K151179 · Zeltiq Aesthetics, Inc. · OOK · Sep 22, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151179
Device Name
ZELTIQ CoolSculpting System
Applicant
Zeltiq Aesthetics, Inc.
Product Code
OOK · General, Plastic Surgery
Decision Date
Sep 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4340
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of the submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the submental area, thigh, abdomen and flank. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort. The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite. The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
Device Story
Portable thermoelectric cooling/heating system; applies controlled thermal energy to treatment sites via vacuum or surface applicators. System includes control unit, applicators, and patient-contact supplies (gelpads, liners, skin wipes). Sensors in applicator panels monitor skin surface; provide feedback to control heat flux. Freeze Detect® system monitors tissue during cooling to minimize damage. Used in clinical settings by trained operators. Output is localized thermal therapy; affects fat appearance via cryolipolysis. Benefits include non-invasive fat reduction and pain relief. Clinicians use system to deliver specific treatment cycles/profiles via cycle cards.
Clinical Evidence
Prospective, multi-center, open-label, non-randomized study (IDE G140083) of 60 subjects (submental area). Primary efficacy: 91% correct identification of pre/post-treatment photos by blinded reviewers (n=58), exceeding 80% success criterion. Secondary efficacy: statistically significant reduction in fat layer thickness (0.20 cm, p<0.0001) via ultrasound. Safety: no SAEs or UADEs; transient side effects (erythema, edema, numbness) resolved by 12-week follow-up.
Technological Characteristics
Thermoelectric cooling/heating device; vacuum and surface applicators. Sensors monitor skin surface for heat flux control. Safety features include Freeze Detect® system. Patient-contact parts: applicators, foam borders, gel/gelpads, liners, skin wipes. Connectivity: cycle cards for profiles. Software-controlled.
Indications for Use
Indicated for cold-assisted lipolysis of submental area, thigh, abdomen, and flank in individuals with BMI ≤ 30; intended to reduce appearance of visible fat bulges. Also indicated for localized thermal therapy (hot/cold) for pain relief (post-trauma/surgery, minor aches/spasms) and as local anesthetic for dermatological procedures. Optional massage for muscle relief and cellulite appearance reduction.
Regulatory Classification
Identification
A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ZELTIQ™ Aesthetics Incorporated Ms. Shruti Jayakumar Regulatory Affairs Manager 4698 Willow Road Pleasanton, California 94588
September 24, 2015
Re: K151179
Trade/Device Name: CoolSculpting System Regulation Number: 21 CFR 878.4340 Regulation Name: Contact cooling system for aesthetic use Regulatory Class: Class II Product Code: OOK Dated: July 13, 2015 Received: July 14, 2015
Dear Ms. Jayakumar:
This letter corrects our substantially equivalent letter of September 22, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S.
For Director
Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151179
Device Name CoolSculpting System
#### Indications for Use (Describe)
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of the submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the submental area, thigh, abdomen and flank.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIO Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY
This 510(k) summary of information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | ZELTIQ™ Aesthetics, Inc.<br>4698 Willow Road<br>Pleasanton, CA 94588 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Shruti Jayakumar<br>Regulatory Affairs Manager<br>ZELTIQ Aesthetics, Inc.<br>Phone: 925-474-2516<br>Fax: 925-474-8028 |
| DATE PREPARED: | August 21, 2015 |
| TRADE NAME: | ZELTIQ CoolSculpting System |
| COMMON NAME: | Skin Cooling Device |
| CLASSIFICATION NAME: | Contact Cooling System for Aesthetic Use |
| DEVICE CLASSIFICATION: | Class II, 21 CFR §878.4340 |
| PRODUCT CODE: | OOK |
| PREDICATE DEVICES: | The ZELTIQ CoolSculpting System (DEN090002, K120023, K13321,<br>K142491) |
#### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The CoolSculpting System is a portable thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The CoolSculpting System consists of a control unit, detachable vacuum and surface applicators, as well as supplies such as liners, gelpads, skin wipes, cycle cards, foam borders and securement system, and software which includes functions to collect patient data, monitor tissue during cooling and minimize the risk of damage to tissue.
#### SUBSTANTIALLY EQUIVALENT TO:
The ZELTIQ CoolSculpting System is substantially equivalent to the ZELTIQ Dermal Cooling Device, also known as the ZELTIQ CoolSculpting System, which has been cleared for the indication of cold-assisted
{4}------------------------------------------------
lipolysis of the flank (love handle) under DEN090002, for the abdomen under K120023, and for the thighs under K133212. It has also been cleared for flexible treatment parameters under K142491.
# INDICATION FOR USE:
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of the submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the submental area, thigh, abdomen and flank.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
# TECHNICAL CHARACTERISTICS:
The CoolSculpting System is a thermoelectric cooling and heating device that applies cooling or heating to a treatment site with option for vacuum to hold tissue and or massage to temporarily improve local circulation. The applicators, foam borders, gel, gelpads, liners, pretreatment skin wipes, and securement systems are patient-applied parts.
Sensors in the applicator panels monitor the skin surface, providing feedback that controls the rate of heat flux. The pretreatment skin wipes and gel/gelpad protect the skin by providing thermal coupling at the interface between the applicator panels and the skin. Cards provide cycles and profiles for use with the system. The system monitors tissue during cooling and employs multiple safety features including the Freeze Detect® system, to minimize the risk of damage to tissue. To accommodate the submental treatment site, the system includes the CoolMini applicator. In addition, proprietary skin wipes have been introduced to improve contact to skin. The CoolSculpting System has vacuum applicators of various sizes and a non-vacuum surface applicator that is intended to provide clinicians with options when treating different areas of the body. The technological characteristics are the same as the predicate devices. All share the same mechanism of cooling and heating for the same intended use.
# PERFORMANCE DATA:
Studies have demonstrated the use of the CoolSculpting System on the submental area (this 510k), thighs (K133212), abdomen (K120023), and flanks (DEN090002).
# ZELTIQ Clinical Studies
The table below describes the results from the submental area IDE study. Data supporting the previously cleared indications can be found in the 510(k) summaries listed above.
# Submental Area Study
| Study Design | The ZELTIQ System is a medical device intended for use as a non-invasive |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | aesthetic treatment for the reduction of subcutaneous fat. This study was FDA<br>approved under IDE G140083 and was intended to evaluate the safety and<br>efficacy of cryolipolysis for non-invasive reduction of submental fat. The study<br>protocol was approved by IRB. |
| | This study was a prospective, multi-center, open label, non-randomized,<br>interventional cohort study. Sixty (60) subjects were enrolled at three clinical<br>sites. Male or female subjects between 22 and 65 years of age, with submental<br>skin fold thickness greater than 1cm were eligible to participate. Subjects were<br>treated on the submental area at least once, with the option of a second<br>treatment 6 weeks after the first. Subject safety was assessed throughout the<br>study, including immediately post treatment, and at the subsequent 1-week, 6-<br>week and 12-week post-final treatment follow-up visits. |
| Inclusion Criteria | a) Male or female subjects > 22 years of age and < 65 years of age.<br>b) Submental skin fold thickness > 1cm (measured by caliper).<br>c) No weight change exceeding 5% of body weight in the preceding month.<br>d) Agreement to maintain his/her weight (i.e., within 5%) by not making any<br>major changes in diet or exercise routine during the course of the study.<br>e) Subject has signed a written informed consent form. |
| Exclusion Criteria | a) Skin laxity in the neck or chin area for which reduction in submental fat may,<br>in the opinion of the investigator, result in an unacceptable aesthetic result.<br>b) Prominent platysmal bands at rest which may interfere with assessment of<br>submental fat |
| | c) Evidence of any cause of enlargement in the submental area other than<br>localized subcutaneous fat, such as swollen lymph nodes or ptotic<br>submandibular glands. |
| | d) Significant enlargement on the anterior neck that may prevent the proper<br>placement of the applicator e.g. enlarged thyroid glands. |
| | e) Treatment with dermal fillers, radiofrequency or laser procedures, or<br>chemical peels in the neck or chin area (below the mandible) within the past<br>6 months. |
| | f) Botulinum toxin or other aesthetic drug injections within the neck or chin<br>area (below the mandible) within the past 6 months. |
| | g) History of facial nerve paresis or paralysis (such as Bell's palsy). |
| | h) History of a fat reduction procedure (e.g., liposuction, surgery, lipolytic<br>agents, etc.) or implant in or adjacent to the area of intended treatment. |
| | i) History of prior neck surgery, or prior surgery in the area of intended<br>treatment. |
| | j) Current dental infection. |
| | k) Known history of cryoglobulinemia, cold urticaria, or paroxysmal cold<br>hemoglobinuria. |
| | l) Known history of Raynaud's disease, or any known condition with a response<br>to cold exposure that limits blood flow to the skin. |
| | m) History of bleeding disorder or is taking any medication that in the<br>investigator's opinion may increase the subject's risk of bruising. |
| | n) Currently taking or has taken diet pills or weight control supplements within<br>the past month. |
| | o) Any dermatological conditions, such as scars in the location of the treatment |
| | area that may interfere with the treatment or evaluation. |
| p) | Active implanted device such as a pacemaker, defibrillator, or drug delivery system. |
| q) | Pregnant or intending to become pregnant in the next 6 months. |
| r) | Lactating or has been lactating in the past 6 months. |
| s) | Unable or unwilling to comply with the study requirements. |
| t) | Currently enrolled in a clinical study of an unapproved investigational drug or device. |
| u) | Any other condition or laboratory value that would, in the professional opinion of the investigator, potentially affect the subject's response or the integrity of the data or would pose an unacceptable risk to the subject. |
| Study protocol | Sixty initial treatments were performed with the prototype CoolMini vacuum applicator; 59 Fifty-nine (59) subjects were re-treated at the 6-week follow-up visit. Treatments were performed at -10°C for 60 minutes. Follow-up data is available through 12 weeks post-treatment. Subject safety was assessed throughout the study. |
| | Two of the subjects in the group who received 2 cooling cycles received a partial cooling cycle due to device interference. Therefore, the per-protocol sample size for the primary photographic efficacy endpoint analysis is 58. |
| Demographics | Average age was 49.3 years (range from 25 and 61). |
| | Forty-eight (48) of the subjects enrolled in the study were female and twelve were male. |
| | Average BMI was 31.8 (range from 23.3 to 46.2). |
| | Average weight was 196.1 pounds (range from 139.2 to 285.6). |
| | Ethnicity: Caucasian, Hispanic, African American, Other |
| Study endpoints | The primary safety endpoint was the incidence of device- and/or procedure-related adverse events. The primary efficacy endpoint involved independent panel review of pre-treatment and 12-week post-final treatment photographs of the treatment area for discernible fat layer reduction. Secondary endpoints included the reduction in fat layer thickness as measured by ultrasound at 12 weeks post-final treatment and subject satisfaction as assessed by a questionnaire administered at 12 weeks post-final treatment. |
| Results | The primary safety endpoint was the measurement of all device- or procedure-related adverse events. All adverse events reported during and after the treatment were included in the safety analysis. The primary safety endpoint was met. Per the protocol definitions, there were no device- or procedure-related adverse events that were categorized as serious adverse events (SAE) or unanticipated adverse device effects (UADE). There were four device- and/or procedure-related adverse events, which consisted of two (2) incidents of prolonged erythema, one (1) incident of hyperpigmentation, and one (1) incident of subject report of fullness sensation in back of throat due to swelling. All adverse events resolved by the final follow-up at 12 weeks post-final treatment visit. Four device- or procedure-related adverse events were reported and have resolved. Clinical safety assessment showed anticipated and transient side-effects, all of which resolved over the course of the study. Immediately post- |
| | treatment, the most common events within the treatment area were erythema,<br>edema, and numbness. At the 1-week follow-up visit, most incidences of<br>erythema and edema had resolved and numbness was the most prevalent side<br>effect. By the 6-week follow-up visit post-treatment #1, four incidences of mild<br>numbness were reported. At the 6-week follow-up visit post-treatment #2, two<br>incidences of mild numbness were reported. By the 12-week post-final treatment<br>visit, all adverse events had resolved. The safety data recorded during this study<br>supports the safety of the treatment parameters and device investigated. |
| | The primary efficacy endpoint was correct identification of pre-treatment vs. 12-<br>week post-final treatment images by 3 blinded independent reviewers. The<br>overall correct identification rate by the 3 reviewers was 91% for the per-protocol<br>population (n=58), which met the pre-established 80% criterion for success. The<br>primary efficacy endpoint was met. |
| | Reduction in subcutaneous fat layer thickness as measured by ultrasound at 12-<br>weeks post-final treatment was a secondary efficacy endpoint for this study.<br>Analysis of the per-protocol data (57 subjects) showed a statistically significant<br>(p<0.0001) reduction of 0.20 cm. Therefore, the secondary efficacy endpoint for<br>statistically significant reduction of fat layer thickness was met. |
| | The secondary efficacy endpoint for subject satisfaction was assessed by a<br>questionnaire administered at 12 weeks post-final treatment. Overall, 83% of<br>subjects enrolled in the study indicated they were satisfied with the treatment<br>and 80% reported that they would recommend the treatment to a friend. |
| Conclusions | These clinical findings demonstrate that use of the CoolSculpting System for<br>treatment of the submental area is substantially equivalent to treatments of the<br>previously cleared areas of the flanks, abdomen, and thighs. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.