Stryker Universal Neuro III System: UN III AXS Screw, UN III AXS Screwdriver Blade
Applicant
Stryker
Product Code
HBW · Neurology
Decision Date
Jun 24, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker Universal Neuro III System is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
Device Story
Stryker Universal Neuro III System consists of titanium alloy screws and stainless steel screwdriver blades used for rigid fixation of non-load-bearing cranial bone. Device functions via mechanical plate fixation with screws to stabilize bone segments post-craniotomy or fracture. Operated by surgeons in clinical settings; screws provide long-term fixation (>30 days) while blades are transient accessories (<24 hours). Modifications involve dimensional changes to screws and screwdriver blades to improve performance. Output is physical stabilization of bone; clinical benefit is structural reconstruction of the cranium. No software or AI components involved.
Clinical Evidence
Bench testing only. No clinical or animal testing required. Bench tests included self-retention, screw insertion, life cycle fatigue (screwdriver blade), torque-to-failure (screwdriver blade), and end-user/end-product testing. All tests met pre-defined acceptance criteria.
Technological Characteristics
Materials: Titanium alloy (ASTM F136) for screws; stainless steel (ASTM F899) for blades. Principle: Mechanical plate fixation with screws. Dimensions: Modified from predicate. Sterilization: Not specified (no change from predicate). Connectivity: None. Software: None.
Indications for Use
Indicated for reconstruction, stabilization, and rigid fixation of non-load-bearing bony areas following craniotomy, craniectomy, and cranial fractures in adults and adolescents aged 12 and older.
Regulatory Classification
Identification
A cranioplasty plate fastener is a screw, wire, or other article made of tantalum, vitallium, or stainless steel used to secure a plate to the patient's skull to repair a skull defect.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of a human figure, represented by three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2015
Stryker Craniomaxillofacial Mr. Jonathan Schell Senior Regulatory Compliance Specialist 750 Trade Centre Way, Suite 200 Portage, Michigan 49002
Re: K151387
Trade/Device Name: Stryker Universal Neuro III System: UN III AXS Screw. UN III AXS Screwdriver Blade Regulation Number: 21 CFR 882.5360 Regulation Name: Cranioplasty Plate Fastener Regulatory Class: Class II Product Code: HBW, GWO, GXR Dated: Mav 22, 2015 Received: May 26, 2015
Dear Mr. Schell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -SFD/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Stryker Universal Neuro III System: UN III AXS Screw, UN III AXS Screwdriver Blade
Indications for Use (Describe)
The Stryker Universal Neuro III System is intended for reconstruction and/or rigid fixation of non loadbearing bony areas subsequent to craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is in black and is set against a white background. A small registered trademark symbol is located to the right of the word.
# Section 5. 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
#### I. SUBMITTER
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Jonathan Schell<br>Sr. Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-5596<br>Fax: 877-648-7114 |
| Date prepared: | May 22, 2015 |
| II.<br>DEVICE | |
| Trade Name: | Stryker Universal Neuro III System (also referred to as Universal<br>Neuro 3 System): UN III AXS Screw, UN III AXS Screwdriver Blade |
| Common or Usual<br>name: | Neuro Plating System |
| Classification<br>name: | Cranioplasty Plate Fastener 21 CFR §882.5360<br>Preformed alterable cranioplasty plate 21 CFR §882.5320<br>Burr hole cover 21 CFR §882.5250 |
| Regulatory Class: | Class II |
| Product Code: | HBW, GWO, GXR |
{4}------------------------------------------------
### III. PREDICATE DEVICE
### Predicate: Stryker Universal Neuro III System - K112557
This predicate has not been subject to a design-related recall.
### IV. DEVICE DESCRIPTION
The Stryker Universal Neuro III System (UN III) is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 or higher). The UN III System consists of an implant module for the respective anatomical and indicated areas with each containing various screw, and plate versions and shapes. This special 510(k) is being submitted due to dimensional modifications to the screw and screwdriver blade. There have been no modifications to the plates, meshes, or other accessory devices. The following Table 5-1 lists the devices subject to this Special 510(k):
| Article Number | Description |
|----------------------------------------------------------------------------------|-----------------------------|
| 62-15035 | AXS Screwdriver Blade Long |
| 62-15036 | AXS Screwdriver Blade Short |
| 56-15034<br>56-15035<br>56-15036<br>56-15933<br>56-15934<br>56-15935<br>56-17334 | AXS Screw |
| 29-56034<br>29-56035<br>29-56933<br>29-56934<br>29-56935 | Preloaded AXS Screw Disc |
### TABLE 5-1 - LIST OF SUBJECT DEVICES
{5}------------------------------------------------
#### V. INDICATIONS FOR USE
The Stryker Universal Neuro III System is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
The proposed modifications do not alter the Indications for Use statement for the proposed device. The Indications for Use is identical to the predicate device Stryker Universal Neuro III System (K112557).
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The modified Universal Neuro III System is compared to its predicate device for substantial equivalence of technological characteristics based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
# A. Principle of Operation
The basic operational principle of the Stryker Universal Neuro III System (UN III) remains the same: the operating principle for the UN III system is to reconstruct, stabilize and/or provide rigid fixation of non load-bearing areas. Subsequently, the principle of operation of the modified screw device and the screwdriver blade accessory remain the same as the predicate device.
# B. Technological Characteristics
The dimensional modification described in this special 510(k) is to the UN III System screw and screwdriver blade accessory. However, the modifications to the UN III screw and screwdriver blade accessory do not alter the technological and operational characteristics of the screw and the screwdriver blade accessory. The technological characteristics remain the same:
- Same operating principle: reconstruct, stabilize and/or provide rigid fixation of non । load-bearing bony areas
- Same mode of fixation: plate fixation with screws -
- Same area of contact and contact duration: the screws have contact to tissue/bone with a duration greater than 30 days; the blades, at worst, may contact breached or compromised surface for less than 24 hours
- Same material: Screw made from titanium alloy (ASTM F136) the blades are made from stainless steel (ASTM F899),
{6}------------------------------------------------
- -Similar design: The modifications to the UN III screw and screwdriver blade accessory are dimensional specifications.
# VII. PERFORMANCE DATA
Based on the Risk Analysis performed on the modifications to the UN III screw device and the modification to the screwdriver blade accessory, bench testing was performed in support of the substantial equivalence determination.
Biocompatibility and sterility testing was not required as a basis for substantial equivalence. There is no change in the material, manufacturing process, duration or location of contact, or reprocessing methods.
### Performance Bench Testing
The following performance bench tests were completed.
- -Self-Retention Test
- Screw Insertion Test -
- Life Cycle Test Fatigue (Screwdriver Blade) -
- Torque to Failure Test Quasi Static (Screwdriver Blade) -
- -End User / End Product Test
The subject device met all pre-defined acceptance criteria. Overall, the results of the performance bench tests support the proposed substantial equivalence of the subject device.
# Animal Testing
Animal testing was not required as a basis for substantial equivalence.
# Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
# VIII. CONCLUSIONS
The results of the non-clinical data demonstrate that the modified UN III screw and screwdriver blade will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.