BiNAPS Nasal Airflow and Snore Transducer; ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal
Applicant
Salter Labs
Product Code
MNR · Anesthesiology
Decision Date
Apr 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BiNAPS Nasal Airflow Pressure and Snore Transducer (using the pressure transducer) is an accessory intended for use with polysomnography equipment during sleep disorder studies for the purpose of detecting and amplifying breathing signals and detection of snoring of a sleeping patient through a Salter Labs nasal cannula. This device is intended for adult and pediatric (excluding neonatal and infant) use. The ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal is a thermal resistive thermistor designed to monitor oral/ nasal thermal airflow temperature and pressure changes during sleep studies. The ThermiSense is intended for use with a nasal cannula and polysomnography headbox or pressure transducers, such as, BiNAPS Nasal Airflow Transducer. This device is intended for adult and pediatric (excluding neonatal and infant) use.
Device Story
BiNAPS and ThermiSense are sleep diagnostic accessories used with polysomnography equipment. BiNAPS uses a nasal cannula to direct respiratory pressure waves to a piezo-electric ceramic element; flexing of element generates proportional voltage signals for breathing and snoring detection. ThermiSense uses a thermistor mounted in a nasal cannula to monitor oral/nasal thermal airflow temperature changes; requires external signal-conditioning unit with battery for current flow. Both devices are used in sleep laboratories by clinicians to acquire respiratory data. Outputs are electrical signals connected to polygraph AC jackboxes. Data assists clinicians in diagnosing sleep disorders. Material modification from PVC (DEHP/DIDP) to PVC (DINCH/DOTP) implemented.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing (irritation, sensitization, cytotoxicity) passed. Performance testing verified flow rate, back pressure, and bond strength.
Technological Characteristics
BiNAPS: Piezo-electric ceramic element, passive electronic circuitry, no power source. ThermiSense: Thermal resistive thermistor, requires battery-powered signal-conditioning unit. Materials: PVC with DINCH/DOTP plasticizers. Non-sterile. Connectivity: 1.5mm safety connectors for polygraph AC jackbox.
Indications for Use
Indicated for adult and pediatric (excluding neonatal and infant) patients undergoing sleep disorder studies for monitoring breathing signals, snoring, and oral/nasal thermal airflow.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2016
Salter Labs Mara Caler Director, Regulatory Affairs 2365 Camino Vida Roble Carlsbad, California 92011
Re: K151874
Trade/Device Name: BiNAPS Nasal Airflow and Snore Transducer: ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR, BZQ Dated: March 17, 2016 Received: March 18, 2016
Dear Ms. Caler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151874
### Device Name
BiNAPS Nasal Airflow Pressure and Snore Transducer
ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal
### Indications for Use (Describe)
The BiNAPS Nasal Airflow Pressure and Snore Transducer (using the pressure transducer) is an accessory intended for use with polysomnography equipment during sleep disorder studies for the purpose of detecting and amplifying breathing signals and detection of snoring of a sleeping patient through a Salter Labs nasal cannula. This device is intended for adult and pediatric (excluding neonatal and infant) use.
The ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal is a thermal resistive thermistor designed to monitor oral/ nasal thermal airflow temperature and pressure changes during sleep studies. The ThermiSense is intended for use with a nasal cannula and polysomnography headbox or pressure transducers, such as, BiNAPS Nasal Airflow Transducer. This device is intended for adult and pediatric (excluding neonatal and infant) use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807) is provided:
## 510 (k) Summary
#### A. Submitter:
Salter Labs 2365 Camino Vida Robles Carlsbad, CA 92011 Telephone: 760-795-7094 Fax: 760-683-6797
## B: Contact Person:
Mara Caler Director, Regulatory Affairs
# C: Date Prepared:
13 April 2016
## D. Device Names:
| Trade Name | Common Name | Classification Name | Product Code |
|----------------------------------------------------------------------------------------------------------------|--------------------------------|---------------------------------|--------------|
| BiNAPS Nasal Airflow<br>Pressure and Snore<br>Transducer | Airflow Pressure<br>Transducer | Ventilatory Effort<br>Recorder | MNR |
| ThermiSense Oral/Nasal<br>Thermal AirFlow Sensor<br>with Airflow Pressure<br>Cannulas, Nasal and<br>Oral/Nasal | Airflow Sensor | Monitor, Breathing<br>Frequency | BZQ |
| Regulation Number: | 868.2375 |
|-----------------------|-----------------|
| Classification: | II |
| Classification Panel: | Anaesthesiology |
## Predicate Devices:
This submission demonstrates substantial equivalence to the predicate devices: K051313, Airflow Pressure Transducer K080922, Airflow Sensor
{4}------------------------------------------------
#### E. Device Descriptions
The Salter Labs BiNAPS Nasal Airflow Pressure and Snore Transducer is a two output channel device used to acquire respiratory low pressure waves and low air flow that are sensed through a Nasal Cannula typically worn by a subject during a sleep diagnostic session. It is used to convert changes in air pressure and flow, occurring during sleep, into electrical signals that can be measured by polysomnography equipment. The Nasal Cannula directs the airflow and pressure waves generated by breathing and snoring from the nares and mouth of a patient through a luer lock fitting and then into a cup shaped plastic cylinder chamber sealed closed at the open end by a fitting and then into a cup shaped plastic cylinder chamber sealed closed at the open end by a piezo-electric ceramic element. The piezo element, when flexed by the impinging air pressure changes, generates a proportional electric voltage. This voltage is attenuated and filtered by subsequent passive electronic circuitry composing the sensor. The Salter Labs Airflow Pressure Transducer does not require a power source.
ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal is composed of a Thermistor as the element which changes resistance as airflow from the patient is delivered across the element of the thermistor. The thermistor elements are located directly underneath the nares and in the airflow of the mouth. In both cases the element is kept from touching the skin of the patient in order to be the most effective change in temperature.
The Thermistor is mounted in the Cannula to position the thermistor properly under the nares and in the airflow path of the mouth. The Thermistor is covered with a heat shrink tubing to protect it from moisture from the patient. The Thermistor requires a battery source in order to derive a current flow through the circuit. This battery source is located in a small container called a signal-conditioning unit. This unit has circuitry that will remove noise, display a smooth signal and is an added feature test for wire continuity.
### F. Intended Use
## Intended / Indications for Use
The BiNAPS Nasal Airflow Pressure and Snore Transducer (using the pressure transducer) is an accessory intended for use with polysomnography equipment during sleep disorder studies for the purpose of detecting and amplifying breathing signals and detection of snoring of a sleeping patient through a Salter Labs nasal cannula. This device is intended for adult and pediatric (excluding neonatal and infant) use.
The ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal is a thermal resistive thermistor designed to monitor oral/nasal thermal airflow temperature and pressure changes during sleep studies. The ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal is intended for use with a nasal cannula and polysomnography headbox or pressure transducers, such as, BiNAPS Nasal Airflow Pressure and Snore Transducer. This device is intended for adult and pediatric (excluding neonatal and infant) use.
{5}------------------------------------------------
### Substantial Equivalence Table: G.
Comparison of Technological Characteristics with the Predicate Device
The proposed BiNAPS Nasal Airflow Pressure and Snore Transducer and ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal are substantially equivalent to the predicate devices.
| Characteristic | Predicate device | Modified device |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Intended use | The BiNAPS Nasal Airflow Pressure and<br>Snore Transducer is an accessory intended<br>for use with polysomnography equipment<br>during sleep disorder studies for the<br>purpose of detecting and amplifying<br>breathing signals and detection of snoring<br>of a sleeping patient through a Salter Labs<br>nasal cannula.<br>The ThermiSense Oral/Nasal Thermal<br>AirFlow Sensor with Airflow Pressure<br>Cannulas, Nasal and Oral/Nasal is used as<br>a cannula accessory with existing recording<br>devices and data acquisition systems in a<br>sleep laboratory setting to support the<br>diagnostic recording of nasal and or oral<br>airflow. The subject device itself performs<br>no diagnostic functions, and only supports<br>the diagnostic recording of airflow for use<br>as an accessory component to a<br>polysomnography recorder. | Unchanged |
| Intended use<br>population | Patients undergoing sleep diagnosis studies | Unchanged |
| Principles of<br>operation | The detection and amplification of<br>breathing signals and detection of snoring<br>of a sleeping patient through a Salter Labs<br>nasal cannula. | Unchanged |
| Airflow output | Output is a nasal pressure airflow<br>waveform. Output is connected to a<br>polygraph's AC jackbox with 1.5mm safety<br>connectors. Snoring can also be<br>superimposed onto the airflow waveform<br>as shown in the example on the next page.<br>This is accomplished by setting higher<br>values for the high frequency filter and the<br>sampling rate for the airflow channel. | Unchanged |
| Snore output | Output is a snoring waveform derived from | Unchanged |
{6}------------------------------------------------
| Characteristic | Predicate device | Modified device |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| | snore vibrations on the nasal pressure<br>signal. The channel has internal low<br>frequency filter to remove the airflow<br>signal and to provide a flat baseline<br>between snores. Output is connected to a<br>polygraph's AC jackbox with 1.5mm safety<br>connectors. | |
| Clinical setting | Prescription use sleep diagnosis study | Unchanged |
| Device characteristics | Non-sterile | Unchanged |
The only difference between the modified BiNAPS Nasal Airflow Pressure and Snore Transducer and the modified ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal and predicate devices is a material modification used in the cannulas from PVC (DEHP and DIDP) to (DINCH and DOPT). The materials used in the remainder of the devices are unchanged.
| Features | Predicate Airflow<br>Pressure<br>Transducer &<br>Airflow Sensor | Modified Airflow<br>Pressure Transducer<br>& modified Airflow<br>Sensor | Performance<br>Testing |
|-------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------|
| Material<br>Formulation | PVC, DIDP<br>PVC, DEHP | PVC, DINCH<br>PVC, DOTP | Biocompatibility<br>and<br>Performance |
### H. Performance Data
The BiNAPS Nasal Airflow Pressure and Snore Transducer and the ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal were tested to verify that the new material bond and performance characteristics of flow rate, back pressure, and tubing bond strength did not impact the strength or performance of the modified Airflo w Pressure Transducer and the modified Airflow Sensor after the material change.
The test results demonstrate that the BiNAPS Nasal Airflow Pressure and Snore Transducer and the ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal are substantially equivalent to the predicate devices
{7}------------------------------------------------
| Criteria | Predicate specifications | Modified devices | Comments |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Back pressure<br>(flow rates): | Shall not have a back<br>pressure that exceeds<br>$3$ psi at a maximum<br>flow rate in ambient<br>of 5°C, 20°C, and<br>40°C. | Maximum back<br>pressure was found<br>to be less than $2$ psi. | The modified<br>devices passed all<br>specifications |
| Bond Strength<br>(tensile<br>strength): | The bonded<br>components of the set<br>will have a bond<br>strength that is $\ge 2$<br>lbs. when pulled at a<br>rate of 5 inches per<br>minute. | The bond strength<br>test achieved over $2$<br>times the minimum<br>allowable value. | The modified<br>devices passed<br>all specifications |
| Device<br>dimensions: | Unchanged. | Unchanged. | The modified<br>devices use the<br>same molds and are<br>unchanged from the<br>predicate devices |
| Aging (shelf<br>life) | Not included | 3 years | The modified<br>devices meet all<br>specifications at the<br>3 year test point |
# Performance Testing Summary
The modified Sleep Diagnostics Cannulas meet established Salter Labs performance specifications.
| Characteristic | Predicate | New Device | Result |
|----------------------------------|-----------|------------|--------|
| Performance | | | |
| Sampling flow rate (max) | .5 lpm | .5 lpm | Same |
| Sampling backpressure @ max flow | < .64 psi | < .64 psi | Same |
{8}------------------------------------------------
### I. Clinical / Non-Clinical
No clinical patient testing was required for this submission.
The following biocompatibility testing was performed. The materials passed all parameters:
- . Irritation
- Sensitization ●
- Cytotoxicity ●
# J. Conclusions
The BiNAPS Nasal Airflow Pressure and Snore Transducer and the ThermiSense Oral/Nasal Thermal AirFlow Sensor with Airflow Pressure Cannulas, Nasal and Oral/Nasal data and test results demonstrate that the devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.