunion of fracture, or previous osteotomy. Cable-Ready 1.8mm Cerclage Cables: The 1.8mm cerclage cable can be used to secure fractures of the olecranon, patella, femur, humerus, and ankle, as well as to reduce and secure acromioclavicular dislocations. The 1.8mm cables can also be used to reattach the greater trochanter after trochanteric osteotomy during total hip arthroplasty.
Device Story
System comprises GTR devices and 1.8mm cobalt chrome alloy cerclage cables; used for bone fixation and stabilization. GTR devices feature two upper prongs for looping around greater trochanter and two lower fins to prevent migration/rotation. Cables used with GTR devices or independently to secure fractures. Operated by orthopedic surgeons in clinical/surgical settings. Provides mechanical stabilization of bone segments; aids healing by maintaining alignment of fractured or osteotomized bone. Benefits include secure fixation and reduced risk of implant migration.
Clinical Evidence
No clinical data provided. Substantial equivalence established via non-clinical performance testing, including shelf-life accelerated aging, biocompatibility testing per ISO 10993-1, construct fatigue testing for the GTR system, and engineering analysis for the cerclage cables.
Technological Characteristics
Materials: Cobalt chrome alloy and stainless steel. Design: GTR devices include upper prongs and lower anti-migration fins. 1.8mm cerclage cables. Sterilization: Sterile. Biocompatibility: Tested per ISO 10993-1.
Indications for Use
Indicated for patients requiring greater trochanter reattachment post-osteotomy or fracture (including nonunion) in total hip arthroplasty, and for patients requiring fixation of olecranon, patella, femur, humerus, or ankle fractures, or reduction/fixation of acromioclavicular dislocations.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Pioneer System GTR (Greater Trochanter Reattachment) (K961267)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with flowing lines suggesting hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 30, 2015
Zimmer, Incorporated Ms. Dalene Binkley Senior Specialist. Trauma Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46580
Re: K151907
Trade/Device Name: Cable-Ready® Cable Grip System: Cable-Ready® GTR System, Cable-Ready® 1.8mm Cerclage Cables Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT, JDQ Dated: December 14, 2015 Received: December 15, 2015
Dear Ms. Binkley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K151907 Page 1 of 2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name
Cable-Ready® Cable Grip System: Cable-Ready® GTR System
Indications for Use (Describe)
Cable-Ready® GTR System - The Cable-Ready® GTR System is indicated for the reattachment of the greater trochanter following osteotomy in total hip arthroplasty. Additionally; the device is used to reattach the greater trochanter following fracture of the greater trochanter, nonunion of fracture, or previous osteotomy.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name
Cable-Ready® Cable Grip System: Cable-Ready®1.8mm Cerclage Cables
Indications for Use (Describe)
Cable-Ready® 1.8mm Cerclage Cables - The 1.8mm cerclage cable can be used to secure fractures of the olecranon, patella, femur, humerus, and ankle, as well as to reduce and secure acromioclavicular dislocations. The 1.8mm cables can also be used to reattach the greater trochanteric osteotomy during total hip arthroplasty.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
#### 510(k) Summary
| Sponsor: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dalene Binkley, MS, RAC<br>Senior Specialist, Trauma Regulatory Affairs<br>Telephone: 574-372-4970<br>Fax: (574) 372-4605 |
| Date: | July 10, 2015 |
| Trade Name: | This is a bundled traditional 510(k). The two systems<br>bundled in this submission are: |
| | Cable-Ready® Cable Grip System: |
| | 1) Cable-Ready GTR System<br>2) Cable-Ready 1.8mm Cerclage Cables |
| Common Name: | Single/multiple component metallic bone fixation<br>appliance and accessories and cerclage, bone fixation |
| Classification Names<br>and References: | Single/multiple component metallic bone fixation<br>appliance and accessories<br>Regulation Number: CFR 888.3030<br>Classification Number: 87 KTT |
| | Cerclage, fixation<br>Regulation Number: CFR 888.3010<br>Classification Number: 87 JDQ |
| Classification Panel: | Orthopedics/87 |
| Predicate Device(s): | Pioneer System GTR (Greater Trochanter Reattachment)<br>(Pioneer Laboratories, Inc., K961267, cleared 6/27/96) |
| | Extended GTR Device (Pioneer Surgical Technology,<br>K000734, cleared 5/5/00) |
{5}------------------------------------------------
Songer Cable System (Pioneer Laboratories, K935481, cleared 1/26/94)
| Device Description: | The Cable-Ready Greater Trochanter Reattachment Short<br>and Long (GTR) devices are used in combination with<br>1.8mm cobalt chrome alloy cerclage cables to stabilize<br>fractures of the greater trochanter or reattach the greater<br>trochanter after osteotomy in total hip replacement. The<br>GTR devices are designed with two upper prongs that loop<br>around the greater trochanter and two small fins on the<br>lower side to prevent migration or rotation.<br><br>In addition to being used with the GTR devices, the cobalt<br>chrome alloy 1.8mm cerclage cables, along with the<br>stainless steel cables, can be used alone to secure bone<br>fractures. |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Cable-Ready GTR System:<br><br>The Cable-Ready GTR System is indicated for the<br>reattachment of the greater trochanter following osteotomy<br>in total hip arthroplasty. Additionally, the device is used to<br>reattach the greater trochanter following fracture of the<br>greater trochanter, nonunion of fracture, or previous<br>osteotomy.<br><br>Cable-Ready 1.8mm Cerclage Cables:<br><br>Cable-Ready 1.8mm Cerclage Cables - The 1.8mm<br>cerclage cable can be used to secure fractures of the<br>olecranon, patella, femur, humerus, and ankle, as well as to<br>reduce and secure acromioclavicular dislocations. The<br>1.8mm cables can also be used to reattach the greater<br>trochanter after trochanteric osteotomy during total hip<br>arthroplasty. |
| Comparison to Predicate Device: | The subject devices incorporate similar or identical<br>materials, similar or identical indications for use, similar<br>or identical sizes of implants, and the same technological<br>characteristics as the predicate devices. |
{6}------------------------------------------------
Performance Data (Nonclinical and/or Clinical):
Non-Clinical Performance and Conclusions:
- Shelf Life Accelerated aging testing conducted . shows that the sterile devices included in this submission have a shelf life of 10 years.
- Biocompatibility Biocompatibility testing of the . subject devices was conducted per ISO 10993-1 and Good Laboratory Practices (21 CFR 58). All testing passed.
- Performance Evaluation A construct fatigue test . for the GTR system and an engineering analysis for the 1.8mm cerclage cables demonstrate the subject devices are safe and effective and substantially equivalent to the predicate devices.
Conclusions: The data presented in this submission demonstrates that the subject devices are substantially equivalent to their respective predicate devices.
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for these devices to show substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.