Arthrex Short Suture Anchors

K151092 · Arthrex, Inc. · MBI · Feb 23, 2016 · Orthopedic

Device Facts

Record IDK151092
Device NameArthrex Short Suture Anchors
ApplicantArthrex, Inc.
Product CodeMBI · Orthopedic
Decision DateFeb 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arthrex Short Suture Anchors are intended to be used for suture (soft tissue) fixation to bone in the hip. Specifically, acetabular labral repair.

Device Story

Arthrex Short Suture Anchors are cannulated bone fixation fasteners used for soft tissue-to-bone attachment in hip surgery; specifically acetabular labral repair. Devices consist of PEEK anchors with integral or separate eyelets; pre-loaded on handle inserters; provided with suture and optional suture threader. Used by surgeons in clinical settings to secure soft tissue to bone. Benefit includes stable fixation of labral tissue to facilitate healing. Operation involves manual insertion into bone; no electronic or software components involved.

Clinical Evidence

Bench testing only. Tensile testing data demonstrates performance meets or exceeds predicate device.

Technological Characteristics

Materials: Polyetheretherketone (PEEK). Dimensions: 2.0mm - 2.4mm diameter; 8.6mm - 9.0mm length. Design: Cannulated anchor with integral or separate eyelet; pre-loaded on handle inserter. Mechanical fixation principle.

Indications for Use

Indicated for suture (soft tissue) fixation to bone in the hip, specifically for acetabular labral repair.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 23, 2016 Arthrex, Incorporated Mr. David Rogers Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945 Re: K151092 Trade/Device Name: Arthrex Short Suture Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: January 15, 2016 Received: January 19, 2016 Dear Mr. Rogers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K151092 Device Name Arthrex Short Suture Anchors Indications for Use (Describe) The Arthrex Short Suture Anchors are intended to be used for suture (soft tissue) fixation to bone in the hip. Specifically, acetabular labral repair. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information, Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | {4}------------------------------------------------ | 2.6 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS | | |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Summary Prepared | April 21, 2015 | | Manufacturer/Distributor/Sponsor | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 USA | | 510(k) Contact | David L Rogers<br>Regulatory Affairs Associate<br>Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 USA<br>Telephone: 239/643.5553, ext. 71924<br>Fax: 239/598.5508<br>Email: david.rogers@Arthrex.com | | Trade Name | Arthrex Short Suture Anchors | | Common Name | Suture Anchor | | Product Code -Classification Name<br>CFR | MBI<br>21 CFR 888.3040: Smooth or threaded metallic bone<br>fixation fastener | | Predicate Device | K140855: Arthrex SutureTak Suture Anchors | | Reference Predicate Device | K110473: Pivot Nanotak Suture Anchor | | Purpose of Submission | This Special 510(k) premarket notification is intended<br>to address the Arthrex Short Suture Anchors as a line<br>extension to the Arthrex SutureTak Suture Anchors,<br>K140855. The intended use, material, and<br>fundamental technological characteristics of the<br>proposed Arthrex Short Suture Anchors are<br>substantially equivalent to the predicate. | | Device Description | The Arthrex Short Suture Anchors share the same<br>design features, materials, and intended use as the<br>predicate. The anchors consist of cannulated anchors<br>with an integral or separate eyelet. They are pre-<br>loaded on a handle inserter. Suture, with or without<br>needles, and a suture threader may be provided. The<br>anchors are made from polyetheretherketone (PEEK)<br>and range from 2.0mm - 2.4mm in diameter and 8.6 -<br>9.0mm in length (including eyelet). | | Indications For Use | The Arthrex Short Suture Anchors are intended to be<br>used for suture (soft tissue) fixation to bone in the hip. | | | Specifically, acetabular labral repair | | Substantial Equivalence Summary | The Arthrex Short Suture Anchors are substantially<br>equivalent to the predicate devices in which the basic<br>design features and intended uses are the same. Any<br>differences between the Arthrex Short Suture Anchors<br>and the predicates are considered minor and do not<br>raise questions concerning safety and effectiveness. | | | The submitted tensile testing data demonstrates that<br>the performance of the proposed devices meets or<br>exceeds the predicate device for the desired<br>indications. | | | Based on the indications for use, technological<br>characteristics, and the summary of data submitted,<br>Arthrex, Inc. has determined that the Arthrex Short<br>Suture Anchors are substantially equivalent to the<br>predicate. | {5}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...