K151617 · Surgical Instrument Service and Savings Inc.(Dba Medline Ren · NLH · Jan 22, 2016 · Cardiovascular
Device Facts
Record ID
K151617
Device Name
Livewire Electrophysiology Catheter
Applicant
Surgical Instrument Service and Savings Inc.(Dba Medline Ren
Product Code
NLH · Cardiovascular
Decision Date
Jan 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Device Story
Reprocessed diagnostic electrophysiology catheter; used for electrogram recording and cardiac stimulation during diagnostic EP studies. Placed at high right atrium, right ventricular apex, His bundle, and coronary sinus. Features steerable design with user-adjustable tip curve and various electrode spacings. Used by clinicians in clinical settings. Reprocessed a maximum of two times; marked and retired after cycle limit. Device provides electrical interface for cardiac mapping; output used by physician to diagnose arrhythmias.
Steerable diagnostic electrophysiology catheter with multiple electrode spacings. Features include adjustable tip curve, shaft/handle flexibility, and electrical continuity. Reprocessed using validated cleaning and ethylene oxide sterilization processes. Maximum of two reprocessing cycles allowed.
Indications for Use
Indicated for patients requiring evaluation of cardiac arrhythmias via endocardial and intravascular electrogram recording and cardiac stimulation.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
St. Jude Medical Livewire Electrophysiology Catheter (K022380)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized human figure in profile, with three overlapping heads suggesting a sense of community and interconnectedness.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2016
Medline ReNewal Brandi Panteleon Director, Quality Assurance/Regulatory Affairs Medline ReNewal 2747 SW 6th St Redmond. Oregon 97756
Re: K151617
Trade/Device Name: Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: December 17, 2015 Received: December 18, 2015
Dear Brandi Panteleon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Arthell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the Medline Renewal logo. The word "Renewal" is written in a combination of green and blue colors, with "Re" in green and "newal" in blue. Below "Renewal" is the text "Full Circle Reprocessing" in a smaller font size. To the left of the text is the Medline logo, which is a blue star-like shape with the word "MEDLINE" written next to it.
Traditional 510(k) Notification Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter
## 4.0 Indications for Use
| | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|--------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K151617 | |
| TBD | | |
Device Name
Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter
Indications for Use (Describe)
The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medline Renewal Full Circle Reprocessing. The Medline logo is on the left, followed by the Renewal logo in green and blue. Below the Renewal logo is the text "Full Circle Reprocessing".
## 510(k) Summary – K151617
| Submitter/<br>Owner | Medline ReNewal<br>2747 SW 6th St.<br>Redmond, OR 97756 | | |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Names | Brandi Panteleon<br>Director, QA/RA<br>P: 541-923-3310<br>F: 541-923-3375<br>E: bpanteleon@medline.com | | |
| Date Prepared | June 19, 2015 | | |
| Device Names | Proprietary Name: Medline ReNewal Reprocessed Livewire<br>Electrophysiology Catheter<br>Common Name: Diagnostic Electrophysiology Catheter, Reprocessed | | |
| Classification | Classification: Class II<br>Regulation Number: 870.1220<br>Product Code: NLH | | |
| Predicate<br>Device | K022380 St. Jude Medical Livewire Electrophysiology Catheter | | |
| Device<br>Description | Medline ReNewal Reprocessed St. Jude Livewire Electrophysiology<br>Catheters are commonly placed at the high right atrium, right ventricular<br>apex and His bundle, and in the coronary sinus, and are used for<br>electrogram recording and cardiac stimulation during diagnostic<br>electrophysiology studies. | | |
| Statement of<br>Intended Use | Medline ReNewal Reprocessed Electrophysiology catheters can be used in<br>the evaluation of a variety of cardiac arrhythmias from endocardial and<br>intravascular sites. | | |
| Technological<br>Characteristics | The Medline ReNewal Reprocessed Electrophysiology Catheters contain<br>various electrode spacing and a steerable design with a tip that can be<br>shaped into a user-desired curve. The technological characteristics of the<br>proposed devices are substantially equivalent to the predicate devices<br>listed in this submission. The proposed devices are a reprocessed version<br>of the predicate devices. Only Medline ReNewal reprocesses the Medline<br>ReNewal Reprocessed St. Jude Livewire Electrophysiology Catheters.<br>Catheters are reprocessed a maximum of two times. Catheters are marked<br>and taken out of service after the maximum number of cycles is reached. | | |
| Performance<br>Testing | The functional characteristics of the subject device have been evaluated<br>and found to be substantially equivalent to the predicate device based on<br>the following tests:<br>Functional performance studies: | ●<br>O<br>O<br>O<br>O | simulated use and artificial soiling;<br>bond strength (tensile testing);<br>torsional strength;<br>leakage current;<br>steerable deflection and continuity; |
| | ○ | catheter/handle joint flexibility and continuity (flex fatigue); | |
| | ○ | electrode adherence; | |
| | ○ | applied force deflection (tip flexibility); | |
| | ○ | shaft flexibility; | |
| | ○ | tip buckling; | |
| | ○ | direct current resistance; | |
| | ○ | impedance at 5 kHz; and | |
| | ○ | corrosion resistance. | |
| ● | | Cleaning: | |
| | ○ | protein, total organic carbon, and endotoxins; | |
| | ○ | visual inspection under magnification; and | |
| | ○ | cleaning performance qualification. | |
| ● | | Biocompatibility: | |
| | ○ | cytotoxicity; | |
| | ○ | sensitization; | |
| | ○ | irritation; | |
| | ○ | acute systemic toxicity; | |
| | ○ | pyrogenicity; | |
| | ○ | hemocompatibility (hemolysis, thrombogenicity; and | |
| | | complement activation). | |
| ● | | Packaging and shelf life validation; sterilization validation: | |
| | ○ | bioburden testing; and | |
| | ○ | ethylene oxide and ethylene chlorohydrin residuals testing. | |
| ● | | Product stability | |
| | Based on comparisons of the indications for use, intended use, | | |
| Conclusion | technological characteristics, and performance data to the predicate<br>devices, Medline ReNewal Reprocessed St. Jude Medical Livewire<br>Electrophysiology Catheters are substantially equivalent to the predicate<br>device. | | |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline ReNewal logo. The word "ReNewal" is in a combination of green and blue, with "Re" in green and "Newal" in blue. Below "ReNewal" is the phrase "Full Circle Reprocessing" in a smaller, gray font. To the left of the text is the Medline logo, which is a stylized star.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.