← Product Code [KPO](/productcode/KPO) · K150966

# HMB32 (K150966)

_Haemopharm Biofluids S.R.L. · KPO · Dec 4, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150966

## Device Facts

- **Applicant:** Haemopharm Biofluids S.R.L.
- **Product Code:** [KPO](/productcode/KPO.md)
- **Decision Date:** Dec 4, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Intended for use as a dialysis solution in Continuous Renal Replacement Therapy.

## Device Story

HMB32 is a bicarbonate-based dialysis solution for Continuous Renal Replacement Therapy (CRRT). It is supplied in a sterile, non-pyrogenic, two-chambered PVC-free container system. One chamber holds an acidic solution; the other holds a basic solution. At the point of use, the internal membrane is broken to mix the contents, creating the final ionic composition. The solution is delivered to the dialysate pathway of the blood tubing set via an administration tube. A medication port allows for the optional addition of substances. The device is intended for clinical use by healthcare professionals in settings where CRRT is performed. It functions as a replacement or dialysate fluid to manage electrolyte and fluid balance in patients with renal failure.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility assessments per ISO 10993-1 and USP Class VI, sterilization validation per EN ISO 17665-1, pyrogenicity testing (LAL test), and real-time stability testing confirming shelf life and container integrity.

## Technological Characteristics

Two-chambered PVC-free solution container system. Sterilization: moist heat (EN ISO 17665-1). Biocompatibility: ISO 10993-1 and USP Class VI. Ionic composition includes Na+, K+, Ca++, Mg++, HCO3-, Cl-, and glucose. Form factor: 500+4500 ml and 500+1500 ml volumes.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Accusol Dialysis Solutions ([K041428](/device/K041428.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 4, 2015

Haemopharm Biofluids Srl (HBiofluids) % Sheila Hemeon-Heyer President Heyer Regulatory Solutions, LLC P.O. Box 2151 Amherst, MA 01004-2151

Re: K150966

> Trade/Device Name: HMB32 500 + 4500 ml and 500 + 1500 ml Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KPO Dated: October 29, 2015 Received: October 29, 2015

Dear Sheila Hemeon-Heyer,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K150966

Device Name

HMB32 500 + 4500 ml and 500 + 1500 ml

Indications for Use (Describe)

Intended for use as a dialysis solution in Continuous Renal Replacement Therapy.

| Type of Use (Select one or both, as applicable)                                                                  | <table><tr><td><span style="font-family: DejaVu Sans, sans-serif">☑</span> Research Use Only (21 CFR 201.128 Subject to 501(b))</td><td>☐ For In-Vitro Use Only (21 CFR 201.128 Subject to 501(b))</td></tr></table> | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Research Use Only (21 CFR 201.128 Subject to 501(b)) | ☐ For In-Vitro Use Only (21 CFR 201.128 Subject to 501(b)) |
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Research Use Only (21 CFR 201.128 Subject to 501(b)) | ☐ For In-Vitro Use Only (21 CFR 201.128 Subject to 501(b))                                                                                                                                                           |                                                                                                                  |                                                            |

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

| 1. SUBMITTER:                             | Haemopharm Biofluids srl<br>Via dell'Industria 6<br>23030 Tovo S. Agata<br>Sondrio, Italy             |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 2. CONTACT PERSON:                        | Giuseppe Tomasini<br>Phone: 011 39 0535 29323<br>Mobile: 011 39 3355733390<br>Fax: 011 39 0535 29177  |
| 3. DATE PREPARED:                         | April 10, 2015                                                                                        |
| 4.   DEVICE TRADE NAME:<br>MODEL NUMBERS: | HMB32<br>(500 + 4500 ml) and (500 + 1500 ml)                                                          |
| 5.   COMMON NAME:<br>CLASSIFICATION NAME: | Haemofiltration and Dialysis Solution<br>Dialysate Concentrate for Hemodialysis<br>(Liquid or Powder) |
| 6. CLASSIFICATION:                        | 21 CFR 876.5820, Product Code KPO<br>Class II                                                         |
| 7. PREDICATE DEVICE(S):                   | K041428, Accusol Dialysis Solutions, Baxter<br>Healthcare                                             |

#### 8. DEVICE DESCRIPTION:

The HBiofluids HMB32 is a bicarbonate based dialysis solution packaged in a ready to mix, sterile, non-pyrogenic, two-chambered solution container system. One chamber contains the acidic solution and one chamber contains the basic solution. At the point of use, the membrane between the two chambers is broken, and the acidic and basic solutions are mixed to form the final ionic composition of the solution to be used in the Continuous Renal Replacement Therapy (CRRT) treatment.

The HBiofluids HMB32 solution will be offered in one formulation, in two different volumes: 500+4500 ml and 500+1500 ml. The chemical compositions of the final solutions for these two volumes are shown in the substantial equivalence table below.

{4}------------------------------------------------

# 9. INDICATION FOR USE:

Intended for use as a dialysis solution in Continuous Renal Replacement Therapy.

# 10. COMPARISON TO PREDICATE DEVICE:

The table below provides a technical comparison to the predicate device. There are no significant differences between the HBiofluids HMB32 solution as compared to the predicate Accusol solution.

| Parameter                                           | Predicate Device<br>ACCUSOL, K041428                                                                                                                                                                                         | HMB32                                                                                                                                                                                                                                                     |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use                                  | For use as a dialysis<br>solution in continuous<br>renal replacement<br>therapy                                                                                                                                              | For use as a dialysis<br>solution in continuous<br>renal replacement<br>therapy                                                                                                                                                                           |
| Overall design                                      | The solution is supplied<br>in a carton box<br>containing two units of 5<br>liter double chamber<br>non-PVC bags.<br>Each bag is over-<br>wrapped in its<br>overpouch.<br>The solution in the bag<br>is clear and colorless. | The solution is supplied<br>in a carton box<br>containing two units of<br>5 liter or four units of 2<br>liter double chamber<br>PVC free bags.<br>Each bag is over-<br>wrapped in its<br>overpouch.<br>The solution in the bag<br>is clear and colorless. |
| lonic composition<br>of final solution<br>(mmol/l): |                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                           |
| Na+                                                 | 140.00                                                                                                                                                                                                                       | 140.00                                                                                                                                                                                                                                                    |
| K+                                                  | 2.00                                                                                                                                                                                                                         | 2.50                                                                                                                                                                                                                                                      |
| Ca++                                                | 1.75                                                                                                                                                                                                                         | 1.50                                                                                                                                                                                                                                                      |
| Mg++                                                | 0.50                                                                                                                                                                                                                         | 0.75                                                                                                                                                                                                                                                      |
| HCO3-                                               | 35.00                                                                                                                                                                                                                        | 32.00                                                                                                                                                                                                                                                     |
| Cl-                                                 | 111.30                                                                                                                                                                                                                       | 115.00                                                                                                                                                                                                                                                    |
| Glucose                                             | 5.55                                                                                                                                                                                                                         | 5.55                                                                                                                                                                                                                                                      |
| Theoretical<br>osmolarity                           | 296 mOsm/l                                                                                                                                                                                                                   | 297 mOsm/l                                                                                                                                                                                                                                                |
| Solution container<br>components and<br>materials   | Supplied in a two<br>chamber container<br>system. At the time of<br>use, the contents of the<br>two chambers are mixed<br>together creating the                                                                              | Supplied in a two<br>chamber container<br>system. At the time of<br>use, the contents of the<br>two chambers are<br>mixed together creating                                                                                                               |

{5}------------------------------------------------

| Parameter                                                                       | Predicate Device<br>ACCUSOL, K041428                                                                                                                                                                                                                                                                                                                                                            | HMB32                                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                 | dialysate solution. The<br>mixed solution is<br>delivered to the<br>dialysate pathway of the<br>blood tubing path<br>through the<br>administration site<br>consisting of a<br>membrane tube and a<br>pull-ring closure and<br>connected to the bottom<br>chamber. A medication<br>port is provided for<br>optional insertion of<br>other substances into<br>the mixed solution prior<br>to use. | the dialysate solution.<br>The mixed solution is<br>delivered to the<br>dialysate pathway of<br>the blood tubing path<br>through the<br>administration tube<br>connected to the<br>bottom chamber, which<br>is opened and closed<br>using a luer lock. A<br>second tube in the large<br>chamber can be used<br>for optional insertion of<br>other substances into<br>the mixed solution prior<br>to use. |
| Biocompatibility<br>of patient<br>contacting<br>container/closure<br>components | Established via testing<br>per ISO 10993-1                                                                                                                                                                                                                                                                                                                                                      | Established via testing<br>per ISO 10993-1                                                                                                                                                                                                                                                                                                                                                               |
| Sterilization                                                                   | moist heat                                                                                                                                                                                                                                                                                                                                                                                      | moist heat                                                                                                                                                                                                                                                                                                                                                                                               |
| Shelf life                                                                      | 12 months                                                                                                                                                                                                                                                                                                                                                                                       | 12 months                                                                                                                                                                                                                                                                                                                                                                                                |

### 11. NONCLINICAL TESTING:

Biocompatibility: The HMB32 solution consists of components found in human plasma and is in compliance with the European Pharmacopeia 8.0, "Haemofiltration and haemodiafiltration, solution for," current edition 07/2013:0861. The materials of the container/closure system come into contact with the solution, but do not come into direct contact with the patient. Biological assessments are provided to demonstrate that the materials in contact with the solution are medical grade, meet ISO 10993-1 and USP Class VI plastics requirements, and are commonly used for medical devices and injectable pharmaceuticals.

Sterilization Validation: The steam sterilization method was validated to an SAL of 10° in accordance with EN ISO 17665-1: 2007 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of sterilization process for medical devices.

Pyrogenicity testing conducted using the LAL test confirmed that the solution is not pyrogenic.

{6}------------------------------------------------

Stability testing conducted after real-time aging out to 18 months confirmed the integrity of both the solution and the container/closure system. The HMB32 product will be labeled with a 12 month shelf life.

# 12. CONCLUSION:

The information and testing presented in this 510(k) demonstrates that the HBiofluids HMB32 is safe and effective for use during Continuous Renal Replacement Therapy as demonstrated by confirmation of substantial equivalence to the Accusol dialysis solution cleared under K041428.

---

**Source:** [https://fda.innolitics.com/device/K150966](https://fda.innolitics.com/device/K150966)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
