← Product Code [MNH](/productcode/MNH) · K151695

# Romeo posterior osteosynthesis system (K151695)

_Spineart · MNH · Sep 10, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151695

## Device Facts

- **Applicant:** Spineart
- **Product Code:** [MNH](/productcode/MNH.md)
- **Decision Date:** Sep 10, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Romeo® posterior osteosynthesis system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, non-pedicle screw fixation system, Romeo® posterior osteosynthesis system is intended for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

## Device Story

Posterior spinal fixation system; includes polyaxial/monoaxial/spondylolisthesis screws, setscrews, rods (Titanium/CoCr), and connectors. Used by surgeons in clinical settings to immobilize/stabilize spinal segments during fusion procedures. Components implanted to provide structural support for thoracic, lumbar, and sacral spine instabilities/deformities. System provides mechanical fixation; surgeon selects components based on patient anatomy. Benefits include spinal stabilization and fusion support.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing included torsion tests for setscrews and axial push-down/torsion tests for connectors per ASTM F1798, demonstrating performance comparable to predicate devices.

## Technological Characteristics

Materials: Titanium alloy Ti6Al4V ELI (ISO 5832.3, ASTM F136). Components: Polyaxial/monoaxial/spondylolisthesis screws, setscrews, rods, connectors. Sterilization: Gamma radiation. Mechanical testing: ASTM F1798. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Ellipse posterior osteosynthesis system ([K081165](/device/K081165.md))
- Romeo® posterior osteosynthesis system ([K111127](/device/K111127.md))
- Romeo® posterior osteosynthesis system ([K141835](/device/K141835.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 10, 2015

SPINEART Mr. Franck Pennesi Director of Industry & Ouality International Center Cointrin 20 route de pré-bois. CP 1813 1215 Geneva 15 Switzerland

Re: K151695

Trade/Device Name: Romeo posterior osteosynthesis system Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: June 22, 2015 Received: June 23, 2015

Dear Mr. Pennesi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. Franck Pennesi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

K151695

Page 1 of 1

510(k) Number (if known) K151695

#### Device Name

Romeo® posterior osteosynthesis system

#### Indications for Use (Describe)

Romeo® posterior osteosynthesis system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

When used as a posterior, non-cervical, non-pedicle screw fixation system, Romeo® posterior osteosynthesis system is intended for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

 Commercial Use (If marked, SEE Additional Instructions)
 

 Any Other use covered by 37 CFR
 
|X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

TRADITIONAL 510k

ROMEO® Posterior Osteosynthesis System

## 510(k) SUMMARY

| Submitted by                          | SPINEART                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                       | International Center Cointrin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                       | 20 route de pré-bois                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                       | CP1813                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                       | 1215 GENEVA 15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                       | SWITZERLAND                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Contacts                              | Franck PENNESI Director of Industry & Quality                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                       | Phone : +41 22 570 1246 Fax : +41 22 799 40 26                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                       | Mail : fpennesi@spineart.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                       | Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                       | Mail : idrubaix@nordnet.fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Date Prepared                         | August 17th 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name                           | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade Name                            | Romeo® posterior osteosynthesis system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name                   | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Class                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Product Code                          | MNH, MNI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| CFR section                           | 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device panel                          | ORTHOPEDIC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Legally marketed<br>predicate devices | Primary predicate: Ellipse posterior osteosynthesis system (K081165) manufactured by<br>Spineart Other predicate: Romeo® posterior osteosynthesis system (K111127, K141835)                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                       | manufactured by Spineart                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for use                   | Romeo® posterior osteosynthesis system is intended to provide immobilization and<br>stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the<br>treatment of the following acute and chronic instabilities or deformities of the thoracic,<br>lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra;<br>degenerative spondylolisthesis with objective evidence of neurologic impairment;<br>fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion<br>(pseudarthrosis). |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                       | When used as a posterior, non-cervical, non-pedicle screw fixation system, Romeo®<br>posterior osteosynthesis system is intended for the following indications: degenerative<br>disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc<br>confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or<br>dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor;<br>pseudoarthrosis; and failed previous fusion                                                   |

{4}------------------------------------------------

| Description of the device        | The Romeo posterior fixation system comprises polyaxial screws, monoaxial screws,<br>spondylolisthesis screws, setscrew, straight and pre-bent Titanium or CoCr rods, and<br>various connectors. The Romeo screws come in various lengths (from 25 to 90 mm) and<br>diameters (4.0, 4.5, 5.0, 5.5, 6.0, 7.0 and 8.0 mm) to accommodate different patient<br>anatomies. Romeo components are delivered sterile (gamma sterilization) in a dedicated<br>packaging or not sterile upon request. The Romeo posterior fixation system is supplied<br>with a set of surgical instruments. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | The purpose of this 510(k) submission is to obtain clearance for a Romeo Setscrew with a<br>modified hexalobe imprint, a Romeo Parallel Rod Connector and a T Rod Connector.<br>These Romeo components are made of Titanium alloy Ti6Al4V ELI conforming to ISO<br>5832.3 and ASTM F136 and are fully compatible with components of the existing Romeo posterior fixation system.                                                                                                                                                                                                   |
| Discussion of Testing            | The following non-clinical tests were conducted: Torsion test for the set-screw according<br>to in-house protocol; Axial Push Down and Axial Torsion in subassembly according to<br>ASTM F1798 for the connectors. Results demonstrate comparable mechanical properties<br>to the predicate devices.                                                                                                                                                                                                                                                                                |
| Conclusion                       | Design comparisons and non-clinical performance testing demonstrate that the added<br>devices (Setscrew, T connector and Parallel Rod Connector O/O) are substantially<br>equivalent to their predicate devices (K081165, K111127 and K141835) in terms of<br>intended use, material, design, mechanical properties and function.                                                                                                                                                                                                                                                   |

---

**Source:** [https://fda.innolitics.com/device/K151695](https://fda.innolitics.com/device/K151695)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
