← Product Code [FMF](/productcode/FMF) · K151777

# Intradermal Adapter (K151777)

_West Pharmaceutical Services, Inc. · FMF · Sep 29, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151777

## Device Facts

- **Applicant:** West Pharmaceutical Services, Inc.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Sep 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K151777 · Sep 29, 2015 | Intradermal Adapter | West Pharmaceutical Services, Inc. | Peer-reviewed clinical literature; Clinical cohort data (485 pediatric patients) | The sponsor used data from peer-reviewed clinical studies involving 485 pediatric patients to demonstrate that the device's geometry maintains appropriate injection depth (≤0.75mm) for the expanded pediatric patient population. | Pediatric population; Injection depth; Clinical literature; Intradermal injection |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Peer-reviewed clinical journal data; Clinical cohort study (retrospective/observational) | 485 pediatric patients; Sample Size: 485 | Not applicable for this study | Injection depth control (preventing depth >0.75mm) |

## Indications for Use

The Intradermal Adapter is an accessary to a 1ml, ½ inch fixed-needle allergy syringe indicated for use on patients ranging from children 2 years old to adults, as a guide for performing intradermal injections.

## Device Story

Intradermal Adapter is a single-use, injection-molded polycarbonate accessory for 1mL allergy syringes with 1/2 inch (27g-29g) fixed needles. Device snaps onto syringe barrel to act as a physical guide; controls needle insertion depth and angle for intradermal injections via Mantoux technique. Used by clinicians in clinical settings. Limits needle exposure to ensure injection remains within epidermal/dermal layers. Benefits include consistent injection depth and reduced risk of improper administration across pediatric (2+ years) and adult populations.

## Clinical Evidence

Leveraged bench, animal, and simulated use data from predicate K123588. Included review of peer-reviewed medical literature involving 485 pediatric patients. Data confirms device geometry limits injection depth to ≤0.75mm, ensuring placement within the epidermis/dermis layer for pediatric patients aged 2 years and older.

## Technological Characteristics

Material: Medical grade polycarbonate. Principle: Physical guide for needle depth/angle control. Energy: Manual. Form factor: Snap-fit accessory for 1mL syringe. Sterilization: Ethylene Oxide (SAL 10^-6). Single-use.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Intradermal Adapter ([K123588](/device/K123588.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, line-art style.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2015

West Pharmaceutical Services Inc. Mr. Jeffrev Ravel Director, Regulatory Affairs 530 Herman O. West Drive Exton, Pennsylvania 19341

Re: K151777

Trade/Device Name: Intradermal Adapter Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: June 30, 2015 Received: July 1, 2015

Dear Mr. Ravel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151777

Device Name

Intradermal Adapter

#### Indications for Use (Describe)

The Intradermal Adapter is an accessary to a 1ml, ½ inch fixed-needle allergy syringe indicated for use on patients ranging from children 2 year to adults, as a guide for performing intradermal injections.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

Device: Intradermal Adapter

#### Company Name:

West Pharmaceutical Services, Inc. 530 Herman O. West Drive Extron, PA 19341

Contact Person: Jeffrey Ravel Director of Regulatory Affairs Phone: 610-594-3482 Fax: 610-717-0668 E-mail: Jeffrey.Ravel@westpharma.com

Preparation date: 29 September 2015

#### Classification:

| Classification name:  | Syringe, Piston (Accessory)      |
|-----------------------|----------------------------------|
| Trade name:           | Intradermal Adapter              |
| Common/usual name:    | Intradermal Adapter (ID Adapter) |
| Product Code:         | FMF                              |
| Regulation No.:       | 21 CFR 880.5860                  |
| Class:                | II                               |
| Panel Identification: | General Hospital                 |

Predicate Devices: K123588 Intradermal Adapter

#### Device Description:

The Intradermal Adapter consists of a single injection molded part manufactured from a medical grade polycarbonate. The Intradermal Adapter is a piston syringe accessory which is designed

{5}------------------------------------------------

## West Intradermal Adapter

for use with 1mL allergy syringes with ½ inch (27g), (28g), (29g) needles, which are commonly used for intradermal injections given with a traditional Mantoux technique.

The Intradermal Adapter has been designed to snap fit on to the end of the syringe forming an injection guide to control the depth and angle of needle insertion exposing only the minimal needle length required to perform a successful intradermal injection. The Intradermal Adapter is provided sterile in an individually packaged configuration.

## Indication for Use:

The Intradermal Adapter is an accessary to a 1ml, ½ inch fixed-needle allergy syringe indicated for use on patients ranging from children 2 years old to adults, as a guide for performing intradermal injections.

## Technological Characteristics and Substantial Equivalence:

The Intradermal Adapter utilized as an accessory to the piston syringe has the same purpose and principle of operation as the predicate device with respect to performing an injection below the surface of the skin including intradermally and is therefore substantially equivalent to the cleared device referenced: K123588 (West Pharmaceutical Services - Intradermal Adapter).

| Areas for comparison          | Claimed Substantially<br>Equivalent Product West<br>Intradermal Adapter<br>K123588                                                                                        | Proposed Device<br>Intradermal Adapter<br>Subject 510(k)                                                                                                                                                                    | Comparison                                                                                                                                                                                                                |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use            | The Intradermal Adapter is<br>an accessory to a 1 ml, ½<br>inch fixed- needle allergy<br>syringe indicated for use as<br>a guide for performing<br>intradermal injections | The Intradermal Adapter is<br>an accessory to a 1ml, ½<br>inch fixed- needle allergy<br>syringe indicated for use<br>on patients ranging from<br>infants to adults, as a guide<br>for performing intradermal<br>injections. | Modified<br>Only patient population<br>has changed. The form<br>and function of the<br>proposed device has not<br>been modified from the<br>predicate device in order to<br>incorporate an expanded<br>patient population |
| Sterilization Method          | Sterile Ethylene Oxide<br>SAL 10-6                                                                                                                                        | Sterile Ethylene Oxide<br>SAL 10-6                                                                                                                                                                                          | Identical                                                                                                                                                                                                                 |
| Single use                    | Yes                                                                                                                                                                       | Yes                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                 |
| Control of Injection<br>Depth | Design of Intradermal<br>Adapter limits the<br>exposed needle of the<br>compatible piston                                                                                 | Design of Intradermal<br>Adapter limits the<br>exposed needle of the<br>compatible piston                                                                                                                                   | Identical                                                                                                                                                                                                                 |

{6}------------------------------------------------

## West Intradermal Adapter

## Traditional 510(k)

West Pharmaceutical Services, Inc.

| Areas for comparison                       | Claimed Substantially<br>Equivalent Product West<br>Intradermal Adapter<br>K123588                           | Proposed Device<br>Intradermal Adapter<br>Subject 510(k)                                                     | Comparison |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------|
|                                            | syringe to control depth<br>and angle of injection                                                           | syringe to control depth<br>and angle of injection                                                           |            |
| Material                                   | Polycarbonate                                                                                                | Polycarbonate                                                                                                | Identical  |
| Type of Injection and<br>needle gauge size | Intradermal Compatible<br>with 27-29 gauge<br>needles                                                        | Intradermal Compatible<br>with 27-29 gauge<br>needles                                                        | Identical  |
| Delivery form and<br>dosing system         | Manual, using syringe<br>with Intradermal<br>Adapter (Utilizes<br>standard syringe-vial<br>dosing system)    | Manual, using syringe<br>with Intradermal<br>Adapter (Utilizes<br>standard syringe vial<br>dosing system)    | Identical  |
| Compatible Syringe<br>type                 | Adapter fits to piston<br>syringe barrel - 1 ml<br>disposable, ½ inch, 27g<br>to 29g fixed needle<br>syringe | Adapter fits to piston<br>syringe barrel - 1 ml<br>disposable, ½ inch, 27g<br>to 29g fixed needle<br>syringe | Identical  |
| Delivery form and<br>dosing system         | Manual, using syringe<br>with Intradermal<br>Adapter (Utilizes<br>standard syringe-vial<br>dosing system)    | Manual, using syringe<br>with Intradermal<br>Adapter (Utilizes<br>standard syringe vial<br>dosing system)    | Identical  |
| Syringe supply                             | Separate from the device                                                                                     | Separate from the device                                                                                     | Identical  |

## Performance Testing:

Performance testing including bench, animal, and simulated use was leveraged from the predicate K123588 submission. The data collected throughout the course of these studies included bevel up and bevel down configurations using 27g to 29g needles as compared to the Mantoux method. Information gathered from medical literature obtained through peer reviewed journal articles was obtained and used to support expanding the indications for use as the injection depth is properly controlled by the geometry of the ID Adapter indicating no compromise to the intended depth of penetration for the expanded patient population. Results of performance testing demonstrated that the Intradermal Adapter, when used as an accessory to the piston syringe, can be indicated for administering intradermal injections to the indicated patient population.

The peer reviewed studies presented in this submission confirm through the acquisition of data from 485 pediatric patients, that the controlled design of the ID Adapter prevents an injection

{7}------------------------------------------------

## West Intradermal Adapter

depth to greater than 0.75mm which is well within the epidermis + dermis layer necessary for successful intradermal injection in a pediatric patient age group. Based on the clinical journal data presented in this submission in conjunction with the visual references of injection application and physical limitations of the device, the ID Adapter is acceptable for use in expanded intended use applications for a wider patient age range including pediatric patients over the age of 2 years.

## Conclusion:

Comparative analysis of the technological characteristics between the proposed and predicate device and results of verification testing performed demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K151777](https://fda.innolitics.com/device/K151777)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
