The LeForte System II is intended for use in selective trauma of the mid-face, reconstruction procedure and selective orthognathic surgery of the maxilla and chin.
Device Story
LeForte System II is a rigid fixation system comprising titanium plates and screws; used for mid-face trauma, reconstruction, and orthognathic surgery. Plates and screws are provided non-sterile; require autoclave sterilization by user prior to surgery. System includes manual surgical instruments (drill bits, driver shafts, depth gauge, bender, hand body). Surgeon selects appropriate plate configuration and screw type (self-tapping, self-drilling, or self-tapping locking) based on clinical need. Devices provide mechanical stabilization of bone segments to facilitate healing. System is used in clinical settings by surgeons.
Clinical Evidence
Bench testing only. Compliance with ASTM F382-99 (metallic bone plates) and ASTM F543-07 (metallic bone screws) verified via dimension, bending strength, driving torque, axial pull-out, and torsion tests. No clinical data provided.
Technological Characteristics
Materials: Pure Titanium (ASTM F67) for plates; Titanium Alloy Ti-6Al-4V (ASTM F136) for screws. Features: Polyaxial locking, non-locking, and compression configurations. Dimensions: Plates 0.5-2.5mm thick; screws 1.3-2.7mm diameter, 2-20mm length. Surface: Anodized. Sterilization: Autoclave (user-performed).
Indications for Use
Indicated for patients requiring fracture fixation in mid-face trauma, reconstruction procedures, and orthognathic surgery of the maxilla and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
OsteoMed Modular Locking Fixation System (K080694)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2015
Jeil Medical Corporation c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 East Chapman Avenue, Suite 110 Fullerton, CA 92831
Re: K150965
Trade/Device Name: LeForte System II Regulation Number: 21 CFR 872.4760 Regulation Name: Bone plate Regulatory Class: II Product Code: JEY Dated: August 14, 2015 Received: August 19, 2015
#### Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for JEIL MEDICAL CORPORATION. The logo consists of a colorful abstract design on the left, followed by the word "JEIL" in orange, and the words "MEDICAL CORPORATION" in gray. The abstract design features a combination of green, blue, orange, and gray shapes.
702·703·704·705·706·804·805·807·812-ho, 55, Digital-ro34-gil, Guro-gu, 152-728, Korea Tel : +82 2 850 3500 / Fax : +82 2 850 3535
# Indication for Use
510(k) Number (if known): K150965
Device Name: LeForte System II
Indication for Use:
The LeForte System II is intended for use in selective trauma of mid-face, reconstruction procedure and selective orthognathic surgery of the maxilla and chin.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CRF 807.92]
#### 1. Date Prepared [21 CRF 807.92(a)(a)]
August 31, 2015
## 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Sponsor: Jeil Medical Corporation Address: 702-703-704-705-706-804-805-807-812-ho, 55 i Digital-ro34-gil, Guro-gu, Seoul, 152-728, Korea
. Contact Name: Jieun KIM (Ms.) / RA Specialist
- Telephone No. : +82 2 850 3500 -
- Fax No. : +82 2 850 3525 ı
- -Email Address : jekim@jeilmed.co.kr
- Registration Number: 3004049923 .
- Name of Manufacturer: Same as Sponsor
#### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] 3.
- Trade Name: • LeForte System II
- Common Name: Bone plates and screws •
- Classification Name: Single/multiple component metallic bone fixation appliances and • accessories
- Classification Panel: Dental
- . Classification Regulation: 21 CFR 872.4760
- Product Code: JEY ●
- . Device Class: II
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
Primary Predicate Device:
| • 510(k) Number: | K112457 |
|------------------|--------------------------|
| • Applicant: | Jeil Medical Corporation |
| • Common Name: | Bone plates and screws |
| • Device Name: | LeForte System |
Reference Predicate Device:
| • 510(k) Number: | K080694 |
|------------------|---------------|
| • Applicant: | OsteoMed L.P. |
{4}------------------------------------------------
- Common Name: Bone plates and screws •
- . Device Name: OsteoMed Modular Locking Fixation System
# 5. Description of the Device [21 CFR 807.92(a)(4)]
The LeForte System II is rigid fixation consisting of plates and screws in various configurations, shapes and sizes.
| | Plate | Self-tapping<br>Screw | Self-drilling<br>Screw | Self-tapping<br>Locking Screw |
|-------------------------|----------------------------|---------------------------------------------|---------------------------------------------|---------------------------------------------|
| Type /<br>Configuration | 242.Series | 241.01 Series | 241.02 Series | 241.03 Series |
| Material | ASTM F 67<br>Pure Titanium | ASTM F 136<br>Titanium Alloy<br>(Ti-6Al-4V) | ASTM F 136<br>Titanium Alloy<br>(Ti-6Al-4V) | ASTM F 136<br>Titanium Alloy<br>(Ti-6Al-4V) |
The LeForte System II is made of Pure Titanium and Titanium Alloy (Ti-6AL-4V), which meet ASTM F67: Standard Specification for Unalloyed Titanium for Surgical Implant Applications, and ASTM F136: Standard Specification for Wrought Titanium-6 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility. The plates vary essentially through different lengths and number of plate holes. The Self-tapping screws, Self-drilling screws & Self-tapping locking screws diameter is from 1.3 to 2.7mm and lengths is from 2mm to 20mm. It also includes various manual surgical instruments such as drill bits, driver shafts, depth gauge, bender and hand body, etc.. The LeForte System II is not provided sterile. It is required to be sterilized via autoclave method to reach a SAL of 10° by the user prior to surgery.
# 6. Indications for Use [21 CFR 807.92(a)(5)]
The LeForte System II is intended for use in selective trauma of the mid-face, reconstruction procedure and selective orthognathic surgery of the maxilla and chin.
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
Le Forte System II. Bone Plates: Based on a technical feature comparison, the subject device was found to be similar to the predicate devices in regards to design and materials. The subject plates also have a polyxial locking feature, similar to the design used in the primary predicate device (K112457).
LeForte System II. Bone Screws: Both the subject and the predicate devices share similar head, neck and thread designs.
## Non-Clinical Test Summary:
Bench tests were conducted to verify that the proposed device met all the design specifications. The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
- ASTM F 382-99, Standard Specification and Test Method for Metallic Bone Plates
- ASTM F 543-07, Standard Specification and Test Methods for Metallic Medical Bone . Screws
The following tests were performed on the subject device and the primary predicate device:
- . Plates
- -Dimension test
- Bending strength test per ASTM F382 -
- Screws
- Dimension test .
- Driving torque test per ASTM F543-07
- । Axial pull-out test per ASTMV F543-07
- . Torsion test per ASTM F543-07
The results of the testing support that the LeForte System II is substantially equivalent to the predicate device.
## 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
| | Subject Device | Primary Predicate<br>Device | Reference<br>Predicate Device | Similarity/Difference<br>Discussion | |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| | | | | | |
| Device Name | LeForte System II | LeForte System<br>Bone Plate & Screw | OsteoMed Modular<br>Locking Fixation<br>System | Different | |
| 510(k) Number | K150965 | K112457 | K080694 | Different | |
| Manufacturer | Jeil Medical<br>Corporation | Jeil Medical<br>Corporation | OsteoMed L.P. | Different | |
| Indications for<br>Use | Intended for use in<br>Selective trauma of<br>the mid-face,<br>reconstruction<br>procedures and<br>selective<br>orthognathic surgery<br>of the maxilla and<br>chin. | Intended for use in<br>Selective trauma of<br>the mid-face,<br>reconstruction<br>procedures and<br>selective<br>orthognathic surgery<br>of the maxilla and<br>chin. | Indicated for fracture<br>fixation in cranio-<br>maxillofacial trauma<br>reconstruction,<br>mandibular<br>reconstruction and<br>orthognathic<br>reconstruction. | Same | |
| | Subject Device | Primary Predicate<br>Device | Reference<br>Predicate Device | Similarity/Difference<br>Discussion | |
| Plate Design | • Non-Locking Type<br>- Curved Type<br>- Straight Type<br>- Square Type<br>- Quad Type<br>- L Type<br>- Double Y Type<br>- X Type<br>- I Type<br>- Chin Type | • Non-Locking Type<br>- Curved Type<br>- Straight Type<br>- Square Type<br>- L Type<br>- Double Y Type<br>- X Type<br>- I Type<br>- Chin Type<br>- Calvarium Type<br>- Mesh Type<br>- H Type<br>- Hexagon Type<br>- Quad Type<br>- T Type<br>- Y Type<br>- RC Type<br>- Z Type<br>- Rigid Straight<br>Type<br>- MG Type<br>- A Type<br>- Angled Type<br>- Angled<br>Reconstruction<br>Type<br>- Multi<br>Reconstruction<br>Type<br>- Reconstruction<br>Type<br>- BSSO Type | • Non-Locking Type<br>- Curved Type<br>- Straight Type<br>- L Type<br>- Double Y Type<br>- X Type<br>- I Type<br>- Chin Type<br>- Mesh Type<br>- H Type<br>- Quad(Square)<br>Type<br>- Y Type<br>- Z Type<br>- BSSO Type<br>- Orbital Type | Similarity<br>There are three same<br>plate type categories in<br>the subject device and<br>the predicate devices.<br>Difference There are<br>a few additions with<br>minor<br>shape changes to each<br>category in the subject<br>device comparing to the<br>primary predicate<br>device. However, the<br>difference in shape is<br>very minor; it is not<br>introducing significantly<br>different design, and the<br>performance test results<br>supported that this<br>difference does not raise<br>an issue in performance.<br>Overall, the subject<br>device plates are<br>designed to enhance<br>strength; the width and<br>the cross-sectional area<br>are larger than the<br>primary predicate<br>devices. | |
| | • Compression Type<br>- Straight Type<br>- Angled Type<br>- Curved Type | • Compression Type<br>- Straight Type<br>- Angled Type | • Compression Type<br>- Straight Type | | |
| | Subject Device | Primary Predicate<br>Device | Reference<br>Predicate Device | Similarity/Difference<br>Discussion | |
| | • Locking Type<br>- AJ Type<br>- Curved Type<br>- L Type<br>- Square Type<br>- Straight Type<br>- X Type<br>- I Type<br>- Double Y Type<br>- Z Type<br>- Angled Type<br>- Angled<br>Reconstruction<br>Type<br>- Multi<br>Reconstruction<br>Type<br>- Reconstruction<br>Type<br>- BSSO Type | • Locking Type<br>- Curved Type<br>- Straight Type<br>- Angled Type<br>- Reconstruction<br>Type<br>- Angled<br>Reconstruction<br>Type<br>- Multi<br>Reconstruction<br>Type<br>- BSSO Type | • Locking Type<br>- Curved Type<br>- Straight Type<br>- Angled Type<br>- Condyle Plate<br>- Comminution<br>Type<br>- Strut Type<br>- Reconstruction<br>Type | | |
| Plate Thickness | 0.5mm<br>0.6mm<br>0.8mm<br>1.0mm<br>1.3mm<br>2.5mm | 0.1mm<br>0.2mm<br>0.3mm<br>0.4mm<br>0.5mm<br>0.6mm<br>0.8mm<br>1.0mm<br>1.3mm<br>2.5mm | 0.25~2.5mm<br>1.0mm~2.5mm | Similarity<br>The plate thickness<br>range of the subject<br>device is within the<br>range of the predicate<br>devices. | |
| Hole Shape | "O" shape (circle) | "O" shape (circle) | "O" shape (circle) | Same | |
| Screw Type and<br>Diameter | • Self-tapping Screw<br>1.3mm<br>1.5mm(Emergency)<br>1.6mm<br>1.8mm(Emergency)<br>2.0mm<br>2.3mm(Emergency)<br>2.4mm<br>2.7mm(Emergency) | • Self-tapping Screw<br>1.2mm<br>1.5mm (Emergency)<br>1.6mm<br>1.9mm (Emergency)<br>2.0mm<br>2.3mm (Emergency)<br>2.4mm<br>2.7mm (Emergency) | • Self-tapping Screw<br>1.2mm<br>1.5mm<br>1.6mm<br>1.9mm<br>2.0mm<br>2.3mm<br>2.4mm<br>2.7mm | Similarity<br>The subject device<br>screws are designed to<br>enhance strength; the<br>core-diameter of the<br>screws is larger than the<br>primary predicate<br>devices.<br>The screw diameter and | |
| | • Self-drilling Screw<br>1.4mm<br>1.6mm<br>2.0mm | • Self-drilling Screw<br>1.4mm<br>1.6mm<br>2.0mm | • Self-drilling Screw<br>1.2mm<br>1.6mm<br>2.0mm | length range of the<br>subject device is within<br>the range of the | |
| | | Subject Device | Primary Predicate<br>Device | Reference<br>Predicate Device | Similarity/Difference<br>Discussion |
| | | • Self-tapping<br>Locking Screw<br>2.0mm<br>2.4mm | • Locking Screw<br>(Self-tapping<br>Locking Screw)<br>2.0mm<br>2.3mm Emergency<br>2.4mm<br>2.7mm Emergency | • Locking Screw<br>(Self-tapping<br>Locking Screw)<br>2.0mm<br>2.3mm<br>2.4mm<br>2.7mm | predicate devices. |
| Screw Length | | • Self-tapping Screw<br>2 - 20mm<br>• Self-drilling Screw<br>3 - 12mm | • Self-tapping Screw<br>2 - 20mm<br>• Self-drilling Screw<br>3 - 12mm | • Self-tapping Screw<br>2 - 20mm<br>• Self-drilling Screw<br>3 - 8mm | |
| | | • Self-tapping<br>Locking Screw<br>3 - 20mm | • Self-tapping<br>Locking Screw<br>4 – 18mm | • Self-tapping<br>Locking Screw<br>4 – 22mm | |
| Bone plates are<br>used with<br>general surgical<br>instrumentation | | Yes | Yes | Yes | Same |
| Material | Plate | Titanium (ASTM<br>F67) | Titanium (ASTM<br>F67) | Titanium (ASTM<br>F67) | Same |
| | Screw | Titanium Alloy<br>(ASTM F136) | Titanium Alloy<br>(ASTM F136) | Titanium Alloy<br>(ASTM F136) | Same |
| Surface casting | | Plate - Anodizing<br>Screw - Anodizing | Plate - Anodizing | Plate - Anodizing<br>Screw - Anodizing | Same |
| Sterilization | | Non-sterile, Steam<br>sterilization prior<br>to use | Non-sterile,<br>Steam sterilization<br>prior to use | Non-sterile,<br>Steam sterilization<br>prior to use | Same |
| Single Use | | Yes | Yes | Yes | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
When compared to the predicate device (K112457), the LeForte System II presented in this submission has similar;
- . Intended Use
- . Technological characteristics
- Operating principle .
- . Design features
- Performance
- . Biocompatibility
- . Materials
- . Size
- . Method of sterilization and sterility assurance level
The subject device has slight differences in design from the predicate devices which are intended to enhance strength with the wider plates and the screws having larger core diameter. However, the performance test results support that this difference does not raise an issue in performance.
{9}------------------------------------------------
OsteoMed Modular Locking Fixation System (K080694) is identified as a reference predicate device which encompasses the length range of the subject device.
# 9. Conclusion [21 CFR 807.92(b)(3)]
In all the respects, the LeForte System II is the substantially equivalent to currently marketed devices. This device is made of the same materials and has similar dimensions and characteristics. Based on the information provided in this submission, we conclude that the LeForte System II is substantially equivalent to the predicates in design, material, and functions.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.