K151629 · Quanta System Spa · GEX · Oct 27, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151629
Device Name
Guardian 1000
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Guardian 1000 with its delivery accessories is intended to emit energy in the visible and infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle. Guardian 1000 with its delivery accessories is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gastroenterology, neurosurgery, general surgery, genitourinary surgery (urology), thoracic surgery, gynecology, pulmonology, ophthalmology, orthopedics, otolaryngology (ENT) and podiatry.
Device Story
Guardian 1000 is a 980 nm GaAlAs infrared laser surgical system; delivers energy via quartz optical fibers and various handpieces. Console includes power electronics, cooling system (Peltier cells/fans), and touch screen interface for controlling laser emission and source temperature. Used in clinical settings by physicians for soft tissue surgery (ablation, vaporization, excision, incision, coagulation) and topical heating for pain relief. Physician operates device via console; monitors output via interface. Laser energy elevates tissue temperature (40-45°C) for therapeutic effect or performs precise tissue interaction. Benefits include minimally invasive surgical capability and temporary pain/stiffness relief.
Clinical Evidence
Bench testing only. Thermal performance testing on 3 subjects at 2 physical locations (shoulder/knee) demonstrated the device maintains a therapeutic topical temperature range of 40-45°C for over 10 minutes, achieving a >5°C increase within 1 minute. Electrical safety and EMC testing confirmed compliance with IEC 60601-1, IEC 60601-2-22, and IEC 60601-1-2 standards.
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, and coagulation across various surgical specialties (gastroenterology, neurosurgery, general surgery, urology, thoracic, gynecology, pulmonology, ophthalmology, orthopedics, ENT, podiatry) or for topical heating to relieve minor muscle/joint pain, stiffness, arthritis, or muscle spasm and increase local circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with a design above them that resembles a bird or abstract shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ouanta System SPA Mr. Francesco Dell'Antonio Compliance Manager Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058
October 27, 2015
Re: K151629 Trade/Device Name: Guardian 1000 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ILY, PDZ Dated: September 25, 2015 Received: September 28, 2015
Dear Mr. Dell' Antonio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 809]); medical device reporting
{1}------------------------------------------------
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S.
For Director
Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151629
# Device Name
Guardian 1000
#### Indications for Use (Describe)
Guardian 1000 with its delivery accessories is intended to emit energy in the visible and infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle.
Guardian 1000 with its delivery accessories is indicated for use in surgical applications requiring the ablation. vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gastroenterology, neurosurgery, general surgery, genitourinary surgery (urology), thoracic surgery, gynecology, ophthalmology, orthopedics, otolaryngology (ENT) and podiatry.
# Gastroenterology
Hemostasis of esophageal varices; palliation of malignant dysphagia; palliative ablation of obstructive neoplasms; hemostasis in colonoscopy.
### Neurosurgery
Tumors adjacent to the spinal cord; tumors adjacent to the cortex.
#### General Surgery
Ablation, vaporization, excision, and coagulation of soft tissue in general surgery including endoscopic and open procedures. Applications include but are not limited to: Laparoscopic: appendectomy, cholecystection. Open: mastectomy, reduction mammoplasty, breast biopsy, rectal and anal hemorrhoidectomy, bowel resection, colectomy, cholecystectomy, liver resection, condyloma, thyroidectomy, cavernous hemangioma.
#### Genitourinary (Urology)
Transurethral: transurethral incision of the prostate (TUIP), bladder neck incisions, urethral strictures, exterior sphincterotomy, laparoscopic lymphadenectomy. Open: condyloma, circumcision, benign and malignant lesions of external genitalia.
#### Thoracic Surgery
Pulmonary resection, coagulation of blebs and bullae, adhesiolysis, pericardiectorny, mediastinal and thoracic lesions and abnormalities, mediastinal lymph node dissection, hemostasis: thoracotomy,
# Gynecology (GYN)
Laparoscopic excision/lysis of adhesions, endometrial lesions, including ablation of endometriosis, laparoscopic assisted hysterectomy (LAVH), laser uterosacral nerve ablation (LUNA); myomectomy; ovarian drilling, tubal fimbrioplasty, appendectomy. Open: conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia VIN, VAIN, condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowen's disease, (Erithroplasia of Queyrat) and Bowenoid papulosa (BP) lesions. Intrauterine: Fibroids/ polyps/adhesions; Resection of septum.
Pulmonology Tracheal bronchial lesions.
{3}------------------------------------------------
#### Ophthalmology
Oculoplastics, open DCR, endo-nasal DCR, tumor excision and biopsy, eyelid reconstruction; blepharoplasty.
Orthopedics Dissect and coagulate.
Otolaryngology (ENT)
Nasal/Sinus: turbinectomy and turbinate reduction; polypectomy of nose and nasal passages; ethmoidectomy; meatal antrostomy;
Laryngo-tracheal: removal of vocal cord/ fold nodules, polyps and cysts; arytenoidectomy; tracheal stenosis; Oropharyngeal: uvulopalatoplasty (LAUP, laser UTPP); tonsillar cryptolysis, neoplasma) and tonsil; hemi glossectomy;
Head & Neck: tumor resection on oral, sub facial and neck tissues; parathyroidectomy;
Podiatry
Matrixectomy, Periungual and subungual warts, Plantar warts, Neuromas.
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T mentagrophytes, and /or yeasts Candida albicans, etc.).
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5. 510(K) SUMMARY
### Introduction:
This document contains the 510(k) Summary for the Guardian 1000 laser device. The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058 |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Francesco Dell'Antonio<br>Compliance Manager<br>Quanta System SPA<br>Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | June 4th 2015 |
| Device Name: | Guardian 1000 |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX, ILY, PDZ |
# Predicate Devices:
Guardian 1000 is substantially equivalent to the following legally marketed predicate devices:
- K110375 Gold Series - Blueshine srl (Italy)
- K101893 Nexus 30- USA Laser Biotech Inc. (USA) .
{5}------------------------------------------------
# Description of the device:
Guardian 1000 delivers an invisible laser beam in the infrared spectrum at 980 nm using a gallium aluminum arsenide (GaAlAs) source. The laser light beam is carried to the focusing lens on the handpiece probe by quartz optical fibers.
Guardian 1000 consists of two main sub-systems: a laser console and its delivery accessories, as follows: The laser console includes:
- i an external enclosure
- । power electronics
- touch screen display with control electronics which controls:
- o power electronics
- O user interface
- o laser source temperature
- laser source emission o
- . a laser system including:
- O laser source
- cooling system (Peltier cells, dissipator and fans) o
- fiber launching system o
- aiming beam o
Laser radiation is delivered to the patient trough an optical fiber and several handpieces depending on the treatment.
The optical fiber is connected to the laser unit trough a SMA905 connector
#### Intended use
Guardian 1000 with its delivery accessories is intended to emit energy in the visible and infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle.
Guardian 1000 with its delivery accessories is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gastroenterology, neurosurgery, general surgery, genitourinary surgery (urology), thoracic surgery, gynecology, pulmonology, ophthalmology, orthopedics, otolaryngology (ENT) and podiatry.
#### Gastroenterology
Hemostasis of esophageal varices; palliation of malignant dysphagia; palliative ablation of obstructive neoplasms; hemostasis in colonoscopy.
#### Neurosurgery
Tumors adjacent to the spinal cord; tumors adjacent to the cortex.
#### General Surgery
Ablation, vaporization, excision, and coagulation of soft tissue in general surgery including endoscopic and open procedures. Applications include but are not limited to: Laparoscopic: appendectomy, cholecystectomy, bowel resection. Open: mastectomy, reduction mammoplasty,
{6}------------------------------------------------
breast biopsy, rectal and anal hemorrhoidectomy, bowel resection, cholecystectomy, liver resection, condyloma, thyroidectomy, thoracotomy; cavernous hemangioma.
# Genitourinary (Urology)
Transurethral: transurethral incision of the prostate (TUIP), bladder tumors, bladder neck incisions, urethral strictures, exterior sphincterotomy, laparoscopic lymphadenectomy. Open: condyloma, circumcision, benign and malignant lesions of external genitalia.
# Thoracic Surgery
Pulmonary resection, coagulation of blebs and bullae, adhesiolysis, pericardiectomy, mediastinal and thoracic lesions and abnormalities, mediastinal lymph node dissection, hemostasis; thoracotomy.
# Gynecology (GYN)
Laparoscopic excision/lysis of adhesions, endometrial lesions, including ablation of endometriosis, laparoscopic assisted hysterectomy (LAVH), laser uterosacral nerve ablation (LUNA); myomectomy; ovarian cystectomy, ovarian drilling, tubal fimbrioplasty, appendectomy. Open: conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia VIN, VAIN, condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowen's disease, (Erithroplasia of Queyrat) and Bowenoid papulosa (BP) lesions. Intrauterine: Fibroids/ polyps/adhesions; Resection of septum.
# Pulmonology
Tracheal bronchial lesions.
# Ophthalmology
Oculoplastics, open DCR, endo-nasal DCR, tumor excision and biopsy, evelid reconstruction; blepharoplasty.
# Orthopedics
Dissect and coagulate.
# Otolaryngology (ENT)
Nasal/Sinus: turbinectomy and turbinate reduction/ablation; polypectomy of nose and nasal passages; ethmoidectomy; meatal antrostomy;
Laryngo-tracheal: removal of vocal cord/ fold nodules, polyps and cysts; arytenoidectomy; tracheal stenosis;
Oropharyngeal: uvulopalatoplasty (LAUP, laser UTPP); tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil; hemi glossectomy;
Head & Neck: tumor resection on oral, sub facial and neck tissues; parathyroidectomy;
# Podiatry
Matrixectomy, Periungual and subungual warts, Plantar warts, Neuromas.
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T mentagrophytes, and /or yeasts Candida albicans, etc.).
# Comparison of Technological Characteristics:
{7}------------------------------------------------
Guardian 1000 has the same technological characteristics (energy source, control mechanism) and specifications as its predicate devices.
### Performance data:
The following performance data are provided in support of the substantial equivalence determination:
#### Safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Guardian 1000 device and its delivery accessories.
The system complies with the IEC 60601-1, IEC 60601-2-22 standards for safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The Guardian 1000 device is capable of achieving therapeutic temperature range of 40 - 45 degrees centigrade as accepted by the FDA. An increase in topical heating of the tissue level by at least 5 degrees centigrade was reached within one (1) minute as demonstrated in the bench testing. The therapeutic temperature range was maintained for over 10 minutes testing time. The temperatures versus time measurements were conducted on 3 subjects at 2 physical locations (shoulder and knee areas). The preexposed topical skin temperature ranged from 31.8 to 34.3 centigrade degrees. The topical temperature during exposure following brief stabilization time ranged from 40.8 to 43.9 centigrade degrees. These data demonstrate the System meets the generally accepted topical temperature range for therapeutic heat of 40 - 45 degrees centigrade during the recommended treatment time of 10 minutes.
# Comparison of Intended Use:
Guardian 1000 device's Intended Use combines the Intended Uses of its predicate devices.
#### Substantial Equivalence:
The Guardian 1000 device and its delivery system is as safe and effective as its predicate devices. The Guardian 1000 device and its delivery system has the same intended use, technological characteristics and specifications as its predicate devices.
Thus, Guardian 1000 device and its delivery system is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.