K151303 · Care Fusion · BZE · Jan 21, 2016 · Anesthesiology
Device Facts
Record ID
K151303
Device Name
AirLife Infant Heated Wire Circuit
Applicant
Care Fusion
Product Code
BZE · Anesthesiology
Decision Date
Jan 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatic population. specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 vears of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.
Device Story
AirLife Infant Heated Wire Circuit acts as airway conduit between breathing machine and patient; delivers and warms breathing gases. Device consists of dual-limb corrugated tubing with internal resistance wires; connects to Fisher and Paykel MR850 humidifier. Used in professional healthcare and intra-hospital transport environments by skilled clinicians under doctor supervision. Resistance wires generate heat to maintain gas temperature and humidity; prevents condensation and ensures optimal gas delivery to patient airway. Output is warmed, humidified gas; helps maintain patient airway moisture and temperature. Benefits include reduced risk of airway drying and improved patient comfort during mechanical ventilation.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 5367 (flow resistance, leakage, compliance, connector security) and ISO 8185 (humidity output, thermal overshoot, surface temperature). Biocompatibility testing performed per ISO 10993-1, -3, -5, -6, -10, -17, and -18.
Technological Characteristics
Dual-limb corrugated tubing; 304 steel resistance wires; copper/Ni alloy conductors. Compatible with Fisher and Paykel MR850 humidifier. Dimensions: 0.45-0.46in ID. Electrical safety per IEC 60601-1. Single-use, non-sterile. Standards: ISO 5367 (breathing tubes), ISO 8185 (humidifiers), ISO 10993 (biocompatibility).
Indications for Use
Indicated for pediatric population, specifically neonates (birth to 1 month) and infants (>1 month to 2 years), requiring warmed breathing gases delivered via Fisher and Paykel MR850 humidifier at flow rates >4 LPM in professional healthcare or intra-hospital transport settings.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
Predicate Devices
Fisher and Paykel Heated Breathing Circuit (K020332)
AirLife Heated Ventilator and Anesthesia Breathing Circuits (K000697)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 21, 2016
CareFusion 2200, Inc. Erika Fernandez Regulatory Affairs Manager 75 North Fairway Drive Vernon Hills, Illinois 60061
Re: K151303
> Trade/Device Name: AirLife Infant Heated Wire Circuit Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: December 20, 2015 Received: December 23, 2015
Dear Ms. Fernandez,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name AirLife Infant Heated Wire Circuit
#### Indications for Use (Describe)
The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatic population. specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 vears of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
## Summary of Safety and Effectiveness
510k summary complying with 21 CFR 807.92
## 1. SUBMITTER
CareFusion 2200, Inc. 75 N Fairway Drive, Vernon Hill, IL 60061
Phone: 847 362-8097 Fax: 312 949-0731
Contact Person: Erika Fernandez
Date Prepared: January 20, 2016
#### 2. Device
| Product Name: | AirLife Infant Heated Wire Circuit |
|----------------------|---------------------------------------------------------------------------|
| Device Name: | Heated Breathing Circuit |
| Common Name: | Heater, breathing system w/wo controller (not humidifier or<br>nebulizer) |
| Classification Name: | Breathing system heater (21 CFR 868.5270) |
| Regulatory Class: | II |
| Product Code: | BZE |
#### 3. Predicate Device
This submission demonstrates substantial equivalence to the Fisher and Paykel Heated Breathing Circuit K020332 cleared on July 700, 2003 and AirLife Heated Ventilator and Anesthesia Breathing Circuits, K000697 cleared on March 30, 2000. The predicate devices have not been subject to a design-related recall.
#### 4. Device Description
The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. It is provided non sterile for single patient use. The heated breathing circuit is used with Fisher and Paykel heated humidifiers.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".
## 5. Indication for use
The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatric population, specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 years of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.
## 6. Comparison of technological characteristics with the predicate device
The fundamental scientific technology is the same for both proposed and predicate devices. It is based on acting as an airway conduit between a breathing machine and the patient (typically attached to an endotracheal or tracheal tube previously insert into the patient's airway). The differences in the design do not raise any different questions of safety and effectiveness.
| Element of comparison | Predicate Device 1<br>Fisher and Paykel heated<br>wire breathing circuit<br>RT131 with flow rate<br>>4L/min | Predicate Device 2<br>AirLife Heated and<br>Anesthesia Breathing<br>Circuits<br>with flow rate >3L/min | Proposed Device<br>AirLife Infant Heated<br>Wire Circuit<br>with flow rate<br>>4L/min |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Principal of Operation | Resistance wires within<br>the tubing generate heat<br>to maintain<br>temperatures and<br>humidity | Resistance wires within<br>the tubing generate heat<br>to maintain<br>temperatures and<br>humidity | Resistance wires<br>within the tubing<br>generate heat to<br>maintain<br>temperatures and<br>humidity |
| Circuit Characteristics | | | |
| Infant Inspiratory Limb with<br>Unheated Extension | 5 ft | 5 ft | 5 ft |
| Dryline Length | 2 ft | 2 ft | 2 ft |
| Inspiratory Limb Length | 4 ft | 4 ft | 4 ft |
| Heated wire | Present | Present | Present |
| Expiratory Limb Length | 5ft 6in | 4 ft 3 in | 4 ft 5in |
| Water trap | Present | Not Present | Not Present |
| Heated Wire | Not Present | Present | Present |
| Patient population | Neonates | Infants, pediatric and<br>Adults | Infants and Neonates |
| Usage | Disposable | Disposable | Disposable |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top right. To the right of the circle is the text "CareFusion" in gray. Below the text is the phrase "has joined BD" in a smaller, lighter gray font.
| Element of comparison | Predicate Device 1<br>Fisher and Paykel heated<br>wire breathing circuit<br>RT131 with flow rate<br>>4L/min | Predicate Device 2<br>AirLife Heated and<br>Anesthesia Breathing<br>Circuits<br>with flow rate >3L/min | Proposed Device<br>AirLife Infant Heated<br>Wire Circuit<br>with flow rate<br>>4L/min |
|----------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Design | Dual Limb | Dual Limb | Dual Limb |
| Tube Specifications | | | |
| ID | 0.44in-0.45in | 0.394in | 0.45in-0.46in |
| Design | Corrugate | Corrugate | Corrugate |
| Wire Design | | | |
| Minimum Resistance at<br>ambient | 22 Ω | 27 Ω | 22 Ω |
| Wire Total Length | 115in | 99in | 121in |
| Constant Power | 22W | 16W | 22W |
| Typical Wattage/foot of<br>wire | 2 | 1.9 | 2 |
| Typical Watt/ foot of<br>circuit | 6 | 4 | 6 |
| Conductor | Copper | Copper/Ni alloy | 304 Steel |
| Anchor Position of Wire | 19 cm | 5-10 cm | 19 cm |
| Compatible Humidifier | Fisher and Paykel MR850 | Fisher and Paykel MR850 | Fisher and Paykel<br>MR850 |
## 7. Performance Data
The proposed device was tested to ensure compliance to the following standards:
### Biocompatibility
Tests for an externally communicating, tissue by way of gas path and direct mucosal contact with prolonged contact (greater than 24 hours but less than 30 days): Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller, lighter gray font.
## Standards
| Performance Characteristic | Standard |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Biological Evaluation of Medical Devices Part 1: Evaluation and<br>Testing<br>FDA Guidance: Use of International Standard ISO- 10993,<br>"Biological Evaluation of Medical Devices Part 1: Evaluation<br>and Testing" | ISO 10993-1:2009 |
| Biological Evaluation of Medical Devices Part 3: Tests for<br>genotoxicity, carcinogenicity and reproductive toxicity | ISO 10993-3:2014 |
| Biological Evaluation of Medical Devices Part 5: Tests for in<br>vitro cytotoxicity | ISO 10993-5:2009 |
| Biological Evaluation of Medical Devices Part 6: Tests for local<br>effects after implantation | ISO 10993-6:2007 |
| Biological Evaluation of Medical Devices Part 10: Tests for<br>irritation and skin sensitization | ISO 10993-10:2010 |
| Biological Evaluation of Medical Devices Part 17:<br>Establishment of allowable limits for leachable substances | ISO 10993-17:2002 |
| Biological Evaluation of Medical Devices Part 18: Chemical<br>characterization of materials | ISO 10993-18:2005 |
## Performance
The following tests were performed for the proposed device to support the substantial equivalence decision.
| Test | Description | Relevant Standard |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Electrical safety<br>Patient Leakage Current | To ensure, under normal system<br>operating conditions, that<br>excessive current does not<br>reach the patient | IEC 60601-1:2005, 3rd<br>Edition (Equivalent to<br>AAMI ES60601-1:2005) |
| Electrical safety<br>Dielectric Strength Means | To check the integrity of the<br>wire insulation under excessive<br>power conditions. | IEC 60601-1:2005, 3rd<br>Edition (Equivalent to<br>AAMI ES60601-1:2005) |
| Electrical Safety<br>Insulation | To ensure IEC requirements for<br>creepage and clearance were<br>met. | IEC 60601-1:2005, 3rd<br>Edition (Equivalent to<br>AAMI ES60601-1:2005) |
| Length | To ensure that the actual length<br>of the device is within ±10% of<br>the labeled length | ISO 5367:2000, 4th Edition |
| Test | Description | Relevant Standard |
| Resistance to Flow/Rated<br>Flow | To ensure, when tested per the<br>standard, that the pressure<br>increase at the rated flow<br>through the breathing tube does<br>not exceed 0.2 kPa | ISO 5367:2000, 4th Edition |
| Resistance to Flow @ 2.5<br>LPM | To ensure, when tested per the<br>standard, that the flow<br>resistance at 2.5 LPM through<br>the breathing tube does not<br>exceed 0.74 cmH2O/l/min | ISO 5367:2014, 5th Edition |
| Increase in Flow Resistance<br>w/Bending @ Rated Flow<br>(10 LPM) | To ensure, when tested per the<br>standard, that the pressure at<br>the rated flow when the<br>breathing tube is suspended<br>over a metal cylinder does not<br>exceed 150% of the value<br>obtained with the tube straight. | ISO 5367:2000, 4th Edition |
| Increase in Flow Resistance<br>w/Bending @ 2.5 LPM | To ensure, when tested per the<br>standard, that the pressure at<br>2.5 LPM when the breathing<br>tube is suspended over a metal<br>cylinder does not exceed 150%<br>of the value obtained with the<br>tube straight. | ISO 5367:2014, 5th Edition |
| Conical Connectors | To ensure, when tested per the<br>standard, that required ports are<br>correctly sized. | ISO 5367:2000, 4th Edition<br>ISO 5356-1:2004, 3rd Edition<br>ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Security of Attachment | To ensure, when tested per the<br>standard, that the adaptor shall<br>not become detached from the<br>tube at a force greater than 45N. | ISO 5367:2000, 4th Edition |
| Leakage | To ensure, when tested per the<br>standard, that the system<br>supplied ready to use does not<br>leak at a rate greater than 50<br>ml/min. | ISO 5367:2000, 4th Edition |
| Compliance per Meter | To ensure, when tested per the<br>standard, that the compliance<br>per meter at a pressure of 6 kPa<br>does not exceed 10 ml*kPa-1<br>per meter length of tube. | ISO 5367:2000, 4th Edition<br>ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Test | Description | Relevant Standard |
| Humidity Output (Invasive<br>Mode) | To ensure, when tested per the<br>standard, that the system<br>delivers at least the minimum<br>humidity of at least 33 mg/L to<br>the patient. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Humidity Output (Non-<br>Invasive Mode) | To ensure, when tested per the<br>standard, that the system<br>delivers at least the minimum<br>humidity of at least 10 mg/L to<br>the patient. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Corrugate Melt (100% Duty<br>Cycle, Zero Gas Flow) | To ensure that the breathing<br>tubes do not collapse on<br>bending, occlude, or otherwise<br>cause a safety hazard at the<br>maximum output of the<br>humidification system. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Security of engagement<br>temperature sensor | To ensure that the temperature<br>probes that accompany the<br>circuit do not become<br>disconnected under the<br>condition of no flow or<br>maximum rated flow. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Leakage from sensing port | To ensure, when tested per the<br>standard, that the sensing port<br>does not leak more than 5<br>ml/min. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Specific enthalpy | To ensure under normal and<br>single fault conditions, a thermal<br>overshoot at the patient<br>connection port shall not exceed<br>an energy equivalent to 43C and<br>100% relative humidity (a<br>specific enthalpy not to exceed<br>194 kJ/kg dry gas) when<br>averaged over 30 seconds. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Surface temperature | To ensure that the accessible<br>temperature within 25 cm of the<br>patient port connection does<br>not exceed 44°C | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Steady state noise | To ensure, when tested per the<br>standard, that the noise at any<br>point within 1 m of the system<br>does not exceed 50 dB. | ISO 8185:2008, 3rd Edition<br>(Corrected Version) |
| Test | Description | Relevant Standard |
| Duration of Use (30 days) | To assess the circuit use over a<br>30 day time period | N/A |
| Longitudinal Temperature<br>Profile | To assess the temperature range<br>of the wire once a nominal<br>voltage is induced across it. | N/A |
| Patient End Temperature<br>Profile | To assess the temperature of the<br>airway gas delivered to the<br>patient when a given humidified<br>circuit is used with a ventilator<br>and the Fisher & Paykel MR850<br>heater base. | N/A |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield-like shape inside. To the right of the logo is the text "CareFusion" in a dark gray, sans-serif font. Below the company name, in a smaller, lighter font, is the phrase "has joined BD".
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield-like shape inside. To the right of the logo is the text "CareFusion" in a dark gray, sans-serif font. Below the company name, in a smaller, lighter font, is the phrase "has joined BD".
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, in smaller gray font, it says "has joined BD."
## 8. Conclusion
The test results demonstrate that the device is as safe and effective as the predicate and therefore substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.