EzSensor Soft, EzSensor Soft i, EzSensor Bio, and EzSensor Bio i
Applicant
Rayence Co., Ltd.
Product Code
MUH · Dental
Decision Date
Jul 19, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for dagnostic use by dentists.
Device Story
Digital dental intraoral sensor system; acquires intraoral x-ray images using CMOS technology. Sensor captures x-ray photons, converts to electronic impulses; transmits via USB 2.0 interface to computer. Used in dental clinics by dentists/staff. Viewer software (Easydent or EzDent-i) manages patient info, image capture, display, and manipulation (windowing, measurement, editing). Provides rapid image acquisition compared to film; ergonomic design for patient comfort. Output used by dentists for diagnostic decision-making; enables lower radiation dose than film.
Clinical Evidence
Bench testing only. Performance evaluated via DQE, MTF, and linear response to x-ray exposure, demonstrating equivalence to the predicate. Electrical, mechanical, and environmental safety verified per IEC 60601-1; EMC testing per IEC 60601-1-2. No clinical data presented.
Technological Characteristics
CMOS-based digital intraoral sensor; amorphous silicon and fluorescent materials. USB 2.0 connectivity. Available in sizes 1.0, 1.5, 2.0. Thickness 5mm. Operates with Easydent or EzDent-i viewer software. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for dentists to collect, store, view, and manipulate dental x-ray images for diagnostic purposes in patients requiring intraoral dental radiography.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
Predicate Devices
EzSensor Soft [Alternative name : EzSensor Bio] (K143753)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2015
Rayence Co. Ltd. % Mr. David Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K151707
Trade/Device Name: EzSensor Soft, EzSensor Soft i, EzSensor Bio, and EzSensor Bio i (Model: 1.0, 1.5, 2.0) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: June 23, 2015 Received: June 24, 2015
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151707
#### Device Name
EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i, Model: 1.0, 1.5, 2.0
Indications for Use (Describe)
EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for dagnostic use by dentists.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Rayence logo. The logo consists of the word "Rayence" in red, block letters. To the right of the word is a graphic of two overlapping circles in shades of orange and yellow. The logo is simple and modern.
ail. Hwaseong-si.
Date: June 23, 2015
To: Department of health & Human Services, Food and Drug Administration
# Special 510(k): Device Modification
## EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i Model: 1.0, 1.5, 2.0
### Reference : K143753, EzSensor Soft [Alternative name- EzSensor Bio] Decision Date : March 10, 2015
The Rayence Co., Ltd. hereby submits this Special 510(k) for EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i, (model: 1.0, 1.5, 2.0) to inform the following device modifications to the predicate device, EzSensor Soft or EzSensor Bio.
1) New device model names.
2) New Imaging Viewer Software Version Update
3) New sensor size 1.5
We believe these modifications are eligible for the Special 510(k) process since they have the same fundamental scientific technology and intended use as the predicate device. We consider our intent to market this device as confidential commercial information and requests that it be treated as such by FDA. We have taken precautions to protect the confidentiality of the intent to market these devices. We understand that the submission to the government of false information is prohibited by 18 U.S.C. 1001 and 21 U.S.C. 331(q).
Thank you in advance for your consideration of our application. If there are any questions, please feel free to contact me at +82-31-8015-6459.
Sincerely yours,
Kee Dock Kim, RA Team manager Rayence Co., Ltd
{4}------------------------------------------------
## 1. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510K summary prepared: June 23, 2015
Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | Rayence Co., Ltd. |
|----------------------------------------------------|---------------------------------------------------------|
| Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-8015-6459 |
| Contact person: | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459 |
| Official Correspondent:<br>(U.S. Designated agent) | Dave Kim (davekim@mtech-inc.net) |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone: | +713-467-2607 |
| Fax: | +713-583-8988 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary Name: | EzSensor Soft, EzSensor Soft i, EzSensor Bio,<br>EzSensor Bio I,<br>Model: 1.0, 1.5, 2.0 |
|-------------------------|------------------------------------------------------------------------------------------|
| Regulation Number : | 21 CFR 872.1800 |
| Regulation Name : | Extraoral source x-ray system |
| Regulatory Class: | II |
| Product Code: | MUH |
| Predicate Device : | |
| Manufacturer | : Rayence Co., Ltd. |
| Device | : EzSensor Soft [Alternative name : EzSensor Bio] |
{5}------------------------------------------------
| 510(k) Number | : K143753 (Decision Date – March 10, 2015 |
|-------------------|-------------------------------------------|
| Regulation Number | : 21 CFR 872.1800 |
| Regulation Name | : Extraoral source x-ray system |
| Regulatory Class | : II |
| Product Code | : MUH |
#### 510(k) Submission - EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i
## 2. Device Description
EzSensor Soft, EzSensor Soft i, EzSensor Bio i are digital dental intraoral sensors which acquire digital intra oral images. Direct digital systems acquire images with a bendable sensor that is connected to a computer to produce an image almost instantaneously following exposure. The primary advantage of direct sensor systems is the image acquisition speed. For patient comfort, the ergonomic design is based on human intraoral anatomy.
- Excellent image quality based on advanced CMOS technology
- · Sensor's ergonomic shape is more comfortable for the human oral structure
- · Lower dose exposure (compared to film sensor)
- Enhanced durability
- · Easy-to-use USB interface
## 3. Indication for use
EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.
### 4. Summary of Design Control Risk management
The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.
## 5. Summary of the technological characteristics of the device compared to the predicate device
EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensor described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, EzSensor Soft [Alternative name : EzSensor Bio] of Rayence Co., Ltd.
{6}------------------------------------------------
### 5. 1 Separate model name
The following table lists the new devise model names and their respective hardware and software combination.
| | Model Name | Component |
|-----------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Predicate | EzSensor Soft<br>(alternate name : EzSensor Bio)<br>Size: 1.0, 2.0 | Intra Oral Sensor + Easydent (Viewer Software)<br>Intra Oral Sensor + EzDent-i 1.0 (Viewer<br>Software, K131594) |
| Proposed | EzSensor Soft<br>(alternate name : EzSensor Bio)<br>Size: 1.0, 1.5, 2.0 | Intra Oral Sensor + Easydent (Viewer Software) |
| Proposed | EzSensor Soft i<br>(alternate name : EzSensor Bio i)<br>Size: 1.0, 1.5, 2.0 | Intra Oral Sensor + EzDent-i 2.0 (Viewer<br>Software, K150747) |
### 5. 2 Characteristic
| Characteristic | Proposed<br>Rayence Co., Ltd. | Predicate<br>Rayence Co., Ltd. |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Image: EzSensor Soft, EzSensor Soft i, EzSensor Bio and EzSensor Bio i | Image: EzSensor Soft [Alternative name : EzSensor Bio] digital dental image processing system |
| 510(k) number | - | K143753 |
| Indications<br>for use | EzSensor Soft, EzSensor Soft i,<br>EzSensor Bio and EzSensor Bio i<br>Digital Dental Intra Oral Sensors are<br>intended to collect dental x-ray<br>photons and convert them into<br>electronic impulses that may be<br>stored, viewed and manipulated for<br>diagnostic use by dentists. | EzSensor Soft [Alternative name :<br>EzSensor Bio] Digital Dental Intra<br>Oral Sensor is intended to collect<br>dental x-ray photons and<br>convert them into electronic impulses<br>that may be stored, viewed and<br>manipulated for diagnostic use by<br>dentists. |
| Sensor Dimension(mm)<br>(±10%) | Size "1.0": 37.8 x 26.6<br>Size "1.5": 40.8 x 30.6<br>Size "2.0": 44.0 x 32.5 | Size "1.0": 37.5 x 26.5<br>Size "2.0": 43.5 x 32.5 |
| Sensor Thickness(mm) | 5 | 5 |
| Active Area(mm) | Size "1.0": 20.01 x 30.01<br>Size "1.5": 23.98 x 33.00<br>Size "2.0": 25.99 x 35.99 | Size "1.0": 20 x 30<br>Size "2.0": 25.99 x 35.99 |
| USB Module | Integrated USB 2.0 module | Integrated USB 2.0 module |
{7}------------------------------------------------
| Pixel<br>Pitch (μm) | Full Resolution | 14.8 | 14.8 |
|-----------------------------|-----------------|-------------------------------------------------|-------------------------------|
| | Binning mode | 29.6 | 29.6 |
| DQE<br>84.64 µGy<br>6 lp/mm | Full Resolution | 0.070 | 0.070 |
| | Binning mode | 0.070 | 0.070 |
| MTF<br>84.64 µGy<br>6 lp/mm | Full Resolution | 0.154 | 0.154 |
| | Binning mode | 0.133 | 0.133 |
| Typical dose range(µGy) | | Incisor & Canine : 300 ~ 500 / Molar: 400 ~ 600 | |
| Viewer Software* | | Easydent or EzDent-i(K150747) | Easydent or EzDent-i(K131594) |
510(k) Submission - EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i
This device is operated using viewer software. With the identical hardware configuration, the intraoral sensor model name is distinguished by the type and version of image viewing software available. The software functions include the patient information management, image capture, and viewer for captured images. Both EzDent-i 1.0 and 2.0 versions have the same indications for use and functionalities like operation software, computer platform, picture archiving and communication format, image format, image processing features, windowing, image edit, measurements and manipulation. The differences are that the EzDent-i 2.0 (K150747) has additional features such as "Hide the patient information", "Export to acquisition information", "Create New Report" and "Print Setting".
In addition, a new sensor size 1.5 is available to the existing sensor sizes 1.0 and 2.0 of the predicate device.
These differences do not raise the questions of safety or effectiveness. Based on the laboratory testing results submitted in this 510K, we conclude that the subject device is substantially equivalent to the predicate device.
### 6. Summary of Performance Testing
The intended use, application and detector type of EzSensor Soft i, EzSensor Soft i, EzSensor Bio, EzSensor Bio i are identical to the predicate device. Both the subject and predicate device use the same amorphous silicon alley and materials for fluorescent as the sensing means. The performance test result indicates that the EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i detector performed equally the predicated device; EzSensor Soft [Alternative name : EzSensor Bio] and no additional safety risk is identified in the bench test: Non-clinical report.
Non-clinical according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-
{8}------------------------------------------------
ray Imaging Devices" was performed.
The DQE, MTF and linear response to X-ray exposure test demonstrated that the subject sensor performed equivalently compared to the predicate device with the same dynamic range.
Electrical, mechanical, environmental safety and performance testing were performed according to standard IEC 60601-1: 2005 + CORR.1(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance). EMC testing was conducted in accordance with standard IEC 60601-1-2:2007.
## 7. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Rayence Co., Ltd. concludes that EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i is safe and effective and substantially equivalent to predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.