K151533 · Varian Medical Systems, Inc. · JAK · Aug 13, 2015 · Radiology
Device Facts
Record ID
K151533
Device Name
Respiratory Gating for Scanners
Applicant
Varian Medical Systems, Inc.
Product Code
JAK · Radiology
Decision Date
Aug 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
Respiratory Gating for Scanners (RGSC) is used to characterize the patient's respiratory patterns, providing the necessary information to diagnostic devices to acquire images synchronized with the breathing motion. RGSC can also be used to monitor the patient position during the image acquisition.
Device Story
RGSC tracks patient respiratory motion to synchronize diagnostic imaging (CT/PET-CT). System components: infrared camera, reflector block (placed on patient chest/abdomen), visual coaching device (VCD), and cabinet containing workstation/real-time units. Infrared camera tracks reflective markers on block; real-time unit processes motion data; workstation manages patient files, calibration, and session review. Operated by radiologic technicians/therapists under physician supervision in clinical imaging settings. VCD provides visual cues to guide patient breathing. Output: respiratory pattern data sent to imaging scanner to trigger gated acquisition. Benefits: reduced motion artifacts in images, improved diagnostic accuracy. System is non-sterile; reflector block is reusable.
Clinical Evidence
No clinical tests included. Evidence consists of bench testing, including biocompatibility (ISO 10993), electrical safety (AAMI ANSI ES 60601-1), EMC (IEC 60601-1-2), software verification/validation, and mechanical testing of mounting hardware.
Technological Characteristics
System includes infrared camera, reflector block (ABS 737 plastic), VCD, and workstation. Sensing: infrared tracking of reflective markers. Connectivity: wireless access point, 3rd party scanner interface. Software: major level of concern. Mounting: wall/ceiling camera, couch-mounted VCD. Sterilization: none (non-sterile).
Indications for Use
Indicated for patients undergoing CT or PET/CT imaging requiring respiratory-synchronized image acquisition or patient position monitoring.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Varian Medical Systems, Inc. % Mr. Peter J. Coronado Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
Re: K151533
Trade/Device Name: Respiratory Gating For Scanners Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 5, 2015 Received: June 8, 2015
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151533
Device Name Respiratory Gating for Scanners
### Indications for Use (Describe)
Respiratory Gating for Scanners (RGSC) is used to characterize the patient's respiratory patterns, providing the necessary information to diagnostic devices to acquire images synchronized with the breathing motion. RGSC can also be used to monitor the patient position during the image acquisition.
Type of Use (Select one or both, as applicable)
| <div> <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 3.375L5.75 10.875L2.75 7.875" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect fill="white" height="16" width="16"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Varian Medical Systems, Inc. Corporate Headquarters 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636
www.varian.com
# 510k Summary
# 510k Submission for Respiratory Gating for Scanners
A partner for life
As required by 21 CFR 807.92; Reference: FDA's Guidance Document "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]"
#### SUBMITTER l.
| Submitter's Name: | Varian Medical Systems<br>3100 Hansen Way, m/s E-110<br>Palo Alto CA 94304-1038 |
|-------------------|---------------------------------------------------------------------------------|
| Contact Name: | Peter J. Coronado |
| Position: | Director, Regulatory Affairs |
| Phone: | 1.650.424.6230 |
| Fax: | 1.650.646.9200 |
| Email: | submissions.support@varian.com |
| Date Prepared: | Friday, June 5, 2015 |
#### ll. DEVICE
| Name of Device: | Respiratory Gating for Scanners |
|----------------------|--------------------------------------------------|
| Common/Usual Name: | System, X-Ray, Tomography, Computed |
| Classification Name: | Computed Tomography X-ray System (21CFR892.1750) |
| Regulatory Class: | Class II |
| Product Code: | JAK |
#### PREDICATE DEVICE lll.
| Name of Predicate: | RPM Respiratory Gating System |
|--------------------|-------------------------------|
| 510(k) Number: | K102024 |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
Respiratory Gating for Scanners (RGSC) consists of the RGSC Cabinet (Containing Workstation Unit & Real-Time Unit). Infrared Camera, Reflector Block, and Visual Coaching Device. RGSC is used in the breathing-synchronized acquisition of images on CT and/or PET CT Scanners. RGSC is an accessory to these systems. RGSC is operated primarily by radiologist technicians and/or radiotherapists, in accordance with the prescription of radiologist or a radiation oncologist and under the general supervision of a chief technologists and/or medical physicists.
The infrared camera tracks the position and motion of the reflector block, which is placed on the patient's chest or abdomen during this process. This light, plastic block has four reflective markers which face the direction of the reflector block is in transitory contact with the patient's skin for a limited amount of time (< 24 hours). The Instructions for Use provides instruction on how to properly clean the block for re-use between patients. This component is not intended for sterilization.
Within the camera housing, an illuminator ring emits the infrared light which reflects off of the markers on the reflector block back to the camera. The camera is mounted either on the wall or ceiling along the central axis of the couch at a maximum distance of 4.0 meters from the block. The camera is directly connected to the RGSC Cabinet through the back panel. The housing also contains the class II laser used to calibrate the camera. This calibration is to be performed after installation and any time the camera's position has changed.
The RGSC Cabinet houses the Workstation Unit and Real-Time Unit. The Workstation contains the RGSC application which has functions for patient file creation and storage, calibration of the system, set up, recording, and review of a reference session. The Real-Time Unit contains software for the real-time image data processing, which is used in the dynamic tracking of the reflector block. The real-time unit controls the interface to the Wireless Access Point, Camera, and the 3d Party Diagnostic Imaging Scanner.
No biologics or drugs are in this device is not intended for single use. No parts of the system are provided sterile or are intended for sterilization.
The associated accessories include:
- . Keyboard and Mount for RGSC Workstation
- Monitor for RGSC Workstation ●
- VCD Couch Mounting Arm .
- VCD Battery Packs, Charging Station, and External DC Power Supply ●
- . Wireless Access Point
#### V. INDICATIONS FOR USE
Indications for Use Statement: Respiratory Gating for Scanners (RGSC) is used to characterize the patient's respiratory patterns, providing the necessary information to diagnostic devices to acquire images synchronized with the breathing motion. RGSC can also be used to monitor the patient position during the image acquisition.
{5}------------------------------------------------
The Indications for Use statement for RGSC is not identical to the predicate device; however, the differences do not extend or add to the intended use of the device, nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have within their intended use the same application in obtaining respiratory-synchronized images and monitoring patient position during image acquisition.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Respiratory tracking and gating is the technological principle for both the subject and predicate device. It is based on using an infrared camera to recognize and monitor the position of a plastic reflector block with reflective markers placed on the patient's chest or abdomen. The collected respiratory pattern can be gated to trigger image acquisition by a CT or PET CT scanner.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Reflector Block Plastic Block with Reflective Markers ●
- . Infrared Tracking Camera - Used to Monitor the Movement and Location of the Reflector Block
- . Placement of the Block on the Abdomen or Chest of the Patient to Measure Respiratory Motion
- Support for Audio or Visual Coaching to guide patient in breathing patterns
- . Support for Phase-Based or Amplitude-Based gating
The following technological differences exist between the subject and predicate devices:
- Use of a wireless Visual Coaching Device (VCD) with couch mounting
- Use of a camera with different resolution .
- Use of a different Reflector Block design ●
- Support for Database Mode (connection to ARIA DB) .
- Support for Wide Screen Monitor Displays ●
- Introduction of Baseline Drift Display and Correction ●
- . Introduction of Additional Image and Data Export Functions
- The predicate device is compatible for use with radiation therapy treatment devices whereas the subject device is not compatible for use with those types of devices
- The predicate device is compatible for use with radiation therapy simulators whereas the subject device is not compatible for use with those types of devices
Further details about the differences between the predicate and subject device are included in the Substantial Equivalence Discussion of this submission.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Non-Clinical Tests: Biocompatibility Testing:
{6}------------------------------------------------
The biocompatibility evaluation for the Reflector Block component of RGSC was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA.
The Reflector Block component of RGSC is considered skin contacting for durations of less than 24 hours. The primary material of contact in this component. ABS 737, has no identified health or environmental effects, and no leachable material can be identified under the conditions specified for use of RGSC.
# Electrical Safety and Electromagnetic Compatibility (EMC):
Electrical safety and EMC testing were conducted on RGSC, consisting of the electronic components: RGSC Cabinet (Workstation Unit and Real-Time Unit), VCD, Wireless Access Point, and Infrared Camera. The system complies with the FDA Recognized Standards AAMI ANSI ES 60601-1 for Basic Safety of Medical Electrical Equipment and IEC 60601-1-2 for EMC.
# Software Verification and Validation Testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "major" level of concern as the device is an accessory to a device with "major" level of concern. The appropriate level software documentation is provided with reference to the relevant guidance.
# Mechanical Testing:
- Testing of the VCD Couch Mounting ●
- Testing of the Camera Wall/Ceiling Mounting ●
- Simulated Use Testing ●
# Animal Study:
No animal studies have been included in this submission.
# Clinical Tests:
No clinical tests have been included in this submission.
#### VIII. CONCLUSIONS
The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrate that Respiratory Gating for Scanners should perform as intended in the specified use conditions. The data demonstrates that RGSC is as safe and effective as and performs as well as or better than the predicate device, RPM Respiratory Gating System (K102024) for the same use in respiratory-synchronized image acquisition and patient position monitoring during image acquisition.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.