NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs
K150973 · Terang Nusa Snd Bhd · KGO · Aug 4, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150973
Device Name
NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs
Applicant
Terang Nusa Snd Bhd
Product Code
KGO · General, Plastic Surgery
Decision Date
Aug 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Intended Use
This Surgeons glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves are tested for use with Chemotherapy Drugs. Testing based on ASTM D6978-05
Device Story
NUZONE X2G is a powder-free, synthetic polychloroprene surgical glove. Designed for use by operating room personnel to provide a barrier against contamination of surgical wounds. Device is also tested for resistance to permeation by specific chemotherapy drugs per ASTM D6978-05. Used in clinical/surgical settings; provides physical protection for both patient and clinician during procedures involving chemotherapy agents.
Clinical Evidence
Bench testing only. Permeation testing performed for ten chemotherapy drugs (Carmustine, Cisplatin, Cyclophosphamide, Dacarbazine, Doxorubicin Hydrochloride, Etoposide, Paclitaxel, Fluorouracil, Thiotepa, Vincristine Sulfate) per ASTM D6978-05. Breakthrough detection times reported for each agent.
Technological Characteristics
Material: Synthetic polychloroprene rubber. Form factor: Powder-free surgical glove. Testing standard: ASTM D6978-05 for chemotherapy drug permeation. Non-sterile/sterile status not explicitly detailed, but intended for surgical use.
Indications for Use
Indicated for use by operating room personnel to protect surgical wounds from contamination. Suitable for use during handling of chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K110698 — NUZONE X2T CHEMO, POLYCHLOROPRENE SYNTHETIC TAN SURGICAL GLOVES POWDER FREE · Terang Nusa Sdn Bhd · Dec 7, 2011
K201774 — Polychloroprene Powder Free Sterile Surgical Gloves, Low Dermatitis Potential and Tested for Use with Chemotherapy Drugs · Pt. Medisafe Technologies · Sep 16, 2020
K173258 — Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy · Pt. Medisafe Technologies · Apr 20, 2018
K142190 — Polychloroprene Powder-free Surgical Glove (White) Tested for Use with Chemotherapy Drugs; Polychloroprene Powder-free Surgical Glove (Green) Tested for Use with Chemotherapy Drugs · Pt. Medisafe Technologies · Jul 14, 2015
K051094 — NUZONE X2T POLYCHLOROPRENE SYNTHETIC TAN SURGICAL GLOVE - POWDERFREE · Terang Nusa Sdn Bhd · Jun 6, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2015
Terang Nusa, Sdn. Bhd. Mr. Robert Hill Regulatory Director 1. Jalan 8 Pengkalan Chepa 2 Industrial Zone 16100 Kota Bharu MALAYSIA
Re: K150973
Trade/Device Name: NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's glove Regulatory Class: I Product Code: KGO, LZC Dated: July 1, 2015 Received: July 5, 2015
Dear Mr. Hill:
This letter corrects our substantially equivalent letter of August 4, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150973
### Device Name
NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
This Surgeons glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
These gloves are tested for use with Chemotherapy Drugs. Testing based on ASTM D6978-05
| DRUG | Minimum Breakthrough Detection Time<br>(Specimen 1/2/3)<br>Minutes |
|--------------------------------------------------|--------------------------------------------------------------------|
| Carmustine (BCNU), 3.3 mg/ml (3,300ppm) | 30.6 (30.6, 30.6, 30.8) |
| Cisplatin, 1.0 mg/ml (1,000ppm) | Up to 240 |
| Cyclophosphamide (Cytoxan), 20 mg/ml (20,000ppm) | Up to 240 |
| Dacarbazine (DTIC), 10.0 mg/ml (10,000ppm) | Up to 240 |
| Doxorubicin Hydrochloride, 2.0 mg/ml (2,000ppm) | Up to 240 |
| Etoposide (Toposar), 20.0 mg/ml (20,000ppm) | Up to 240 |
| Paclitaxel (Taxol), 6.0 mg/ml (6,000ppm) | Up to 240 |
| Fluorouracil, 50.0 mg/ml (50,000ppm) | Up to 240 |
| Thiotepa, 10.0 mg/ml (10,000ppm) | 31.2 (31.2, 60.8, 61.3) |
| Vincristine Sulfate, 1.0 mg/ml (1,000ppm) | Up to 240 |
Please note that the following drugs have low permeation times: Carmustine (BCNU) 30.6 Thiotepa 31.2
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.