# Rainbow Shine (K151842)

_Genoss · EIH · Jan 14, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151842

## Device Facts

- **Applicant:** Genoss
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Jan 14, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

rainbow™ Shine is used in the manufacture of a dental core through milling by machine (MAD/ MAM or CAD/CAM) followed by sintering.

## Device Story

rainbow™ Shine is a partially sintered dental ceramic block composed of colored ZrO2 (Y-TZP). The device is used by dental technicians in a laboratory setting to manufacture dental cores. The process involves milling the block into the desired core shape using CAD/CAM or MAD/MAM machinery, followed by sintering in a furnace to harden the zirconia. The resulting core is then layered with porcelain to create a finished dental restoration. The device is provided as a non-sterile preformed block in various shapes (cylinders, prisms, disks) to ensure compatibility with multiple milling systems. By providing a high-translucency zirconia material, the device enables the fabrication of aesthetic dental restorations, benefiting patients requiring tooth replacement or repair.

## Clinical Evidence

Bench testing only. Performance testing conducted per ISO 6872:2008 included bending strength, linear thermal expansion coefficient, and chemical solubility. Biocompatibility testing performed per ISO 10993-1:2009 included cytotoxicity (ISO 10993-5), irritation/intracutaneous reactivity (ISO 10993-10), sensitization (ISO 10993-10), acute systemic toxicity (ISO 10993-11), and genotoxicity (ISO 10993-3). All tests demonstrated compliance with standards.

## Technological Characteristics

Partially sintered dental ceramic (ZrO2/Y-TZP). Form: preformed blocks (cylinders, prisms, disks). Sintering density: 6 g/cm3. Bending strength: 706 MPa. Chemical solubility: 3 µg/cm². Radioactivity: < 1.0 Bq/g. Complies with ISO 6872:2008 (Type II, Class 5). Non-sterile. No software or electronic components.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- RAINBOW BLOCK ([K092513](/device/K092513.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles facing to the right. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 14, 2016

GENOSS Co., Ltd. c/o Mr. Dachan Kwon iCT America, Inc. 180 Sylvan Avenue, 2nd Floor Englewood Cliffs, New Jersey 07632

Re: K15842

Trade/Device Name: rainbow™ Shine Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: December 7, 2015 Received: December 9, 2015

#### Dear Mr. Kwon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GEN" are solid black, while the letters "OSS" appear to be composed of a textured, almost grainy pattern. The overall effect gives the word a dynamic and somewhat distressed appearance.

# Indication for use

к151842 510(k) Number:

rainbow™ Shine Device Name:

Indication for use:

rainbow™ Shine is used in the manufacture of a dental core through milling by machine (MAD/ MAM or CAD/CAM) followed by sintering.

Prescription Use __ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Indication for use

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The word "GEN" is in dark green, and the word "OSS" is in light green. The letters are all capitalized and in a sans-serif font. The logo is simple and modern.

# 510(k) Summary

12/07/2015

# 1. Company

|                | Submitter                                                                                                |
|----------------|----------------------------------------------------------------------------------------------------------|
| Name           | GENOSS Co., Ltd.                                                                                         |
| Address        | 1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105<br>Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Phone/Fax      | +82-31-888-5100/ +82-31-888-5105                                                                         |
| Contact person | Byungsun Kim / RA<br>bskim@genoss.com                                                                    |
| Summary Date   | 12/07/2015                                                                                               |

# 2. Device Name

| Proprietary name    | : rainbowTM Shine                   |
|---------------------|-------------------------------------|
| Regulation number   | : 21 CFR 872.6660                   |
| Classification name | : Porcelain powder for clinical use |
| Product code        | : EIH                               |
| Device class        | : Class II                          |

# 3. Predicated Device

K092513 RAINBOW BLOCK

## 4. Description

rainbow™ Shine is a partially sintered dental ceramic made out of colored ZrO2(Y-TZP). rainbow™ Shine is milled into cores of artificial teeth and then is finally sintered in a furnace to harden the ZrO2. Then, the core is layered with porcelain to make a finished tooth.

{4}------------------------------------------------

## 510(K) Number: K151842

Image /page/4/Picture/1 description: The image shows the word "GENOSS" in a stylized font. The letters "GENO" are in a dark green color, while the letters "SS" are in a lighter green color. The font is sans-serif and has a modern look.

# 5. Indication for use

rainbow™ Shine is used in the manufacture of a dental core through milling by machine (MAD/ MAM or CAD/CAM) followed by sintering.

## 6. Technological Characteristics

The following comparison table of the technological characteristics of the subject device and the predicate devices outlines and provides the similarities and the substantial equivalency of the rainbow™ Shine and the predicate.

| Device name                                   | rainbowTM Shine                                                                                                                                    | RAINBOW BLOCK                                                                                                                                   | Comparison                                                                                                               |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                  | Genoss Co., Ltd.                                                                                                                                   | Genoss Co., Ltd.                                                                                                                                | Same                                                                                                                     |
| 510(k) Number                                 | New Device                                                                                                                                         | K092513                                                                                                                                         | -                                                                                                                        |
| Materials                                     | Tosoh Powder                                                                                                                                       | Tosoh Powder                                                                                                                                    | Same                                                                                                                     |
| Form                                          | Preformed block                                                                                                                                    | Preformed block                                                                                                                                 | Same                                                                                                                     |
| Type, class of<br>dental ceramic              | Type II - Class 5                                                                                                                                  | Type II - Class 6                                                                                                                               | The subject and predicate<br>devices satisfy dental<br>ceramic classification<br>requirement by<br>ISO 6872:2008         |
| Sterilization                                 | Non-sterile                                                                                                                                        | Non-sterile                                                                                                                                     | Same                                                                                                                     |
| Indication for<br>use                         | rainbowTM Shine is used in<br>the manufacture of a dental<br>core through milling by<br>machine (MAD/ MAM or<br>CAD/CAM) followed by<br>sintering. | Rainbow Block is used in<br>the manufacture of a dental<br>core through milling by<br>machine (MAD/MAM or<br>CAD/CAM) followed by<br>sintering. | Same                                                                                                                     |
| Use                                           | Prescription                                                                                                                                       | Prescription                                                                                                                                    | Same                                                                                                                     |
| Technical characteristics                     |                                                                                                                                                    |                                                                                                                                                 |                                                                                                                          |
| Bending<br>Strength<br>(Flexural<br>strength) | 706MPa                                                                                                                                             | 1144MPa                                                                                                                                         | Bending strength is<br>higher than required by<br>ISO 6872:2008 for<br>Class 5 dental ceramics.<br>( > 500 Mpa )         |
| Sintering Density<br>(g/cm3)                  | 6                                                                                                                                                  | 6.00~6.04                                                                                                                                       | Same                                                                                                                     |
| Radioactivity<br>(Bq/g)                       | < 1.0 Bq/g                                                                                                                                         | < 1.0 Bq/g                                                                                                                                      | Same                                                                                                                     |
| Chemical<br>Solubility<br>(µg/cm²)            | 3                                                                                                                                                  | 14.47                                                                                                                                           | Chemical solubility<br>satisfies requirement by<br>ISO 6872:2008 for<br>Class 5 dental ceramics.<br>(< 2000 µg/cm²)      |
| Biocompatibility                              | Biocompatible                                                                                                                                      | Biocompatible                                                                                                                                   | Same                                                                                                                     |
| Shape                                         | Single<br>Image: Single cylinder with diameter and height labeled                                                                                  | Image: Single cylinder                                                                                                                          | Similar<br>The subject and predicate                                                                                     |
|                                               | 2~9<br>bridge<br>Image: Rectangular prism with width, depth, and height labeled                                                                    | Image: Rectangular prism                                                                                                                        |                                                                                                                          |
|                                               | 10~16<br>bridge<br>Image: Curved shape with width, depth, and height labeled                                                                       | Image: Curved shape                                                                                                                             | devices are available in<br>numerous shapes and sizes<br>in order to be compatible<br>with multiple milling<br>machines. |
|                                               | Disk<br>Image: Disk with diameter and height labeled                                                                                               | Image: Disk                                                                                                                                     |                                                                                                                          |

## Comparison of Characteristics

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GEN" are in a dark green color, while the letters "OSS" are in a lighter green color. The font is sans-serif and has a modern look.

# 510(K) Number: K151842

{6}------------------------------------------------

## 510(K) Number: K151842

Image /page/6/Figure/2 description: The image shows a table with three rows and three columns. The first column is labeled "Shape" and contains the text "Step block". The second column contains images of white, cylindrical objects with varying diameters and heights, specifically diameters of 95mm and 98mm, and heights ranging from 10mm to 28mm. The third column is mostly empty, with only a dash in the last row.

There are two minor differences that are worth discussing:

- 1) The subject and predicate device have slight difference in materials. The ZrO2 and Y2O3 are major materials of general zirconia blank. The differences in the composition between the subject and predicate device relate to the percentages of Y2O3 and they have effect on the crystal structure and consequently on the translucency of the device. SiO2, Fe2O3, Na2O and Er2O3 are trace elements used as colorants, but they have no material effects on the biocompatibility of the device. The non-clinical tests(bench test and biocompatibility test) demonstrate the Technical characteristics.
- 2) The differences of technological characteristics(Bending strength, chemical solubility) are within what is expected of this type of device. The rainbow Shine has a lower bending strength than the predicate device, due to the fact that the material is completely stabilized with Y2O3 in order to achieve higher translucency. However, the performance test results satisfy the requirement by ISO 6872:2008

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for GENOSS. The word "GENOSS" is written in a sans-serif font. The letters "GENO" are in a dark green color, while the letters "SS" are in a lighter green color. The logo is simple and modern.

## 7. Summary of non-clinical testing

Non-clinical device testing was conducted to confirm the performance of the subject device. Bench testing was conducted in accordance with the FDA recognized consensus standard(Recognition number : 4-178: ISO 6872 Third edition 2008-09-01, dentistry - ceramic materials) Bench tests for performance comparison of the subject device and the predicate device includes the following testing:

- Bending Strength - Linear thermal expansion coefficient - Chemical Solubility
Technical characteristics of both devices satisfy the requirements by ISO 6872:2008. Slight differences between the subject and predicate devices do not raise any new issues of safety and effectiveness.

Biocompatibility testing was conducted on the device pursuant to the ISO 10993-1:2009 Biological evaluation of medical device - Part 1: Evaluation and testing within a risk management process.

- Cytotoxicity test(ISO 10993-5)
- Irritation or intracutaneous reactivity(ISO 10993-10)
- Sensitization(ISO 10993-10)
- Acute systemic toxicity(ISO 10993-11)
- Genotoxocity(ISO 10993-03)

The result of biocompatibility testing demonstrated that no issue of biocompatibility arises.

#### 8. Conclusion

Based on the information provided in this premarket notification of GENOSS Co., Ltd. Concludes that rainbow™ Shine is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K151842](https://fda.innolitics.com/device/K151842)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).