K151102 · Bausch & Lomb, Inc. · MSS · May 22, 2015 · Ophthalmic
Device Facts
Record ID
K151102
Device Name
Bausch + Lomb Injector System
Applicant
Bausch & Lomb, Inc.
Product Code
MSS · Ophthalmic
Decision Date
May 22, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Bausch + Lomb Injector System is indicated for the folding and insertion of Bausch + Lomb intraocular lenses cleared or approved for use with this IOL Injector system.
Device Story
System consists of reusable handpiece (titanium body, plunger tip, knob; stainless steel plunger shaft/threads) and single-use, sterile, disposable coated polyamide cartridge. Used by ophthalmologists in clinical settings for cataract surgery. IOL is placed in cartridge; cartridge snapped into handpiece; screw plunger mechanism advances IOL through cartridge, folding lens for delivery into eye. Facilitates controlled IOL insertion; reduces surgical incision size requirements.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Reusable handpiece: titanium body, plunger tip, knob; stainless steel plunger shaft/threads. Single-use cartridge: coated polyamide with hydrophilic coating (Medicoat A). Sterilization: handpiece via user-performed steam; cartridge via manufacturer-performed ethylene oxide.
Indications for Use
Indicated for folding and injection of Bausch + Lomb intraocular lenses that identify the Bausch + Lomb Injector System in their approved labeling.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2015
Bausch + Lomb Inc. Ms. Joyce Zhong Regulatory Affairs Specialist 50 Technology Drive Irvine, CA 92618
Re: K151102
Trade/Device Name: Bausch + Lomb Injector System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: April 23, 2015 Received: April 24, 2015
Dear Ms. Zhong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a logo for Valeant Pharmaceuticals International, Inc. The logo features a purple and gray geometric shape on the left, followed by the word "VALEANT" in a serif font. Below the word "VALEANT" is the text "Pharmaceuticals International, Inc." in a smaller, sans-serif font. The logo is simple and professional, and the colors are eye-catching.
Valeant Pharmaceuticals International, Inc. 50 Technology Drive, Irvine CA 92618 Tel: 949-398-5594, Fax: 949-398-5764
#### 5 Indications for Use Statement
510(k) Number (if known):
Device Name: Bausch + Lomb Injector System
Indications for Use:
The Bausch + Lomb Injector System is indicated for the folding and insertion of Bausch + Lomb intraocular lenses cleared or approved for use with this IOL Injector system.
Prescription Use ___X_________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Valeant Pharmaceuticals International, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The text "VALEANT" is in a serif font, with "Pharmaceuticals International, Inc." in a smaller font below.
Valeant Pharmaceuticals International, Inc. 50 Technology Drive, Irvine CA 92618 Tel: 949-398-5594, Fax: 949-398-5764
#### 6 510(k) Summary
This summary document has been prepared in accordance with section 21 CFR 807.92(c).
The submitter of the 510(k) is:
Joyce Zhong Regulatory Affairs Specialist Bausch & Lomb, Inc. 50 Technology Drive Phone: 949-398-5594 Fax: 949-398-5764
Date Summary Prepared: April 23, 2015
#### 1. Subject Device:
| Trade name: | Bausch + Lomb Injector System |
|----------------------|-------------------------------|
| Common Name: | Intraocular lens Guide |
| Classification Name: | 21 CFR 886.4300 |
#### 2. Predicate Device:
K131958, Bausch + Lomb Injector System
#### 3. Device Description:
The Bausch + Lomb Injector System is used for folding and delivering validated IOLs into the eye. The system is composed of two items: a single-use, sterile, disposable cartridge and a reusable handpiece. The handpiece has two elements: a plunger and a body. The cartridge snaps into the handpiece.
#### Indications for Use: 4.
The BLIS is indicated for the folding and injection of Bausch + Lomb intraocular lenses identifying the BLIS in their approved labeling.
This is the same exact indications for use as the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Valeant Pharmaceuticals International, Inc. The logo features a stylized "V" shape in purple and gray, followed by the company name in a serif font. Below the company name is the text "Pharmaceuticals International, Inc." in a smaller font size.
Valeant Pharmaceuticals International, Inc. 50 Technology Drive, Irvine CA 92618 Tel: 949-398-5594, Fax: 949-398-5764
#### 5. Brief Summary of Nonclinical Test and Results:
Performance bench testing was completed on the Bausch + Lomb Injector System and results show the system to be in conformance with the requirements set forth in respective standards.
#### Comparative Analysis 6.
A table comparing the proposed device to the predicate device is provided below.
| Characteristic | K131958, (Predicate Device)<br>Bausch + Lomb Injector System | (Proposed Device)<br>Bausch + Lomb Injector System |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The BLIS is indicated for the folding<br>and injection of Bausch + Lomb<br>intraocular lenses identifying the BLIS<br>in their approved labeling. | The BLIS is indicated for the folding<br>and injection of Bausch + Lomb<br>intraocular lenses identifying the BLIS<br>in their approved labeling. |
| Contraindications | None | None |
| Anatomical site | Eye | Eye |
| Injector configuration<br>(reusable) | Titanium body, plunger tip, knob;<br>stainless steel plunger shaft, plunger<br>threads | Titanium body, plunger tip, knob;<br>stainless steel plunger shaft, plunger<br>threads |
| Injector configuration<br>(single use) | Coated polyamide cartridge | Coated polyamide cartridge |
| How is the device<br>used | The IOL is placed in the loading<br>cartridge. The cartridge is snapped into<br>the handpiece. The screw plunger<br>advances the IOL through the cartridge,<br>which folds the IOL and advances it<br>into the eye. | The IOL is placed in the loading<br>cartridge. The cartridge is snapped into<br>the handpiece. The screw plunger<br>advances the IOL through the cartridge,<br>which folds the IOL and advances it<br>into the eye. |
| Single use? | Handpiece: No<br>Cartridge: Yes | Handpiece: No<br>Cartridge: Yes |
| Is the product sterile | Handpiece: shipped nonsterile, to be<br>cleaned and sterilized/resterilized by user<br>Cartridge: shipped sterile | Handpiece: shipped nonsterile, to be<br>cleaned and sterilized/resterilized by user<br>Cartridge: shipped sterile |
| How sterilized | Handpiece: steam (by user)<br>Cartridge: Ethylene oxide (by<br>manufacturer) | Handpiece: steam (by user)<br>Cartridge: Ethylene oxide (by<br>manufacturer) |
### Table 6-1: Comparison of Predicate Device to the Proposed Bausch + Lomb Injector System
{5}------------------------------------------------
## BAUSCH+LOMB See better. Live better.
Image /page/5/Picture/1 description: The image shows the logo for Valeant Pharmaceuticals International, Inc. The logo features a stylized purple and gray geometric shape on the left, followed by the word "VALEANT" in purple, sans-serif font. Below the word "VALEANT" is the text "Pharmaceuticals International, Inc." in a smaller, sans-serif font.
Valeant Pharmaceuticals International, Inc. 50 Technology Drive, Irvine CA 92618 Tel: 949-398-5594, Fax: 949-398-5764
| Characteristic | K131958, (Predicate Device)<br>Bausch + Lomb Injector System | (Proposed Device)<br>Bausch + Lomb Injector System |
|----------------|--------------------------------------------------------------|----------------------------------------------------|
| Coating | Handpiece: None | Handpiece: None |
| | Cartridge: Hydrophilic coating<br>(Medicoat A) | Cartridge: Hydrophilic coating<br>(Medicoat A) |
| How Supplied | Handpiece: single | Handpiece: single |
| | Cartridge: Packs of 10 | Cartridge: Packs of 10 |
#### 7. Conclusion
The Bausch + Lomb Injector System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.