K151450 · Intuitive Surgical, Inc. · KCT · Jan 12, 2016 · General Hospital
Device Facts
Record ID
K151450
Device Name
Endoscope Sterilization Tray
Applicant
Intuitive Surgical, Inc.
Product Code
KCT · General Hospital
Decision Date
Jan 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect da Vinci Xi endoscopes (Model #'s 470026 and 470027) for sterilization in any of the following sterilization machines/cycles: • STERRAD 100NX sterilization system using the Express cycle • STERRAD 100S sterilization system using the Standard cycle • Steris V-PRO max using the Non Lumen, Flexible, or Lumen cycles • Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles • Steris V-PRO 1 using the V-PRO/Lumen Cycle The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of tray and endoscope is 8.9 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.
Device Story
Thermoformed plastic tray with silicone inserts and clear lid; designed to encase/protect da Vinci Xi endoscopes during sterilization. Perforated design allows penetration of sterilization gases. Used in hospital settings; operated by clinical staff. Device does not maintain sterility independently; requires use with FDA-cleared sterilization wrap. Subject device is identical to predicate in design, materials, and dimensions; updated to accommodate increased maximum weight load (8.9 lbs vs 7.7 lbs).
Clinical Evidence
Bench testing only. Sterilization validation and verification testing for weight and functional specifications were performed. Cleaning, biocompatibility, and reuse testing were not repeated as the device design, materials, and manufacturing processes are identical to the predicate.
Technological Characteristics
Thermoformed plastic tray with silicone inserts and clear lid. Materials: ULTEM, Elastosil R401/70 silicone, stainless steel. Perforated for gas sterilization. Compatible with STERRAD 100NX, 100S, and Steris V-PRO (maX, 1 Plus, 1) low-temperature hydrogen peroxide gas sterilization systems. Dimensions: 25in x 15in.
Indications for Use
Indicated for encasing and protecting da Vinci Xi endoscopes (Model #'s 470026 and 470027) during sterilization in specified STERRAD and Steris low-temperature hydrogen peroxide gas sterilization systems. Maximum load: one endoscope; maximum weight: 8.9 lbs.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2016
Intuitive Surgical, Inc. Kunal Gunjal Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086
Re: K151450
Trade/Device Name: Endoscope Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: December 15, 2015 Received: December 16, 2015
Dear Mr. Gunjal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below |
|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K151450 | |
| Device Name<br>Endoscope Sterilization Tray | |
| Indications for Use (Describe) | |
| | The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect da Vinci Xi endoscopes<br>(Model #'s 470026 and 470027) for sterilization in any of the following sterilization machines/cycles: |
| | • STERRAD 100NX sterilization system using the Express cycle<br>• STERRAD 100S sterilization system using the Standard cycle<br>• Steris V-PRO max using the Non Lumen, Flexible, or Lumen cycles<br>• Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles<br>• Steris V-PRO 1 using the V-PRO/Lumen Cycle |
| | The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum<br>product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm<br>and the diameter of the shaft is 8.8 mm. The maximum weight of tray and endoscope is 8.9 lbs. |
| | The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility<br>of the enclosed medical instrument. |
| Type of Use (Select one or both, as applicable) | ☐ Prescription Use (Part 21 CFR 801 Subpart D) <span style="padding-left:20px"></span> ☒ Over-The-Counter Use (21 CFR 801 Subpart C) |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: |
| | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." |
| FORM FDA 3881 (8/14) | Page 1 of 1 |
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Owner: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kunal Gunjal<br>Regulatory Affairs Specialist, Regulatory Affairs<br>Phone Number: 408-523-8017<br>Fax Number: 408-523-8907<br>Email: Kunal.Gunjal@intusurg.com |
| Date Summary Prepared: | December 15, 2015 |
| Trade Name: | Endoscope Sterilization Tray |
| Common Name: | Sterilization Tray |
| Classification: | Class II<br>21 CFR 880.6850, Sterilization Wrap |
| Product Codes: | KCT |
| Classification Advisory<br>Committee: | General Hospital |
| Predicate Device: | Intuitive Surgical Endoscope Sterilization Tray (K142937) |
{4}------------------------------------------------
#### Device Description
The Endoscope Sterilization Tray is a thermoformed plastic tray with silicone inserts and a clear lid. The tray and lid contain perforations to allow sterilization gases to penetrate the tray and sterilize the endoscope. The Endoscope Sterilization Tray is designed to encase and protect da Vinci Xi endoscopes (Model #'s 470026 and 470027) during transport and sterilization. The tray is compatible with the STERRAD 100NX (Express cycle), STERRAD 100S (Standard cycle) and the following Steris low temperature hydrogen peroxide gas sterilization systems:
- o Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles
- o Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles
- Steris V-PRO 1 using the V-PRO/Lumen Cycle o
The subject Endoscope Sterilization Tray and predicate tray are identical in terms of design, materials, dimensions and weight. The purpose of this 510(k) is to update the Indications for Use to reflect that the tray can accommodate a heavier endoscope. The previous Indications for Use reflected that the tray could accommodate a maximum of 7.7 lbs for the combined weight of the endoscope and tray, while the Indications for Use in this 510(k) reflect that the tray can accommodate a maximum of 8.9 lbs for the combined weight of the endoscope and tray.
#### Indications for Use:
The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect da Vinci Xi endoscopes (Model #'s 470026 and 470027) for sterilization in any of the following sterilization machines/cycles:
- o STERRAD 100NX sterilization system using the Express cycle
- STERRAD 100S sterilization system using the Standard cycle ●
- o Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles
- Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles
- Steris V-PRO 1 using the V-PRO/Lumen cycle
The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 8.9 lbs.
The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.
#### Technological Characteristics:
The subject Intuitive Surgical Endoscope Sterilization Tray is identical to the predicate Intuitive Surgical Endoscope Sterilization Tray (K142937) in design, materials and technological characteristics; it is the same tray.
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray | Predicate Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray K142937 |
|-------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Manufacturer | Intuitive Surgical, Inc. | Intuitive Surgical, Inc. |
| Trade Name | Endoscope Sterilization Tray | Endoscope Sterilization Tray |
| 510(k) No. | K151450 | K142937 |
| 510(k) Decision<br>Date | Not Applicable | February 24, 2015 |
| Common Name | Sterilization Tray | Sterilization Tray |
{5}------------------------------------------------
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray | Predicate Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray K142937 |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation No. | 21 CFR 880.6850 | 21 CFR 880.6850 |
| Product Code | KCT | KCT |
| Device Class/<br>Regulation Name | Class II/ Sterilization Wrap | Class II/ Sterilization Wrap |
| Classification<br>Advisory<br>Committee | General Hospital | General Hospital |
| Sterilization<br>Method | SAME | H2O2 chemical sterilization |
| Sterility /<br>Disposable or<br>Multiple use | SAME | Multiple use |
| Intended Use | SAME | The Intuitive Surgical Endoscope<br>Sterilization Tray is intended to encase<br>and protect da Vinci Xi endoscopes for<br>sterilization in the STERRAD 100NX<br>sterilization system using the Express<br>cycle setting, the STERRAD 100S<br>sterilization system using the Standard<br>cycle setting, and the Steris low<br>temperature hydrogen peroxide gas<br>sterilization systems (V-PRO maX, V-<br>PRO 1 Plus, and V-PRO 1). |
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray | Predicate Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray K142937 |
| Indications for Use | The Intuitive Surgical Endoscope<br>Sterilization Tray is intended for use to<br>encase and protect da Vinci Xi<br>endoscopes (Model #'s 470026 and<br>470027) for sterilization in any of the<br>following sterilization machines/cycles:<br>• STERRAD 100NX sterilization<br>system using the Express cycle<br>• STERRAD 100S sterilization<br>system using the Standard cycle<br>• Steris V-PRO maX using the<br>Non Lumen, Flexible, or Lumen<br>cycles<br>• Steris V-PRO 1 Plus using the<br>Non Lumen or Lumen cycles<br>• Steris V-PRO 1 using the V-<br>PRO/Lumen Cycle<br>The sterilization cycle parameters of the<br>sterilizers are preset by the manufacturers<br>and are not adjustable. The maximum<br>product load per tray is one da Vinci Xi<br>Endoscope. The length of the da Vinci Xi<br>Endoscope is approximately 600 mm and<br>the diameter of the shaft is 8.8 mm. The<br>maximum weight of tray and endoscope<br>is 8.9 lbs.<br>The Endoscope Sterilization Tray is not<br>intended to maintain sterility; it is<br>intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>STERRAD and Steris compatible<br>sterilization wrap in order to maintain<br>sterility of the enclosed medical<br>instrument. | The Intuitive Surgical Endoscope<br>Sterilization Tray is intended for use to<br>encase and protect da Vinci Xi<br>endoscopes (Model #'s 470026 and<br>470027) for sterilization in any of the<br>following sterilization machines/cycles:<br>• STERRAD 100NX sterilization<br>system using the Express cycle<br>• STERRAD 100S sterilization<br>system using the Standard cycle<br>• Steris V-PRO maX using the<br>Non Lumen, Flexible, or Lumen<br>cycles<br>• Steris V-PRO 1 Plus using the<br>Non Lumen or Lumen cycles<br>• Steris V-PRO 1 using the V-<br>PRO/Lumen Cycle<br>The sterilization cycle parameters of the<br>sterilizers are preset by the manufacturers<br>and are not adjustable. The maximum<br>product load per tray is one da Vinci Xi<br>Endoscope. The length of the da Vinci Xi<br>Endoscope is approximately 600 mm and<br>the diameter of the shaft is 8.8 mm. The<br>maximum weight of tray and endoscope<br>is 7.7 lbs.<br>The Endoscope Sterilization Tray is not<br>intended to maintain sterility; it is<br>intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>STERRAD and Steris compatible<br>sterilization wrap in order to maintain<br>sterility of the enclosed medical<br>instrument. |
| Where used<br>(hospital, home,<br>ambulance, etc) | SAME | Hospital |
| Dimensions<br>(LxW) | SAME | 25in. x 15in. |
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray | Predicate Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray K142937 |
| Materials; | SAME | ULTEM™, Elastosil R401/70 silicone,<br>stainless steel |
| Compatibility<br>with the<br>environment and<br>other devices | SAME | Compatible with da Vinci Xi endoscopes |
| Sterilization<br>system<br>compatibility | SAME | STERRAD 100NX and STERRAD 100S<br>and Steris V-PRO maX, V-PRO 1 Plus,<br>and V-PRO 1 low temperature hydrogen<br>peroxide gas sterilization systems |
{6}------------------------------------------------
{7}------------------------------------------------
### Performance Data:
Performance test data demonstrates that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. The testing conducted consisted of sterilization validation and bench testing, included verification testing for weight and functional specifications. Cleaning, biocompatibility, and limits of reuse testing were not repeated as the subject device design, materials, and manufacturing processes are identical to the predicate device (Intuitive Surgical Endoscope Sterilization Tray, K142937).
## Substantially Equivalent Conclusion:
Based on the intended use, indications for use, technological characteristics, performance data and nonclinical tests performed, the subject device, Intuitive Surgical Endoscope Sterilization Tray (K151450) is substantially equivalent and is as safe, as effective, and performs as well as the legally marketed predicate device, Intuitive Surgical Endoscope Sterilization Tray (K142937), Class II (21 CFR 880.6850), Product code KCT.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.