← Product Code [MAX](/productcode/MAX) · K151276

# DePuy Synthes T-PAL Ti Spacer (K151276)

_Synthes USA Products, LLC · MAX · Jun 18, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151276

## Device Facts

- **Applicant:** Synthes USA Products, LLC
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jun 18, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DePuy Synthes T-PAL Ti Spacer is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the T-PAL Spacers should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The T-PAL Spacers are intended to be used with DePuy Synthes supplemental fixation.

## Device Story

Interbody fusion device providing structural stability post-discectomy; used in lumbar spine (L2-S1). Device features bullet-nose design for insertion, pyramidal teeth for stabilization, and open architecture for autogenous bone graft packing. Manufactured from titanium alloy (TAN); available in various heights/footprints with 5° lordotic angulation. Used by surgeons in clinical settings; requires supplemental fixation. Benefits include stabilization of spinal segments in DDD patients.

## Clinical Evidence

Bench testing only. Finite Element Analysis (FEA) performed to simulate static and dynamic axial compression and compression shear testing to evaluate the impact of material and design modifications.

## Technological Characteristics

Material: Ti-6Al-7Nb (TAN) per ASTM F1295. Form factor: Interbody spacer with pyramidal teeth, bullet-nose design, and 5° lordotic angle. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified (modality change noted).

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Synthes T-PAL Spacer ([K100089](/device/K100089.md))

## Reference Devices

- Patriot Signature® Ti Spacer ([K122097](/device/K122097.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

June 18, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Synthes USA Products, LLC Mr. Eric Zhu Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts 02767

Re: K151276

Trade/Device Name: DePuy Synthes T-PAL Ti Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: May 12, 2015 Received: May 13, 2015

Dear Mr. Zhu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director, Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use Statement

## 510(k) Number: K151276

Device Name: DePuy Synthes T-PAL Ti Spacer

Indications For Use:

The DePuy Synthes T-PAL Ti Spacer is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the T-PAL Spacers should be packed with autogenous bone graft (i.e. autograft).

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment.

The T-PAL Spacers are intended to be used with DePuy Synthes supplemental fixation.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

### A. Submitter Information Manufacturer:

Submitter:

Mezzovico, CH-6805 Switzerland

Zona Industriale 4

Synthes Production GmbH

Synthes USA Products, LLC 325 Paramount Drive, Raynham MA 02767

Contact Person:

Telephone number: Fax number: Email:

B. Date Prepared

C. Device Name

Trade/Proprietary Name:

Common/Usual Name:

Classification Name:

#### D. Predicate Device Name

Predicate Trade name: Reference Device:

Device Description E.

Eric Zhu 325 Paramount Drive Raynham, MA 02767 508-828-3730 508-828-3797 Ezhu@its.jnj.com

June 17, 2015

DePuy Synthes T-PAL Ti Spacer

Spinal Intervertebral Body Fusion Device

Intervertebral Fusion Device with Bone Graft, Lumbar Intervertebral Body Fusion Device per 21 CFR §888.3080 Product Code MAX (Intervertebral body fusion device)

Synthes T-PAL Spacer (K100089) Patriot Signature® Ti Spacer (K122097) 

{4}------------------------------------------------

The DePuy Synthes T-PAL Ti Spacer is an interbody fusion device used in conjunction with supplemental fixation to provide structural stability in skeletally mature individuals following total or partial discectomy. Previously, only T-PAL Spacers manufactured from PEEK (With TAN marker pins) material were available. The subject T-PAL Ti spacers will be manufactured entirely of TAN per ASTM F1295. The T-PAL Ti Spacer is available in two footprints and a range of heights and is angulated 5° to accommodate the lordotic curve (except for the smallest height of each footprint, which does not have a lordotic angle). Pyramidal teeth that assist in stabilization of the construct are located on the inferior and superior surfaces of the spacers. These teeth are oriented along a contour that follows the curve of the implant to assist in implantation. A bullet-nose design also facilitates self-distraction and ease of insertion. The open architecture of the devices allows them to be packed with autogenous bone graft material, i.e. autograft.

#### F. Intended Use

The DePuy Synthes T-PAL Ti Spacer is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the T-PAL Spacers should be packed with autogenous bone graft (i.e. autograft).

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The T-PAL Spacers are intended to be used with DePuy Synthes supplemental fixation.

## Summary of Similarities and Differences in Technological Characteristics, F. Performance and Intended Use

The proposed modifications to the Synthes T-PAL Spacer are identical to the predicate device (K100089) except for the material change, sterilization modality change, and a minor modification to the design. The indications, performance, and technology remain identical to the predicate systems. We are not replacing the predicate device with the proposed Titanium Spacer device, but we are making additional T-PAL Ti Spacers available for the market.

#### G. Materials

T-PAL Ti Spacers will be manufactured from Ti-6Al-7Nb (TAN) ASTM F-1295 implant grade titanium alloy.

{5}------------------------------------------------

#### H. Performance Data

The minor modification to the design of the device and material change do not change the technological characteristics of the device. A Finite Element Analysis simulating Static/Dynamic Axial Compression and Compression Shear testing demonstrated that the proposed modifications do not introduce a new worst case.

#### I. Conclusion

The Substantial Equivalence Justification demonstrate that the subject devices are substantially equivalent to the predicate devices because the intended use remain unchanged and technological characteristics of the subject device are similar to the predicate. The Finite Element Analysis and post market data further demonstrates that the material change and modification to the design do not introduce any new questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K151276](https://fda.innolitics.com/device/K151276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
