The ZelanteDVT™ Thrombectomy Set is intended for the removal of thrombus from and infusion of fluids into the peripheral vasculature.
Device Story
ZelanteDVT Thrombectomy Set is a component of the AngioJet Ultra Thrombectomy System; requires AngioJet Ultra Console for operation. Device uses high-velocity saline jets to create a low-pressure zone at the catheter tip; thrombus is drawn into the catheter, fragmented by jets, and evacuated through waste tubing to a collection bag. Catheter is 105 cm, introduced via 8F sheath, and tracks over a 0.035" guidewire. Manifold includes a port for contrast or fluid injection. Used in clinical settings by physicians to treat peripheral vascular thrombus. Benefits include mechanical thrombus removal and targeted infusion of thrombolytic agents, potentially restoring blood flow in patients with DVT or other peripheral venous occlusions.
Clinical Evidence
No human clinical data provided. Evidence consists of extensive nonclinical bench testing, including biocompatibility (cytotoxicity, sensitization, hemolysis, thromboresistance, pyrogenicity), and in vitro performance testing (clot removal, net evacuation, tensile strength, guidewire compatibility, pressure testing). Safety and performance were further evaluated in acute and chronic GLP animal studies in peripheral veins.
Technological Characteristics
Catheter-based thrombectomy system; 105 cm length; 8F introducer sheath compatible; 0.035" guidewire compatible. Operates via high-velocity saline jet (Bernoulli principle) to create low-pressure zone for thrombus aspiration. Includes manifold with stopcock for fluid/contrast injection. Sterilized device. System requires external AngioJet Ultra Console for saline delivery and vacuum generation.
Indications for Use
Indicated for use with AngioJet Ultra Console to break apart and remove thrombus, including DVT, from iliofemoral, lower extremity, and upper extremity peripheral veins ≥ 6.0 mm in diameter. Also indicated for controlled, selective infusion of physician-specified fluids, including thrombolytic agents, into the peripheral vascular system.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines above them that resemble hair or a stylized representation of the human spirit.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 21, 2015
Boston Scientific Corporation c/o Ms. Anne V. Rossi Regulatory Affairs Manager One Scimed Place Maple Grove. MN 55311
Re: K151313
Trade/Device Name: ZelanteDVT Thrombectomy Set Regulation Number: 21 CFR 870.5150 Regulation Name: Catheter, Embolectomy Regulatory Class: Class II Product Code: DXE, KRA Dated: August 21, 2015 Received: August 24, 2015
Dear Ms. Rossi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
of Surveillance and Biometrics/Division of Postmarket Surveillance.
Sincerely yours,
# Kenneth J. Cavanaugh -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K151313
Device Name ZelanteDVT™ Thrombectomy Set
Indications for Use (Describe)
The ZelanteDVT™ Thrombectomy Set is intended for use with the AngioJet® Ultra Console to break apart and remove thrombus, including deep vein thrombus (DVT), from:
· Iliofemoral and lower extremity veins ≥ 6.0 mm in diameter and
· Upper extremity peripheral veins ≥ 6.0 mm in diameter.
The ZelanteDVT Thrombectomy Set is also intended for use with the AngioJet Ultra Power Pulse® technique for the controlled and selective infusion of physician specified fluids, including thrombolytic agents, into the peripheral vascular system.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
per 21 CFR §807.92
| Date Prepared | 15 May 2015 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311 |
| Contact Name and<br>Information | Anne V. Rossi<br>Regulatory Affairs Fellow<br>Phone: 763-494-1094<br>Fax: 763-494-2222<br>Email: Anne.Rossi@bsci.com |
| Trade Name(s) | ZelanteDVT™ Thrombectomy Set |
| Common or Usual Name | Catheter, Embolectomy |
| Product Code | DXE |
| Secondary Product<br>Code | KRA (Catheter, Continuous Flush) |
| Classification of Device | Class 2 according to 21 CFR 870.5150 |
| Predicate Device(s) | Solent™ Omni Thrombectomy Set K111182 24 May 2011 |
# Device Description
The ZelanteDVTT™ Thrombectomy Set is one component of the AngioJet® Ultra Thrombectomy System (AngioJet Ultra System). The other component, packaged and sold separately, is the multiple-use AngioJet® Ultra Console. The ZelanteDVT Thrombectomy Set can only be used in conjunction with the AngioJet Ultra Console.
The ZelanteDVT Thrombectomy Set uses high-velocity saline to create a low pressure zone at the catheter tip. Thrombus is drawn into the catheter where it is fragmented by the jets and evacuated from the body. The waste tubing transports the thrombus debris from the catheter to the waste collection bag for ultimate disposal.
The 105 cm ZelanteDVT Thrombectomy Catheter is introduced through an 8F introducer sheath and tracks and operates over a 0.035" (0.89 mm) guidewire. The manifold is equipped with a port with an attached stopcock. This port can be used to inject contrast media and other fluids into the bloodstream.
# Intended Use
The ZelanteDVT™ Thrombectomy Set is intended for the removal of thrombus from and infusion of fluids into the peripheral vasculature.
{4}------------------------------------------------
# Indications for Use
The ZelanteDVTT™ Thrombectomy Set is intended for use with the AngioJet® Ultra Console to break apart and remove thrombus, including deep vein thrombus (DVT), from:
- Iliofemoral and lower extremity veins ≥ 6.0 mm in diameter and
- · Upper extremity peripheral veins ≥ 6.0 mm in diameter.
The ZelanteDVT Thrombectomy Set is also intended for use with the AngioJet Ultra Power Pulse® technique for the controlled and selective infusion of physician specified fluids, including thrombolytic agents, into the peripheral vascular system.
# Comparison of Required Technological Characteristics
The ZelanteDVT Thrombectomy Set incorporates substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate AngioJet Solent Omni Thrombectorny Set (K111182).
# Summary of Nonclinical Tests
Nonclinical testing was performed on the ZelanteDVT Thrombectomy Set to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the device testing.
The following biocompatibility tests were completed on the ZelanteDVT Thrombectomy Set:
| Cytotoxicity | Acute Systemic Toxicity | In Vivo Thromboresistance |
|---------------------------|----------------------------|---------------------------------|
| Sensitization | Direct Contact Hemolysis | Materials Mediated Pyrogenicity |
| Intracutaneous Reactivity | In Vitro Hemocompatibility | USP Physicochemical |
The following in vitro performance tests were completed with the ZelanteDVT Thrombectomy Set:
| Shaft Outer Diameter | Hemolysis Ratio | Device Tensile |
|-------------------------|----------------------|---------------------------------|
| Effective Length | Contrast Injection | Tip Compression |
| Guidewire Compatibility | Extended Use | Manifold Torque |
| Operating Pressure | Thermal Dose | Torsional Buckling Strength |
| Net Evacuation | Torque Transference | Proximal Saddle Torque |
| Infusion Rate | Corrosion & FM | Manifold Fluid Immersion |
| Particulate | Freedom from Leaks | U-Bend |
| Clot Removal | Guidewire Lumen Leak | Distal Pressure Test |
| | | Packaging Bubble and Peel Tests |
Additionally Acute and Chronic GLP studies were performed to evaluate the safety and performance of the ZelanteDVT Thrombectomy Set in peripheral veins.
# Conclusion
Based on the indications for use, technological characteristics, safety and performance testing, the ZelanteDVT Thrombectomy Set is appropriate for the intended use and is considered to be substantially equivalent to the AngioJet Solent Omni Thrombectomy Set (K111182).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.