AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup, AssureTech Secobarbital/Oxycodone Turn Key-Split Cup

K151211 · Assure Tech. Co., Ltd. · DIS · Jun 4, 2015 · Clinical Toxicology

Device Facts

Record IDK151211
Device NameAssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup, AssureTech Secobarbital/Oxycodone Turn Key-Split Cup
ApplicantAssure Tech. Co., Ltd.
Product CodeDIS · Clinical Toxicology
Decision DateJun 4, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3150
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K151211 · Jun 4, 2015AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup, AssureTech Secobarbital/Oxycodone Turn Key-Split CupAssure Tech. Co., Ltd.Unaltered clinical urine samples; Lay-user study participantsClinical samples were used to validate the device performance against GC/MS (gold standard) and to assess the performance of the device by lay users in a simulated real-world setting.Method comparison; Lay-user study; Clinical urine samples

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Method comparison study; In-house method comparison using clinical samplesUnaltered clinical urine samples; Sample Size: 80 samples per device; Number of Sites: In-houseGC/MSQualitative detection of Secobarbital and Oxycodone
Lay-user study; Prospective lay-user performance evaluation1060 lay persons with diverse backgrounds; Sample Size: 1060; Number of Sites: 3GC/MSCorrect results percentage

Indications for Use

The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL. The test may yield preliminary positive results when prescription drug Secobarbital is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use. The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL. The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use. The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxyoodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use. The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use. The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

Device Story

AssureTech drug test series (Strips, Panel Dips, Quick Cups, Turn-Key Split Cups) are lateral flow immunochromatographic assays for qualitative detection of Secobarbital and/or Oxycodone in human urine. Devices function via competitive binding: target drugs in urine compete with immobilized drug-conjugates for binding sites on monoclonal mouse antibody-coated particles. Absence of a test line indicates drug presence above cut-off; presence of a line indicates drug absence or levels below cut-off. A control line confirms proper flow. Used by lay users (OTC) or clinicians (prescription) for preliminary screening. Results are visual. Preliminary positive results require confirmatory testing via GC/MS. Benefits include rapid, point-of-care identification of potential drug use, facilitating clinical decision-making regarding further diagnostic or therapeutic steps.

Clinical Evidence

Bench testing only. Precision studies (n=50 per concentration per lot) across 25 days confirmed performance at cut-off levels. Interference and specificity studies evaluated cross-reactivity with various compounds. Method comparison study (n=80 clinical samples) compared device results to GC/MS. Lay-user study (n=1060) demonstrated high accuracy across diverse demographics and ease of use (Flesch-Kincaid Grade Level 7).

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle. Uses monoclonal mouse antibodies. Qualitative visual readout. Specimen: human urine. Cut-offs: 300ng/mL (Secobarbital), 100ng/mL (Oxycodone). Stable at 4-30°C for 24 months. No electronic components or software.

Indications for Use

Indicated for qualitative detection of Secobarbital (300ng/mL) and/or Oxycodone (100ng/mL) in human urine. Intended for over-the-counter and prescription use as a preliminary screening tool. Not for quantitative measurement. Positive results require confirmation by GC/MS.

Regulatory Classification

Identification

A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) #: K151211 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...