K151337 · Centre Light Solutions, LLC · OHS · Feb 24, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151337
Device Name
Strialite
Applicant
Centre Light Solutions, LLC
Product Code
OHS · General, Plastic Surgery
Decision Date
Feb 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Strialite is an over-the-counter device indicated for use in the temporary reduction of redness in striae rubra (red stretch marks). Strialite does not remove the stretch marks but may improve their appearance.
Device Story
Strialite is a handheld, compact, OTC phototherapy/thermal device for home use by adults. It treats striae rubra (red stretch marks) on the abdomen, breasts, back, hips, and thighs. The device features an array of 35 red LEDs (633 nm ± 5nm) and a heating element that reaches up to 40°C (105°F). The user applies the device directly to the skin for 20-minute sessions, twice weekly for at least four weeks. The combination of red light and heat stimulates the skin at the cellular level to repair damaged collagen and increase local circulation, thereby reducing the visible redness of stretch marks. The device includes an over-temperature sensor for safety, shutting off if the plate exceeds 41°C. The user monitors the appearance of their stretch marks to assess improvement. The device is non-sterile and reusable.
Clinical Evidence
Clinical study (n=30) assessed temporary reduction of redness in striae rubra over 8 treatments/4 weeks. Primary endpoint: self-assessed improvement of at least 2 grades on a 1-10 scale; 16/30 subjects reported this improvement. Secondary endpoint: blinded photographic assessment identified 18/30 subjects with improvement. No adverse effects reported. Usability study (n=40) per AAMI ANSI IEC 62366:2007 showed 97.5% comprehension of instructions and 95% correct demonstration of critical use steps. Self-selection study (n=41) confirmed users could correctly identify device purpose and suitability from labeling.
Technological Characteristics
Handheld LED/thermal device. Materials: anodized aluminum and plastic. Light source: 35 red LEDs (633 nm ± 5nm). Thermal: heating element (max 40°C). Safety: over-temperature sensor (shuts off at >41°C). Power: corded. Standards: AAMI ES60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2. No software used.
Indications for Use
Indicated for adults under 50 years of age for the temporary reduction of redness in striae rubra (red stretch marks).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2016
Centre Light Solutions, LLC % Ms. Audrey Swearingen Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 August, Texas 78701
Re: K151337 Trade/Device Name: Strialite Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONE Dated: January 8, 2016 Received: January 12, 2016
Dear Ms. Swearingen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151337
Device Name Strialite
#### Indications for Use (Describe)
Strialite is an over-the-counter device in the temporary reduction of redness in striae rubra (red stretch marks). Strialite does not remove the stretch marks but may improve their appearance
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
for
# Strialite™
#### 1. Submission Sponsor
Centre Light Solutions, LLC 611 East Douglas Road Suite 108 Mishawaka Indiana, 46545 USA Phone: 574.612.3381 Contact: Ms. Carrie Colborn, Director of Operations
### 2. Submission Correspondent
Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: 512.818.3811 Office Phone: 512.327.9997 Fax: 512.327.9998 Contact: Audrey Swearingen, Director Regulatory Affairs Email: project.management@emergogroup.com
## 3. Date Prepared
February 19, 2016
### 4. Device Identification
| Trade/Proprietary Name: | Strialite™ |
|----------------------------|-----------------------------------------------------------------------------------------|
| Common/Usual Name: | Light Emitting Diode (LED) device |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. |
| Classification Regulation: | 21 CFR 878.4810 |
| Product Code: | ONE |
| Device Class: | Class II |
| Classification Panel: | General and Plastic Surgery |
#### 5. Legally Marketed Predicate Device(s)
Candela V Beam Pulse Dye Laser System, K013748
{4}------------------------------------------------
## 6. Device Description
The Strialite is an over the counter (OTC) device utilizing a combination of light emitting diode (LED) lights and a heating element as a phototherapy/thermal treatment for the end users' stretch marks (striae). It is a hand-held, compact, lightweight device, designed to specifically treat and reduce the visible appearance of an individual's red stretch marks. A stretch mark is a defect in the skin caused by damaged collagen that occurs from rapid skin stretching, that may have a red appearance. The red LED light and heat work together to stimulate the skin at the cellular level to produce and repair collagen. As the damaged collagen tissue is gradually replaced, over time the appearance of the stretch mark is reduced. Typical areas for treatment are the abdomen, breasts, back, hips, and thighs. The recommended treatment schedule is two (2) 20-minute sessions a week for at least four (4) weeks. Strialite is intended for home use by adults under the age of 50 years.
The Strialite device consists of a housing enclosure with an ergonomic handle, electronic circuit boards, treatment plate with a specific array of LED lights, thermal heating elements that generate temperature up to 105 degree F (40 degree C), touch control power button (On/Off), and a power cord. The flat LED panel, consisting of thirty-five (35) visible red LEDs in the 633 nm (±5nm) spectrum, is located on the base of the device, and directly contacts the area of skin to be treated. The device is provided non-sterile and reusable.
## 7. Indications for Use Statement
Strialite is an over-the-counter device indicated for use in the temporary reduction of redness in striae rubra (red stretch marks). Strialite does not remove the stretch marks but may improve their appearance.
### 8. Substantial Equivalence Discussion
As shown in Table 5A below, Strialite has the same intended use and similar technological characteristics and principles of operation as the predicate device, the Candela V Beam, and therefore is substantial equivalent to the predicate device.
| Manufacturer | Centre Light Solutions | Candela Corporation | |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Strialite | Candela V Beam Pulse Dye<br>Laser System | Comparison of<br>Significant Differences |
| 510(k) Number | N/A | K013748 | N/A |
| Product Code | ONE | ONE | Same |
| Regulation Number | 878.4810 | 878.4810 | Same |
| Regulation Name | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology | Same |
| Intended Use | Over the counter light<br>device for use in<br>dermatology | Prescription light device<br>for use in dermatology | Strialite is for OTC use,<br>which is supported by<br>the usability and self-<br>selection studies, and<br>clinical data. |
| Manufacturer | Centre Light Solutions | Candela Corporation | |
| Trade Name | Strialite | Candela V Beam Pulse Dye<br>Laser System | Comparison of<br>Significant Differences |
| Indications for Use | An over the counter device<br>indicated for use in the<br>temporary reduction of<br>redness in striae rubra.<br>Strialite does not remove<br>the stretch marks but may<br>improve their appearance. | A prescription device<br>indicated for use in<br>dermatology for the<br>treatment of striae. | The fundamental<br>technology and the<br>intended use to treat<br>stria rubra using and<br>heat and red light are<br>the same. |
| Hand-Held | Yes | Yes | Same |
| Patient Contact<br>Materials | Anodized aluminum<br>Plastic | Stainless steel<br>Plastic | Both devices utilize<br>materials commonly<br>used in medical devices<br>having limited skin<br>contact. |
| Sterile | No | No | Same |
| Single-Use | No | No | Same |
| Peak Emission<br>Wavelength | 633 ± 5nm | 595 ± 5nm | Both devices utilize<br>light in the red<br>spectrum. |
| Mechanism of<br>Action | Red light energy and dermal<br>heating increases local<br>circulation and collagen<br>production | Red light energy and<br>dermal heating increases<br>local circulation and<br>collagen production | Both devices use a<br>combination of red<br>light energy and heat |
| Software | No | Yes | To keep the user<br>functions simple, no<br>software is used in the<br>Strialite. |
| Recommended<br>Operating<br>Conditions | Room temperature | Room temperature | Both operate within<br>normal room<br>temperatures |
| Electrical Safety /<br>EMC Testing Passed | Yes | Yes | Same |
| Clinical Study | Yes | unknown | Strialite has been<br>shown to be safe and<br>effective for its<br>intended use. |
#### Table 5A - Comparison of Characteristics
{5}------------------------------------------------
## 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of Strialite and to show substantial equivalence to the predicate device, Centre Light Solutions completed a number of tests. The Strialite passed all testing stated in Table 5B below as shown by the acceptable results obtained. These tests confirm that the output meets the design inputs and specifications and support Strialite's safety, performance and substantial equivalence to the predicate device.
{6}------------------------------------------------
| Testing Type | Test Performed / Standard | Result |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design Verification | Strialite Over-Temperature Sensor<br>Verification Study | The Strialite shuts off when the plate<br>temperature exceeds 41 °C. |
| | Strialite Power Switch Reliability<br>Verification | The power switch of the device<br>successfully performed 500 ON/OFF<br>cycles. |
| | Strialite Overtemperature Sensor with<br>Runaway Resistors | In a single fault failure mode, the over-<br>temperature protection sensor shut-off<br>the heat well before the treatment<br>plate reached a hazardous temperature<br>(<51°C). |
| | Strialite Use Life – LED Specification | The LEDs exceeded the specification of<br>1000 hours of continuous use. |
| | Strialite Temperature Verification | The temperature of the treatment<br>plate met the specification at 1 minute<br>and at 20 minutes after the device is<br>turned on. The device operated<br>successfully within the specified<br>environmental temperature for the<br>stated time period. |
| Electrical Safety | AAMI ES60601-1:2005/A1:2012 (Ed.<br>3.1) - Medical electrical equipment<br>Part 1: General requirements for basic<br>safety and essential performance | Strialite complies with the stated<br>performance standard for electrical<br>safety for medical equipment used in a<br>home healthcare environment. |
| | IEC 60601-1-11: 2010, first ed. -<br>Medical electrical equipment Part 1-2:<br>General requirements for basic safety<br>and essential performance Collateral<br>Standard: Requirements for medical<br>electrical equipment and medical<br>electrical systems used in the home<br>healthcare environment | |
| | IEC 60601-2-57: 2011 - Medical<br>electrical equipment Part 2-57:<br>particular requirements for basic<br>safety and essential performance on<br>non-laser light source equipment<br>intended for therapeutic, diagnostic,<br>monitoring and cosmetic/aesthetic use | |
| | Electromagnetic<br>Compatibility | IEC 60601-1-2: 2007, 3rd ed. - Medical<br>electrical equipment Part 1-2: General<br>requirements for basic safety and<br>essential performance Collateral<br>Standard: Electromagnetic<br>Compatibility |
| | | Photobiological Safety |
Table 5B – Summary of Non-clinical Performance Testing
{7}------------------------------------------------
## 10. Clinical Performance Data
A clinical study was conducted to assess the effect of Strialite on the temporary reduction of redness of the subjects' red stretch marks. Thirty of the 33 subjects enrolled completed the full series of 8 treatments over 4 weeks. Sixteen out of 30 subjects reported, upon selfassessment, improvement of the severity of the redness of their stretch marks of at least 2 grades when graded on a scale of 1 to 10 from baseline to post-treatment. As a secondary endpoint blinded photographic assessment was performed and the blinded evaluation identified 18 of the 30 subjects as demonstrating improvement. Two subjects that selfassessed as having the worst appearing lesions (grade 10) did not show any improvement throughout the study. No adverse effects were observed or reported during the study.
As the Strialite is for OTC use, a usability study, designed in accordance with AAMI ANSI IEC 62366:2007, assessed whether subjects, after reading the labeling provided with the Strialite, were able to properly use the device without assistance. The usability test consisted of two components:
- Demonstrating the ability to read and comprehend the information in the Instructions 1) for Use and Quick Start Guide, based on survey questions asked by an interviewer.
- 2) Demonstrating correct simulated use of the device by following the Instructions for Use and Quick Start Guide, as observed and recorded by the trained interviewer.
The study showed that 39 of 40 (97.5%) subjects were able to read and comprehend the information in the Instructions for Use manual, and that 38 of 40 (95.0%) subjects were able to correctly demonstrate the six critical steps in the use of the Strialite.
In addition, a Self-selection study was performed with forty-one (41) subjects who were each given the device carton labeling to read in private. When interviewed, all responders correctly determined from the carton labeling that the device is used for stretch marks, and all but one correctly self-selected.
### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when it has the same intended use and the same technological characteristics as the previously cleared predicate device; or the device has the same intended use and different technological characteristics, but it can be demonstrated that the differences do not raise different questions regarding its safety and effectiveness as compared to the predicate device.
Strialite is substantially equivalent to the predicate device, Candela V Beam Pulse Dye Laser, in that they both have the same intended use and share similar technological characteristics and principles of operation. Both devices are intended for the same use which is the treatment of striae rubra (red stretch marks) by combining heat and light to reduce the redness of the stretch mark. It has been shown in this 510(k) submission that the differences between the Strialite and the Candela V Beam predicate device do not raise new questions regarding Strialite's safety and effectiveness. The Strialite is determined to be substantially equivalent to the predicate device.