K151628 · Medline Industries, Inc. · PIO · Jul 24, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K151628
Device Name
Medline ENFit Connectors
Applicant
Medline Industries, Inc.
Product Code
PIO · Gastroenterology, Urology
Decision Date
Jul 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The Medline ENFit Transition Connector is intended for connecting an enteral feeding set with an ENFit connector to a gastronomy tube with funnel. The Medline ENFit G-Tube Connector is intended for connecting an enteral feeding set equipped with a stepped enteral distal end to a gastronomy tube equipped with an ENFit connector.
Device Story
The Medline ENFit Transition Connector and Medline ENFit G-Tube Connector are single-use adapters designed to bridge the compatibility gap between legacy enteral feeding devices and new ISO 80369-compliant ENFit connectors. The devices function as physical adapters; one end features an ENFit connector (female or male depending on the model) and the other features a legacy stepped distal tip or funnel port. These adapters are used in clinical settings to ensure continuity of enteral feeding when legacy and new-standard equipment must be interconnected. By facilitating these connections, the device prevents interruptions in patient nutrition during the transition period of adopting uniform ENFit standards. The device is operated by healthcare providers during the setup of enteral feeding systems.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including enteral connector misconnection assessment, human factors validation, FMEA, bioburden, metrology, shelf-life, leachables/extractables, biocompatibility (ISO 10993-5, ISO 10993-10), and ISO 80369-20 performance testing (leakage, stress cracking, axial load, unscrewing, and overriding).
Technological Characteristics
Non-sterile, single-use adapters made of non-DEHP PVC. Features rigid ENFit connectors designed per AAMI/CN3:2014 (PS) and AAMI/CN20:2014 (PS) standards. Mechanical interface design facilitates connection between legacy stepped distal tips/funnel ports and ENFit-compliant components.
Indications for Use
Indicated for patients requiring enteral feeding who need to connect legacy non-ISO 80369-1 compliant enteral feeding devices to new AAMI/CN3:2014 (PS) compliant ENFit enteral devices.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Cedic ENFit Transition Connectors for Enteral Applications (K140581)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24, 2015
Medline Industries, Inc. Stephanie Blair Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060
Re: K151628
> Trade/Device Name: The Medline ENFit G-Tube Connector, Medline ENFit Transition Connector Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIO Dated: June 12, 2015 Received: June 17, 2015
Dear Stephanie Blair,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K151628
Device Name The Medline ENFit Transition Connector
Indications for Use (Describe)
The Medline ENFit Transition Connecting an enteral feeding set with an ENFit connector to a gastronomy tube with funnel.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|| | Prescription Use (Part 21 CFR 801 Subpart D)
|__| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K151628
#### Device Name The Medline ENFit G-Tube Connector
#### Indications for Use (Describe)
The Medline ENFit G-Tube Connector is intended for connecting an enteral feeding set equipped with a stepped enteral distal end to a gastronomy tube equipped with an ENFit connector.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="2,2 14,2 14,14 2,14" style="fill:none;stroke:black;stroke-width:3"></polygon> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in bold, sans-serif font, and a white starburst-like symbol is above the text. The starburst has four points, with two pointing upwards and two pointing downwards.
edline Industries. Inc One Medline Place Mundelein, IL 60060
510(k) Premarket Notification Medline ENFit Transition Connector Medline ENFit G-Tube Connector
# TRADITIONAL 510(k) SUMMARY [AS REQUIRED BY 21 CFR 807.92(a)]
### SUMMARY PREPARATION DATE
July 21st, 2015
### SUBMITTER / 510(k) SPONSOR
Medline Industries, Inc. One Medline Place Mundelein, IL 60060 Registration Number: 1417592
## CONTACT PERSON
Stephanie Blair Regulatory Affairs Specialist Phone: 847-643-3690 Email: SBlair(@medline.com
### TYPE OF SUBMISSION
Traditional 510(k)
#### DEVICE NAME / CLASSIFICATION
- Trade/Proprietary Name(s):
- Common Name:
- . Product Code:
- Classification Name:
- Device Class: ●
- Classification Regulation:
- Classification Panel:
Medline ENFit Transition Connector Medline ENFit G-Tube Connector Transition Connectors for Enteral Applications PIO Gastrointestinal Tube and Accessories Class II 21 CFR 876.5980 Gastroenterology and Urology
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, block letters on the left side of the logo. To the right of the word is a white, stylized cross symbol.
ledline Industries. Inc One Medline Place Mundelein, IL 60060
510(k) Premarket Notification Medline ENFit Transition Connector Medline ENFit G-Tube Connector
## PREDICATE DEVICE
Cedic ENFit Transition Connectors for Enteral Applications (K140581) are identified herein as the predicate device of the Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector. Please note that there are multiple design configurations of Cedic ENFit Transition Connectors for Enteral Applications cleared under K140581; therefore:
- . The specific Cedic ENFit Transition Connector for Enteral Applications cleared under K140581 that is selected for comparison with the Medline ENFit Transition Connector is: The Cedic Enteral Distal End ENFit Transition Connector.
- . The specific Cedic ENFit Transition Connector for Enteral Applications cleared under K140581 that is selected for comparison with the Medline ENFit G-Tube Connector is: The Cedic Enteral Funnel ENFit Transition Connector.
#### INDICATIONS FOR USE
The Medline ENFit Transition Connector is intended for connecting an enteral feeding set with an ENFit connector to a gastronomy tube with funnel.
The Medline ENFit G-Tube Connector is intended for connecting an enteral feeding set equipped with a stepped enteral distal end to a gastronomy tube equipped with an ENFit connector.
#### DEVICE DESCRIPTION
In an effort to mitigate the risk of misconnections associated with enteral feeding devices, manufacturers have recently began to adopt a uniform connector for enteral devices, called an ENFit connector, which is based upon the ISO 80369 series of standards and is in accordance with FDA Final Guidance for Industry: Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications (February 11, 2015). This uniform ENFit connector provides a locking feature that signals the appropriate connection to stay in place. In addition, the ENFit connector features a female mating component end for enteral feeding sets and a male mating component end for gastronomy feeding tubes, which allows for the interconnection of these devices.
However, it will take time for enteral devices with this uniform ENFit connector to become widely available and fully adopted across healthcare settings, as existing enteral devices with previouslydesigned "legacy" connectors, which are not compliant with ISO 80369-1, still remain in inventory. Therefore, to prevent an interruption or delay in patient care when an enteral device with an
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in bold, white font and is stacked on top of a white star-like symbol. The star symbol has four points, with the top and bottom points being longer than the side points.
ledline Industries. One Medline Place Mundelein, IL 60060
510(k) Premarket Notification Medline ENFit Transition Connector Medline ENFit G-Tube Connector
AAMI/CN3:2014 (PS) compliant ENFit connector must be connected to an existing enteral device with a non-AAMI/ANSI/ISO 80369-1 compliant connector, a temporary need has emerged for an adapter that can facilitate connection between enteral devices with these incompatible end connectors.
The Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector address this temporary need and facilitate enteral-specific connections between AAMI/CN3:2014 (PS) compliant enteral connectors and non-AAMI/ANSI/ISO 80369-1 compliant connectors to ensure that patient enteral feeding is not impeded when an enteral feeding device with an AAMI/CN3:2014 (PS) compliant ENFit connector must be connected to a previously-designed, existing enteral feeding device with a non-AAMI/ANSI/ISO 80369-1 compliant end connector.
The Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector, which are described in further detail below, allow specific connections between enteral feeding devices with AAMI/CN3:2014 (PS) compliant ENFit connectors (also known as "PGLock" connectors) and enteral feeding devices with non-AAMI/ANSI/ISO 80369-1 compliant, legacy connectors that are still available for use in healthcare settings. To facilitate this connection, the Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector are both designed with the mating component of an AAMI/CN3:2014 (PS)-compliant ENFit connector at one end and the mating component of a previously-used, legacy connector at its other end.
Specifically, the two designs of ENFit transition adapters that are within the scope of this 510(k) Premarket Notification submission are:
#### 1.) The Medline ENFit Transition Connector
The Medline ENFit Transition Connector allows fitment of newly-released enteral feeding sets with AAMI/ CN3:2014 (PS) compliant ENFit connectors to existing gastronomy tubes with non-AAMI/ ANSI/ISO 80369-1 compliant end connectors, referred to as "funnel feeding ports," which are still available for use in healthcare settings. Specifically, to facilitate the interconnection between newlyreleased feeding sets, which are equipped with female ENFit connectors, and the funnel feeding ports of previously-designed gastronomy tubes, this adapter provides the mating component of an ENFit female connector, designed and tested in accordance with AAMI/ CN3:2014 (PS) and AAMI/CN20:2014 (PS), at one of its ends, and the mating component of a non-AAMI/ANSV/ISO 80369-1 compliant funnel feeding port, a stepped Distal Tip connector, at its other end.
#### 2.) The Medline ENFit G-Tube Connector
The Medline ENFit G-Tube Connector allows fitment of newly-released gastronomy tubes (or "Gtubes") with AAMI/CN3:2014 (PS) compliant ENFit connectors to previously-designed enteral feeding sets with non-AAMI/ANSI/ISO 80369-1 compliant connectors, which still may be available for use in healthcare settings. To facilitate the compatibility between gastronomy tubes with ENFit male
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, bold, sans-serif font, stacked on top of a white, stylized starburst shape. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.
ledline Industries. In One Medline Place Mundelein, IL 60060
510(k) Premarket Notification Medline ENFit Transition Connector Medline ENFit G-Tube Connector
connectors and the stepped Distal Tip connector of existing enteral feeding sets, the Medline ENFit G-Tube Connector provides the mating component of an ENFit male connector, designed and tested based upon AAMI/ CN3:2014 (PS) and AAMI/CN20:2014 (PS), at one of its ends, and the mating component of a stepped Distal Tip connector, a funnel feeding port, at its other end.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector are made of non-DEHP PVC and are both non-sterile, single-use devices. In addition, the ENFit connector featured on both configurations of the proposed device is made of rigid materials to reduce the likelihood of forced fits between flexible connectors that are not intended to connect with each other, as recommended in FDA Final Guidance for Industry: Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications (February 11, 2015).
Additionally, the Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector are similar in design, intended use, function and internal and external characteristics to the predicate device cleared under K140581. Cedic ENFit Transition Connectors for Enteral Applications. Both the proposed and predicate devices operate by providing a means of interconnecting incompatible enteral feeding device end-fittings together, so that patient enteral feeding can take place when enteral feeding devices with AAMI/CN3:2014 (PS) compliant ENFit connectors must be connected to enteral feeding devices with previously-designed, non-AAMI/ANSI/ISO 80369-1 compliant enteral connectors. To facilitate this connection, both the proposed and predicate devices act as an adapter by providing the mating component of an AAMI/CN3:2014 (PS) compliant ENFit connector at one end and the mating component of a previously-used, legacy connector at its other end.
### SUMMARY OF PERFORMANCE TESTING
Non-clinical verification of the Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector has been conducted to evaluate the safety, performance and functionality of these devices. The results of all testing were ultimately found to support a substantial equivalence determination.
Specifically, the proposed Medline ENFit Transition Connector and Medline ENFit G-Tube Connector were evaluated through the following tests:
- Enteral Connector Misconnection Assessment Study
- Human Factors Validation Study: Enteral Connectors Final Report ●
- PG-Lock Misconnection Data with Failure Modes and Effects Analysis (FMEA) ●
- Internal Risk Analysis
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medline. The logo is set against a blue square background. The word "MEDLINE" is written in white, block letters, stacked on top of a white starburst symbol. The starburst has four points, two pointing horizontally and two pointing vertically.
edline Industries. Inc One Medline Place Mundelein, IL 60060
510(k) Premarket Notification Medline ENFit Transition Connector Medline ENFit G-Tube Connector
- Bioburden Testing ●
- Metrology Testing
- Stability (Shelf-Life) Testing
- Leachables and Extractables (L&E) Analysis with Toxicological Health Risk Assessment
- Biocompatibility Testing:
- Cytotoxicity MEM elution per ISO 10993-5 O
- Irritation Intracutaneous reactivity per ISO 10993-10 O
- Delayed-Type Hypersensitivity (Sensitization) Guinea Pig Maximization Test per о ISO 10993-10
- ISO 80369-20 Testing: ●
- Positive Pressure Liquid Leakage O
- Stress Cracking O
- Resistance to Separation from Axial Load O
- Resistance to Separation from Unscrewing O
- Resistance to Overriding O
- Disconnection by Unscrewing O
#### SUMMARY OF CLINICAL TESTING
Not applicable.
#### SUMMARY OF ANIMAL TESTING
Not applicable.
#### CONCLUSION
In accordance with 21 CFR Part 807, and based on a comparison of 'Indications for Use,' technological characteristics and performance data. Medline Industries, Inc. concludes that the proposed Medline ENFit Transition Connector and Medline ENFit G-Tube Connector are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.