The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.
Device Story
Disposable and reusable blood pressure cuffs; accessory to noninvasive blood pressure measurement systems. Disposable version: non-woven fabric and PVC; single-patient use. Reusable version: 210D Nylon TPU and PVC; multi-patient use. Both versions include bladder and air hose (single or double tube). Used in clinical settings to facilitate blood pressure monitoring; cuff inflates to occlude artery; system measures pressure oscillations. Output provided to host monitor for clinical decision-making regarding patient hemodynamic status.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 81060-1 (non-invasive sphygmomanometers) verified.
Technological Characteristics
Materials: Non-woven fabrics/PVC (disposable), 210D Nylon TPU/PVC (reusable). Pressure range: 0-300 mmHg. Configuration: Single or double tube. Sizes: Neonatal, infant, pediatric, adult (AHA compliant). Non-sterile. Mechanical cuff/bladder system; no electronic or software components.
Indications for Use
Indicated for neonatal, infant, pediatric, and adult patient populations requiring noninvasive blood pressure measurement as an accessory to noninvasive blood pressure measurement systems.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2015
Xuzhou Maicuff Technology Co.,ltd. c/o Mr. Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 1-202, Build 3, Beijing New World, No.5 Chaoyang Rd., Chaoyang District Beijing, 100024 CN
Re: K151290
> Trade/Device Name: Disposable Blood Pressure Cuff, Reusable Blood Pressure Cuff -Models: M5541-1#, M5541-2#, M5541-3#, M5514-4#. M5542. M5543. M5544. M5545. M5546: M5301. M5302, M5303, M5304, M5305; M5641-1#, M5641-2#, M5641-3#, M5614-4#, M5642, M5643, M5644, M5645, M5646; and M5401, M5402, M5403, M5404, M5405 Regulation Number: 21 CFR 870.1120 Regulation Name: Blood-Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: April 29, 2015 Received: May 15, 2015
Dear Mr. Ray Wang,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 - Mr. Ray Wang
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151290
### Device Name
Disposable/Reusable Blood Pressure Cuff, Models -M5541-1#,M5541-2#,M5541-3#,M5542,M5543,M5544,M5545,M5546;M5301,M5302,M5304,M5305 M5641-1#,M5641-2#,M5641-3#,M5614-4#,M5642,M5643,M5646; and M5401,M5402,M5403,M5404,M5405
Indications for Use (Describe)
The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 2 L14 14 L2 14 Z M3 3 L13 3 L13 13 L3 13 Z M4 7 L6 9 L12 3 L13 4 L6 11 L3 8 Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 2 L14 14 L2 14 Z M3 3 L13 3 L13 13 L3 13 Z" fill="white"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K151290
# Page 1 of 6
# 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________
- 1. Date of Preparation:2015/5/5
- 2. Sponsor Identification
## XUZHOU MAICUFF TECHNOLOGY CO., LTD.
8#-2-1106 Jinxiujiayuan, Jianguo West Road, Xuzhou City, Jiangsu Province, China
Establishment Registration Number: Pending
Contact Person: Li Zhaoqian Position: General Manager Tel: +86-516-85908810 Fax:+86-516-85908830 Email: info@maicuff.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
## Beijing Believe Tech. Service Co., Ltd
Tel: +86-21-50313932, Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com
{4}------------------------------------------------
K151290 Page 2 of 6
#### Identification of Proposed Device 4.
Trade Name: Disposable/Reusable Blood Pressure Cuff Common Name: Blood Pressure Cuff Model(s): M5541-1#,M5541-2#,M5541-3#,M5514-4#,M5542,M5543,M5544,M5545,M5546; M5641-1#,M5641-2#,M5641-3#,M5614-4#,M5642,M5643,M5644,M5645,M5646; M5301,M5302,M5303,M5304,M5305; M5401,M5402,M5403,M5404,M5405;
### Regulatory Information
Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number: 870.1120 Review Panel: Cardiovascular
### Indication For Use Statement:
The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.
The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.
### Device Description
The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, pediatric and adult sizes.
The proposed device includes disposable blood pressure cuff and reusable blood pressure, both disposable and reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.
The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics & PVC (Cuff) and PVC (Air Hose), thereinto, the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient.
The disposable blood pressure cuff has 18 models with different size for different population with different arm size, and the 9 models of 18 are use single tube of air hose and other 9 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
The reusable blood pressure cuff is reusable device, and which is made of 210D Nylon TPU (Cuff)
{5}------------------------------------------------
K151290
and PVC (Air Hose), thereinto, the 210D Nylon TPU (Cuff) is the material used to contacting with the patient.
The reusable blood pressure cuff has 10 models with different size for different population with different arm size, and the 5 models of 10 are use single tube of air hose and other 5 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
Both disposable and reusable blood pressure cuff are provided as non-sterile.
- Identification of Predicate Device(s) 5.
Predicate Device 510(k) Number: K112544 and K120364
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro > Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- > ISO 81060-1:2012 Non-Invasive Sphygmomanometers - Part 1: Requirements And Test Methods For Non-Automated Measurement Type
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
{6}------------------------------------------------
K151290 Page 4 of 6
#### Substantially Equivalent (SE) Comparison 8.
| ITEM | Proposed Device<br>(Disposable Blood Pressure Cuff) | Predicate Device (K120364) | Remark |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Intended Use | The disposable blood pressure cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and for<br>single-patient use. It is available in<br>neonatal, pediatric and adult sizes. | The Unimed Blood Pressure Cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and for<br>single-patient use. It is available in<br>neonatal, pediatric and adult sizes. The cuff<br>is no designed, sold, or intended for use<br>except as indicated. | SE |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | |
| Tube<br>Configuration | One or two tube | One or two tube | |
| Size | Conform to AHA bladder sizes<br>recommendations<br>Neonatal 1 (3-5.5 cm)<br>Neonatal 2 (4-8 cm)<br>Neonatal 3 (6-11 cm)<br>Neonatal 4 (7-13 cm)<br>Infant (9-14.5 cm)<br>Pediatric (13-21.5 cm)<br>Small Adult (21-27 cm)<br>Adult (26-35.5 cm)<br>Large Adult(35-45 cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal 1 (3-6 cm)<br>Neonatal 2 (4-8 cm)<br>Neonatal 3 (6-11 cm)<br>Neonatal 4 (7-13 cm)<br>Neonatal 5 (8-15 cm)<br>Infant (9-14.8 cm)<br>Child (13.8-21.5 cm)<br>Small Adult (20.5-28.5 cm)<br>Adult (27.5-36.5 cm)<br>Adult Long (27.5-36.5 cm)<br>Large Adult(35.5-46 cm)<br>Large Adult Long(35.5-46 cm)<br>Thigh (45-56 cm) | SE |
| Single Use | Y | Y | |
| Sterile | No | No | |
| Pressure Limits | 0-300 mmHg | 0-300 mmHg | |
# Table 1 General Comparison (Disposable BP Cuff)
{7}------------------------------------------------
## K151290
| ITEM | Proposed Device<br>(Reusable Blood Pressure Cuff) | Predicate Device (K112544) | Remark |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Intended Use | The reusable blood pressure cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and may be<br>reused. It is available in infant, pediatric<br>and adult sizes. | The Unimed Blood Pressure Cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and may be<br>reused. It is available in pediatric and adult<br>sizes. The cuff is no designed, sold, or<br>intended for use except as indicated. | SE |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | SE |
| Tube<br>Configuration | One or two tube | One or two tube | SE |
| Size | Conform to AHA bladder sizes<br>recommendations<br>Infant (9-14.5 cm)<br>Pediatric (13-21.5 cm)<br>Small Adult (21-27 cm)<br>Adult (26-35.5 cm)<br>Large Adult (35-45 cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11 cm)<br>Infant (10-19 cm)<br>Pediatric (18-26 cm)<br>Small Adult (20-28 cm)<br>Adult (25-35 cm)<br>Adult Long(25-35 cm)<br>Large Adult (33-47 cm)<br>Large Adult Long(33-47 cm)<br>Adult Thigh (46-66 cm) | SE |
| Single Use | No | No | SE |
| Sterile | No | No | SE |
| Pressure Limits | 0-300 mmHg | 0-300 mmHg | SE |
Table 2 General Comparison (Reusable BP Cuff)
Table 3 Biocompatibility Comparison
| ITEM | Proposed Device | Predicate Device | Remark |
|---------------|----------------------------------------------------------------|--------------------------|--------|
| Cytotoxicity | Under the conditions of the study, not<br>cyteotoxicity effect | Comply with ISO 10993-5 | SE |
| Irritation | Under the conditions of the study, not an<br>irritant | Comply with ISO 10993-10 | SE |
| Sensitization | Under conditions of the study, not a<br>sensitizer. | | SE |
{8}------------------------------------------------
K151290 Page 6 of 6
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.