K151273 · Inmode MD , Ltd. · GEI · Jan 4, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151273
Device Name
InMode FRF Applicator
Applicant
Inmode MD , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InMode FRF Applicator is intended for use in Dermatologic and General Surgical procedures for Electrocoagulation and Hemostasis. At higher energy levels greater than 62 mJ/pin, use of the FRF applicator is limited to Skin Types I-IV.
Device Story
The InMode FRF Applicator is a handpiece attachment for the InMode WMFace system; it delivers fractional radiofrequency (RF) energy to skin via a disposable 24-electrode pin array. The system includes an RF generator, controller, and LCD interface; a footswitch triggers energy delivery. Used in clinical settings by healthcare providers for electrocoagulation and hemostasis. The device monitors RF parameters to enhance safety. By delivering controlled thermal energy, it facilitates tissue coagulation, benefiting patients requiring dermatologic or general surgical procedures. The provider uses the interface to set parameters and monitors the procedure to ensure appropriate clinical outcomes.
Clinical Evidence
No clinical data. Evidence consists of bench performance testing (RF output accuracy/consistency) and a pre-clinical swine study evaluating thermal effects and tissue healing via histology at 0, 7, 14, and 21 days post-treatment.
Technological Characteristics
Fractional RF multielectrode technology; 24-electrode pin array tip. RF output frequency: 1 MHz ± 2%; Max power: 65 Watts. System includes AC/DC power supply, RF generator, controller, and LCD interface. Biocompatible tip materials. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-2 (HF surgical equipment).
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis. Use at energy levels >62 mJ/pin is restricted to patients with Skin Types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center of the logo is an abstract symbol of a human figure, represented by three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 4, 2016
InMode MD Ltd. Ms. Ahava Stein A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 44425 Israel
Re: K151273
Trade/Device Name: InMode FRF Applicator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: OUH Dated: December 1, 2015 Received: December 4, 2015
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K151273
Device Name InMode FRF Applicator
#### Indications for Use (Describe)
The InMode FRF Applicator is intended for use in Dermatologic and General Surgical procedures for Electrocoagulation and Hemostasis.
At higher energy levels greater than 62 m.Jpin, use of the FRF applicator is limited to Skin Types I-IV.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(K) SUMMARY INMODE FRF APPLICATOR
## 510(k) Number K151273
#### Applicant Name:
| Company Name: | InMode MD Ltd. |
|---------------|-------------------------------|
| Address: | Tabor Building, Shaar Yokneam |
| | Yokneam 20692 |
| | Israel |
| | Tel: +972-4-9097470 |
| | Fax: +972-4-9097471 |
| | E-mail: ahava@asteinrac.com |
#### Contact Person:
#### Official Correspondent: Ahava Stein
| Company Name: | A. Stein – Regulatory Affairs Consulting Ltd. |
|----------------|---------------------------------------------------------|
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 44425<br>Israel |
| Tel: | +972-9-7670002 |
| Fax: | +972-9-7668534 |
| E-mail: | ahava@asteinrac.com |
| Date Prepared: | December 29, 2015 |
Trade Name: InMode FRF Applicator
Classification Name: CFR Classification section 878.4400; Classification Product Code: OUH Subsequent Product Code: GEI
Classification: Class II Medical Device
#### Predicate Device:
The InMode FRF Applicator is substantially equivalent to the following predicate device;
| Manufacturer | Device | 510(k) No. |
|-------------------------|---------------------|------------|
| EndyMed<br>Medical Ltd. | Intensif Applicator | K130501 |
{4}------------------------------------------------
#### Device Description:
The InMode FRF Applicator is a treatment hand piece attached the FDA cleared InMode WMFace treatment system (K140926).
The InMode WMFace device with the FRF Applicator employs fractional RF multielectrode technology for procedures requiring electrocoagulation and hemostasis. The FRF Applicator is designed to deliver radiofrequency energy to the skin in a fractional manner, via an array of multi-electrode pins. The Device provides enhanced safety while minimizing possible side effects by monitoring RF parameters.
The InMode WMFace device with the FRF Applicator consists of an AC/DC power supply unit, RF generator, controller and user interface including a LCD screen and functional buttons. The FRF Applicator is connected to the console via a cable and a foot switch activates the energy delivery to the hand piece. The hand piece comprises a disposable, single use, 24 electrode pin array tip.
Following are the InMode WMFace device with the FRF Applicator specifications: FRF Applicator RF Max Output Power: 65 Watt FRF Applicator RF Output Frequency: 1[MHz] ± 2% Dimension: 46cm W x 46cm D x 100cm H (14.2'' W x 14.2'' D x 40'' H) Weight: 30 Kg (66 lbs) Main Line Frequency (nominal): 50-60 Hz Input Voltage (nominal): 100-240 VAC
#### Intended Use/Indication for Use:
The InMode FRF Applicator is intended for use in Dermatologic and General Surgical procedures for Electrocoagulation and Hemostasis .
At higher energy levels greater than 62 mJ/pin, use of the FRF applicator is limited to Skin Types I-IV.
#### Performance Standards:
The InMode FRF Applicator has been tested and complies with the following voluntary recognized standards:
- AAMI/ANSI 60601-1:2005/(R)2012 And A1:2012,, C1:2009/(R)2012 And ● A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod).
- IEC 60601-1-2 Edition 3: 2007-03. Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
{5}------------------------------------------------
- IEC 60601-2-2 Edition 5.0 2009-02, Medical Electrical Equipment Part 2-2: ● Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.
#### Biocompatibility & Sterilization
All of the InMode FRF Applicator tip materials are biocompatible.
The InMode FRF Applicator tip is for single use and provided non-sterile to the user. The user must follow the sterilization procedures as specified in the user manual and sterilize the Applicator tip in approximation to the treatment.
#### Non-Clinical (Bench) Performance Data:
Performance bench tests were conducted to measure the accuracy and consistency of the RF output parameters of the InMode FRF Applicator and compare them to the specific design requirements and to the RF output parameters of the predicate device. The results of the bench tests demonstrate that the InMode FRF Applicator complies with the design requirements and consists of similar RF output specifications as the predicate device and therefore, is substantially equivalent to the predicate device.
#### Animal Performance Data / Histology Data:
The thermal effects of the InMode FRF Applicator and tissue healing processes were evaluated in a pre-clinical study. The study was conducted on swine model and included a single RF treatment followed by histology analysis performed immediately. 7, 14 and 21 days post treatment. The animal study results show that the InMode FRF Applicator is safe for use and effective in achieving the specified indications of dermatological and general electrocoagulation and hemostasis.
## Clinical Performance Data:
Not Applicable
## Substantial Equivalence:
The indications for use and technological characteristics of the InMode FRF Applicator are substantially equivalent to the indications for use and technological characteristics of the EndyMed Intensif Applicator.
The design and components in the InMode FRF System, including the console (with power supply, RF generator, controller and display panel) and the hand piece Applicator (with cable, connector to console, handle and tip) are similar to the design and components found in the predicate Endymed Intensif System. The performance specifications of the InMode FRF Applicator were shown to be similar and yielded RF energy per pin values in the range of the EndyMed Intensif Applicator specifications. The safety features and compliance with safety standards in the InMode FRF Applicator are similar to the safety features and compliance with safety standards found in the
{6}------------------------------------------------
predicate device. Patient contact materials are also similar. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new InMode FRF Applicator underwent performance testing, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2, performance bench testing and animal pre-clinical testing to evaluate the thermal effect of the device and the tissue healing process. These performance tests demonstrated that the minor differences in the device design and specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the InMode FRF Applicator is substantially equivalent to the predicate EndyMed Intensif Applicator, cleared under 510(k) K130501, and therefore, may be legally marketed in the USA.
#### Conclusions:
Conclusively, based on the performance testing, pre-clinical testing and comparison to predicate device, the InMode FRF Applicator is substantially equivalent to the EndyMed Intensif Applicator predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.