← Product Code [DQY](/productcode/DQY) · K151638

# Phenom 17 Catheter, Phenom 21 Catheter, Phenom 27 Catheter, Phenom Plus Catheter (K151638)

_Cathera, Inc. · DQY · Nov 13, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151638

## Device Facts

- **Applicant:** Cathera, Inc.
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Nov 13, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Phenom™ Catheters are intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the neuro, peripheral, and coronary vasculatures.

## Device Story

Phenom™ Catheter is a variable stiffness, single-lumen catheter designed for accessing small, tortuous vasculature. Device features an outer lubricious coating and an internal liner to facilitate movement of interventional devices. Distal tip includes radiopaque markers for visualization under fluoroscopy. Used by physicians in clinical settings to navigate neuro, peripheral, and coronary vessels. Device acts as a conduit for delivering therapeutic/diagnostic agents or introducing other interventional devices (e.g., coils, stents). Performance is monitored by the physician via fluoroscopic imaging. Benefits include improved access to complex vascular anatomy for minimally invasive procedures.

## Clinical Evidence

Bench testing only. Testing included dimensional inspection, material verification, accessibility/trackability, device compatibility, shaft stiffness, chemical compatibility, tip shapeability, kink resistance, hub fitting, dead space volume, corrosion resistance, tensile strength, leakage testing, burst pressure, particulate testing, coating lubricity/durability, flow rate, flexural fatigue, torque strength, radiopacity, biocompatibility, and shelf-life.

## Technological Characteristics

Variable stiffness, single-lumen catheter. Features lubricious outer coating and internal liner. Distal tip includes radiopaque markers. Sterilized via Ethylene Oxide per ISO 11135. Mechanical design optimized for trackability in tortuous vasculature.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Prowler® 14 Infusion Catheter ([K021591](/device/K021591.md))
- Prowler® Plus Infusion Catheter ([K993266](/device/K993266.md))
- Marksman™ Catheter ([K091559](/device/K091559.md))
- Tracker® Infusion Catheter ([K862117](/device/K862117.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one in front of the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13, 2015

Cathera, Inc. Victor Ham Chief Regulatory & Quality Officer 627 National Ave. Mountain View, California 94043

Re: K151638

Trade/Device Name: Phenom™ Catheters Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 9, 2015 Received: October 13, 2015

Dear Mr. Ham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

M.A. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151638

Device Name Phenom™ Catheter

Indications for Use (Describe)

The Phenom™ Catheters are intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the neuro, peripheral, and coronary vasculatures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(c).

### Submitted by:

Cathera, Inc. 627 National Ave. Mountain View, CA 94043 Tel.: (650) 388-5088 Fax: (650) 390-0107

Contact Person: Victor Ham

Date summary prepared: May 29, 2015

Trade Name: Phenom™ Catheter

Common Name: Catheter

| Classification Name & Product Codes: Continuous Flush Catheter |                                     |
|----------------------------------------------------------------|-------------------------------------|
|                                                                | (21 CFR 870.1210, Product Code KRA) |

Predicate Devices: Prowler® 14 Infusion Catheter (510K # K021591) Prowler® Plus Infusion Catheter (510K # K993266) Marksman™ Catheter (510K # K091559) Tracker® Infusion Catheter (510k # K862117)

# Device Description:

The Phenom™ Catheters are variable stiffness, single lumen catheters designed to access small, tortuous vasculature. They are available in a variety of lengths, stiffness and inner and outer diameters. The outer surface of the catheter is coated to enhance navigation in the vessel. The catheter also incorporates a liner to facilitate movement of introduction devices passing through its lumen. The distal tip has radiopaque marker(s) to aid visualization and positioning under fluoroscopy.

### Indications for Use:

The Phenom™ Catheters are intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the neuro, peripheral, and coronary vasculatures.

# Performance Data:

{4}------------------------------------------------

The following bench testing was performed in support of the Phenom™ Catheter and to establish substantial equivalence to the predicate devices:

- Dimensional Inspection (OD, ID, Length, Distal Tip Configuration) ●
- Material Verification ●
- . Accessibility/ Trackability
- Device Compatibility (with Guide Catheter, Guide Wire, RHV) ●
- Shaft Stiffness ●
- Chemical Compatibility (Saline, Contrast Medium) .
- Tip Shapeability ●
- Kink Resistance ●
- Conical Fitting for Hub ●
- Dead Space Volume .
- Corrosion Resistance
- Tensile Strength of Catheter (Body & Hub Attachment) ●
- Liquid Leakage at Air Leakage During Aspiration Hub ●
- Burst Pressure ●
- Particulate Testing
- Outer Surface Coating Lubricity and Durability ●
- Catheter Flow Rate ●
- Flexural Fatigue ●
- Torque strength
- Tip Mark Radiopacity testing ●
- Biocompatibility Testing ●
- Shelf-life Testing ●

{5}------------------------------------------------

# Substantial Equivalence Determination

The information presented in the 510k shows that the Phenom™ Catheters are substantially equivalent to predicate endovascular catheters in regards to the following aspects:

| Design:        | The subject and predicate devices are substantially equivalent with<br>respect to design characteristics. The slight variations in flexibility,<br>length of coated segments and lubricity are what differentiate these<br>catheters. Each manufacturer optimizes these design variations<br>towards a more specific application (e.g. infusion of diagnostic and<br>therapeutic agents) or for the introduction of a specific devices such<br>as embolic agents, coils and stents. |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Function:      | The subject and predicate devices are substantially equivalent with<br>respect to functional characteristics.                                                                                                                                                                                                                                                                                                                                                                       |
| Manufacturing: | The subject and predicate devices are similar with respect to<br>technological manufacturing processes.                                                                                                                                                                                                                                                                                                                                                                             |
| Materials:     | The subject and predicate devices are composed of similar<br>materials, all of which have an extensive clinical history of safe use<br>in medical devices.                                                                                                                                                                                                                                                                                                                          |
| Indications:   | The subject and predicate devices maintain similar indications.                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Packaging:     | The subject and predicate devices utilize similar packaging<br>configurations.                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterilization: | The subject and predicate devices are both sterilized utilizing an<br>Ethylene Oxide sterilization cycle validated in accordance with<br>ISO 11135 - Medical Devices - Validation and Routine Control of<br>Ethylene Oxide Sterilization.                                                                                                                                                                                                                                           |
| Labeling:      | Both the subject and predicate devices have similar labeling.                                                                                                                                                                                                                                                                                                                                                                                                                       |

Testing was conducted to show that no new risks were identified and that the performance profile is similar to well-established predicate devices cleared for the market. Evaluation was performed in the more complex and higher risk neurovascular anatomy, which is the worst case representation of the cardiac and peripheral vascular anatomies.

---

**Source:** [https://fda.innolitics.com/device/K151638](https://fda.innolitics.com/device/K151638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
