Leksell GammaPlan

K151666 · Elekta Instrument AB · MUJ · Sep 3, 2015 · Radiology

Device Facts

Record IDK151666
Device NameLeksell GammaPlan
ApplicantElekta Instrument AB
Product CodeMUJ · Radiology
Decision DateSep 3, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning.

Device Story

Leksell GammaPlan is a computer-based treatment planning system for Leksell Gamma Knife stereotactic radiosurgery/radiotherapy. It processes inputs from neurosurgeons, radiation therapists, and physicists to calculate dosimetry for precise radiation delivery to defined volumes. New features include stereotactic reference definition from stand-alone Cone Beam CT (CBCT), mask fixation support with automatic patient position correction via CBCT, and guided treatment delivery evaluation. The system enables clinicians to plan, evaluate, and verify radiation dose delivery, facilitating targeted treatment of intracranial or other lesions while sparing surrounding healthy tissue.

Clinical Evidence

No clinical testing on patients was required. Evidence consists of bench testing, including module, integration, and system-level verification, regression testing, and design/usability validation performed by qualified personnel to ensure requirements and user needs were met.

Technological Characteristics

Computer-based treatment planning system for stereotactic radiosurgery. Features include CBCT integration for stereotactic referencing and patient position correction. Software-based dosimetry calculation and treatment planning. Operates as a workstation-based system for clinical use.

Indications for Use

Indicated for use by neurosurgeons, radiation therapists, and radiation physicists for planning the dosimetry of stereotactic radiosurgery and stereotactic radiation therapy treatments using the Leksell Gamma Knife.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 3, 2015 Elekta Instrument AB % Ms. Matilda Forsberg Regulatory Affairs Engineer Kungstensgatan 18 Stockholm, 10393 SWEDEN Re: K151666 Trade/Device Name: Leksell GammaPlan Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: June 14, 2015 Received: June 19, 2015 Dear Ms. Forsberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. ## 510(k) Number (if known) K151666 Device Name Leksell GammaPlan® Indications for Use (Describe) Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary As Required by 21 CFR 807.92(c) 510 (k) Summary #### Subscribers Name & Address 1. Elekta Instrument AB Kungstensgatan 18, P.O. Box 7593 SE-103 93 Stockholm, Sweden Tel: (011) 46 8 587 254 00 Fax: (011) 46 8 587 255 00 Official Correspondent: Ms Louise Wachtmeister Date summary prepared: 2015-06-10 #### 2. Trade Name Leksell GammaPlan® #### 3. Device Classification | Common Name | Product Code | Class | Regulation Number | |--------------------|--------------|-------|-------------------| | Leksell GammaPlan® | MUJ | II | 21 CFR 892.5050 | #### 4. Predicate Device Identification | Legally marketed devices to which equivalence is being claimed | 510(k) # | |----------------------------------------------------------------|----------| | Leksell GammaPlan® | K103093 | #### 5. Other relevant submissions | Devices | 510(k) # | |--------------------|----------| | Leksell GammaPlan® | K090972 | #### 6. Device Description Leksell GammaPlan® is designed for use with the Leksell Gamma Knife and is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgery and stereotactic radiation therapy. It processes the inputs of health care professionals (e.g. neurosurgeons, radiation therapists, radiation physicists) such that the desired radiation dose is delivered by the Leksell Gamma Knife to a precisely defined volume. {4}------------------------------------------------ #### Summary of Clinical Testing 7. Clinical testing on patients was not required to demonstrate substantial equivalence or safety and effectiveness of the device. #### 8. Summary of Non Clinical and Performance testing Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against requirement specification. Regression and re-test of unchanged functionalities in the developed system was done to ensure that the new and updated functionalities did not introduce any undesirable effects. Design and usability validation of the system has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs. The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) for the new functionality was effective. Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification and user needs have been met and safety & effectiveness have been achieved. #### 9. Intended Use Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning. #### 10. Technological Characteristics Leksell GammaPlan® is designed for use with the Leksell Gamma Knife and is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgery and stereotactic radiation therapy. The fundamental functionality and technical characteristics of the device are similar to those of the predicate device K103093. #### 11. Substantial Equivalence The functionality for Leksell GammaPlan® is equivalent to its predicate device Leksell GammaPlan® (K103093) in safety and effectiveness. The following workflows have been added to the new version: - Stereotactic reference can be defined from stand-alone Cone Beam CT ● (СВСТ) - Mask fixation support with automatic patient position correction from CBCT - Guided treatment delivery evaluation after CBCT with review of delivery dose {5}------------------------------------------------ The indication for use is the same as its predicate device and the fundamental technical characteristics are similar.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...