K151475 · Endochoice, Inc. · FBK · Feb 8, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K151475
Device Name
EndoChoice Select Injection Needle
Applicant
Endochoice, Inc.
Product Code
FBK · Gastroenterology, Urology
Decision Date
Feb 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EndoChoice Select injection needle is intended for endoscopic injection into the gastrointestinal mucosa.
Device Story
EndoChoice Select Injection Needle is a single-use endoscopic accessory used for local injection into gastrointestinal mucosa. Device consists of a needle, cap, boosting tube, positioning ring (304 stainless steel), outer tube/nut (PTFE), spring/spring tube (302 stainless steel), shell/handle (ABS), bushing (TPE), inner tube (nylon), and color band (CNBR). Operated by a physician via a standard endoscope; device is inserted through the endoscope channel to deliver therapeutic agents to the digestive tract mucosa. Single-use, ETO-sterilized, and intended for contact durations under 1 hour. Modification involves a new patient-contacting material in the Teflon sheath configuration. Clinical benefit derived from enabling targeted mucosal injection during endoscopic procedures.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility testing for the new patient-contacting material, accelerated aging, and sterilization validation. All tests passed.
Indicated for endoscopic injection into the gastrointestinal mucosa for patients requiring local injection therapy via endoscope.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 8, 2016
EndoChoice, Inc. Daniel Hoefer Regulatory Affairs Manager 11810 Wills Rd. Alpharetta, GA 30009
Re: K151475
> Trade/Device Name: Endochoice Select Injection Needle Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FBK Dated: January 6, 2016 Received: January 8, 2016
Dear Daniel Hoefer,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151475
Device Name EndoChoice Select Injection Needle
Indications for Use (Describe)
The EndoChoice Select injection needle is intended for endoscopic injection into the gastrointestinal mucosa.
Type of Use (Select one or both, as applicable):
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Special 510(k) Summary
### 1. Company Identification
EndoChoice, Inc. 11810 Wills Road Alpharetta, GA 30009 Telephone (678) 708 4743 FAX (678) 878 3373 Establishment Registration: 3007591333
#### 2. Contact Person
Daniel Hoefer Regulatory Affairs Manager
#### 3. Device Name
| Trade name: | EndoChoice Select Injection Needle |
|--------------------|------------------------------------|
| Common/Usual Name: | Injection Needle |
#### 4. Device Classification
| Injection Needle |
|---------------------------------------|
| Endoscope accessories, 21CFR 876.1500 |
| FBK |
| Gastroenterology/Urology |
| |
#### 5. Indications for Use
The EndoChoice Select injection needle is intended for endoscopic injection into the gastrointestinal mucosa.
#### 6. Device Description
The EndoChoice Select Injection Needle mainly consists of a cap, needle, boosting tube, and positioning ring, all made of 304 stainless steel. It also contains an outer tube and nut made of PTFE. The spring and spring tube are made of 302 stainless steel. The shell and injection handle are made of ABS. The bushing is made of TPE (SBS), the inner tube is made of nylon, and the color band indicator is made of CNBR.
The products are intended for single use; an individual device is packed in a sealed pouch following ETO sterilization. The device is used for local injection via endoscope, with the average contact time of the products and the mucosa of the human digestive tract of less than1 hour.
{4}------------------------------------------------
# 7. Substantial Equivalence
#### 7.1 Predicate devices
The modified EndoChoice Select Injection needle is substantially equivalent to the predicate devices. EndoChoice Injection Needle (K132065) manufactured by EndoChoice, Inc. The intended use, design and materials and labeling are all substantially equivalent.
The predicate device as described in K132065 included 2 catheter configurations. This modification applies only to the version utilizing a Teflon sheath and does not apply to the catheter design, which incorporates a stainless steel spring. This submission has been prepared to address a change, in which a new patient contacting material has been introduced, as identified in the Needle select design assembly attached.
#### 7.2 Intended Use
The indications for use of the modified EndoChoice Select Injection Needle is unchanged from the predicate device, the EndoChoice Injection Needle (K132065) manufactured by EndoChoice. In each case, the injection needle is intended for endoscopic injection into tissues of the digestive system.
#### 7.3 Technical Characteristics
The injection needle is available in multiple sizes and two styles. The needle gauge sizes are available in 22 gauge and 25 gauge needles. Two styles of sheaths are available; one style is a metal stainless steel spring sheath and the other is a Teflon sheath. The modified EndoChoice Select Injection Needle only applies to the Teflon sheath configuration.
#### 7.4 Performance Characteristics
The steps for operator use of each of the devices are equivalent. The instructions for use describe how to use the device with a standard endoscope, which is unchanged from the predicate.
{5}------------------------------------------------
| Substantial Equivalence Comparison | | | |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| | EndoChoice Injection Needle<br>Predicate (K132065) | Modified EndoChoice Select<br>Injection Needle | Substantial<br>equivalence |
| Manufacturer | EndoChoice, Inc. | EndoChoice, Inc. | Equivalent |
| Compatible with<br>currently<br>available<br>endoscopes | Yes | Yes | Equivalent |
| Supplied Sterile | Yes | Yes | Equivalent |
| Sheath diameter | 2.3mm | 2.3mm | Equivalent |
| Needle size | 22-25 gauge | 22-25 gauge | Equivalent |
| Outer Tubing | Thermoplastic polymer-<br>Teflon | Thermoplastic polymer-<br>Teflon | Equivalent |
| Length | 240 cm | 240 cm | Equivalent |
| Needle extension<br>length | 5 mm | 5 mm | Equivalent |
| Indications for<br>use | The EndoChoice Select<br>Injection Needle is intended<br>for endoscopic injection into<br>the gastrointestinal mucosa. | The EndoChoice Select<br>Injection Needle is intended<br>for endoscopic injection into<br>tissues of the digestive<br>system. | Equivalent |
| Packaging | Single-use EO sterilized<br>Tyvek pouch with one device<br>per pouch. | Single-use EO sterilized<br>Tyvek pouch with one<br>device per pouch. | Equivalent |
## 7.5 Substantial Equivalence Table
#### 8. Non-Clinical testing
Biocompatibility, accelerated aging and sterilization testing is unchanged from the predicate Injection Needle, which demonstrates that this device is safe and effective for use. Testing was based on a formal risk analysis. Biocompatibility testing was performed for the new patient contacting material of the modified device.
All test results passed, demonstrating that the safety and efficacy is equivalent to the predicate device.
#### 9. Conclusion
The EndoChoice Select Injection Needle is substantially equivalent to the predicate device listed above. It is the same or equivalent in terms of design, intended use, materials and labeling.