K150999 · Edan Instruments, Inc. · IYO · May 6, 2015 · Radiology
Device Facts
Record ID
K150999
Device Name
Acclarix Diagnostic Ultrasound System
Applicant
Edan Instruments, Inc.
Product Code
IYO · Radiology
Decision Date
May 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include: Abdominal Gynecology (including endovaginal) Obstetric Cardiac Small parts (Breast, Testes, Thyroid, etc.) Urology Musculoskeletal Peripheral vascular Intra-operative
Device Story
Portable laptop-based diagnostic ultrasound system for point-of-care environments (emergency departments, interventional procedures, mobile imaging). Inputs: acoustic echoes via piezoelectric transducers. Processing: high-voltage burst excitation; multi-beam-forming; speckle resistance imaging (eSRI); frequency/spatial compounding; HPRF. Outputs: real-time ultrasound images on 15-inch LCD with 10-inch touch screen and 5-inch trackpad. Operated by physicians or sonographers. Output used for diagnostic imaging and fluid flow analysis; supports clinical decision-making via visualization of anatomy and pathology, biopsy guidance, and measurement tools (distance, volume, velocity, etc.). Benefits: portable, high-resolution imaging for rapid assessment in diverse clinical settings.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (per FDA 2008 guidance), and biocompatibility (ISO 10993-1, -5, -10).
Indicated for ultrasound evaluations in adults, pregnant women, and pediatric patients. Clinical applications include abdominal, gynecological (including endovaginal), obstetric, cardiac, small parts (breast, testes, thyroid), urological, musculoskeletal (conventional and superficial), peripheral vascular, and intra-operative exams.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem with three abstract human profiles facing right, stacked on top of each other. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 6, 2015
Edan Instruments, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K150999
Trade/Device Name: Acclarix Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: April 14, 2015 Received: April 15, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K150999
Device Name Acclarix Diagnostic Ultrasound System
#### Indications for Use (Describe)
The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include:
Abdominal Gynecology (including endovaginal) Obstetric Cardiac Small parts (Breast, Testes, Thyroid, etc.) Urology Musculoskeletal Peripheral vascular Intra-operative
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### Acclarix AX8 Diagnostic Ultrasound System
| | Clinical Application | | | Mode of Operation | | | | |
|-----------------------------|---------------------------------|---|---|-------------------|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | N | N | N | | N | N | N |
| | Abdominal | N | N | N | N | N | N | N |
| | Intra-operative (Specify) | N | N | N | | N | N | N |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | N | N | N | | N | N | N |
| Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | | N | N | N |
| | Trans-vaginal | N | N | N | | N | N | N |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | N | N | N | | N | N | N |
| | Musculo-skeletal (Superficial) | N | N | N | | N | N | N |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | N | N | N | | N | N | N |
| | Adult Cardiac | N | N | N | N | N | N | N |
| Cardiac | Pediatric Cardiac | N | N | N | N | N | N | N |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | N | N | N | | N | N | N |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PD//DPDI +PW
______________________________________________________________________________________________________________________________________________________________________________
Note * Small Organ includes Thyroid, Testes, Breast - Caracter - Career - Career - Career - Career
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{6}------------------------------------------------
#### Acclarix AX8 with L10-4Q Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | N | N | N | N | N | N | N |
| Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | N | N | N | N | N | N | N |
| | Musculo-skeletal (Superficial) | N | N | N | N | N | N | N |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | | | | | | | |
| | Adult Cardiac | | | | | | | |
| Cardiac | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | N | N | N | N | N | N | N |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW
* Small Organ includes Thyroid, Testes, Breast. Note
| ** Other use includes Urology, Gynecology. | | |
|--------------------------------------------|--|--|
| | | |
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{7}------------------------------------------------
#### Acclarix AX8 with E8-4Q Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|--------------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | N | N | N | | N | N | N |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | | | | | | | |
| Fetal | Neonatal Cephalic | | | | | | | |
| Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | | N | N | N |
| | Trans-vaginal | N | N | N | | N | N | N |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | | | | | | | |
| Cardiac | Adult Cardiac | | | | | | | |
| | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | | | | | | | |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW
* Small Organ includes Thyroid, Testes, Breast Note
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging ,DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{8}------------------------------------------------
### Acclarix AX8 with L17-7HQ Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | N | N | N | N | N | N | N |
| Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | N | N | N | N | N | N | N |
| | Musculo-skeletal (Superficial) | N | N | N | N | N | N | N |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | | | | | | | |
| | Adult Cardiac | | | | | | | |
| | Pediatric Cardiac | | | | | | | |
| Cardiac | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | N | N | N | N | N | N | N |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW
Note * Small Organ includes Thyroid, Testes, Breast
| ** Other use includes Urology, Gynecology | | | | |
|-------------------------------------------|--|--|--|--|
|-------------------------------------------|--|--|--|--|
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{9}------------------------------------------------
## Acclarix AX8 with C5-2XQ Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | N | N | N | N | N | N | N |
| | Abdominal | N | N | N | N | N | N | N |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal | Small Organ (Specify) * | | | | | | | |
| Imaging | Neonatal Cephalic | | | | | | | |
| & Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | N | N | N | N | N | N | N |
| Cardiac | Adult Cardiac | | | | | | | |
| | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | | | | | | | |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW
______________________________________________________________________________________________________________________________________________________________________________
* Small Organ includes Thyroid, Testes, Breast Note
| ** Other use includes Urology, Gynecology | | |
|-------------------------------------------|--|--|
| | | |
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{10}------------------------------------------------
## Acclarix AX8 with P5-1XQ Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | | | | | | | |
| | Abdominal | N | N | N | N | N | N | N |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal<br>Imaging<br>& Other | Small Organ (Specify) * | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | | | | | | | |
| | Adult Cardiac | N | N | N | N | N | N | N |
| Cardiac | Pediatric Cardiac | N | N | N | N | N | N | N |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M,B+PW,B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast Note
______________________________________________________________________________________________________________________________________________________________________________
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
「2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{11}------------------------------------------------
### Acclarix AX8 with L17-7SQ Transducer
| | Clinical Application | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|----|----|-------|---------------------------|---------------------------|
| General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) [1] | Other<br>(Specify) [2][3] |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | N | N | N | | N | N | N |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | | | | | | | |
| Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | N | N | N | | N | N | N |
| | Musculo-skeletal (Superficial) | N | N | N | | N | N | N |
| | Intravascular | | | | | | | |
| | Other (Specify) ** | | | | | | | |
| Cardiac | Adult Cardiac | | | | | | | |
| | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW
Note * Small Organ includes Thyroid, Testes, Breast
| | ** Other use includes Urology, Gynecology |
|--|-------------------------------------------|
| | |
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{12}------------------------------------------------
## 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc. | | | | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|--|--|--|--|
| | 3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou,<br>Nanshan Shenzhen, 518067 P.R. China | | | | |
| | Tel .: (0755) 26858739 | | | | |
| | Fax: +1 (408) 418-4059 | | | | |
| Contact Person: | Queena Chen | | | | |
| Date prepared: | May 4, 2015 | | | | |
| 2. Device name<br>and classification: | Device Name: Diagnostic Ultrasound System<br>Model: Acclarix AX8 | | | | |
| | Classification Name: | | | | |
| | System, Imaging, Pulsed Doppler, Ultrasonic<br>892.1550 | | | | |
| | Product code: IYN | | | | |
| | Ultrasonic, Pulsed echo, Imaging<br>892.1560 | | | | |
| | Product code: IYO | | | | |
| | 892.1570<br>Transducer, Ultrasonic, Diagnostic | | | | |
| | Product code: ITX | | | | |
| | Regulatory Class: Class II | | | | |
| 3.Premarket<br>Notification Class<br>III Certification<br>and Summarv | Not applicable, the subject device is Class II. | | | | |
| 4. Predicate<br>Device(s): | U50 Diagnostic ultrasound system/ K123249/ Shenzhen EDAN<br>Instruments CO., Ltd | | | | |
| | M7 Diagnostic Ultrasound System / K131690/ Shenzhen Mindray<br>Bio-medical Electronics Co., Ltd | | | | |
| 5. Reason for<br>Submission | Acclarix AX8 is a new device. | | | | |
| 6. Pre-Submission,<br>IDE | Not applicable, there is no prior submissions. | | | | |
{13}------------------------------------------------
- The Acclarix AX8 is a portable laptop diagnostic ultrasound system, 7. Device intended for use by a qualified physician or sonographer for ultrasound Description: evaluation in Point of Care environments such as Emergency Departments, Interventional procedures and mobile imaging centers, which applies advanced technologies such as HPRF. Dual screen display, Panorama, Spatial Compounding Imaging, Frequency Compounding Imaging, Multi-Beam-Forming, Speckle Resistance Imaging (eSRI), One-Key Optimization, B-Steer, Digital Zoom, Needle Visualization, etc. Various image parameter adjustments, 15 inch LCD display with 10 inch touch screen and 5 inch track pad and diverse probes are configured to acquire and display clear and stable ultrasound images, following Track 3 for all applications in the 510(k) submission.
- The Edan Acclarix AX8 Ultrasound System is intended for use by a 8. Intended Use: qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include: Abdominal, Gynecology (including endovaginal), Obstetric, Cardiac, Small parts (Breast, Testes, Thyroid, etc.), Urology, Musculoskeletal, Peripheral vascular, and Intra-operative.
## 9. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table:
| Item | Acclarix AX8 Ultrasound Imaging System (Edan Instruments) | U50 Diagnostic Ultrasound System (Edan Instruments) | M7 Diagnostic Ultrasound System (Shenzhen Mindray) |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Current Submission | K123249 | K131690 |
| Intended Use | Diagnostic ultrasound imaging or fluid flow analysis of the human body | Same | Same |
| Item | Acclarix AX8<br>Ultrasound<br>Imaging System<br>(Edan<br>Instruments) | U50 Diagnostic<br>Ultrasound<br>System<br>(Edan<br>Instruments) | M7 Diagnostic Ultrasound<br>System<br>(Shenzhen Mindray) |
| Indications<br>for Use | The Acclarix<br>AX8 Ultrasound<br>Imaging System<br>is intended for<br>use by a<br>qualified<br>physician or<br>sonographer for<br>ultrasound<br>evaluation.<br>Clinical<br>applications<br>include:<br>Abdominal,<br>Gynecology<br>(including<br>endovaginal),<br>Obstetric,<br>Cardiac, Small<br>parts (Breast,<br>Testes, Thyroid,<br>etc.), Urology,<br>Musculoskeletal,<br>Peripheral<br>vascular, and<br>Intra-operative. | The diagnostic<br>ultrasound<br>system (U50) is<br>applicable for<br>adults, pregnant<br>women,<br>pediatric<br>patients'<br>ultrasound<br>evaluation in<br>hospitals and<br>clinics. It is<br>intended for use<br>in abdominal,<br>obstetrics,<br>gynecology,<br>pediatric, small<br>parts, urology,<br>peripheral<br>vascular,<br>musculoskeletal<br>(conventional<br>and superficial),<br>endovaginal<br>and cardiac<br>clinical<br>applications, by<br>or on the order<br>of a physician<br>or similarly<br>qualified health<br>care<br>professional. | The M7/M7T diagnostic<br>ultrasound system is applicable for<br>adults, pregnant women, pediatric<br>patients and neonates. It is<br>intended for use in gynecology,<br>obstetric, abdominal, pediatric,<br>small parts (breast, testes, thyroid),<br>neonatal cephalic, transcranial,<br>cardiac, transvaginal, transrectal,<br>peripheral vascular, urology,<br>orthopedic, and musculoskeletal<br>(conventional and superficial),<br>intraoperative and transesophageal<br>(cardiac) exams. |
| Installation<br>and Use | Portable (laptop)<br>Mobile<br>Equipment | Same | Same |
| Item | Acclarix AX8<br>Ultrasound<br>Imaging System<br>(Edan<br>Instruments) | U50 Diagnostic<br>Ultrasound<br>System<br>(Edan<br>Instruments) | M7 Diagnostic Ultrasound<br>System<br>(Shenzhen Mindray) |
| Safety<br>Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-37<br>ISO 10993-1, -5,<br>-10, -12<br>AIUM, NEMA<br>UD 2, UD 3 | Same | Same |
| Patient<br>Contact<br>Materials | Complies with<br>ISO 10993 | Same | Same |
| Mode of<br>Operations | B-Mode.,<br>M-Mode, Color,<br>PDI/DPDI, PW,<br>CW | Same | B-Mode, M-Mode, Color,<br>PDI/DPDI, PW, CW Smart 3D,<br>Static 3D, 4D, iScape, and TDI |
| Measurements | B-Mode:<br>Distance,<br>Circ/Area,<br>Angle, Volume,<br>Stenosis<br>M-Mode:<br>Distance, Time,<br>Slope and Heart<br>Rate<br>D-Mode:<br>Velocity, RI,<br>Time, PI, Heart<br>Rate, Auto Trace<br>PG, S/D, AV,<br>Acceleration,<br>PHT, VTI | B-Mode:<br>Distance, Area,<br>Volume, Ratio,<br>Histogram and<br>Angle<br>M-Mode: Same<br>D- Mode:<br>Time, Heart<br>Rate, Velocity,<br>Acceleration,<br>RI, RI and Auto<br>trace | B-Mode: Distance, Depth, Angle,<br>Are, Volume, Cross Line, Parallel<br>Line, Trace Length, Distance,<br>Ratio, Area Ratio, B Histogram<br>M-Mode: Same<br>D-Mode: Time, HR, D Velocity,<br>Acceleration, D Trace, PS/ED |
| Principle of<br>Operation | Applying high<br>voltage burst to<br>the Piezoelectric<br>material in the<br>transducer and<br>detect reflected<br>echo to construct<br>diagnostic image | Same | Same |
| Item | Acclarix AX8 | U50 Diagnostic | M7 Diagnostic Ultrasound |
| | Ultrasound | Ultrasound | System |
| | Imaging System | System | (Shenzhen Mindray) |
| | (Edan | (Edan | |
| | Instruments) | Instruments) | |
| Acoustic | Track 3: MI, | Same | Same |
| Output | TIS, TIC, TIB | | |
| | (TI Range 0-6.0) | | |
| | Derated IspTA:…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.