The intended use is to deliver oxygen to the patient nasally, controlled by a demand system
Device Story
The NOP Demand and Modified Demand Nasal Oxygen Cannulas are dual-port devices designed for use with oxygen conservation delivery systems. One tube delivers oxygen to the patient, while the second tube senses inspiratory/expiratory effort. The device is intended for use by adult and pediatric patients. It functions as a passive interface between the patient and an external demand-based oxygen delivery system. Healthcare providers or patients use the cannula to facilitate oxygen therapy; the output is the delivery of oxygen triggered by the patient's respiratory effort, which can improve oxygen conservation and patient comfort. The device is constructed from PVC with DINCH or DOTP plasticizers.
Clinical Evidence
Bench testing only. Testing verified performance characteristics including back pressure (max < 2 psi), tubing bond strength (all samples > 2 lbs), and environmental storage/transport stability (-29°C to 50°C). Biocompatibility testing confirmed the materials passed irritation, sensitization, cytotoxicity, particulate, and volatile organic compound requirements. Sample size (n=29) supported a 95% confidence level at 91% reliability.
Technological Characteristics
Dual-port nasal cannula; materials: PVC with DINCH or DOTP plasticizers. Sensing/actuation: passive dual-lumen tubing for oxygen delivery and respiratory effort sensing. Dimensions/form factor: standard nasal cannula interface. Connectivity: none (passive). Sterilization: not specified. Software: none.
Indications for Use
Indicated for adult and pediatric patients requiring nasal oxygen delivery controlled by a demand system.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with a stylized design that suggests flowing hair or water beneath the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
Salter Labs Margaret Caler Regulatory Associate 2365 Camino Vida Robles Carlsbad, California 92011
Re: K151506
Trade/Device Name: NOP Demand Nasal Oxygen Cannula, Models 4804, 4805, 4807, 4824. & 4827 and NOP Modified Demand Nasal Oxygen Cannula, Models 4904, 4905, & 4907 Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: NFB Dated: February 5, 2016 Received: February 11, 2016
Dear Ms. Caler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151506
### Device Name
NOP Demand Nasal Oxygen Cannula, Models 4804, 4807, 4824, & 4827 and NOP Modified Demand Nasal Oxygen Cannula, Models 4904, 4905, & 4907.
Indications for Use (Describe)
The intended use is to deliver oxygen to the patient nasally, controlled by a demand system
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 9.0: 510 (k) Summary
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), a the following summary information is provided:
### 510 (k) Summary
#### A. Submitter:
Salter Labs 2365 Camino Vida Robles Carlsbad, CA 92011 Telephone: 760-795-7100 Fax: 760-683-6797
Contact Person: Mara Caler Director, Regulatory Affairs
Date Prepared: 11 March 2016
#### B. Device Names:
| Trade Name | Common Name |
|----------------------------------------------------------------------------|------------------------------------------|
| NOP Demand Nasal Oxygen Cannula<br>models: 4804, 4805, 4807, 4824, & 4827. | NOP Demand Nasal Oxygen Cannula |
| NOP Modified Demand Nasal Oxygen<br>Cannula models: 4904, 4905, & 4907. | NOP Modified Demand Nasal Oxygen Cannula |
| Classification Name: | Conserver, Oxygen |
|-----------------------|-------------------|
| Product Code: | NFB |
| Regulation Number: | 868.5905 |
| Classification: | II |
| Classification Panel: | Anaesthesiology |
#### Predicate Devices: C.
This submission demonstrates substantial equivalence to the predicate devices:
K890298, Demand Nasal Oxygen Cannula
K892407, Modified Demand Nasal Oxygen Cannula
{4}------------------------------------------------
#### D. Device Descriptions
The demand cannula is a special cannula designed to give adult and pediatric users of dual port oxygen conservation delivery system all the comfort and conveniences of a Salter-Style® cannula. Unique dual tubing delivers oxygen on demand through one tube while inspiratory / expiratory effort is also sensed through the other tube. Meets dual port device manufacturers' specifications.
#### Indications for Use E.
The intended use is to delivery oxygen to patient nasally, controlled by a demand system.
#### Comparison of Technological Characteristics with the Predicate Device F.
The proposed NOP Demand Nasal Oxygen Cannula and Modified NOP Demand Nasal Oxygen Cannula is substantially equivalent to the predicate device listed above in that the the indications for use, the intended use, and fundamental scientific technology remain unchanged.
The differences between the NOP Demand Nasal Oxygen Cannula and Modified NOP Demand Nasal Oxygen Cannula and the predicate device are:
| Features | Predicate Demand<br>Nasal Oxygen<br>Cannula | NOP Demand Oxygen<br>Cannula | Performance<br>Testing |
|-------------------------|---------------------------------------------|------------------------------|------------------------|
| Material<br>Formulation | PVC, DIDP | PVC, DINCH | Biocompatibility |
| | PVC, DEHP | PVC, DOTP | and Performance |
#### ﻦ Performance Data
The NOP Demand Nasal Oxygen Cannula and Modified NOP Demand Nasal Oxygen Cannula were tested to verify that the new material bond and performance characteristics of flow rate, back pressure, and tubing bond strength did not impact the strength or performa nce of the modified demand devices after the material change. Statistical methods were used to determine, the minimum required sample size was 29 to support a 95% confidence level at 91% reliability.
The test results demonstrate that the NOP Demand Nasal Oxygen Cannula and Modified NOP Demand Nasal Oxygen Cannula is substantially equivalent to the predicate devices.
{5}------------------------------------------------
The tests performed are summarized below:
| Criteria | Results | Comments |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Shall not have a back<br>pressure that exceeds 3 psi at<br>a maximum flow rate in<br>ambient of 5°C, 20°C, and<br>40°C | The NOP Demand/Modified<br>Demand Nasal Oxygen<br>Cannula met the objective back<br>pressure requirements | Maximum back<br>pressure was<br>found to be less<br>than 2 psi. |
| The bonded components of<br>the set will have a bond<br>strength that is $≥$ 2 lbs. when<br>pulled at a rate of 5 inches<br>per minute. | The NOP Demand/Modified<br>Demand Nasal Oxygen<br>Cannula passed all bond strength<br>requirements. All test samples<br>measured above the bond interface<br>tensile load limit. | The worst case<br>cannula sample<br>was able<br>to achieve over<br>2 times the<br>minimum<br>allowable<br>value. |
| The cannula shall withstand<br>storage and transport<br>temperatures. A high<br>temperature limit of 50°C<br>and a low temperature limit<br>of -29°C. | The NOP Demand/Modified<br>Demand Nasal Oxygen<br>Cannula met the above<br>performance criteria after storage<br>and transport environmental<br>conditioning. | N/A |
The NOP Demand Nasal Oxygen Cannula and Modified NOP Demand Nasal Oxygen
Cannula meet established Salter Labs performance specifications.
#### Clinical / Non-Clinical H.
The following biocompatibility testing was performed. The materials passed all parameters:
- Irritation
- · Sensitization ·
- · Cytotoxicity
- Particulate
- Volatile organic compounds
#### I. Conclusions
The NOP Modified Demand and NOP Demand Nasal Oxygen Cannula data and test results demonstrate that the devices are substantially equivalent to the predicate devices
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.