Finger Type Pulse Oximeter

K151024 · Taidoc Technology Corporation · DQA · Mar 2, 2016 · Cardiovascular

Device Facts

Record IDK151024
Device NameFinger Type Pulse Oximeter
ApplicantTaidoc Technology Corporation
Product CodeDQA · Cardiovascular
Decision DateMar 2, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The Finger Type Pulse Oximeter is indicated for use in measuring oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for patients during no-motion condition. The patients are limited to adults with weight above 40 kg. This device is indicated for non-invasive spot checking.

Device Story

Finger-worn pulse oximeter; uses spectrophotometry with dual-wavelength LEDs to measure arterial hemoglobin oxygen saturation and pulse rate; inputs are light absorption changes in perfused tissue; outputs are SpO2 percentage, pulse rate, and pulse amplitude displayed on an LCD; includes Bluetooth for wireless transmission of readings to smartphones or tablets; features visual alarm via blinking red backlight; intended for non-invasive spot checking by adults; provides real-time physiological data to assist in monitoring; benefits include convenient, portable access to vital sign measurements.

Clinical Evidence

Clinical evaluation conducted on 14 healthy adult volunteers (diverse skin pigmentations) per ISO 9919 and FDA guidance. SpO2 accuracy validated over 70-100% range: ±2% for 80-100% saturation, ±3% for 70-79% saturation. Root-mean-square (Arms) value was 2.0. Heart rate accuracy met acceptance criteria of ±1 bpm difference.

Technological Characteristics

Spectrophotometry using dual-wavelength LEDs; powered by two AAA alkaline batteries; LCD display; Bluetooth connectivity for data transmission; silicone skin-contacting materials compliant with ISO 10993-5 and ISO 10993-10; IP22 water resistance; dimensions 63mm x 37mm x 32mm; weight 40g (without batteries).

Indications for Use

Indicated for non-invasive spot checking of SpO2 and pulse rate in adults weighing over 40 kg during no-motion conditions.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 2, 2016 TaiDoc Technology Corporation Sharon Peng Regulatory Affairs Specialist 6F, No. 127, Wugong 2nd Rd. Wugu District New Taipei City, Taiwan 24888 Re: K151024 Trade/Device Name: Finger Type Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: December 7, 2015 Received: February 29, 2016 Dear Sharon Peng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, *Tejashri Purohit-Sheth, M.D.* Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K151024 Device Name Finger Type Pulse Oximeter Indications for Use (Describe) The Finger Type Pulse Oximeter is indicated for use in measuring oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for patients during no-motion. The patients are limited to adults with weight above 40 kg. This device is indicated for non-invasive spot checking. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This is being submitted in accordance with the requirements of 21 CFR 807.92. The Assigned 510(k) number is:__k151024 ### 1. Submitter Information | Company Name: | TaiDoc Technology Corporation | |-----------------|-------------------------------------------------------------------------------| | Contact Person: | Sharon Peng | | Title: | Regulatory Affairs Specialist | | Address: | 6F, No. 127, Wugong 2nd Rd., Wugu District, New Taipei City,<br>24888, Taiwan | | Phone: | +886-2-6625-8188 # 1176 | | Fax: | +886-2-6625-0288 | | E-mail: | sharon_peng@taidoc.com.tw | | Prepared Date: | March 2, 2016 | ### 2. Device name: | Proprietary Name: | Finger Type Pulse Oximeter | |-----------------------|----------------------------| | Common Name: | Pulse Oximeter | | Product Code: | DQA | | Classification Panel: | Anesthesiology | | Classification: | Class II | | Regulation Citation: | 21 CFR §870.2700, Oximeter | ### 3. Primary Predicate Device | Proprietary Name: | VTRUST Finger Type Pulse Oximeter | |-------------------|-----------------------------------| | 510(k) Number: | k110893 | ## 4. Intended Use The Finger Type Pulse Oximeter is indicated for use in measuring oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for patients during no-motion condition. The patients are limited to adults with weight above 40 kg. | TaiDoc Tech. Corp. | CONFIDENTIAL | C1 - 1 of 6 | |--------------------|--------------|-------------| |--------------------|--------------|-------------| {4}------------------------------------------------ This device is indicated for non-invasive spot checking. ### 5. Device Description: The main function of Finger Type Pulse Oximeter is same as the predicate's to monitor the oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. Except for the general feature, Finger Type Pulse Oximeter also has the wireless data transmission function which transmits the readings of SpO2 and pulse rate from the Finger Type Pulse Oximeter to the personnel device via Bluetooth pairing. The personnel device may include the smart phone, tablet PC and so on. ### 6. Test Principle: The Finger Type Pulse Oximeter determines functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passing through perfused tissue. Changes in absorption caused by the pulsation of blood in the vascular bed are used to determine oxygen saturation and pulse rate. #### 7. Substantial Equivalence Information: | Predicate device name: | VTRUST Finger Type Pulse Oximeter | |------------------------|-----------------------------------| | Predicate K number: | k110893 | The modified Finger Type Pulse Oximeter have the following Comparison with similarities to the predicate device: predicated device: - same operating principle . - same fundamental scientific technology . - . same intended use The modified Finger Type Pulse Oximeter encompass the following modification to the predicate device: - . accuracy - operating and storage specification . - . physical appearance - . data transmission function - software . - water resistance level . {5}------------------------------------------------ | Item | Predicate device | Proposed device | | | | |---------------------|-------------------------------------|-------------------------------------|--|--|--| | Product Name | VTRUST Finger Type Pulse | Finger Type Pulse Oximeter | | | | | | Oximeter (K110893) | | | | | | | Similarity | | | | | | Detection method | Spectrophotometry | Spectrophotometry | | | | | Power source | Two AAA alkaline batteries | Two AAA alkaline batteries | | | | | Detection mode | Spot checking | Spot checking | | | | | Range of oximetry | 0% to 100% | 0% to 100% | | | | | Resolution of | 1% | 1% | | | | | oximetry | | | | | | | Method | Dual wavelength LED | Dual wavelength LED | | | | | Measurement | 30 to 250 beats per minutes (bpm) | 30 to 250 beats per minutes (bpm) | | | | | Range of pulse rate | | | | | | | Resolution of pulse | 1 bpm | 1 bpm | | | | | rate | | | | | | | Accuracy of pulse | ±1bpm or ±1%, whichever is greater | ±1 bpm or ±1% whichever is greater | | | | | rate | | | | | | | Type | LCD Display with white backlight | LCD Display with white backlight | | | | | Parameters | SpO2; pulse rate; pulse bar | SpO2; pulse rate; pulse bar | | | | | Status | Indicate a stable reading is taken; | Indicate a stable reading is taken; | | | | | | Battery status indication | Battery status indication | | | | | Memory | None | None | | | | | Power button | Yes | Yes | | | | | SpO2 % display | Yes | Yes | | | | | Pulse amplitude | Yes | Yes | | | | | indicator | | | | | | | Pulse rate display | Yes | Yes | | | | | Low battery | Yes | Yes | | | | | indicator | | | | | | | Sensor error | Yes | Yes | | | | | indicator | | | | | | | Display | LCD | LCD | | | | A comparison table that described similarities and modifications is briefly provided below: CONFIDENTIAL {6}------------------------------------------------ | Modification | | | |--------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------| | Accuracy of<br>oximetry | 70% to 100%: ± 2%<br><69%: unspecified | 100% ~ 80% ±2%;<br>79% ~ 70% ±3%;<br>others are undefined | | Operating<br>Condition | 50°F to 104°F (10°C to 40°C);<br>below 90% R.H. (non-condensing) | 50°F to 104°F (10°C to 40°C);<br>below 95% R.H. (non-condensing) | | Storage Condition | -4°F to 122°F (-20°C to 50°C);<br>15% to 90% R.H. (non-condensing) | -13°F to 158°F (-25°C to 70°C);<br>below 95% R.H. (non-condensing) | | Water resistant<br>level | IP56 | IP22 | | Uploading | NA | Instant readings uploading via<br>Bluetooth | | Rotation | None | 180 degrees | | Medium Priority<br>Alarm | None | Visual alarm (Blinking RED<br>backlight) | | Dimensions: | 62.0mm (L) x 37.7mm (W) x 37.4mm<br>(H) | 63mm (L) x 37mm (W) x 32mm<br>(H) | | Weight (g) | <39g (without batteries) | Without batteries: 40g | For testing according to the above modification, please see below table | Modification | Validation Report | |---------------------------------|-----------------------------------------------------------| | accuracy | Accuracy study report | | operating and storage condition | Meter reliability report | | device dimension and weight | Electrical safety, EMC, and biocompatibility test reports | | software | Software validation report | | water resistance level | Electrical safety test report | According to these reports, it demonstrates that the Finger Type Pulse Oximeter does not raise new questions of safety and effectiveness. {7}------------------------------------------------ #### 8. Functional and Safety Testing: The Finger Type Pulse Oximeter is designed and tested in accordance with following standards. The accuracy study was performed to demonstrate the Finger Type Pulse Oximeter meet the criteria of ISO 80601-2-61:2011. The evaluation study is compliance with the standard. The laboratory testing of electrical safety, electromagnetic compatibility, and shock and vibration meet the relevant requirements of the applicable recognized standard: IEC/EN 60601-1, IEC/EN 60601-1-2 and ISO 80601-2-61. The software validation was performed to verify and validate the proposed device works functionally and is in compliance with FDA Guidance for the Content of the Premarket Submissions for Software Contained in Medical Devices. The skin contacting material (silicone) included in the SpO2 probe meets the requirement of the biocompatibility standard, ISO 10993-5:2009 and ISO 10993-10:2010. The tests were performed by MedGaea Life Sciences Institute compiling with U.S. FDA (21 CFR Part 58) on principles of Good Laboratory Practice for Nonclinical Laboratory Studies. The cytotoxicity test is performed to investigate that the skin contacting material is considered non-cvtotoxicity and met the requirements of ISO 10993-5:2009. The skin contacting material revealed no evidence of causing cytotoxicity and did not induce the morphologic changes of cells. The irritation and skin sensitization tests are executed to evaluate the possibility of local inflammation after a single topical application and the possible response of skin sensitization of skin contacting material. Both studies met the requirements of ISO 10993-10:2010 and the results indicated that the skin contacting material did not cause skin irritation and sensitization. The clinical evaluation was conducted and the result of Finger Type Pulse Oximeter is in compliance with the criteria of ISO 9919 standard and FDA guidance - "Pulse Oximeter Premarket Notification Submissions". The functional oxygen saturation (SpO2) measurement has been validated on a total of 14 healthy adult male and female volunteers with 7 dark, 6 intermediate, and 1 light skin pigmentations in the range of 70% to 100%. The SpO2 accuracy result include that the SpO2 value range from 70% to 79% is within CONFIDENTIAL C1 - 5 of 6 TaiDoc Tech, Corp. {8}------------------------------------------------ 3% and the SpO2 range from 80% to 100% is within 2%. The accuracy of SpO2 to be ±3 dig % was verified over the range of 70% to 100%. The heart rate is within the acceptance criteria of ±1 dig bpm difference. And the root-mean-square (Arms) value is 2.0 with the saturations form 70% to 100%. #### 9. Conclusion The proposed devise of Finger Type Pulse Oximeter has the same classification information, same intended use, same test principle and similar product design and specifications with the predicate device except the data transmission and alarm function. Based on the information provided in this submission, the Finger Type Pulse Oximeter is substantially equivalent to the predicate VTRUST Finger Type Pulse Oximeter (k110893).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...