The OsteoMed ICON IMF Screw System is intended and indicated for the temporary ligature and wire lock fixation for temporary constriction and stabilization of fractured bone segments in the oral cavity in conjunction with primary fixation devices.
Device Story
System consists of 2.0mm and 2.4mm titanium alloy screws and stainless steel ligature wire; used for temporary stabilization of fractured jaw bone segments. Implantation involves either pilot hole creation (standard screws) or self-drilling (AutoDrive screws). Instruments include drill bits, wire cutters, pliers, drill guides, and screwdrivers. Used in oral cavity by clinicians; screws provide rigid fixation points for wire to prevent bone segment movement. Benefits include temporary stabilization of fractures during healing. Non-sterile; requires sterilization before use.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included failure torque, insertion torque, pull-out force, and bending force testing per ASTM F543. Results demonstrated equivalence to primary predicate (K962774) and reference predicate (K974785).
Technological Characteristics
Materials: Titanium alloy (ASTM F136) for screws, stainless steel for wire. Instrumentation: stainless steel, anodized aluminum, medical grade polymer. Principle: Rigid fixation and wire lock stabilization. Connectivity: None. Sterilization: Non-sterile, requires user sterilization. Form factor: 2.0mm/2.4mm diameter screws, 10-20mm lengths.
Indications for Use
Indicated for temporary ligature and wire lock fixation for temporary constriction and stabilization of fractured bone segments in the oral cavity in conjunction with primary fixation devices.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 21,2015
Osteomed Piedad Pena Regulatory Affairs Manager 3885 Arapaho Rd Addison, Texas 75001
Re: K151195
Trade/Device Name: Osteomed Imf Screw Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw Or Wire Regulatory Class: Class II Product Code: DZL Dated: September 4, 2015 Received: September 10, 2015
Dear Piedad Pena:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin Keith Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K151195
Device Name: OsteoMed ICON IMF Screw System
Indications for Use:
The OsteoMed ICON IMF Screw System is intended and indicated for the temporary ligature and wire lock fixation for temporary constriction and stabilization of fracture bone segments in the oral cavity in conjunction with primary fixation devices.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OsteoMed. The logo is in an oval shape with a blue top and a gray bottom. The word "OsteoMed" is written in blue and gray inside the oval.
# 510(k) SUMMARY
#### OsteoMed ICON IMF Screw System
| Name of Submitter: | OsteoMed<br>3885 Arapaho Road<br>Addison, TX 75001<br>972 677 4600<br>972 677 4601 (fax) |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mrs. Piedad Pena, M.S. |
| Date Prepared: | September 4, 2015 |
| Device Proprietary Name:<br>Device Common Name:<br>Classification Name:<br>Product Code: | OsteoMed ICON IMF Screw System<br>IMF Fixation Screw and wire<br>21CFR 872.4880: Intraosseous fixation screw or wire<br>DZL |
Primary Predicate: OsteoMed QUICK-FIX, K962774 Classification Name: 21CFR 872.4880: Intraosseous fixation screw or wire; Product Code: DZL; Class: II
Reference Predicate: Auto-Drive Bone Screw System, K974785 Classification Name: 21CFR 888.3040, Smooth or threaded metallic fixation fastener, Product Code HWC, Device Class: II
Summary:
#### Device Description:
The OsteoMed ICON IMF (Intermaxillary Fixation) Screw System is offered in 2.4mm and 2.0mm standard IMF screws in lengths from 10mm (thread lengths from 5mm to 15mm) and 2.0mm ICON AutoDrive IMF Screws in lengths from 12mm to 18mm (thread lengths of 8mm).
The implants of the OsteoMed ICON IMF Screw System are made from Titanium alloy (ASTM F-136) and the ligature wire is made from Stainless Steel.
The system instruments include drill bits, wire cutters, pliers, drill guides and screwdrivers to facilitate the placement of screws and ligature wires. The instrumentation is made from various grades of stainless steel, anodized aluminum, and/medical grade polymer.
#### Intended Use/Indications for Use:
The OsteoMed ICON IMF Screws System is intended and indicated for the temporary ligature and wire lock fixation for temporary constriction and stabilization of fractured bone segments in the oral cavity in conjunction with primary fixation devices.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 (973) 677-4600 FAX: (800) 390-2620 Customer Service: (800) 456-7779
Image /page/3/Picture/14 description: The image contains the logo for Colson Associates. The logo consists of the word "Colson" inside of a circle with a vertical line through it. To the right of the circle is the text "A Colson Associate".
{4}------------------------------------------------
Summary of Technological Characteristics:
| System / Device<br>Name/510(k)/<br>Purpose | OSTEOMED ICON IMF Screw<br>System (New) | OSTEOMED Quick-fix System<br>K962774<br>Primary Predicate | OsteoMed Auto-Drive<br>Bone Screw K974785<br>(Reference Predicate)<br>self-driving tip features |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II<br>21 CFR 872.4880<br>DZL | Class II<br>21 CFR 872.4880<br>DZL | Class II<br>21 CFR 888.3040<br>HWC |
| Device Name<br>Classification: | Intraosseous fixation screw or<br>wire. | Intraosseous fixation screw or<br>wire. | Smooth or threaded<br>metallic bone fixation<br>fastener. |
| | a) Identification. An<br>intraosseous fixation screw or<br>wire is a metal device intended<br>to be inserted into fractured<br>jaw bone segments to prevent<br>their movement. | a) Identification. An<br>intraosseous fixation screw or<br>wire is a metal device intended<br>to be inserted into fractured jaw<br>bone segments to prevent their<br>movement. | |
| Indication for<br>Use/Intended<br>Use | The device is intended and<br>indicated for the temporary<br>ligature and wire lock fixation<br>for temporary constriction and<br>stabilization of fracture bone<br>segments in the oral cavity in<br>conjunction with primary<br>fixation devices. | The device is intended and<br>indicated for the temporary<br>ligature and wire lock fixation<br>for temporary constriction and<br>stabilization of fracture bone<br>segments in the oral cavity in<br>conjunction with primary<br>fixation devices. | Fixation secondary to<br>trauma or<br>reconstruction of the<br>craniofacial and<br>maxillofacial skeleton and<br>bones of the<br>hand |
| Technology | The screws fixate on bone in<br>conjunction with ligature wire<br>to allow fixation and<br>stabilization of the fracture. | The screws fixate on bone in<br>conjunction with ligature wire to<br>allow fixation and stabilization<br>of the fracture. | The screws fixate on bone<br>in conjunction with plates<br>for rigid fixation.<br>(N/A, this device does not<br>used plates) |
| Screw Driving<br>technology for<br>implantation | The standard screws are<br>implanted after a pilot hole has<br>been created in the desired<br>location.<br>AutoDrive screws do not<br>require pilot hole for<br>implantation, they are self-<br>drilling screws. | The standard screws are<br>implanted after a pilot hole has<br>been created in the desired<br>location. | AutoDrive screws do not<br>require pilot hole for<br>implantation, they are self-<br>drilling screws. |
| Screw Material | Ti6Al4V per ASTM F136 | Ti6Al4V per ASTM F136 | Ti6Al4V per ASTM F136 |
| Principles of<br>Operation | Rigid Fixation and Wire Lock<br>stabilization | Rigid Fixation and Wire Lock<br>stabilization | Rigid Fixation |
| | Table 1: Technology Comparison to the Primary Predicate and Reference Device |
|--|------------------------------------------------------------------------------|
| | |
{5}------------------------------------------------
| System / Device<br>Name/510(k)/<br>Purpose | OSTEOMED ICON IMF Screw<br>System (New) | OSTEOMED Quick-fix System<br>K962774<br>Primary Predicate | OsteoMed Auto-Drive<br>Bone Screw K974785<br>(Reference Predicate)<br>self-driving tip features |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Screw<br>(type and length<br>range according<br>to screw<br>diameter) | 2.4mm IMF Screw (standard)<br>Length range: 10mm-20mm<br>Thread range: 5mm to 15mm<br>(Same as Primary Predicate;<br>cleared K962774)<br><br>2.0mm IMF Screw (standard)<br>Length range: 10mm-20mm<br>Thread range: 5mm to 15mm<br><br>2.0mm IMF AutoDrive Screw<br>Length range: 12mm-18mm<br>Thread range: 8mm-14mm | 2.4mm IMF Standard Screw<br>Length range: 10mm-20mm<br>Thread range: 5mm to 15mm<br>(Cleared) | 2.0mm AutoDrive Screw<br>Length range: 4mm-8mm<br>Thread range: 3mm-7mm<br>(Cleared) |
| Instrumentation<br>(Materials) | • Stainless Steel<br>• Anodized aluminum<br>• Medical grade polymer | • Stainless Steel<br>• Anodized aluminum<br>• Medical grade polymer | • Stainless Steel<br>• Anodized aluminum<br>• Medical grade polymer |
| Instrumentation | drill bits, wire cutters, pliers,<br>drill guides and screwdrivers to<br>facilitate the placement of<br>screws and ligature wire | drill bits, wire cutters, pliers,<br>drill guides and screwdrivers to<br>facilitate the placement of<br>screws and ligature wire | screwdrivers to facilitate<br>the placement of screws |
| Sterilization | Packaged and sold Non-sterile,<br>requires sterilization prior to<br>use | Packaged and sold Non-sterile,<br>requires sterilization prior to use | Packaged and sold Non-<br>sterile, requires<br>sterilization prior to use |
| Biocompatibility | Ti alloy per ASTM F136 is a<br>biocompatible material | Ti alloy per ASTM F136 is a<br>biocompatible material | Ti alloy per ASTM F136 is a<br>biocompatible material |
# Substantial Equivalence Discussion:
The OSTEOMED 2.0mm ICON Standard IMF Screw and the 2.0mm ICON AutoDrive IMF Screw are the same as the predicate device (2.4mm QuickFix MMF Screw, K962774) in the following ways:
- The screws have the same Regulatory Classification (21 CFR 872.4880 DZL)
- The screws have the same indication for use.
- The technology and features of the screws are the same (screws to fixate and allow ligature with wire)
- The range of the screw length is the same as, or encompassed by, that of the predicate:
- o Overall Length = 10mm 20mm
- o Thread Length = 5mm 15mm
- . The length of the screw head, the head diameter, and cross-sectional areas through the wire holes are the same as the predicate device.
- . The 2.0mm ICON Standard IMF Screw has the same number of wire fixation holes (1) as the predicate
{6}------------------------------------------------
- The screws are manufactured from the same materials (Titanium alloy per ASTM F136) as the predicate and therefore have the same biocompatibility.
- The instrumentation and materials used in conjunction with these screws are the same
- The screws are packaged and sterilized in the same manner and same parameters as the predicate device.
Differences between the new devices and the predicate, do not affect safety and effectiveness
- The 2.0mm ICON AutoDrive IMF Screw has more wire fixation holes (2) than the predicate (1) ●
- The 2.0mm ICON Standard IMF Screw has a smaller cross-sectional area through the wire fixation holes (1.43 mm²) than the predicate (2.04 mm²)
- The 2.0mm ICON AutoDrive IMF Screw has a different tip (sharp, pointed, single threadcutting flute) than the predicate (blunt, rounded) but was equivalent to the reference predicate.
The performance testing comparison listed below showed the new devices were equivalent and did not raise new safety and effectiveness issues.
## Performance/Clinical Data:
The OsteoMed ICON IMF Screws (2.0mm Standard IMF screws and the 2.0mm AutoDrive IMF screws) were compared to the OsteoMed QuickFix 2.4mm standard MMF Screws (K962774-primary predicate) and the 2.0mm AutoDrive Screws (K974785-reference predicate). The 2.0mm AutoDrive IMF screws and the 2.0mm standard IMF screws underwent verification testing to ensure that the design features meet the required mechanical strength criteria for their intended use compared to the predicate device. The intended use of the OsteoMed ICON IMF Screw System is the same as OsteoMed 2.4mm QuickFix screw system (K962774).
The performance testing included failure torque testing of the new devices and the predicate devices in accordance with ASTM F543. All devices met the minimum failure torque requirement based on screw diameter according to the standard. Also, insertion torque testing in accordance with ASTM F543 was performed for all devices met the required failure torque to insertion torque (FT/IT) ratio of FT/IT > 1.0 according to the standard. Lastly, pull-out force and bending force testing was performed on the new devices using equivalent methods to those shown in the primary predicate 510k K962774 and compared against the predicate device results in K962774. The performance testing showed that the new devices were equivalent to the predicate device. Simulated use was performed as part of validation for the new screws and were found to be equivalent to the predicate device.
Biocompatibility and Sterilization testing was justified to the already tested predicate device. No changes to materials, packaging, or sterilization were made, therefore the new devices are equivalent to the predicate device.
OsteoMed believes that these non-clinical tests have demonstrated that these devices are safe and effective as the predicate device; therefore, clinical testing is not required to support substantial equivalence.
## Substantial Equivalence:
The substantial equivalence of the OsteoMed ICON IMF Screw System is based on similarities in intended use, indications for use, material, function, technology, performance, design, sterilization, and operational principles to the OsteoMed QuickFix (K962774, primary predicate) and design, material and operating principles as the AutoDrive self-drilling features for the OsteoMed AutoDrive Bone Screw (K974785) reference predicate.
{7}------------------------------------------------
## Conclusion:
Substantial equivalence was shown through failure torque testing, insertion torque testing, pull-out force testing, and bending testing when compared to known standards and to the predicate devices. The indications, design, technology and operational principles are similar between the subject devices and primary predicate OsteoMed 2.4mm QuickFix MMF Screw (K962774), and reference predicate OsteoMed 2.0mm AutoDrive Screw (K974785), therefore OsteoMed believes that the OSTEOMED 2.0mm ICON Standard IMF Screw and OSTEOMED 2.0mm ICON AutoDrive IMF Screw do not raise any new safety or effectiveness issues.
The assessment of the non-clinical data demonstrated the devices were as safe and effective as the predicate devices OsteoMed QuickFix (K962774) and AutoDrive Bone Screw (K974785). In conclusion, the OsteoMed IMF Screw System have been evaluated and proven to be substantially equivalent to the predicate device.
Image /page/7/Picture/4 description: The image shows the logo for Colson Casters. The logo consists of the word "Colson" in a rounded sans-serif font, enclosed in a black circle. To the right of the circle, the word "Casters" is written in a smaller, italicized font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.