← Product Code [HSB](/productcode/HSB) · K151010

# Piccolo Composite Nailing System (K151010)

_CarboFix Orthopedics , Ltd. · HSB · Jul 28, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K151010

## Device Facts

- **Applicant:** CarboFix Orthopedics , Ltd.
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Jul 28, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.

## Device Story

Intramedullary nailing system for proximal femur fracture fixation; consists of cylindrical carbon fiber reinforced polymer (CFRP) nails, lag screws, and titanium alloy distal screws. Tantalum markers embedded in CFRP for radiographic visualization. Device implanted by orthopedic surgeons in clinical/OR settings to stabilize fractures. Provides mechanical support to bone segments via lag and interlocking screws; facilitates bone healing. Benefits include structural stability for various fracture types and revision scenarios.

## Clinical Evidence

Bench testing only. Performance characteristics evaluated included static and dynamic bending of construct, rotational stiffness, lag screw cutout and pullout, shell attachment strength, and locking screw pullout and torque to failure. Results were comparable to predicate devices.

## Technological Characteristics

Intramedullary nail (11mm mid-shaft, 17mm proximal diameter; 180-460mm length). Materials: Carbon fiber reinforced polymer (CFRP) with titanium alloy components and embedded tantalum markers for radiopacity. Mechanical fixation via lag and interlocking screws. Non-software device.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Piccolo Composite Nailing System ([K091425](/device/K091425.md), [K100497](/device/K100497.md), [K102369](/device/K102369.md), [K111056](/device/K111056.md), [K123810](/device/K123810.md), [K132774](/device/K132774.md))
- Gamma3® Nail System ([K043431](/device/K043431.md))
- Trochanteric Fixation Nail (TFN) ([K011857](/device/K011857.md))
- Fixion® Interlocking Proximal Femur Intramedullary Nailing System ([K010988](/device/K010988.md), [K012967](/device/K012967.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.

July 28, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Carbofix Orthopedics Limited Yael Rubin 11 Ha'hoshlim Street 46724 Herzeliya Israel

Re: K151010

Trade/Device Name: Piccolo Composite® Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: June 25, 2015 Received: June 29, 2015

Dear Yael Rubin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 – Yael Rubin

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151010

Device Name Piccolo Composite® Nailing System

Indications for Use (Describe)

Piccolo Composite Proximal Femur Nails

The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> </span> </div> |
|----------------------------------------------|-----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> <span> </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

# CarboFix Orthopedics, Ltd.

# Piccolo Composite® Nailing System - Proximal Femur

### Applicant Name

CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 46724, Israel

#### Contact Person

Yael Rubin CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzliya 46724, Israel Tel: +972 9 9511511, Fax: +972 9 9548939

# Date Prepared

April 2015

# Trade/Proprietary Name

Piccolo Composite Nailing System

# Common Name

Intramedullary Nailing System

# Classification Name

Rod, Fixation, Intramedullary and Accessories (21 CFR §888.3020; Product Code HSB)

# Predicate Devices

- Piccolo Composite Nailing System (CarboFix Orthopedics Ltd.; K091425, K100497, K102369, K111056, K123810, K132774);

{4}------------------------------------------------

- Gamma3® Nail System (Stryker (Howmedica Osteonics Corp.); K043431 and more);
- Trochanteric Fixation Nail (TFN) (Synthes; K011857 and more).
- 트 Fixion® Interlocking Proximal Femur Intramedullary Nailing System (CarboFix; K010988, K012967 and more).

#### Intended Use/Indications for Use

### Piccolo Composite Proximal Femur Nails

The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.

### System Description

The Piccolo Composite Nailing System includes nails, screws and a set of instruments.

The Piccolo Composite nail indicated for treatment of the proximal femur is a cylindrical rod. Nail mid-shaft diameter is 11mm. with the proximal end diameter being 17mm. Nail lengths are 180mm, 200mm, and in the range of 300 - 460mm. The nail provides for holes at the proximal and distal sections, designed for the insertion of a lag screw and interlocking screws. The lag screw is of 10mm diameter, with its length being in the range of 80mm to 110mm. The nails and lag screws are made of carbon fiber reinforced polymer and incorporate small amount of titanium alloy. Tantalum markers are embedded within the carbon fiber reinforced polymer, where applicable, to enable visualization during imaging. The distal screws are made of titanium alloy.

{5}------------------------------------------------

#### Substantial Equivalence

The Piccolo Composite Proximal Femur Nailing System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable.

Performance characteristics for the Piccolo Composite Proximal Femur Nailing System components included static and dynamic bending of construct, rotational stiffness of construct, lag screw cutout, lag screw pullout, proximal and distal shell attachment strength, locking screw pullout and torque to failure, and are comparable to those of predicate devices (as applicable), thus demonstrating that the device is safe and effective for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K151010](https://fda.innolitics.com/device/K151010)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
