PRECICE Intramedullary Limb Lengthening System

K151131 · Ellipse Technologies, Incorporated · HSB · Jun 19, 2015 · Orthopedic

Device Facts

Record IDK151131
Device NamePRECICE Intramedullary Limb Lengthening System
ApplicantEllipse Technologies, Incorporated
Product CodeHSB · Orthopedic
Decision DateJun 19, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

Device Story

The PRECICE Intramedullary Limb Lengthening System is an orthopedic device for lengthening the tibia or femur. It consists of an intramedullary nail, locking screws, surgical instruments, and an external remote controller (ERC/ERC2P). The nail features an internal rare earth magnet, a telescoping lead screw/nut assembly, and planetary gearing. During surgery, the nail is inserted into the intramedullary canal and secured with locking screws. Post-operatively, the patient or clinician uses the external remote controller to non-invasively adjust the nail length via magnetic coupling. This controlled distraction facilitates gradual bone lengthening. The system is designed to accommodate various patient anatomies through different nail lengths and diameters. The device provides a mechanical solution for limb length discrepancy, allowing for controlled, incremental bone growth without additional invasive procedures for adjustment.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is supported by static torsion testing (ASTM F1264) and tensile strength testing.

Technological Characteristics

Intramedullary fixation rod; rare earth magnet; telescoping lead screw/nut assembly; planetary gearing. Materials: unspecified metal (implants). Sterilization: Gamma radiation (nail), non-sterile (screws/accessories). Connectivity: Non-invasive magnetic coupling via external remote controller (ERC/ERC2P). Standards: ASTM F1264 (static torsion).

Indications for Use

Indicated for limb lengthening of the tibia and femur in patients requiring such procedures.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 19, 2015 Ellipse Technologies, Incorporated Ms. Rebecca Shelburne Regulatory Affairs Project Manager 13900 Alton Parkway, Suite 123 Irvine, California 92618 Re: K151131 Trade/Device Name: PRECICE® Intramedullary Limb Lengthening System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: June 1, 2015 Received: June 2, 2015 Dear Ms. Shelburne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. 510(k) Number (if known) K151131 #### Device Name PRECICE® Intramedullary Limb Lengthening System #### Indications for Use (Describe) The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur. Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) _ | Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## PRECICE® Intramedullary Limb Lengthening System 510(k) Summarv - K 151131 April 2015 | 1. | Company: | Ellipse Technologies, Incorporated<br>13900 Alton Parkway, Suite 123<br>Irvine, CA 92618 | |----|----------|------------------------------------------------------------------------------------------| |----|----------|------------------------------------------------------------------------------------------| Rebecca Shelburne Contact: Regulatory Affairs Project Manager Phone: (949) 837-3600 x227 Fax: (949) 837-3664 Date Summary Prepared: April 27, 2015 - 2. Proprietary Trade Name: PRECICE Intramedullary Limb Lengthening System - Common Name: Intramedullary Nail 3. - 4. Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020) - Product Code: HSB (Rod, Fixation, Intramedullary and Accessories) ട്. - 6. Product Description: The Ellipse PRECICE System is composed of the PRECICE nail (supplied sterile), locking screws, surgical instruments and an external remote controller (ERC, ERC2P). The nail is available in tibia or femur models with various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE nail is supplied sterile by gamma radiation while the locking screws and PRECICE specific accessories are supplied non-sterile and must be sterilized prior to use. The nail contains an enclosed rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing. - 7. Indications for Use: The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur. - 8. Substantial Equivalence: Documentation that includes a performance assessment and detailed comparison to the predicate device demonstrates that the Ellipse PRECICE System is substantially equivalent to the following 510(k) cleared device: Trade Name: Ellipse PRECICE System Common Name: Intramedullary Nail 510(k) Clearance Number: K141023, K133289, K131677 Substantial equivalence is based on identical indications for use, technological characteristics and in vitro testing performed. The Risk Management file was updated to include the design modifications and no new risks were identified. The additional length PRECICE nails that are subject of this premarket notification and the predicate device have the same indications for use. Specifically, both the predicate PRECICE {5}------------------------------------------------ System and the PRECICE System that is subject of this Special 510(k) is indicated for limb lengthening of the tibia or femur. The additional length PRECICE nails are available in the same diameters and application. No changes are being made to the materials, technological characteristics or principles of operation as a result of this premarket notification. Both devices are inserted into the intramedullary canal of the femur or tibia and secured with locking screws. Both devices are adjusted non-invasively by the Ellipse external remote controller (ERC or ERC2P). The differences between the subject device and the predicate device are as follows: - Shorter overall Nail lengths of 160-190 mm, - Modified distal screw hole configuration, - . Modified locking rod interface on the proximal end of the housing tube, - Chamfered screw holes on the targeting side of the proximal screw holes and a bulleted tip of the distraction rod, and - Addition of new implantation instruments. Data relied upon to determine substantial equivalence of the PRECICE System with the device modifications described in this submission to the cleared PRECICE System include the following: | Test Description | Applicable Test Standard | |--------------------------|--------------------------| | Static Torsion Testing | ASTM F1264 | | Tensile Strength Testing | N/A | There are no changes being made to the External Remote Controller (ERC or ERC 2P) or Locking Screws as a result of this submission. Conclusions can be drawn from the tests that the modifications to the PRECICE System are substantially equivalent and meet the performance specifications.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...