← Product Code [GEX](/productcode/GEX) · K151569

# Tranberg CLS Laser Fiber (K151569)

_Clinical Laserthermia Systems, AB · GEX · Feb 12, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151569

## Device Facts

- **Applicant:** Clinical Laserthermia Systems, AB
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Feb 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The TRANBERGCLS Laser fiber is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology, and urology, at a wavelength of 1064nm.

## Device Story

The Tranberg CLS Laser fiber is an optical fiber accessory used to deliver laser energy from a compatible laser unit to a target tissue site. It features a 550-micron core, a radial diffusor, and an SMA 905 connector. The device is used by clinicians in various surgical specialties to perform interstitial laser therapy, resulting in tissue necrosis or coagulation. The fiber is delivered sterile for single-use. The clinician operates the laser system to control energy delivery, which is transferred through the fiber to the treatment location. The output is thermal energy, which creates a round lesion in the soft tissue. This allows for minimally invasive thermal ablation of target tissues, providing a therapeutic benefit for patients requiring localized tissue destruction.

## Clinical Evidence

No clinical or animal data were submitted. Substantial equivalence is supported by non-clinical engineering bench testing, including comparative temperature testing and biocompatibility assessments, confirming the device meets applicable safety and performance standards.

## Technological Characteristics

Optical fiber with 550-micron core, 0.22 numerical aperture, and 1mm radial diffusor. 3m length with SMA 905 connector. Operates at 1064nm wavelength in continuous wave mode. Maximum power 8W. Biocompatible materials. Single-use, sterile. Accessory to powered surgical laser instrument.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Visualase Cooled Laser Application System (VCLAS) ([K053087](/device/K053087.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, often associated with healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2016

Clinical Laserthermia Systems AB % Mr. David Makanani OMEDtech, LLC 1725 Signal Ridge Drive, Suite 150 Edmond, Oklahoma 73013

Re: K151569 Trade/Device Name: Tranberg CLS Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: January 4, 2016 Received: January 7, 2016

Dear Mr. Makanani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151569

Device Name TRANBERG CLS Laser fiber

#### Indications for Use (Describe)

The TRANBERG CLS Laser fiber is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the pericardial sac), dermatology, ear nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, and urology at a wavelength of 1064nm.

| Type of Use (Select one or both, as applicable)                                                   |                                                                                                  |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

| Date                                                                                                         | June 1, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER                                                                                                    | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381                                                                                                                                                                                                                                                                                                                                                                                             |
| CONTACT PERSON                                                                                               | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381<br>Tel: +4646152100<br>Email: lee@clinicallaser.se                                                                                                                                                                                                                                                                                                                                          |
| DEVICE NAME<br>Classification<br>Trade Name<br>Common Name<br>Classification<br>Product Code<br>Review Panel | Class II<br>TRANBERGCLS Laser fiber<br>TRANBERGCLS Laser fiber<br>21 CFR 878.4810<br>GEX - Powered Laser Surgical Instrument<br>General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                      |
| PREDICATE DEVICE:                                                                                            | K053087, Visualase Cooled Laser Application System (VCLAS)                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| INTENDED USE:                                                                                                | The TRANBERGCLS Laser fiber is indicated for use to necrotize or<br>coagulate soft tissue through interstitial irradiation or thermal therapy<br>in medicine and surgery in cardiovascular thoracic surgery (excluding<br>the heart and the vessels in the pericardial sac), dermatology, ear-nose<br>throat surgery, gastroenterology, general surgery, gynecology, head<br>and neck surgery, neurosurgery, plastic surgery, pulmonology,<br>radiology, and urology, at a wavelength of 1064nm. |

#### DEVICE DESCRIPTION:

The TRANBERG46|Laser fiber is used to transfer laser energy from the laser unit to the location for the treatment.

Clinical Laserthermia Systems, AB ● 223 81 Lund, Sweden

{4}------------------------------------------------

The laser fiber is an optical fiber with a core of 550 mic and radial diffusor. The length is 3m and it has a standard connector SMA 905 to fit the laser unit. The numerical aperture is at 0.22. The material in contact with human tissue is biocompatible.

The TRANBERGCS|Laser fiber is delivered sterile and for single use only.

### TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE:

The following table provides more detailed information regarding the basis for the determination of substantial equivalence:

| Laser fiber                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name                           | Tranberg CLS Laser fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Visualase Laser fiber LDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Manufacturer                           | Clinical LaserThermia Systems CLS,<br>Sweden                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | BioTex Inc., US                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended use /<br>Indications for use. | The Tranberg CS Laser fiber is indicated<br>for use to necrotize or coagulate soft<br>tissue through interstitial irradiation or<br>thermal therapy in medicine and surgery<br>in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in<br>the pericardial sac), dermatology, ear-<br>nose-throat surgery, gastroenterology,<br>general surgery, gynecology, head and<br>neck surgery, neurosurgery, plastic<br>surgery, pulmonology, radiology, and<br>urology, at a wavelength of 1064nm. | The LDF is indicated for use to necrotize<br>or coagulate soft tissue through<br>interstitial irradiation or thermal therapy<br>in medicine and surgery in cardiovascular<br>thoracic surgery (excluding the heart and<br>the vessels in the pericardial sac),<br>dermatology, ear-nose-throat surgery,<br>gastroenterology, general surgery,<br>gynecology, head and neck surgery,<br>neurosurgery, plastic surgery,<br>pulmonology, radiology, and urology, for<br>wavelengths 800nm through 1064nm. |
| Device Regulatory<br>Classification    | Accessory to powered surgical laser<br>instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Accessory to powered surgical laser<br>instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

Table 5-1 Equivalence Comparison

Clinical Laserthermia Systems, AB ● 223 81 Lund, Sweden

{5}------------------------------------------------

|                                                | FDA 878.4810                                              | FDA 878.4810                                              |
|------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| Product code                                   | GEX                                                       | GEX                                                       |
| Device class                                   | Accessory to powered surgical laser instrument<br>Class 2 | Accessory to powered surgical laser instrument<br>Class 2 |
| 510(k) No                                      |                                                           | K053087                                                   |
| Fiber core diameter:                           | $550 \mu m$                                               | $400 \mu m$ (200 – 1000)                                  |
| Numerical aperture:                            | 0.22                                                      | 0.37                                                      |
| Fiber length:                                  | 3 m                                                       | 3-12m standard                                            |
| Proximal connector:                            | SMA 905                                                   | SMA 905                                                   |
| Wavelength:                                    | 1064 nm                                                   | 532-1064nm                                                |
| Laser operation mode:                          | Continuous Wave                                           | Continuous Wave                                           |
| Diffusing region length:                       | 1 mm                                                      | 7,5-30mm                                                  |
| Diffusing tip assembly diameter:               | 1.55 mm                                                   | 0.6-1.4mm                                                 |
| Lesion Shape:                                  | Round                                                     | Ellipsoidal / Round                                       |
| Max power:                                     | 8 W for $550 \mu m$                                       | 8 W for $400\mu m$                                        |
| Lesion volume (refer to<br>Report DV-2015-024) | $0,8 cm^3$                                                | $0,8 cm^3$                                                |

# PERFORMANCE TESTING - (NON-CLINICAL) BENCH

The Tranberg 65 Laser fiber has been determined through engineering bench testing to support substantial equivalence with this device and the predicates. This testing showed the Tranberg CSS Laser fiber to meet applicable ISO, IEC and FDA safety and performance standards,

Non-clinical bench performance testing completed:

- Engineering comparative temperature testing ●
- Biocompatibility .

## PERFORMANCE TESTING - ANIMAL/CLINICAL

There are no animal or clinical data submitted with this Notification.

#### CONCLUSION:

Based on the results of non-clinical testing, the TRANBERG-55|Laser Fiber performs according to specifications, and as intended, and the comparative discussion of intended use, principle of operation, and technological characteristics, has determined that the Tranberg 955 Laser fiber is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K151569](https://fda.innolitics.com/device/K151569)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
