K151341 · Norfolk Medical · FMI · Aug 20, 2015 · General Hospital
Device Facts
Record ID
K151341
Device Name
Turbo High-Flow Non-coring Needle
Applicant
Norfolk Medical
Product Code
FMI · General Hospital
Decision Date
Aug 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tidal High-Flow Non-coring Needle is intended for the access of totally implantable vascular access ports (VAPs) to administer I.V. fluids, infusions drugs, and other fluids, and for the administration of blood products, and/or withdrawal of blood as part of the therapy regimen
Device Story
Tidal High-Flow Non-coring Needle is a sterile, single-use, disposable hypodermic needle designed to access totally implantable vascular access ports (VAPs). The device features a 16Ga or 18Ga 304 stainless steel needle with an integrated stylet to ensure non-coring performance during septum puncture. Hubs are constructed from K-Resin and LDPE. Used by clinicians in clinical settings to facilitate fluid infusion or blood withdrawal. The device functions by piercing the VAP septum; the stylet prevents coring of the silicone septum, maintaining port integrity. Benefits include high-flow capability for efficient therapy delivery. Performance is verified through septum puncture and port leak testing.
Clinical Evidence
Bench testing only. Evaluated septum puncture and port leak performance using representative market ports. Testing included leak-to-failure analysis with 16Ga needles and 19Ga Huber point needles following 25 punctures with the 16Ga needle. Materials biocompatibility assessed per ISO 10993 and USP Class VI standards.
Technological Characteristics
16Ga and 18Ga non-coring needles; 304 stainless steel needle; K-Resin and LDPE hubs. Stylet-based non-coring mechanism. Sterile, single-use, ETO sterilized. Complies with ISO 10555-6:2015 and ISO 10993 biocompatibility standards.
Indications for Use
Indicated for patients requiring access to totally implantable vascular access ports (VAPs) for administration of IV fluids, drugs, blood products, or blood withdrawal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2015
Norfolk Medical Mr. Natan Pheil Product Development Manager 7350 N. Ridgeway Avenue Skokie, Illinois 60076
Re: K151341
Trade/Device Name: Tidal High-Flow Non-coring Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: May 13, 2015 Received: May 19, 2015
Dear Mr. Pheil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Natan Pheil
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151341
Device Name
Tidal High-Flow Non-coring Needle
Indications for Use (Describe)
The Tidal High-Flow Non-coring Needle is intended for the access of totally implantable vascular access ports (VAPs) to administer I.V. fluids, infusions drugs, and other fluids, and for the administration of blood products, and/or withdrawal of blood as part of the therapy regimen
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the words "Norfolk" and "Medical" in a serif font, with a red heart-shaped graphic in between. Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA •" followed by the phone number "847-674-7075".
# Section 5
# 510(k) Summary
## Tidal High-Flow Non-Coring Needle
# 510(k) Summary 21 CFR 807.92
### Part I General Information
#### 5.1 Submitter/Sponsor Information
5.2
5.3
| Submitter Name:<br>FDA Establishment | Norfolk Medical Products, Inc. |
|--------------------------------------|------------------------------------------|
| Registration Number: | 1450392 |
| Address: | 7350 N. Ridgeway<br>Skokie, IL 60076 |
| Telephone Number: | (847) 674-7075 |
| Fax Number: | (847) 674-7066 |
| Contact Person: | Natan Pheil, Product Development Manager |
| Device Name | |
| Trade Name: | Tidal High-Flow Non-coring Needle |
| Common Name: | Straight Non-coring Needle |
| Classification Panel: | 80 General Hospital |
| Classification: | 21 CFR 880.5570 |
| Class: | Class II |
| Product Code: | FMI - Needle, Hypodermic Single Lumen, |
| Predicate Device Name | |
| Device Name: | Arrow Non-coring Needle |
| Premarket Notification: | K961088 |
| Device Name: | Norfolk Medical Posi-Grip Needle |
| Premarket Notification: | K863721 |
| | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in a serif font, followed by a red heart-shaped symbol, and then the word "Medical" in a serif font. Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075".
#### 5.4 Reference Device Name
Device Name:
Premarket Notification:
Lucent Non-coring Needle Infusion Set (reference device) K11101
#### ર્સ્ટ Device Description
The 16Ga and 18Ga Tidal High-Flow Non-coring Needles are an extension of our 19, 20, and 22 gauge non-coring (Huber) Needle line used to access our line of totally implantable access devices. These large gauge non-coring needles incorporate a stylet to render them non-coring. The needle is made from 304 series stainless steel and the hubs are molded from K-Resin and LDPE. All materials are certified USP Class VI and are the exact same materials used to manufacture the products approved under K863721 and K111101. The needles are a disposable product that will be packaged sterile and non-pyrogenic in a single use peel pouch. ETO sterilization will be used. The sterilization will be done in-house using a validated sterilization procedure that is identical to the sterilization method used in previous 510(k) clearances.
#### ર્સ્વ Indications for Use Statement
The Tidal High-Flow Non-coring Needles are intended for the access of totally implantable vascular access ports (VAPs) to administer I.V. fluids, infusions drugs, and other fluids, and for the administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.
#### 5.7 Technological Characteristics and Substantial Equivalence
The proposed devices are equivalent to Arrow International's current legally marketed device cleared under 510(k) premarket notification K961088 (cleared on July 2nd, 1996). The difference consists of a different needle size. Whereas K961088 sought clearance for 14G and 16G needles, this application seeks clearance for 16G and 18G needles. The intended use, the basic design, and the function for the proposed devices are equivalent to the predicate device. They differ from the other predicate device, K863721 and the reference device, K111101, in that these large gauge non-coring needles incorporate a stylet rather than a deflected tip to render them non-coring. These non-coring devices (K863621 and K111101) are included, as we believe this application is an extension of K863721. The materials used are identical, and the tests performed on these devices can readily be referenced.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in a serif font, followed by a red heart-shaped graphic, and then the word "Medical" in a serif font. Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075".
#### ని.8 Discussion of Non-clinical Tests
Norfolk Medical develops product specifications based on design input and risk analysis activities related to the intended use of the product. These product specifications are used to create appropriate design verification tests with reference/guidance to established standards (listed below). For this application, the tests conducted were based on the following standards, specifically in reference to septum puncture / port leak information:
- FDA Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990
- ISO 10555-6:2015 - Intravascular catheters -- Sterile and single-use catheters -- Part 6: Subcutaneous implanted ports
### Performance Data:
The following bench tests were performed to evaluate the performance of the Tidal High-Flow Non-coring Needles:
- 1. Septum puncture / Port Leak to failure with the 16Ga high-flow, noncoring needles needles (worst case scenario) using a representative sample of ports found in the marketplace
- 2 Septum Puncture / Port Leak to failure with a 19Ga Huber point needle following 25 punctures with the 16Ga high-flow needle using a representative sample of ports found in the marketplace
The data collected from the non-clinical tests demonstrated that the functionality and performance characteristics of the Tidal High-Flow Non-coring Needles are comparable to the currently marketed needles based on available literature.
### Biocompatibility:
The materials used in the Tidal High-Flow Non-coring Needles are regularly used in the medical device industry. These materials have been well characterized chemically and physically in the published literature and have a long history of safe use.
The raw materials that make up device are as follows: 304 Stainless Steel, K-Resin and Low-density Polyethylene (LDPE). These materials comply with biocompatible requirements ISO 10993 and Pharmacopeia E.P/F.U.I/USP.
Referencing the FDA General Program Memorandum-#G95-1, "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing," we have been able to assess the biocompatibility
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Norfolk Medical. The logo consists of the word "Norfolk" in large, bold, black font, followed by a red heart-shaped symbol, and then the word "Medical" in the same font as "Norfolk". Below the logo is the address "7350 N. Ridgeway • Skokie, IL, USA • 847-674-7075".
requirements of the Tidal High-Flow Non-coring Needle.
According to the Blue Book Memorandum-#G95-1, attachment C, "Biocompatibility Flow Chart for the Selection of Toxicity Tests for 510(k)'s", the Tidal High-Flow Needle has met the biocompatibility requirements.
#### 5.9 Conclusion
The Tidal High-Flow Non-coring Needles have met all predetermined acceptance criteria of design verification evaluations through testing examination. Based on the FDA's decision tree, it is logically concluded through evidence that the Tidal High-Flow Non-coring Needles are substantially equivalent to the predicate devices listed above in Section 5.3.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.