Q2 Intravenous Adminstration Sets and Extension Sets
K151079 · Quest Medical, Inc. · FPA · Sep 30, 2015 · General Hospital
Device Facts
Record ID
K151079
Device Name
Q2 Intravenous Adminstration Sets and Extension Sets
Applicant
Quest Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Sep 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
For administration of intravenous fluids to a patient's vascular system utilizing needleless components and an I.V. manifold for multiple simultaneous intravenous therapy via gravity, syringe, or infusion pump. Use of a needle-free system may aid in the prevention of needle-stick injuries.
Device Story
Q2 Multiport IV Administration Sets and Extension Sets are sterile, single-use, non-pyrogenic devices for intravenous fluid delivery. Components include a multiport IV manifold, integrated back-check valves, pre-attached needleless injection sites, drip chamber, and roller clamps. Operated by clinicians in clinical settings (OR, ICU, etc.) to deliver single or multiple infusates to a patient's vascular system. The device facilitates simultaneous therapy via gravity, syringe, or infusion pump. The needle-free design aims to reduce needle-stick injuries. The current submission reflects a material change in the male luer component to Eastman Tritan MX-711. The device does not provide therapeutic claims itself but serves as a conduit for fluids.
Clinical Evidence
Bench testing only. Functional performance testing included high pressure, bond strength, and solvent exposure testing. Biocompatibility testing was conducted per ISO 10993-1:2009. Ethylene oxide residuals were tested per ISO 10993-7:2008. Results demonstrate the modified device performs similarly to the predicate.
Technological Characteristics
Sterile, single-use, non-invasive IV administration set. Materials: Male luer (Eastman Tritan MX-711), tubing (PVC), manifold (polycarbonate/silicone/polyisoprene), filter (copolyester/PVDF/polyestersulfone), swabable Y-site (polycarbonate/silicone), checkvalve (Plexi-Glas), drip chamber (PVC), spike (PVC/ABS/PP/LDPE). Principle of operation: Luer activation. Energy source: User-operated (gravity/pump). Sterilization: 100% EtO, SAL 1x10^-6. Packaging: Tyvek/polyethylene.
Indications for Use
Indicated for patients requiring intravenous fluid administration, including anesthesia drugs, chemotherapeutics, total parenteral nutrition, and antibiotics, via gravity, syringe, or infusion pump in clinical settings such as operating rooms and ICUs.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Quest Medical, Inc. IV Administration Sets (K040385)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 30, 2015
Quest Medical, Inc. Ms. Amy Clendening-Wheeler Senior Regulatory Affairs Specialist One Allentown Parkway Allen, Texas 75002
Re: K151079
Trade/Device Name: Q2® Multiport IV Administration Sets and Extension Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA, FPK Dated: April 20, 2015 Received: April 22, 2015
Dear Ms. Clendening-Wheeler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K151079
Device Name
Q2 Multiport IV Administration Sets and Extension Sets
Indications for Use (Describe)
For administration of intravenous fluids to a patient's vascular system utilizing needleless components and an I.V. manifold for multiple simultaneous intravenous therapy via gravity, syringe, or infusion pump. Use of a needle-free system may aid in the prevention of needle-stick injuries.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| DATE PREPARED | September 29, 2015 | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| SUBMITTER: | Quest Medical, Inc.<br>One Allentown Parkway<br>Allen, TX 75002 USA<br>972-390-9800/800-627-0226 | |
| | Contact: Amy Clendening-Wheeler, Sr. Regulatory Affairs Specialist<br>Phone: (972) 332-6228<br>Fax: (469) 795-2338<br>Email: awheeler@questmedical.com | |
| DEVICE NAME: | Q2® Multiport IV Administration Sets and Extension Sets | |
| COMMON NAME: | IV Administration Set | |
| CLASSIFICATION NAME | Intravascular Administration Set | |
| Product Code: | FPA, FPK | |
| Regulation Number: | 880.5440 | |
| Class: | II | |
| PREDICATE DEVICES: | Quest Medical, Inc. Extension Sets (K800825)<br>Quest Medical, Inc. IV Administration Sets (K040385) | |
| DESCRIPTION: | Sterile, single use non-pyrogenic intravenous fluid administration sets<br>with a multiport IV manifold and integrated back-check valves, pre-<br>attached needleless injection sites, drip chamber and roller clamps. The<br>reason for the submission was a material change to the male luer<br>component. | |
| INTENDED USE: | For administration of intravenous fluids to a patient's vascular system<br>utilizing needleless components and an I.V. manifold for multiple<br>simultaneous intravenous therapy via gravity, syringe, or infusion pump. | |
| | Use of a needle-free system may aid in the prevention of needle-stick<br>injuries. | |
| SUBSTANTIAL EQUIVALENCE | | |
## Performance Testing
{4}------------------------------------------------
#### Sterilization
There is no change to the sterilization process for the proposed Q2® Multiport IV Administration Sets and Extension Sets. Ethylene Oxide residuals testing performed for the modified Male Luer made from Tritan MX711 complies with ISO 10993-7:2008 "Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals."
#### Shelf Life
Shelf life for the proposed Q2® Multiport IV Administration Sets and Extension Sets was verified to remains the same as for the current Q2® Multiport IV Administration Sets and Extension Sets at 3 years.
#### Bench Testing
Functional performance testing including high pressure testing, bond strength testing, and solventexposure testing was completed with the proposed IV Administration Sets to demonstrate that the sets perform as intended. Results of testing successfully demonstrated that the proposed devices perform similarly to the predicate device.
#### Biocompatibility
The materials of construction of a fully assembled IV Administration Set were tested according to ISO 10993-1:2009. Test results successfully verified that the IV Administration Set materials of construction, including the male luer comprised of the Tritan MX-711 material, are biocompatible for their clinical application.
#### Comparison to Predicate:
The following table shows a comparison between the device components of the currently marketed Q2® Multiport IV Administration Sets and Extension Sets with the current Male Luer connector to the proposed Q2® Multiport IV Administration Sets and Extension Sets with new material Male Luer.
| | Predicate Devices | | Modified<br>Device |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|--------------------|
| 510(k) | K800825 | K040385 | Under Review |
| Brand Name | Q2 Extension Set | Q2 Multiport® Manifold IV Set with<br>Swabable Valves | SAME |
| Model # | 95902 | 9520, 9525A, 9526B, 9527B, 22-201-V | SAME |
| Manufacturer | Quest Medical, Inc. | | SAME |
| Device Description | Q2 extension sets and iv administration sets are sterile, non-pyrogenic, single-use intravenous fluid delivery devices. Some models have a multiport IV manifold with backcheck valves, pre-attached needleless injections sites, drip chamber and roller clamps. They are non-invasive devices for short-term use. They deliver either a single infusate or multiple infusates | | SAME |
{5}------------------------------------------------
| | Predicate Devices | Modified<br>Device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Clinical Use | based on the clinical need of the customer.<br>The devices are used by clinicians in a variety of clinical<br>settings such as operating rooms, chemotherapy regimens,<br>ICUs, ext. The devices have direct patient contact due to the<br>administration of fluids to the vascular system. A variety of<br>infusates such as anesthesia drugs, chemotherapeutics, total<br>parental nutrition (TPN) drugs, antibiotics, etc. The devices<br>themselves do not have any intended therapeutic claim. | SAME |
| Intended Use/<br>Indications for Use | For administration of intravenous fluids to a patient's vascular<br>system utilizing needleless components and an I.V. manifold<br>for multiple simultaneous intravenous therapy via gravity,<br>syringe, or infusion pump.<br>Use of a needle-free system may aid in the prevention of<br>needle-stick injuries. | SAME |
| Materials | | |
| Male Luer | Acrylic Cyro MED-2 plastic | Eastman Tritan<br>MX-711 |
| Tubing | PVC | Identical |
| 6 port Manifold | Polycarbonate housing, silicone stem, polyisoprene checkvalve | Identical |
| Filter | Copolyester housing, Durapel PVDF, Polyestersulfone | Identical |
| Swabable Y-site | Polycarbonate and silicone | Identical |
| Inline checkvalve | Plexi-Glas | Identical |
| Drip Chamber | PVC | Identical |
| Spike | PVC, ABS, PP, LDPE | Identical |
| Technology | | |
| Energy Source | User Operated | Identical |
| Principle of<br>Operation | Luer activation | Identical |
| Sterilization/Pkg | | |
| Method | EtO, 100% | Identical |
| Minimum SAL | $1 \times 10^{-6}$ | Identical |
| Packaging | Tyvek polyethylene; heat-sealed | Identical |
| Shelf Life | Three (3) Years | Identical |
| Disposable or<br>Reusable | Disposable | Identical |
## CONCLUSION:
Results of functional performance and biocompatibility testing conducted with the proposed and predicate devices demonstrate that the Q2® Multiport IV Administration Sets and Extension Sets are substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.