K151391 · B.T.I. Biotechnology Institute, Sl. · DZE · May 2, 2016 · Dental
Device Facts
Record ID
K151391
Device Name
BTI Dental Implant System UnicCa
Applicant
B.T.I. Biotechnology Institute, Sl.
Product Code
DZE · Dental
Decision Date
May 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K151391 · May 2, 2016
BTI Dental Implant System UnicCa
B.T.I. Biotechnology Institute, Sl.
Retrospective observational clinical study
A retrospective study was conducted to evaluate the clinical performance (implant and subject-based survival rates) of the BTI Dental Implant System UnicCa in patients treated in a routine clinical setting.
Observational retrospective study for BTI Dental Implant System UnicCa; Observational retrospective study; Follow-up/Duration: At least 6 months
Patients over 18 years old, partially or completely edentulous, treated with one or more BTI UnicCa implants.; Sample Size: 170 patients (430 implants)
Not applicable for this study
Implant-based survival rate (99.5%) and subject-based survival rate (98.8%) at 6 months.
Indications for Use
The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function. In the case of 5.5 – 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations . In the case of Tiny® 3.0 UnicCa® implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.
Device Story
BTI Dental Implant System UnicCa® is a self-tapping, threaded, root-form titanium dental implant. It features a surface treatment using soluble calcium chloride salt to enhance hydrophilicity. Used in oral implant surgery by clinicians to replace missing teeth; implants serve as anchors for fixed or removable prosthetic solutions. The device is implanted into the jawbone to restore mastication function. The system includes various diameters (3.0–6.0 mm) and lengths (5.5–18.0 mm) with Interna® or Externa® connections. Clinical benefit includes improved or restored chewing capacity. The device is supplied sterile via gamma radiation.
Clinical Evidence
Observational retrospective study of 170 patients with 430 UnicCa® implants. Subjects >18 years old, partially or completely edentulous, with minimum 6-month follow-up. Primary endpoint: implant survival (stable prosthesis, no pain/infection/pathology, no bone loss >2-3mm). Results: 99.5% implant-based survival and 98.8% subject-based survival at 6 months. No adverse events reported. Survival rates compare favorably to literature range (90-98%).
Technological Characteristics
Material: Commercially pure titanium grade 4 (ISO 5832-2, ASTM F67). Design: Threaded, root-form. Surface: Calcium chloride salt-treated for enhanced hydrophilicity. Connections: Interna® and Externa®. Dimensions: 3.0–6.0 mm diameter, 5.5–18.0 mm length. Sterilization: Gamma radiation (SAL 10^-6).
Indications for Use
Indicated for partial or total tooth replacement in edentulous patients. Tiny® 3.0 implants indicated only for maxillary lateral incisors and mandibular lateral/central incisors. 5.5–6.5mm implants indicated for two-stage surgery and delayed loading with straight abutments for fixed restorations.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an emblem with three stylized human profiles facing right, stacked on top of each other. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 2, 2016
B.T.I. Biotechnology Institute, S.L. Ms. Fernanda Olabarria Ros Regulatory Affairs Manager Parque Tecnologico De Álava, Leonardo DaVinci 14 01510 Miñano (Álava) SPAIN
Re: K151391
Trade/Device Name: BTI Dental Implant System UnicCa® Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: II Product Code: DZE Dated: February 19, 2016 Received: February 22, 2016
Dear Ms. Ros:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
#### Page 2 - Ms. Fernanda Olabarria Ros
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMS No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K151391
Device Name BTI Dental Implant System UnicCa®
Indications for Use (Describe)
The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function.
In the case of 5.5 – 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations .
In the case of Tiny® 3.0 UnicCa® implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div><span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of
this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office
of Chief Information Officer Paperwork
Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
'An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{4}------------------------------------------------
## 5.0 510(k) Summary
## I. SUBMITTER
# B.T.I. Biotechnology Institute, SL.
Parque Tecnológico de Álava C/Leonardo Da Vinci 14 01510 Miñano (Alava) Spain
Phone: (+34) 945 297 030 Fax: (+34) 945 297 031
Contact Person: Ms. Fernanda Olabarría Date Prepared: April 20, 2016
### II. DEVICE
Name of Device: BTI Dental Implant System UnicCa® Common or Usual Name: Root-form Endosseous Dental Implant Classification Name: Endosseous Dental Implant (21 CFR 872.3640) Regulatory Class: II Product Code: DZE
## III. PREDICATE DEVICE
| Primary Predicate | Reference Predicates | | |
|----------------------------------------|---------------------------------------------|--|--|
| | K053355; BTI Interna Dental Implant System. | | |
| K022258;<br>BTI Dental Implant System. | K091387; BTI Dental Implant 5.5 – 6.5. | | |
| | K092112; BTI Dental Implant Tiny® Ø 3.0. | | |
{5}------------------------------------------------
## IV. DEVICE DESCRIPTION
BTI Dental Implant System UnicCa® is a self-tapping, threaded, root form dental titanium implant. Comprises two types of connections: Externa® and Interna® with a range of diameters (3.0 - 6.0 mm) and lengths (5.5 - 18.0 mm) for various platforms, we refer to Table 5-1. BTI Dental Implant System UnicCa® features an implant surface treatment that improves the hydrophilicity of the implant.
Table 5-1. Overview of BTI Dental Implant System UnicCa® connections, platforms, diameters and lengths:
| Connection | Platform | Diameter<br>(mm) | Length (mm) |
|------------|----------------|------------------|-----------------------------------------------|
| Interna® | Wide | 5.0 | [6.5 / 7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Wide | 5.5 | [5.5 / 6.5] |
| | Wide | 6.0 | [5.5 / 6.5] |
| | Universal Plus | 4.5 | [5.5 / 6.5 / 7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal Plus | 5.0 | [5.5 / 6.5 / 7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal Plus | 5.5 | [5.5 / 6.5] |
| Interna® | Universal | 3.3 | [8.5 / 10 / 11.5 / 13 / 15] |
| | Universal | 3.5 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal | 3.75 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal | 4.0 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15 / 18] |
| | Universal | 4.25 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| Externa® | Wide | 5.0 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Wide | 5.5 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal Plus | 4.5 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal Plus | 5.0 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal | 3.75 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Universal | 4.0 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15 / 18] |
| Externa® | Tiny® | 3.0 | [11.5 / 13 / 15] |
| | Tiny® | 3.3 | [8.5 / 10 / 11.5 / 13 / 15] |
| | Tiny® | 3.5 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
| | Tiny® | 3.75 | [7.5 / 8.5 / 10 / 11.5 / 13 / 15] |
{6}------------------------------------------------
## V. INDICATIONS FOR USE
The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function.
In the case of 5.5 - 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations .
In the case of Tiny® 3.0 UnicCa® implants: These implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.
# VI. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject dental implants are identical to the currently marketed BTI Dental Implant Systems in indications for use, materials and design.
The BTI Dental Implant System UnicCa® is manufactured using the same procedures and processes as currently utilized for the predicate BTI Dental Implants, with exception that an additional manufacturing step has been added which consists of applying a soluble calcium chloride salt which is deposited at the outermost part of the surface of the implants. The predicate implant surface is known as Optima®, the implant with surface treatment is referred to as UnicCa®. This treatment enhances hydrophilic profile of the implant, easing its wetting characteristics to polar liquids.
A comparison of the device features, indications for use, laboratory data and other information demonstrate that the modified BTI Dental Implant System UnicCa® is substantially equivalent to the predicate devices.
In the following pages, a comparison table has been provided, Table 5-2.
{7}------------------------------------------------
| Characteristics | Current Submission<br>BTI Dental Implant System<br>UnicCa®<br>(K151391) | Primary Predicate<br>BTI Dental Implant<br>System<br>(K022258) | BTI Dental<br>Implant System<br>(K053355) | Reference Predicates<br>BTI Dental<br>Implant 5.5-6.5<br>(K091387) | BTI Dental<br>Implant Tiny®<br>(K092112) | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Indications for Use | The BTI Dental Implant System<br>UnicCa® for oral implant surgery<br>is to be used for the partial or<br>total replacement of teeth in<br>edentulate patients. Once attached<br>to the bone, the implants act as an<br>anchor for various fixed or<br>removable prosthetic solutions<br>that can be used to improve or<br>restore a patient's mastication<br>function.<br>In the case of 5.5 - 6.5mm long<br>UnicCa® implants should be used<br>in a two-stage surgical procedure.<br>These implants are indicated for<br>delayed loading.<br>These implants are indicated only<br>for straight abutments and to<br>support permanently fixed<br>restorations. | The intended use of the<br>system is the<br>restoration of missing<br>teeth in partially or<br>fully edentulous<br>patients and/or the<br>fixation of<br>overdentures to restore<br>or enhance the chewing<br>capacity of patients. | Dental implant<br>system comprising<br>endosseous<br>titanium implants<br>and prosthetic<br>elements to be<br>attached to the<br>implants, as well<br>as auxiliary<br>elements for<br>surgical and<br>prosthetic<br>procedures. The<br>intended use of the<br>system is the<br>restoration of<br>missing teeth in<br>partially or fully<br>edentulous patients<br>and/or the fixation<br>of overdentures to<br>restore or enhance<br>the chewing<br>capacity of<br>patients. | BTI Dental implants<br>5.5 - 6.5 mm are<br>intended to be used<br>to restore missing<br>teeth in partially or<br>fully edentulous<br>patients and/or for<br>the fixation of<br>overdentures to<br>restore or enhance<br>the chewing capacity<br>of patients. The<br>device should be<br>used in a two-stage<br>surgical procedure.<br>These implants are<br>not indicated for<br>immediate loading.<br>These implants are<br>not indicated to<br>support removable<br>resilient retained<br>restorations or angled<br>abutments. | BTI Dental Implant<br>Tiny® 3.0 are<br>intended to be used to<br>restore missing teeth<br>in partially or fully<br>edentulous patients<br>and/or the fixation of<br>overdentures'to restore<br>or enhance the<br>chewing capacity of<br>patients.<br>These implants shall<br>be used only to<br>replace maxillary<br>lateral incisors and<br>mandibular lateral and<br>central incisors.<br>Immediate loading is<br>recommended when<br>there is good primary<br>stability and an<br>appropriate occlusal<br>load. | |
| | Current Submission | Primary Predicate | Reference Predicates | | | |
| Characteristics | BTI Dental Implant System<br>UnicCa®<br>(K151391) | BTI Dental Implant<br>System<br>(K022258) | BTI Dental<br>Implant System<br>(K053355) | BTI Dental<br>Implant 5.5-6.5<br>(K091387) | BTI Dental<br>Implant Tiny®<br>(K092112) | |
| | In the case of Tiny® 3.0 UnicCa®<br>implants: These implants shall be<br>used only to replace maxillary<br>lateral incisors and mandibular<br>lateral and central incisors.<br>Immediate loading is<br>recommended when there is good<br>primary stability and an<br>appropriate occlusal load. | | | | | |
| | The indications for use statement for the BTI Dental Implant System UnicCa® has been written to encompass the<br>indications for use statements provided with the primary predicate system (K022258) and other previously cleared<br>connections, platforms, diameters and lengths acting as a single consolidated system.<br>In this sense, no labeling change that affects the indications for use of the device has been implemented. Information<br>relevant to a specific use of an implant included in previously cleared premarket notifications for BTI Dental Implants<br>are carried forward and included within the labeling for the BTI Dental Implant System UnicCa®. The product continues<br>to be specifically indicated for patients undergoing oral implant surgery for the partial or total replacement of teeth in<br>edentulate patients. | | | | | |
| Characteristics | Current Submission<br>BTI Dental Implant System<br>UnicCa®<br>(K151391) | Primary Predicate<br>BTI Dental Implant<br>System (K022258) | Reference Predicates | | | |
| | | | BTI Dental<br>Implant System<br>(K053355) | BTI Dental<br>Implant 5.5-6.5<br>(K091387) | BTI Dental<br>Implant Tiny®<br>(K092112) | |
| Product Classification | Device Class II<br>Regulation No.:<br>21 CFR 872.3640.<br>Product code: DZE;<br>Endosseous dental implant | Device Class II<br>Regulation No.:<br>21 CFR 872.3640.<br>Product code: DZE;<br>Endosseous dental<br>implant | Device Class II<br>Regulation No.: 21 CFR 872.3640.<br>Product code: DZE; Endosseous dental implant | | | |
| Implant Design/Geometry | Threaded, root form | Threaded, root form | Threaded, root form | | | |
| Material | Commercially pure titanium<br>grade 4 | Commercially pure<br>titanium grade 4 | Commercially pure titanium grade 4 | | | |
| Abutment Compatibility/<br>Connection | Internal (Interna) and External<br>(Externa) | External (Externa) | External<br>(Externa) | Internal (Interna) | External (Externa) | |
| Dimensions (mm) | Interna | Diameter: 3.3<br>to 6.0<br>Lengths:<br>5.5 to 18.0 | Diameter:<br>3.3 to 5.5<br>Length:<br>7.0 to 18.0 | Diameter:<br>3.3 to 5.0<br>Length:<br>7.0 to 18.0 | Diameter:<br>4.5 to 6.0<br>Lengths:<br>5.5 to 6.5 | Diameter:<br>3.0<br>Lengths:<br>11.5 to 15.0 |
| | Externa | Diameter: 3.0<br>to 5.5<br>Lengths:<br>7.0 to 18.0 | Diameter:<br>3.3 to 5.5<br>Length:<br>7.0 to 18.0 | Diameter:<br>3.3 to 5.0<br>Length:<br>7.0 to 18.0 | Diameter:<br>4.5 to 6.0<br>Lengths:<br>5.5 to 6.5 | Diameter:<br>3.0<br>Lengths:<br>11.5 to 15.0 |
| Roughness | Same | $Neck: Sq = 0.7 ± 0.1 µm; Sdr^2= 50± 10%$<br>$Thread: Sq≥ 1.2 µm;$<br>$Sdr≥ 200%$ | Same | | | |
| Characteristics | | Current Submission | Primary Predicate | Reference Predicates | | |
| | | BTI Dental Implant System<br>UnicCa®<br>(K151391) | BTI Dental Implant<br>System (K022258) | BTI Dental<br>Implant System<br>(K053355) | BTI Dental<br>Implant 5.5-6.5<br>(K091387) | BTI Dental<br>Implant Tiny®<br>(K092112) |
| | | | Valleys: Sq= $1.0 \pm 0.2$ µm; Sdr= $85 \pm 15$ % | | | |
| Mechanical<br>Properties | Material<br>(Titanium) | Same; in compliance with ISO<br>5832-2 and ASTM F67. | Same; in compliance<br>with ISO 5832-2 and<br>ASTM F67. | Same; in compliance with ISO 5832-2 and ASTM F67. | | |
| | Fatigue | Equivalent; platform<br>(diameter and length)<br>dependent. | Equivalent; platform<br>(diameter and length)<br>dependent. | Equivalent; platform (diameter and length) dependent. | | |
| Hydrophilicity | | Enhanced | | | | |
| Supplied Sterile | | Yes | Yes | Yes | | |
| Sterilization | | Gamma Radiation | e- beam | e- beam | | |
| SAL | | 1 x 10-6 | 1 x 10-6 | 1 x 10-6 | | |
| Packaging |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.