← Product Code [FGA](/productcode/FGA) · K151308

# Schoelly Nephroscope Set, Schoelly Ultra-Mini Nephroscope Set (K151308)

_Schoelly Fiberoptic GmbH · FGA · Feb 9, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151308

## Device Facts

- **Applicant:** Schoelly Fiberoptic GmbH
- **Product Code:** [FGA](/productcode/FGA.md)
- **Decision Date:** Feb 9, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters. The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.

## Device Story

Schoelly Nephroscope Family consists of rigid, reusable endoscopes, sheaths, obturators, and bridges for percutaneous nephrolithotomy (PCNL) and Mini PCNL. Device inputs include light from external sources; optical images transmitted via rod lens and fiber optic systems. Physician operates device under endoscopic control to visualize kidney stones; uses auxiliary instruments (lithotripters) through working channels or accessory setups to disintegrate and extract stones. Output is visual feedback on external monitor for surgeon guidance during stone removal. Used in clinical settings (OR/urology suite). Benefits include minimally invasive access for stone removal.

## Clinical Evidence

Bench testing only. No clinical data. Testing included temperature measurements per IEC 60601-2-18, optical parameter verification per ISO 8600, and biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, acute systemic toxicity). Reprocessing validation performed per AAMI TIR12, AAMI TIR30, ANSI/AAMI ST15883-1, ISO 14937, ANSI/AAMI ST81, ISO 17664, and ANSI/AAMI/ISO 17665-1.

## Technological Characteristics

Rigid, reusable endoscopes (6Fr-19Fr diameter, 220mm length). Materials: stainless steel, glass, glass fibers, adhesive, silicone, tin/silver alloy. Optical: 0°-8° direction of view, 85° field of view. Light transmission: fiber optics. Image transmission: rigid rod lens + fiber optics. Connectivity: external light source/camera/monitor. Sterilization: steam. Compliance: IEC 60601-2-18, ISO 8600.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Nephroscope Set ([K994223](/device/K994223.md))

## Reference Devices

- Compact Operating Fiber Uretero-Renoscopes ([K963855](/device/K963855.md))
- Karl Storz Adult and Pediatric Nephroscope ([K940594](/device/K940594.md))
- Cook Fiber Optic Bundle and Flexor Deflecting Access Sheath ([K072521](/device/K072521.md))
- Schoelly Sinuscope ([K142249](/device/K142249.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus-like design with three human profiles facing right. 

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 9, 2016

Schoelly Fiberoptic, GmbH Dr. Sandra Baumann Senior Manager Regulatory Affairs Robert-Bosch-Str. 1-3 79211 Denzlingen Germany

Re: K151308

Trade/Device Name: Schoelly Nephroscope Set, Schoelly Ultra-Mini Nephroscope Set Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Codes: FGA, FEC, FED Dated: December 23, 2015 Received: December 28, 2015

Dear Dr. Sandra Baumann,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151308

Device Name Schoelly Nephroscope Set Schoelly Ultra-Mini Nephroscope Set

#### Indications for Use (Describe)

The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters.

The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

02/08/2016 Schoelly Nephroscope Family

#### Section 5 – 510(k) Summary

#### I. SUBMITTER

Owner's Name: Address:

Telephone Number: Fax Number: Contact Person:

Schoelly Fiberoptic GmbH (Registration: 8043903) Robert-Bosch-Str. 1 - 3 79211 Denzlingen Germany +49-7666-980-0 +49-7666-908-380 Dr. Sandra Baumann

II. DEVICE Subject Device Name: Trade Name:

Common/Usual Name: Classification Name:

Schoelly Nephroscope Family Schoelly Nephroscope Set Schoelly Ultra-Mini Nephroscope Set Nephroscope Set FGA – Kit, Nephroscope 21 CFR 876.1500; Class II FEC - Obturator, for Endoscope 21 CFR 876.1500; Class II FED - Endoscopic Access Overtube, Gastroenterology-Urology 21 CFR 876.1500; Class II

#### III. PREDICATE/REFERENCE DEVICES

| Predicate Device Name:  | Nephroscope Set                                                                                                                                                                                                                                                                 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:             | Nephroscope Set                                                                                                                                                                                                                                                                 |
| Common/Usual Name:      | Nephroscope Set                                                                                                                                                                                                                                                                 |
| Classification Name:    | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II<br>FEC - Obturator, for Endoscope<br>21 CFR 876.1500; Class II<br>FED - Endoscopic Access Overtube,<br>Gastroenterology-Urology<br>21 CFR 876.1500; Class II<br>FFL - Ureteral stone dislodger<br>21 CFR 876.4680; Class II |
| Premarket Notification: | K994223, Richard Wolf Medical Instruments Corp.,<br>SE date June 5, 2000                                                                                                                                                                                                        |
| Recalls:                | This predicate has not been subject to a design-<br>related recall                                                                                                                                                                                                              |
| Reference Device Name:  | 12 Ureteroscopes/Ureterorenoscopes                                                                                                                                                                                                                                              |
| Trade Name:             | Compact Operating Fiber Uretero-Renoscopes                                                                                                                                                                                                                                      |
| Common/Usual Name:      | Ureteroscopes                                                                                                                                                                                                                                                                   |
| Classification Name:    | FGB - Ureteroscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II                                                                                                                                                                                                 |
| Premarket Notification: | K963855, Richard Wolf Medical Instruments Corp,<br>SE date April 3, 1997                                                                                                                                                                                                        |
| Reference Device Name:  | Karl Storz Adult and Pediatric Nephroscope                                                                                                                                                                                                                                      |
| Trade Name:             | Karl Storz Adult and Pediatric Nephroscope                                                                                                                                                                                                                                      |
| Common/Usual Name:      | Nephroscope                                                                                                                                                                                                                                                                     |
| Classification Name:    | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II                                                                                                                                                                                                                             |
| Premarket Notification: | K940594, Karl Storz Endoscopy America,<br>SE date September 09, 1994                                                                                                                                                                                                            |
| Reference Device Name:  | Cook Fiber Optic Bundle and Flexor Deflecting Access Sheath                                                                                                                                                                                                                     |
| Trade Name:             | Cook Fiber Optic Bundle and Flexor Deflecting Access Sheath                                                                                                                                                                                                                     |
| Common/Usual Name:      | Access Sheath                                                                                                                                                                                                                                                                   |
| Classification Name:    | FED - Endoscopic Access Overtube, Gastroenterology-<br>Urology<br>21 CFR 876.1500; Class II                                                                                                                                                                                     |
|                         | FFS - Image, illumination, fiberoptic, for endoscope<br>21 CFR 876.1500; Class II                                                                                                                                                                                               |
|                         | FAJ - Cystoscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II                                                                                                                                                                                                   |
|                         | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II                                                                                                                                                                                                                             |
|                         | GCJ - Laparoscope, general & plastic surgery<br>21 CFR 876.1500; Class II                                                                                                                                                                                                       |
|                         | FGB - Ureteroscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II                                                                                                                                                                                                 |
| Premarket Notification: | K072521, Cook Urological, Inc.,<br>SE date November 20, 2007                                                                                                                                                                                                                    |
| Reference Device Name:  | Schoelly Sinuscope                                                                                                                                                                                                                                                              |
| Trade Name:             | Schoelly Sinuscope                                                                                                                                                                                                                                                              |
| Common/Usual Name:      | Sinuscope                                                                                                                                                                                                                                                                       |
| Classification Name:    | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II                                                                                                                                                                                                        |
| Premarket Notification: | K142249, Schoelly Fiberoptic GmbH,<br>SE date January 27, 2015                                                                                                                                                                                                                  |

In addition to the Richard Wolf predicate device, the following reference devices were used in this submission to support substantial equivalence and to further outline technological characteristics:

{4}------------------------------------------------

{5}------------------------------------------------

#### IV. DEVICE DESCRIPTION

The proposed Schoelly Nephroscope Family comprises two different application sets: One set for Percutaneous Nephrolithotomy (PCNL) with the trade name "Schoelly Nephroscope Set'' and one set for Mini PCNL with the trade name "Ultra-Mini Nephroscope Set".

Both sets include a rigid, reusable endoscope (nephroscope) that is used in conjunction with a commercially available and approved light guide, light source. video camera, monitor, and printer. Light guide, light source, video camera, monitor, and printer are not included in the scope of delivery and are further not within the scope of this application.

In case of both sets, the nephroscope is accompanied by corresponding accessories to allow for access and passage of the nephroscope and retrieval instruments and for irrigation. Those include compatible endoscopic sheaths, obturators, and bridges.

The Schoelly Nephroscope Set and the Schoelly Ultra-Mini Nephroscope set are delivered in non-sterile conditions and have already obtained CE mark.

# Technological Characteristics

#### Endoscopes (Nephroscopes):

Light that is created by an external light source is transmitted from the nephroscope light guide connector through the nephroscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid rod lens system and a fiber optic bundle. The image can be displayed by a camera/monitor system which can be connected to the nephroscope eyepiece.

The main technical parameters that characterize the optical view of the nephroscopes of both application sets are the direction of view (0°-8°) and the field of view (85°). The insertion tube diameter of the nephroscopes is 6Fr - 19Fr and the insertion tube working length is 220mm.

The nephroscope of the Schoelly Nephroscope Set incorporates a working channel with a max. inner lumen capacity of 9Fr and is used in conjunction with an instrument bridge that is mounted to the nephroscope's proximal end (see "Bridges" Section below).

The nephroscope of the Schoelly Ultra-Mini Nephroscope Set does not have a working channel and is not equipped with an instrument bridge but is accompanied by additional accessories instead. The design of those accessories ensures that if the nephroscope is used in conjunction with them, the setup provides exactly the same functionality. Thus, if assembled appropriately, both application sets incorporate the number of inner lumens required to perform each single step of the intervention and to accommodate auxiliary instruments and irrigation.

{6}------------------------------------------------

Like other currently marketed nephroscopes, all Schoelly configurations have outer surfaces mainly made from stainless steel and further comprise fiber optics for light transmission and rigid rod lenses and fiber optics for image transmission.

# Sheaths:

The endoscopic sheaths included in this submission are rigid reusable instruments with inner lumens and mainly made from stainless steel; the sheaths serve as the most outer part of the whole device setup in PCNL and Mini PCNL procedures. The proximal end of the endoscopic sheaths is equipped with one side port for irrigation. The outer diameter of the sheaths in this submission ranges from 11Fr-24Fr to comply with the different nephroscopes and obturators of the proposed Family.

# Obturators

The obturators included in this submission are rigid reusable instruments mainly made from stainless steel with an inner lumen sized to accept the needed guide wire (not part of this submission) for correct placement of the sheath in the kidney. During insertion, the obturator fills the space inside the sheath to provide a smooth surface. The overall length of the obturators and their diameter comply with the length and the diameter of the available sheaths.

# Bridges:

This submission includes a standard instrument bridge, which is a rigid reusable instrument with an inner lumen and mainly made from stainless steel. It includes ports to allow for insertion of instruments and irrigation through the working channel of the nephroscope of the Schoelly Nephroscope Set. Channels can be sealed with a self-sealing membrane and a stop cock, respectively. It comprises an automatic locking mechanism for easy connection to the proximal end of the nephroscope.

# V. INDICATIONS FOR USE

The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters.

The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.

# VI. COMPARISON OF TECHNOLOGCIAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Indications for Use for the proposed Schoelly Nephroscope Family comprises a subset of the Indications for Use of the predicate device. Since the proposed Schoelly Nephroscope Family is not intended to be used via transurethral access as is the case for the Richard Wolf predicate device, the removal/extraction of bladder stones is not included in the Indications for Use for this submission. Remaining contents of the Indications for Use are identical for the proposed and the predicate device.

{7}------------------------------------------------

| A W Y A A A WA .                                                                                                                                                              |  |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
| AMAAAA                                                                                                                                                                        |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
| 1                                                                                                                                                                             |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
| НААН С С НАНАНИ О С                                                                                                                                                           |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
|                                                                                                                                                                               |  |
| 7                                                                                                                                                                             |  |
| 1                                                                                                                                                                             |  |
|                                                                                                                                                                               |  |
| . It also allerial and the seat the second the second the section of the section of the section of the section of the section of the section of the section of the section of |  |
|                                                                                                                                                                               |  |
| of the property and any and the property of the program and<br>A A A MA A                                                                                                     |  |
| -<br>1                                                                                                                                                                        |  |

| Technological Characteristics: Similarities and Differences between the Proposed Device and the Predicate Devices |                                                              |                                                              |                                                              |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| Attribute                                                                                                         | Proposed<br>Nephroscope Family<br>(current submission)       | Endoscope (Nephroscope)                                      | Predicate<br>Richard Wolf Nephroscope Set (K994223)          |
| Light transmission                                                                                                | Fiber optics                                                 | Fiber optics                                                 | Fiber optics                                                 |
| Light source                                                                                                      | External, connected via light guide to light guide connector | External, connected via light guide to light guide connector | External, connected via light guide to light guide connector |
| Image transmission                                                                                                | Rigid rod lenses + fiber optics                              | Rigid rod lenses, rigid rod lenses + fiber optics            | Rigid rod lenses, rigid rod lenses + fiber optics            |
| Eyepiece orientation                                                                                              | angled, straight                                             | angled, angled offset                                        | angled, angled offset                                        |
| Direction of view                                                                                                 | 0°-8°                                                        | 12°-25°                                                      | 12°-25°                                                      |
| Field of view                                                                                                     | 85°                                                          | 95° (Richard Wolf Long Miniature Endoscope)                  | 95° (Richard Wolf Long Miniature Endoscope)                  |
| Image display                                                                                                     | Camera/monitor connected via the endoscope eyepiece          | Camera/monitor connected via the endoscope eyepiece          | Camera/monitor connected via the endoscope eyepiece          |
| Insertion tube working length                                                                                     | 220mm                                                        | 175-225mm                                                    | 175-225mm                                                    |
| Insertion tube outer diameter                                                                                     | 6Fr-19Fr                                                     | 15Fr-27Fr                                                    | 15Fr-27Fr                                                    |
| Inner lumen capacity                                                                                              | max.3 Fr- max. 9 Fr                                          | max. 6Fr - max.12Fr                                          | max. 6Fr - max.12Fr                                          |
| Single Use / Reusable                                                                                             | Reusable                                                     | Reusable                                                     | Reusable                                                     |
| Reprocessing                                                                                                      | Cleaning, sterilization (steam)                              | Cleaning, sterilization (steam, H2O2)                        | Cleaning, sterilization (steam, H2O2)                        |
| Patient contacting materials                                                                                      | Stainless steel, glass, glass fibers, adhesive               | Stainless steel, glass, glass fibers, adhesive               | Stainless steel, glass, glass fibers, adhesive               |
| Electrical safety                                                                                                 | IEC 60601-2-18 compliant                                     | IEC 60601-2-18 compliant                                     | IEC 60601-2-18 compliant                                     |
| Instruments                                                                                                       |                                                              |                                                              |                                                              |
| Sheath outer diameter                                                                                             | 11Fr-24Fr                                                    | 15Fr-27Fr                                                    | 15Fr-27Fr                                                    |
| Obturator outer diameter                                                                                          | Compatible with 11Fr-24Fr sheath                             | Compatible with 15Fr-27Fr sheath                             | Compatible with 15Fr-27Fr sheath                             |
| Single Use / Reusable                                                                                             | Reusable                                                     | Reusable                                                     | Reusable                                                     |
| Reprocessing                                                                                                      | Cleaning, sterilization (steam)                              | Cleaning, sterilization (steam, H2O2)                        | Cleaning, sterilization (steam, H2O2)                        |
| Patient contacting materials                                                                                      | Stainless steel, silicone, tin/silver alloy                  | Stainless steel, silicone                                    | Stainless steel, silicone                                    |

{8}------------------------------------------------

# VII. PERFORMANCE DATA

The nephroscopes of the proposed Schoelly Nephroscope Family were subjected to temperature, optical parameter and biocompatibility testing.

Temperature testing: the devices were measured for surface temperatures at various locations as long as a steady state temperature was reached using different light sources and found to meet requirements as specified in IEC 60601-2-18.

Optical parameter testing: The devices were tested for all relevant optical parameters, e.g. field of view and direction of view accuracy and found to meet requirements as specified in ISO 8600.

Biocompatibility testing: A series of biocompatibility testing according to ISO 10993, including cytotoxicity, sensitization, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.

#### Reprocessing

The Schoelly Nephroscope Family was the subject of completed reprocessing validations including manual cleaning, automated cleaning and steam sterilization.

Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers), AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices), and ANSI/AAMI ST15883-1: 2009 (Washer-disinfectors - Part 1 - General requirements, terms, and definitions and tests).

Sterilization studies have been performed in accordance with ISO 14937:2009 (Sterilization of health care products - General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process for medical devices), ANSI/AAMI ST81:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), ISO 17664:2004 (Sterilization of medical devices -Information to be provided by the manufacturer for the processing of resterilizable medical devices), and ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of sterilization process for medical devices).

# VIII. CONCLUSION

The Schoelly Nephroscope Family meets all the pre-determined testing and acceptance criteria to effectively demonstrate substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K151308](https://fda.innolitics.com/device/K151308)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
