K151465 · Konica Minolta, Inc. · MQB · Jun 29, 2015 · Radiology
Device Facts
Record ID
K151465
Device Name
AeroDR SYSTEM 2
Applicant
Konica Minolta, Inc.
Product Code
MQB · Radiology
Decision Date
Jun 29, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The AeroDR SYSTEM 2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpse diagnostic procedures. The AeroDR SYSTEM 2 is not in mammography. fluoroscopy, tomography and angiography applications.
Device Story
AeroDR System 2 is a digital radiography system using flat panel detectors (FPDs) (models P-51, P-52) to capture X-ray images of human anatomy. FPDs output digital signals to an image processing console (Console CS-7) via wired Ethernet or wireless LAN (IEEE802.11a/n, WPA2-PSK encryption). The console applies image processing, including 'Intelligent Grid' (IG) processing to reduce scatter radiation effects, and transmits images to filing systems, printers, or displays. Used in clinical settings by healthcare professionals to replace film/screen systems. IG processing allows contrast adjustment; clinicians must verify effects before final image determination. Benefits include digital workflow efficiency and scatter reduction.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including image quality comparisons between the proposed P-52 detector and the predicate P-12 detector, risk management analysis, and biocompatibility testing per ISO 10993.
Technological Characteristics
Digital X-ray system with FPDs (P-51, P-52). Connectivity: Wired Ethernet, Wireless LAN (IEEE802.11a/n, WPA2-PSK/AES). Software: Console CS-7 with 'Intelligent Grid' post-processing. Compliance: AAMI/ANSI ES 60601-1 (Ed.3), IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for generating radiographic images of human anatomy for general-purpose diagnostic procedures. Not indicated for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2015
Konica Minolta, Inc. % Mr. Russell D. Munves U.S. Agent Storch Amini & Munves, PC 140 East 45th Street, 25th Floor NEW YORK NY 10017
Re: K151465 Trade/Device Name: AeroDR System 2 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB, LLZ Dated: May 28, 2015 Received: June 1, 2015
Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151465
Device Name AeroDR System 2
Indications for Use (Describe)
The AeroDR SYSTEM 2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpse diagnostic procedures.
The AeroDR SYSTEM 2 is not in mammography. fluoroscopy, tomography and angiography applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter's Name: | KONICA MINOLTA, INC. |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1 Sakura-machi,<br>Hino-shi, 191-8511 Japan |
| Contact: | Shigeyuki Kojima |
| Telephone: | +81 42 589 8429 |
| Date: | May 28, 2015 |
| Trade Name: | AeroDR System 2 |
| Model No: | AeroDR P-51, AeroDR P-52 |
| Common Name: | Digital Radiography |
| Regulation Name / Number: | Stationary x-ray system / 21 CFR 892.1680 |
| Regulatory Class: | Class II |
| Product Code(s): | 90-MQB, 90-LLZ |
| Predicate Device(s): | K141271 - AeroDR SYSTEM 2 (KONICA MINOLTA, INC.)<br>Regulation Name: Stationary x-ray system (21CFR 892.1680),<br>Product Codes: 90-MQB, 90-LLZ |
| Device Description: | The AeroDR SYSTEM 2 including AeroDR P-51 and AeroDR P-52 is<br>intended for use in generating radiographic images of human anatomy.<br>It is intended to replace radiographic film/screen systems in general-<br>purpose diagnostic procedures. The AeroDR SYSTEM 2 is not indicated for<br>use in mammography, fluoroscopy, tomography and angiography<br>applications.<br><br>The AeroDR P-51 and AeroDR P-52 (hereinafter AeroDR detector) are<br>lightweight, mobile FPD those are that is compatible with the size of ISO<br>standard.<br><br>The AeroDR SYSTEM 2 performs X-ray imaging of the human body using<br>an X-ray planar detector (FPD) that outputs a digital signal, which is then |
{4}------------------------------------------------
input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data. The image processing device displays the image on the screen after applying image processing to the raw data of image provided by FPD. The AeroDR detector can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.
The operator console software, Console CS-7, is a software program for installation on a PC. It performs image processing of X-ray images received from an AeroDR FPDs. The modified post processing function named 'Intelligent Grid' (IG Processing) is introduced as a part of the Image post processing function of console CS-7 software. The IG processing are fundamentally the same as those functions of other post processing of the Image Processing module of Console CS-7. The IG Processing is NOT a function for substitution of Grid's effects. It's a post processing of image adjustment to reduce the effects of scatter radiation. The user can adjusts to improve contrast by reducing effects of scatter radiation, and then the user MUST confirm the effects of the processing prior to determining the final images as well.
The AeroDR System 2 is designed to comply with the following standards: AAMI/ANSI ES 60601-1 (Ed.3), IEC 60601-1-2, and ISO 10993-1.
Indications for Use:
The AeroDR SYSTEM2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. The AeroDR SYSTEM 2 is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
Summary of Technological Characteristics Compared to Predicate Device:
> The modified AeroDR SYSTEM 2 employs the same fundamental scientific technologies as the predicate device (K141271). The indications for use are identical to the predicate device, and the operating principles are not changed from K141271.
{5}------------------------------------------------
The summary of comparisons of technological characteristics for both systems is provided as follows;
Operational principles and designing:
Nothing to be changed from the operational principles of the predicate device. The changes in operation and weight by means of different size of battery are NOT essential for the intended use of this device. The specification for the image performance of proposed device is same as AeroDR P-12, which is a component of AeroDR SYSTEM 2(K141271) and originally cleared as K102349.
#### Non-clinical test:
The qualified persons have affirmed and have concluded that both nonclinical images of proposed AeroDR P-52 and predicate AeroDR P-12 are equivalent and have sufficient capabilities for the intended purpose of the device. Besides, the results of risk management, the proposed modification does NOT require clinical evaluation to demonstrate the substantial equivalency of the modified device.
#### Safety:
The system is in conformance with the standards described above, which are same standards to those of predicate device.
### Biocompatibility:
The patient contact materials of human body surface are evaluated under ISO 10993 and determined as acceptable for these usages. The both systems achieve same acceptance level for biocompatibility.
#### Modification under the Design Control as per 21CFR 820.30:
The risk analysis was performed on the proposed device including two major differences from the predicate device, that differences are the addition of AeroDR P-52 and the modification of the post processing function named 'Intelligent Grid' of console CS-7 software, and the conclusion of risk analysis is that two above-mentioned changes don't pose any significant risk to device safety or effectiveness. The modified device has been verified and validated according to the same procedure as the predicate device. All of the verification activities, as required by the risk analysis, for the modification to AeroDR SYSTEM 2 were performed and the results demonstrated that the predetermined acceptance criteria were met.
{6}------------------------------------------------
Conclusion:
The clinical studies as a performance testing are not required to support substantial equivalence for the proposed device. In addition to that, as discussed in the above technological comparison, the technological characteristics of the modified AeroDR SYSTEM 2 are deemed to be substantially equivalent to the predicate device that have already been cleared for USA distribution with 510(k) premarket notification number K141271.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.