← Product Code [HRS](/productcode/HRS) · K151277

# Airlock® osteosynthesis plate system (K151277)

_Novastep · HRS · Aug 13, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151277

## Device Facts

- **Applicant:** Novastep
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Aug 13, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Airlock® osteosynthesis plate systems are single use devices intended for fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients.

## Device Story

Airlock® osteosynthesis plate system consists of titanium (Ti-6Al-4V ELI) bone plates and 3mm/3.5mm locking and non-locking screws. Intended for permanent implantation to provide fixation and stabilization of small bone fractures, revisions, joint fusions, and reconstructions in hands, feet, wrists, ankles, fingers, and toes. Used by surgeons in clinical settings. Plates feature varying curvatures, lengths, and hole configurations to accommodate anatomical requirements; drill holes are aligned to prevent screw conflict. Device provides mechanical stability to bone segments during healing.

## Clinical Evidence

No clinical data or animal studies were required for this submission. Biocompatibility was evaluated per ISO 10993-1 and Blue Book Memorandum #G95-1. Substantial equivalence is supported by bench testing and mechanical performance data.

## Technological Characteristics

Materials: Wrought titanium 6-aluminium 4-vanadium alloy (ISO 5832-3). Mechanical standards: ASTM F382-99 (plates), ASTM F543-13 (screws). Components: Plates of varying shapes/lengths, 3mm/3.5mm locking and non-locking screws. Single-use, permanently implanted.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Anchorage® Bone Plate System ([K083447](/device/K083447.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 13, 2015

NOVASTEP Mr. Gilles Audic, QA/RA Director Espace performance Alphasis Batiment C1-C2 35769 SAINT GREGOIRE France

Re: K151277

Trade/Device Name: Airlock® osteosynthesis plate system Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: May 21, 2015 Received: May 26, 2015

Dear Mr. Audic:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Mr. Gilles Audic

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Section

### 4: Indications for Use Statement

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Form Approved: OMB No. 0910-0120                              |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| <i>Indications for Use</i>                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Expiration Date: January 31, 2017<br>See PRA Statement below. |
| 510(k) Number (if known)                                                | K151277                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                               |
| Device Name                                                             | Airlock® osteosynthesis plate systems                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                               |
| Indications for Use (Describe)                                          | Airlock® osteosynthesis plate systems are indicated for fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients.                                                                                                                                                                      |                                                               |
| Type of Use (Select one or both, as applicable)                         | <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="2,2 10,2 10,10 2,10" style="fill:black"></polygon> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |                                                               |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  
Food and Drug Administration  
Office of Chief Information Officer  
Paperwork Reduction Act (PRA) Staff  
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for "novastep". The logo is in a pink color. To the left of the word "novastep" is a pink design that looks like two curved lines. To the right of the word "novastep" is a registered trademark symbol.

#### "510(k) Summary" as required by section 807.92(c)

| Submitter        | NOVASTEP<br>Espace performance Alphasis<br>Batiment C1-C2<br>35769 SAINT GREGOIRE<br>France<br>Phone : + 33 (0)2 99 33 86 50<br>Fax : + 33 (0)9 70 29 18 95 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person   | Mister Gilles AUDIC<br>QA / RA Director<br>Cell phone: +33 (0)6 30 93 96 08<br>e-mail: gilles.audic@novastep-ortho.com                                      |
| Preparation date | May 12th 2015                                                                                                                                               |

| Trade name          | Airlock® osteosynthesis plate system                                                                                |
|---------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name         | Plate, Fixation, Bone                                                                                               |
| Classification Name | Single / Multiple component metallic bone fixation appliances and<br>accessories (21CFR 888.3030, product code HRS) |
| Regulatory class    | II                                                                                                                  |

| Legally marketed<br>predicate devices | 510(k) number: K083447<br>Device name: Anchorage® Bone Plate System<br>Original applicant: MEMOMETAL TECHNOLOGIES.<br>This predicate has not been subject to a design-related recall.                                                                                                   |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description                           | Airlock® osteosynthesis plate systems are single-use bone fixation<br>devices intended to be permanently implanted. Plates are designed with<br>different shapes and are made of Titanium (Alloy Ti-6Al-4V ELI).<br>The system uses either 3mm or 3,5mm locking and non-locking screws. |

vastep - Siège Social / Service Client

| n o v a s t e p - S iège Social / Service Client
 Espace Performanoe Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIRE Cedex - FFRANCE
 Tél. +33 (0)2 99 33 86 50 - Fa

Tél. +33 (0)2 99 33 86 50 - Fax: +33 (0)2 30 96 35 61

Code A.P.E.: 3250A - N° SIRET: 752 292 797 00020-N° TVA Intracommunautaire FR 64 752 292 797

{4}------------------------------------------------

|                                                                                    | The drill hole are aligned to make sure there is no risk of conflict between<br>the screws.<br>The plates vary essentially through different curvatures, lengths, number<br>of holes and shape.                                                                                                                                                                                                                                                                                                                                 |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use                                                                       | Airlock® osteosynthesis plate systems are single use devices intended for<br>fixation and stabilization of fresh fractures, revision procedures, joint<br>fusion and reconstruction of small bones of the hand, feet, wrist, ankles,<br>fingers and toes. The system may be used in both adult and pediatric<br>patients.                                                                                                                                                                                                       |
| Comparison of the<br>technological<br>characteristics with the<br>predicate device | The new devices Airlock® osteosynthesis plate systems plates have similar<br>technological characteristics in terms of material (ISO5832-3 Implants for<br>surgery - Metallic material - Part 3: Wrought titanium 6-aluminium 4-<br>vanadium alloy) and mechanical characteristics (ASTM F382-99 Standard<br>Specification and Test Methods for Metallic Bone plates) and thus are<br>believed to be substantially equivalent to the predicate MEMOMETAL<br>Anchorage® Bone Plate System plates (K083447).                      |
|                                                                                    | The new devices Airlock® osteosynthesis plate systems screws have<br>similar technological characteristics in terms of material (ISO5832-3<br>Implants for surgery - Metallic material - Part 3: Wrought titanium 6-<br>aluminium 4-vanadium alloy) and mechanical characteristics (ASTM F543-<br>13 Standard Specification and Test Methods for Metallic Bone screws Part<br>A1, A2 & A3) and thus are believed to be substantially equivalent to the<br>predicate MEMOMETAL Anchorage® Bone Plate System screws<br>(K083447). |
| Performance data                                                                   | The biocompatibility evaluation for new devices Airlock® osteosynthesis<br>plate systems was conducted in accordance with Blue Book<br>Memorandum #G95-1 "Use of International Standard ISO-10993,<br>'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"<br>and International Standard ISO 10993-1 "Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and Testing Within a Risk Management<br>Process" as recognized by FDA.                                                              |
|                                                                                    | The new devices Airlock® osteosynthesis plate systems have similar<br>technological characteristics in terms of design and mechanical<br>characteristics and thus are believed to be substantially equivalent to the<br>predicate MEMOMETAL Anchorage® Bone Plate System (K083447).                                                                                                                                                                                                                                             |
| Indication for use                                                                 | The Airlock® osteosynthesis plate systems are single use devices indicated<br>for fixation and stabilization of fresh fractures, revision procedures, joint<br>fusion and reconstruction of small bones of the hand, feet, wrist, ankles,<br>fingers and toes. The system may be used in both adult and pediatric<br>patients.                                                                                                                                                                                                  |
| Clinical studies                                                                   | Clinical studies was not required for this submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Animal Study                                                                       | Animal Study was not required for this submission<br>1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1                                                                                                                                                                                                                                                                                              |
| Conclusion                                                                         | The Airlock® osteosynthesis plate systems are substantially equivalent to<br>their predicate devices MEMOMETAL Anchorage® Bone Plate System<br>(K083447), in terms of intended use and indications for use, material,<br>design and function. Any minor differences between these two devices<br>do not raise new questions of safety and effectiveness.                                                                                                                                                                        |

vastep - Siège Social / Service Client

| n o v a st e p - Siège Social / Service Client
| Espace Perbrimance Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIRE Cedex - FFRANCE
| Tél. +33 (0)2 99 33 86 50 - Fax

Code A.P.E.: 3250A-N° SIRET 752 292 797 00020 - N° TVA Intracommunautaire FR 64 752 292

{5}------------------------------------------------

#### Traditional 510(k) premarket notification

# step®

| n o v a s t e p - S iège Social / Service Client
| Espace Performance Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIFE Cedex - FFANCE
| Tél. +33 (0)2 99 33 86 50 - F

---

**Source:** [https://fda.innolitics.com/device/K151277](https://fda.innolitics.com/device/K151277)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
