← Product Code [KTT](/productcode/KTT) · K151580

# D.N.E. External Fixation System (K151580)

_Dne, LLC · KTT · Feb 16, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151580

## Device Facts

- **Applicant:** Dne, LLC
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Feb 16, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The D.N.E. Ring Fixator and its components are indicated for open and closed fracture fixation, pseudoarthrosis or nonumions of long bones, limb lengthening by epiphyseal or metaphyseal distraction or bony or soft tissue deformities, and correction of segmental or nonsegmental bony or soft tissue defects. The D.N.E. Ring Fixator is for use on all long bones including: tibia, fibula, femur, humerus, radius and ulna.

## Device Story

D.N.E. External Fixation System is a mechanical orthopedic device consisting of bone anchorage, bridge, and connection elements. It provides stable fixation of bone fragments, allowing axial loading and immediate range of motion for adjacent joints. The system is used by surgeons in clinical settings to treat long bone fractures, nonunions, and deformities. The subject device introduces an Angular Correction Clamp and associated Struts to facilitate ring size adjustments and enhance system stability and strength. The device is applied externally to the patient's limb.

## Clinical Evidence

Bench testing only. Static and dynamic axial compression testing was performed per ASTM F1541 to demonstrate mechanical equivalence to the predicate device.

## Technological Characteristics

Mechanical external fixation system comprising bone anchorage, bridge, and connection elements. Includes Angular Correction Clamp and struts. Testing performed per ASTM F1541 for static and dynamic axial compression.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- D.N.E. External Fixation System ([K113106](/device/K113106.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2016

DNE, LLC % Christine Scifert Official Correspondent Memphis Regulatory Consulting, LLC 3416 Roxee Run Cove Bartlett, Tennessee 38133

Re: K151580

Trade/Device Name: D.N.E. External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: January 20, 2016 Received: January 21, 2016

Dear Ms. Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K151580

Device Name

D.N.E. External Fixation System

Indications for Use (Describe)

The D.N.E. Ring Fixator and its components are indicated for open and closed fracture fixation, pseudoarthrosis or nonumions of long bones, limb lengthening by epiphyseal or metaphyseal distraction or bony or soft tissue deformities, and correction of segmental or nonsegmental bony or soft tissue defects. The D.N.E. Ring Fixator is for use on all long bones including: tibia, fibula, femur, humerus, radius and ulna.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## Special 510(k) Summary

D.N.E. External Fixation System February 3, 2016

- D.N.E., LLC Company: 2225 Park Place Drive Slatington, Pa. 18080 Christine Scifert Primary Contact: 3416 Roxee Run Cove Bartlett, TN 38133 scifert@memphisregulatory.com Trade Name: D.N.E. External Fixation System Common Name: External Fixation Frame Classification: Class II Regulation Number: 21 CFR 888.3030 Panel: Orthopedic Product Code: KTT Primary Predicate: K113106 - D.N.E. External Fixation System
## Device Description:

The D.N.E. External Fixation System assembly consists of three basic types of elements: 1) bone anchorage elements, 2) bridge elements, and 3) connection elements. The design allows freedom of pin placement, ease of assembly and stable fixation of bone fragments with the possibility of axial loading of the extremity and immediate range of motion of all adjacent joints.

An Angular Correction Clamp and associated Struts have been added to the system in order to facilitate a surgeons ability to step down in ring sizes.

## Indications for Use:

The D.N.E. External Fixation System and its components are indicated for open and closed fracture fixation, pseudoarthrosis or nonunions of long bones, limb lengthening by epiphyseal or 

{5}------------------------------------------------

metaphyseal distraction, correction or bony or soft tissue deformities, and correction of segmental or nonsegmental bony or soft tissue defects. The D.N.E. External Fixation System is for use on all long bones including: tibia, fibula, femur, humerus, radius and ulna.

#### Substantial Equivalence:

The D.N.E. External Fixation Angular Correction Clamp and associated Struts are substantially equivalent to the predicate plates, nuts, washers and rods previously cleared in K113106 in terms of intended use, mechanical safety and performances. The clamp and thicker struts add stability and strength to the system when compared with the existing components. Static and Dynamic Axial Compression testing was performed per ASTM F1541 and showed the subject device to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K151580](https://fda.innolitics.com/device/K151580)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
