ABL80 FLEX CO-OX with AQURE connectivity

K151856 · Sendx Medical, Inc. · CHL · Aug 6, 2015 · Clinical Chemistry

Device Facts

Record IDK151856
Device NameABL80 FLEX CO-OX with AQURE connectivity
ApplicantSendx Medical, Inc.
Product CodeCHL · Clinical Chemistry
Decision DateAug 6, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1120
Device ClassClass 2

Intended Use

The ABL80 FLEX CO-OX analyzer is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and oximetry in whole blood. The ABL80 FLEX CO-OX analyzer system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Device Story

Portable, automated blood gas analyzer; measures pH, blood gases, electrolytes, glucose, and oximetry in whole blood. Uses potentiometry, amperometry, conductometry, and spectrophotometry. Operated by trained clinicians in lab or point-of-care settings. Integrates with AQURE data management software for remote result display, device status monitoring, and remote initiation of system cycles (rinse, calibration, lock/unlock). Output provided to clinicians for diagnosis and treatment of metabolic and acid-base disorders. Benefits include rapid, point-of-care diagnostic data for life-threatening conditions.

Clinical Evidence

Bench testing only. Verification and validation activities performed based on risk analysis to ensure software modification functions as intended without impacting core analyzer performance.

Technological Characteristics

Portable analyzer; potentiometry, amperometry, conductometry, and spectrophotometry sensing principles. Connectivity via AQURE system (Ethernet/RS232). Hardware includes 8.4" TFT-LCD touchscreen. Software runs on Windows XP Embedded. Risk management per ISO 14971.

Indications for Use

Indicated for use in clinical laboratory settings for blood gas, electrolyte, and co-oximetry analysis.

Regulatory Classification

Identification

A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 6, 2015 SENDX MEDICAL INC. LASSE MOLLER REGULATORY AFFAIRS SPECIALIST AAKANDEVEJ 21 BROENSHOEJ 2700 DENMARK Re: K151856 Trade/Device Name: ABL80 FLEX CO-OX with AQURE Connectivity Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: II Product Code: CHL, JGS, CEM, JFP, CGZ, CGA, GKR, DQA, GHS, JPI Dated: July 3, 2015 Received: July 8, 2015 Dear Lasse Moller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Courtney H Lias -S Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k151856 Device Name ABL80 FLEX CO-OX with AQURE connectivity ## Indications for Use (Describe) Intended Use: The ABL80 FLEX CO-OX analyzer is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and oximetry in whole blood. The ABL80 FLEX CO-OX analyzer system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-ofcare setting. Indications for use: These tests are only performed under a physicians order: pH, pO2.and pCO2 . pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acidbase disturbances. Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Sodium (cNa+): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance. Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. Chloride (cCl-): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis. Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia. sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin plus reduced hemoglobin. FO2Hb: oxyhemoglobin as a fraction of total hemoglobin FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning. FMetHb: methemoglobin as a fraction of total hemoglobin. FHHb: reduced hemoglobin as a fraction of total hemoglobin. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Date Prepared: August 5, 2015 # Section 5. 510(k) Summary #### 1. Administrative #### Submitter | Company Name: | SenDx Medical Inc | |-----------------|----------------------------------------------------| | ER Number: | 2027541 | | Address: | 1945 Palomar Oaks Way<br>Carlsbad, CA 92011<br>USA | | Contact Person: | Vibeke Agerlin, Director of QA/RA | | Phone: | +1 760 930 6300 | | Fax: | +1 760 930 6310 | #### Application Correspondent | Name: | Lasse Post Møller | |-----------|-------------------------------| | Function: | Regulatory Affairs Specialist | | E-mail: | lpm@radiometer.dk | | Phone: | +45 3827 3436 | | Fax: | +45 3827 2727 | #### 2. Device description #### Device Information | Device Name: | ABL80 FLEX CO-OX with AQURE connectivity | |-----------------------|---------------------------------------------------| | Common Name: | Blood gas analyzer | | Product Code: | CHL (JGS, CEM, JFP, CGZ, CGA, GKR, DQA, GHS, JPI) | | Code description: | Blood gases and blood pH test system | | Regulation Number: | 21 CFR 862.1120 | | Classification: | Class II | | Classification Panel: | Clinical Chemistry | The ABL80 FLEX CO-OX with AQURE connectivity is a portable, automated system intended for in vitro testing of samples of whole blood. The device is manufactured by SenDx Medical Inc, Carlsbad, CA, USA. ABL80 FLEX CO-OX consists of an instrument with a sensor cassette and a solution pack as the main accessories. Multiple models of sensor cassettes and solution packs are available. The various sensor cassette models include models for the different parameter configurations. {5}------------------------------------------------ Four measuring principles are employed in ABL80 FLEX analyzers. - Potentiometry: The potential of a sensor chain is recorded using a voltmeter, and related ● to the concentration of the sample (the Nernst equation). - . Amperometry: The magnitude of an electrical current flowing through a sensor chain is proportional to the concentration of the substance being oxidized or reduced at an electrode in the chain. - . Conductometry: Specific impedance of a sample as measured by two conducting electrodes held at a constant voltage is directly proportional to the conductive properties of that sample. - . Spectrophotometry: Light passes through a cuvette containing a hemolyzed blood sample. The specific wavelengths absorbed and their intensity generates an absorption spectrum used to calculate oximetry parameters. This measuring principle is used in the ABL80 FLEX CO-OX analyzer only. ## 3. Device Modification The modification consists of integration with the Data Management software called AQURE system. The software enables the initiation of device actions on connected ABL80 FLEX analyzers. ## 4. Intended Use The ABL80 FLEX CO-OX analyzer is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and oximetry in whole blood. The ABL80 FLEX CO-OX analyzer system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. ## Indications for use: These tests are only performed under a physicians order: pH, pO2,and pCO2, : pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances. Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Sodium (cNa+): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance. Calcium (cCa2+): calcium measurements are used in the diaqnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. Chloride (cCI-): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis. Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoqlycemia, and idiopathic hypoqlycemia, and of pancreatic islet cell carcinoma. Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia. sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoqlobin plus reduced hemoqlobin. FO2Hb: oxyhemoqlobin as a fraction of total hemoqlobin {6}------------------------------------------------ FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning. FMetHb: methemoglobin as a fraction of total hemoglobin. FHHb: reduced hemoglobin as a fraction of total hemoglobin. #### 5. Substantial Equivalence The modified ABL80 FLEX CO-OX analyzer is substantially equivalent in Intended Use, fundamental scientific technology, features, and characteristics to the predicate (K080370). {7}------------------------------------------------ | Similarities | | | |--------------------------------|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Issue | SE Device | Predicate Device<br>(K080370) | | Intended Use | Same | The ABL80 FLEX CO-OX analyzer is a portable,<br>automated analyzer that measures pH, blood gases,<br>electrolytes, glucose, and oximetry in whole blood. The<br>ABL80 FLEX CO-OX analyzer system is intended for use<br>by trained technologists, nurses, physicians and<br>therapists. It is intended for use in a laboratory<br>environment, near patient or point-of-care setting. | | Blood Gas<br>Measurement | Same | pH, pCO2 by potentiometry,<br>pO2 by amperometry | | Electrolyte<br>Measurement | Same | cK+, cNa+, cCa2+, cCl- by potentiometry | | Metabolite<br>Measurement | Same | cGlu by amperometry | | Oximetry<br>Measurement | Same | ctHb, sO2, FO2Hb, FHHb, FCOHb, FMetHb by<br>spectrophotometry | | Performance<br>Characteristics | Same | Identical Performance Characteristics | | Calibration | Same | Two-Point liquid calibration at configured intervals, tHb<br>calibration every 90 days | | Computer<br>specifications | Same | Microsoft Windows®XP Embedded operating system<br>Minimum 1 GB hard drive<br>ETX single board CPU<br>Minimum 512 MB EDO-RAM | | Sample Introduction | Same | Aspiration | | Interfaces | Same | 8.4" color TFT-LCD, resolution 800 × 600 SVGA Touch<br>screen, Barcode reader, Serial line RS232, RJ45,<br>Ethernet port, 2 USB 1.1, PS2 keyboard | | Software version | Same | Software version 3.12 | {8}------------------------------------------------ | Differences | | | |--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Issue | SE Device | Predicate Device<br>(K080370) | | AQURE system | Data management software (MDDS) functionality:<br>Remote display of test results Receival of data from connected devices at the point-of-care or laboratory Transfer of test results to the HIS/LIS Centralized operator administration – Transmission of operator data to ABL80 FLEX CO-OX Remote approval of patient results | Analyzer functionalities are:<br>Local display of test results Direct transfer of test results to the HIS/LIS Local Operator Administration at the analyzer | | | Remote initiation of the following device actions on ABL80 FLEX CO-OX from AQURE system:<br>System Cycle Rinse Lock/Unlock analyzer Lock/Unlock parameter Set analyzer message Enter/Exit standby 2 point calibration | Local initiation of device actions through device touch screen interface | ## 6. Registration history ABL80 FLEX was originally cleared with K051804 (measures pH, blood gases, electrolytes, glucose, hematocrit (ABL80 FLEX analyzer only), and then modified in K080370 with the addition of an oximetry Module - a configuration of the analyzer adding the oximetry parameters ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, and FHHb. ## 7. Performance Data Remote functionality was verified and validated on both ABL80 FLEX and ABL80 FLEX CO-OX analyzers, and the acceptance criteria was met. No performance characteristics are affected by the change. The performance data submitted in the original submission (K051804 as modified by K080370) still apply. ## 8. Summary of Design Control activities {9}------------------------------------------------ We conducted an FMEA risk analysis and mitigated all identified hazards to an acceptable level as per ISO 14971, and verified software mitigations by using test protocols. Results met predefined acceptance criteria. ### 9. Conclusion The ABL80 FLEX CO-OX with AQURE connectivity described above is substantially equivalent in Intended Use, fundamental scientific technology, safety, effectiveness and characteristics to the predicate ABL80 Flex CO-0X (K080370). Implementation of the change design control principles (risk management, verification and validation) have been applied which indicated that the modification is of no impact to the performance of the device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...