PICO Single Use Negative Pressure Wound Therapy System
K151436 · Smith & Nephew Medical, Ltd. · OMP · Jan 28, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151436
Device Name
PICO Single Use Negative Pressure Wound Therapy System
Applicant
Smith & Nephew Medical, Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Jan 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include: - Chronic - Acute - Traumatic - Subacute and dehisced wounds - Partial-thickness burns - Ulcers (such as diabetic or pressure) - Flaps and grafts - Closed surgical incisions PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both hospital and homecare setting.
Device Story
PICO is a portable, single-use negative pressure wound therapy (NPWT) system. It consists of an electric, twin-diaphragm vacuum pump connected to a super-absorbent adhesive dressing. The system is used in hospital and homecare settings to treat various wound types. The pump is battery-operated (two AA lithium batteries) and provides up to 168 hours (7 days) of therapy. It creates negative pressure to draw excess fluid from the wound into the dressing. The device is programmed to stop automatically after 168 hours. Healthcare providers or patients apply the dressing and fixation strips; the pump operates continuously until the service life ends. The system benefits patients by managing exudate and promoting wound healing. Modifications in this version include a microfuse, pump unit design changes, increased tube length, and software updates.
Clinical Evidence
Bench testing only. Studies included microfuse characterization, accelerated life testing, and service life characterization. Wound model testing demonstrated that the system maintains negative pressure (-60 to -100 mmHg) for at least 95% of the test duration at moderate and low flow rates, confirming performance is unaffected by design modifications. Biocompatibility was established via reference to previous testing (K112127) per ISO 10993-1.
Technological Characteristics
Portable, battery-powered (2x AA lithium) twin-diaphragm suction pump. Includes super-absorbent adhesive dressing and secondary fixation strips. Materials comply with ISO 10993-1. Electrical safety/EMC compliance per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-1-11. Features a microfuse for circuit protection. Single-use, sterile.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing by removing low to moderate exudate and infectious materials. Applicable to chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), flaps, grafts, and closed surgical incisions.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
PICO Single Use Negative Pressure Wound Therapy System (K112127)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 28, 2016
Smith & Nephew Medical Ltd % Samantha Neilson Regulatory Affairs Manager Smith & Nephew, Inc. 970 Lake Carillon Drive Suite 110 St Petersburg, Florida 33716
Re: K151436
Trade/Device Name: Pico Single Use Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: May 27, 2015 Received: May 28, 2015
Dear Samantha Neilson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151436
#### Device Name
PICO Single Use Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
- Closed surgical incisions
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both hospital and homecare setting.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) SUMMARY
### PICO Single Use Negative Pressure Wound Therapy System
| General Information | |
|------------------------------------|----------------------------------------------|
| Submitter Name/Address: | Smith & Nephew Medical Inc |
| | 970 Carillon Drive, Suite 110 |
| | Saint Petersburg, Florida, 33716 |
| Establishment Registration Number: | 3006760724 |
| Contact Person: | Terry McMahon, Director Regulatory Affairs & |
| | Quality, North America |
| Phone Number | +727 399 3468 |
| Date Prepared: | May 27, 2015 |
| Device Description | |
| Trade Name: | PICO Single Use Negative Pressure Wound |
| | Therapy System |
| Common or Usual Name: | Negative Pressure Wound Therapy powered |
| | suction pump |
| Classification Name: | Powered suction pump (21 CFR 878.4780) |
| Regulatory Class: | Class II |
| Product Code: | OMP |
| Predicate Device Information | |
| 510(k) Number: | K112127 |
| Device: | PICO Single Use Negative Pressure Wound |
| | Therapy System |
| Clearance Date: | May 16, 2012 |
| Device Description | |
{5}------------------------------------------------
PICO Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable suction device consisting of an electric motor driven, twin-diaphragm, vacuum pump connected to a super-absorbent, gentle adhesive dressing. The pump, dressing and secondary fixations strips are supplied sterile and single use. The dressing is applied to the wound and secondary fixation strips are placed over the outside edges to help hold the dressing in place. When the suction pump is turned on, air is pulled out of the dressing creating negative pressure and drawing excess fluid from the wound into the dressing. The pump is battery operated and is supplied with two AA lithium batteries which provide up to 168 hours (7 days) of battery life depending upon leak rate. The batteries can be replaced if required. The pump is programmed to stop working after 168 hours (7 days) of use and will not re-start after this time, even with new batteries. Negative pressure will not be applied at this point. PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting. A carry bag is not provided with the system however will be available to order separately.
### Indications for Use
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- · Partial-thickness burns, Ulcers (such as diabetic or pressure)
- Flaps and grafts
- Closed surgical incisions
{6}------------------------------------------------
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
#### Comparison between New and Predicate Devices
The Indications for Use statement of the PICO Single Use Negative Pressure Wound Therapy System is identical to the predicate devices. The technological principal for delivering the negative pressure wound therapy for both the subject and predicate devices are identical. Modifications to the cleared device include:
- (a) Changes to the pump unit as follows: integration of reverse polarity protection into battery cover, modification of battery cover, removal of pump weights, change in acoustic dampening foam, modification of inlet connector, addition of travel limiters, modification of pump cases, change to transistor
- (b) Increase in the length of tube connecting the dressing to the pump
- (c) Increase in shelf life
- (d) Changes to instructions for use to comply with 3td Edition of IEC 60601
- (e) Software change
- (f) Inclusion of microfuse
### Non-Clinical Tests (Bench)
The following bench tests have successfully been completed:
- . PICO 1.5 Micofuse Characterization Test Report
The results of the tests showed that the temperature was within the allowable tolerance limits as per the requirements of IEC 60601-1 and also showed that the fuses did not break and the soldering was intact after the drop tests. All the tests met the acceptance criteria. Therefore it was concluded that the microfuse was fit for purpose.
#### . PICO 1.6 Microfuse Accelerated Life Test Summary Report
The results showed that the current consumption attributed to the fuses was comparable over the accelerate life. It was concluded the effect of the microfuse on the overall power
{7}------------------------------------------------
consumption of the device will be unchanged over the shelf-life of the device and will therefore not cause unacceptable power draw which would limit the life of the device below the nominal service life. All fuse interventions occurred well within the nominal fuse failure time (0.21 seconds) at all time points. The fuse intervention can therefore be expected to occur as designed throughout a 3 year shelf life of the device. Based on the above, the tests have been considered a PASS and meet the set acceptance criteria.
- PICO 1.6 Microfuse Service Life Characterisation Report (12 ml/min nominal leak) .
At the end of execution of the FULL TEST suite, the following results were observed (a) No batteries replacement was required over the seven days of Service Life; (b) The End of Life countdown timer was not reached before 7 days of operation; (c) All the microfuse were still assembled and in function after the seven days of Service Life; and (d) No pump leak was observed during while the pump was operational. Based on the above, it was concluded that the functionality of the pump, over the service life of PICO system, is unaffected by the addition of the microfuse.
- PICO 1.6 Wound Model Testing of PICO pumps with microfuse 25 x 25 cm dressing at . moderate flow rate over 3 days
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a moderately exuding wound (1.1 g/cm/24 hours) for a simulated wound of 25% of the dressing absorbent pad area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the rests it can be concluded that the inclusion of the microfuse in PICO pumps did not affect the performance of the PICO System over a simulated use period of 3 days.
- PICO 1.6 Wound Model Testing of PICO pumps with microfuse 10 x 20 cm dressing at . moderate flow rate over 3 days
{8}------------------------------------------------
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a moderately exuding wound (1.1 g/cm²/24 hours) for a simulated wound of 25% of the dressing absorbent pad area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the rests it can be concluded that the inclusion of the microfuse in PICO pumps did not affect the performance of the PICO System over a simulated use period of 3 days.
- PICO 1.6 Wound Model Testing of PICO pumps with microfuse 25 x 25 cm dressing at . low flow rate over 4 days
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a moderately exuding wound (0.6 g/cm²/24 hours) for a simulated wound of 25% of the dressing area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the results of the tests it can be concluded that the inclusion of the microfuse in PICO pumps did not affect the performance of the PICO System over a simulated use period of 4 days.
- PICO 1.6 Wound Model Testing of PICO pumps with microfuse 10 x 20 cm dressing at . low flow rate over 4 days
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a moderately exuding wound (0.6 g/cm²/24 hours) for a simulated wound of 25% of the dressing area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the results of the tests it can be concluded that the inclusion of the microfuse in PICO pumps did not affect the performance of the PICO System over simulated use period of 4 days.
### Biocompatibility and Sterilization Testing
{9}------------------------------------------------
There were no modifications to the patient contacting components or materials of the predicate device K112127 where additional safety tests were required to meet the requirements of ISO 10993-1. Therefore the results of the previous tests mentioned below are still valid and can be used to show that the PICO Single Use Negative Pressure Wound Therapy System is still safe for its intended use. In the previous cleared 510(k), K112127, the biocompatibility of PICO Single Use Negative Pressure Wound Therapy System has been demonstrated through assessment according to ISO 10993-1 Biological Evaluation of Medical Devices, and appropriate tests had been conducted using final product that has been packaged and sterilised. These tests included cytotoxicity, sensitization and cover the PICO dressing and secondary fixation strips. These studies indicated that PICO Single Use Negative Pressure Wound Therapy System is safe for its intended use. In addition, residual testing had been performed in accordance with ISO 10993-7 which confirms that PICO dressings and fixation strips comply with the allowable residual limits for ethylene oxide (EO) and ethylene chlorohydrin (ECH).
#### Electrical Safety and Electromagnetic Compatibility (EMC)
Following the inclusion of the microfuse, PICO Single Use Negative Pressure Wound Therapy System has been tested and assessed respectively by the electrical safety testing agency CSA in accordance with the following electrical safety standards and has been found to be in compliance with these standards:
• IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) Medical Electrical Equipment -
Part 1: General Requirements for Basic Safety and Essential Performance
· IEC 60601-1-2: 2007 (Third Edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, Interpretation Sheet (Note: Following an assessment against the requirements of IEC 60601-1-2, it was concluded that no new EMC testing was required to show compliance with the later standard following the inclusion of the microfuse. Therefore no new EMC report was released.)
{10}------------------------------------------------
· IEC 60601-1-6: 2010 (Third Edition) Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability)
· IEC 60601-1-11: 2010 (First Edition) General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
#### Conclusions
In establishing substantial equivalence to the predicate devices, Smith & Nephew Medical Ltd evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing, service life testing and electrical safety testing has been completed to demonstrate that the modified PICO Single Use Negative Pressure Wound Therapy System is substantially equivalent to the predicate for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.